RENU MEDICAL VASOPRESS CALF , THIGH GARMENTS- ALL SIZESD, RENU MEDICAL VASOPRESS FOOT GARMENTS

K093658 · Renu Medical, Inc. · JOW · Jan 14, 2010 · Cardiovascular

Device Facts

Record IDK093658
Device NameRENU MEDICAL VASOPRESS CALF , THIGH GARMENTS- ALL SIZESD, RENU MEDICAL VASOPRESS FOOT GARMENTS
ApplicantRenu Medical, Inc.
Product CodeJOW · Cardiovascular
Decision DateJan 14, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ReNu Medical Reprocessed Vaso Press CTC calf, thigh and foot DVT garments are to be used by patients in both the home and institutional settings as a non-invasive therapeutic method to prevent deep vein thrombosis and resulting pulmonary embolism.

Device Story

Reprocessed Vaso Press CTC DVT garments (calf, thigh, foot) used for non-invasive deep vein thrombosis (DVT) and pulmonary embolism prevention. Device functions as a compression sleeve; used in home and institutional settings. Reprocessing methodology ensures performance equivalence to original devices. No active electronic components or software described.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Compression sleeve for DVT prevention. Identical to predicate devices in materials, design, and mode of operation. No energy source or software described.

Indications for Use

Indicated for patients in home and institutional settings for non-invasive prevention of deep vein thrombosis and pulmonary embolism.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the text "K093658 pg 1 of 2" written in black ink on a white background. The text appears to be handwritten. The text "K093658" is larger than the text "pg 1 of 2". The text "pg 1 of 2" indicates that this is page 1 of a 2-page document. ## 510(k) Summary Submitters Name and Address: ReNu Medical, Inc. 9800 Evergreen Way Everett, WA 98024 Phone: 425-353-1110 Fax: 425-353-9116 JAN 1 4 2010 FDA Registration Number: 3034520 Contact Person: L. Bruce Pierson Chief Operating Officer Date Summary Prepared: 01/06/10 Trade or Proprietary Name(s): ReNu Medical Reprocessed Vaso Press CTC calf, thigh and foot DVT garments. (All sizes) Common Name: Sleeve, limb, compression Product Code: JOW Panel: Cardiovascular 870.5800 Classification: Class II Equivalent Device(s) The ReNu Medical Reprocessed ALP DVT garments. 510k# K051227 ReNu Medical Reprocessed many different manufacturers types/sizes DVT garments 510k# K031159 Device Description: The ReNu Medical Reprocessed Vaso Press CTC calf, thigh and foot DVT garments. (All sizes) Indications for Use: The ReNu Medical Reprocessed Vaso Press CTS Calf, Thigh and foot DVT garments are to be used by patients in both the home and institutional settings as a non-invasive therapeutic method to prevent deep vein thrombosis and resulting pulmonary embolism. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a handwritten text that appears to be a combination of alphanumeric characters and some additional text. The alphanumeric part reads 'K093658', and next to it, there is 'Pg 2 of 2'. The text is written in black ink on a white background, and the handwriting style is cursive. ## Technological characteristics of the ReNu Medical Reprocessed Vaso Press CTC DVT garments: The predicate devices and the ReNu Medical Vaso Press CTC DVT garments are identical in intended use, fundamental scientific technology, overall design, materials, energy source, mode of operation, performance techniques and reprocessing methodology. ## Summary of Comparison Tests Bench testing was conducted to ensure that reprocessing did not compromise the performance of the device in a manner that is substantially equivalent to that of the predicate devices. ### Biocompatibility Reprocessing does not affect the biocompatibility of the device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 JAN 1 4 2010 ReNu Medical, Inc. c/o Mr. Bruce Pierson, COO 9800 Evergreen Way Everett, WA 98204 Re: K093658 ReNu Medical Reprocessed Vaso Press CTC Calf, Thigh and Foot DVT garments. (ALL Sizes) Regulation Number: 21 CFR §870.5800 Regulation Name: Sleeve, Limb, Compressible Regulatory Class: Class II Product Code: JOW Dated: November 20, 2009 Received: November 25, 2009 Dear Mr. Pierson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Bruce Pierson, COO Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. sma R. Viher Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K093658 # Indications for Use 510(k) Number (if known): K093658 Device Name: ReNu Medical Reprocessed Vaso Press CTC Calf, Thigh and Foot DVT garments. (All Sizes) Indications for Use: The ReNu Medical Reprocessed Vaso Press CTC calf, thigh and foot DVT garments are to be used by patients in both the home and institutional settings as a non-invasive therapeutic method to prevent deep vien thrombosis and resulting pulmonary embolism. Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) R.V. Luner Division Sign-Off) (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_Kog3(e5) 5
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...