K080232 · Zoll Medical Corporation, World Wide Headquarters · MKJ · Apr 23, 2008 · Cardiovascular
Device Facts
Record ID
K080232
Device Name
ZOLL E SERIES 12SL, ANALYSIS OPTION
Applicant
Zoll Medical Corporation, World Wide Headquarters
Product Code
MKJ · Cardiovascular
Decision Date
Apr 23, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The ZOLL E Series with 12SL™ is intended for the recording and automated analysis of 12-Lead ECG signals acquired from adult and pediatric patients in the supine, resting position. This product may be used for monitoring various patient vital signs, including: electrocardiogram (ECG), Pulse Oximetry (SpO2), End Tidal CO2, 12-Lead ECG, and Non-Invasive Blood Pressure (NIBP). 12 Lead ECG analysis is indicated for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction.
Device Story
ZOLL E Series is a multi-parameter monitor/defibrillator; integrates defibrillation, ECG, pacing, SpO2, EtCO2, and NIBP. 12SL™ Analysis Option uses GE/Marquette 12SL™ algorithm to process 12-lead ECG signals; provides automated interpretation for AMI diagnosis and arrhythmia documentation. Includes ACI-TIPI and TPI decision aids for clinicians. Used in emergency/clinical settings by qualified medical personnel. Device acquires signals via 10-wire cable/electrodes; stores/transmits data via cellular, RS232, or Bluetooth. Output displayed to clinician for diagnostic/therapeutic decision-making; assists in identifying cardiac ischemia/arrhythmias. Benefits include rapid diagnostic support and integrated patient monitoring in acute care.
Clinical Evidence
Bench testing only. Performance and safety testing conducted to ensure compliance with recognized industry standards and functional specifications. No clinical trial data presented.
Technological Characteristics
Multi-parameter monitor/defibrillator. 12-lead ECG acquisition via 10-wire cable and standard electrodes. Connectivity: cellular, RS232, Bluetooth. Software: GE/Marquette 12SL™ algorithm for ECG interpretation. Energy: DC defibrillator (up to 200J). Pacing: Non-invasive transcutaneous pacing. Materials: Disposable MFE pads or paddles.
Indications for Use
Indicated for adult and pediatric patients in supine, resting position requiring 12-lead ECG recording and analysis for diagnosis/treatment of acute myocardial infarction (AMI) and interpretation of transient cardiac arrhythmias. Contraindications: Rhythm analysis unreliable with implantable pacemakers; do not use during patient movement (stretcher/vehicle).
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
K051475 — ZOLL M SERIES 12SL ANALYSIS OPTION · ZOLL Medical Corporation · Jun 24, 2005
K112761 — ZOLL PROPAQ MD · Zoll Medical Corporation, World Wide Headquarters · Apr 25, 2012
K042007 — ZOLL E SERIES DEFIBRILLATOR · Zoll Medical Corp · Apr 7, 2005
K991556 — ZOLL M SERIES 12SL ANALYSIS OPTION · Zoll Medical Corp · Oct 21, 1999
K092598 — ZOLL E SERIES WITH SPCO/SPMET OPTION · Zoll Medical Corporation, World Wide Headquarters · Dec 3, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K080232
# 510(k) Summary:
APR 2 3 2008
Submitter's Name and Address:
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105 (978) 421-9655
Contact Person:
Eileen M. Boyle
(978) 421-9171
Date Summary Prepared:
January 23, 2008
Device:
ZOLL E Series 12SL™ Analysis Option
Classification:
Electrocardiograph: Class II (21 CFR 870.2340)
Automatic External Defibrillators: Class III (21 CFR 870.5310)
Description:
The ZOLL E Series products (K042007) combine a defibrillator, ECG Monitor, Noninvasive Transcutaneous Pacing, Pulse Oximetry, End Tidal CO2 , 12-Lead ECG Monitoring, Non-Invasive Blood Pressure measurement Invasive Blood Pressure and Temperature and data printing and recording in a single instrument.
The 12SL™ Analysis Option using the GE/Marquette 12SLTM Algorithm is useful in the diagnosis of patients with acute myocardial infarction (AMI) and is useful in the interpretation and documentation of other transient cardiac arrhythmias that may occur. The 12-Lead ECG Analysis is indicated for the recording and analysis of 12 Lead ECG signals acquired from adult and pediatric patients in the supine, resting position. The Acute Cardiac Ischemia-Time Insensitive Predictive Instrument (ACI-TIPI) and Thrombolytic Predictive Instrument (TPI) are decision aids for qualified clinicians who may currently wish to calculate ACI-TIPI and TPI in adult patients.
{1}------------------------------------------------
K080d32
### Intended Use:
The ZOLL E Series with 12SL™ is intended for the recording and automated analysis of 12-Lead ECG signals acquired from adult and pediatric patients in the supine, resting position.
# Substantial Equivalence:
The features and functions of the proposed enhancement to the E Series 12SL™ Analysis Option are substantially equivalent to the current features and functions of the E Series 12SL™ Analysis Option (K042007) including the ACI-TIPI feature, cleared for use on 04/07/2005.
# Comparison of Technological Characteristics
The ZOLL E Series 12SLTM Analysis Option utilizes the same interpretive features and functions to those of the currently marketed ZOLL E Series 12SL™ Analysis Option (K042007). The device acquires an ECG signal through a 10wire cable assembly and commonly used patient electrodes placed in a standard 12-Lead configuration. The device is also capable of storing that data to memory and/or transmitting that data via cellular telephone, RS232 port, or wirelessly using Bluetooth™ technology.
# Performance Testing:
Extensive performance testing ensures that the ZOLL E Series 12SL™ Analysis Option performs as well as the indicated predicate devices and meets all of its functional requirements and performance specifications. Safety testing assures the device complies with applicable sections of recognized industry and safety standards.
# Conclusion
Performance and safety testing of the ZOLL E Series 12SL™ Analysis Option demonstrates that its features and functions are substantially equivalent to those of the indicated commercially distributed predicate devices with regard to performance, safety and effectiveness.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/11 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 3 2008
Zoll Medical Corporation c/o Ms. Eileen M. Boyle Regulatory Affairs Specialist 269 Mill Road Chelmsford, MA 01824
Re: K080232
Trade/Device Name: ZOLL E Series 12SL Analysis Option Regulation Number: 21 CFR 870.5310 Regulation Name: Automated external defibrillator Regulatory Class: Class III (three) Product Code: MKJ Dated: March 7, 2008 Received: March 27, 2008
Dear Ms. Eileen M. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Eileen M. Boyle
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Scctions 531-542 of the Act); 21 CMR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilitics under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bfimmerman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Ko8oj3d
510(k) Number (if known):
#### Device Name: ZOLL E Series
# Defibrillator Function
The E Series products contain a DC defibrillator capable of delivering up to 200 joules of energy. It may be used in synchronized mode to perform synchronized cardioversion by using the R-wave of the patient's ECG as a timing reference. The unit uses paddles or disposable, pre-gelled, MFE Pads for defibrillation.
#### Intended Use - Manual Operation
Use of the E Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- Unconsciousness t
- Absence of breathing
- Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
#### Intended Use - Semiautomatic Operation (AED)
The E Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.
They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.
The E Series products must be prescribed for use by a physician or medical advisor of an emergency response team.
Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- . Unconsciousness
- Absence of breathing
- Absence of pulse.
#### (continued on next page)
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Ol)
Division of Cardiovascular Devices
510(k) Number K080232
{5}------------------------------------------------
K080232
Indications for Use (continued from previous page)
. ..... .
### Semiautomatic Operation Contraindications for Use
The rhythm analysis function may not reliably identify ventricular fibrillation in the presence of an implantable pacemaker. Inspection of the electrocardiogram and clinical evidence of cardiopulmonary arrest should be the basis for any treatment of patients with implantable pacemakers.
Do not use the rhythm analysis function during patient movement on a stretcher or in an ambulance or other conveyance. A patient must be motionless during ECG analysis. Do not touch the patient during analysis. Cease all movement via stretcher or vehicle prior to analyzing the ECG. If using the device in an emergency vehicle, bring the vehicle to a halt before activating the analysis function.
#### Defibrillator Complications
Inappropriate defibrillation or cardioversion of a patient (e.g., with no malignant arrhythmia) may precipitate ventricular fibrillation, asystole, or other dangerous arrhythmias.
Defibrillation without proper application of electrode pads or paddle electrolyte gel may be ineffective and cause burns, particularly when repeated shocks are necessary. Erythemia or hypermia of the skin under the paddles or MFE Pads often occurs; this effect is usually enhanced along the perimeter of the paddle or electrode. This reddening should substantially lessen within 72 hours.
### External Pacemaker (Pacer Version Only)
Non-invasive Transcutaneous Pacing (NTP) is an established and proven technique. This therapy is easily and rapidly applied in both emergency and non-emergency situations when temporary cardiac stimulation is indicated.
#### Intended Use - Pacemaker
This product may be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.
Note: This device must not be connected to internal pacemaker electrodes.
The purposes of pacing include:
# Resuscitation from standstill or bradycardia of any etiology:
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, quinidine, digitails, b-blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.
## As a standby when standstill or bradycardia might be expected:
Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous therapy that avoids the risks of displacement, infection, hemorrhage, embolization,
{6}------------------------------------------------
K080J3d
perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.
#### Suppression of tachycardia:
Increased heart rates in response to external pacing often suppress ventricular ectopic activity and my prevent tachycardia.
#### Pacemaker Complications
Ventricular fibrillation will not respond to pacing and requires immediate defibrillation. The patient's dysrhythmia must therefore be determined immediately, so that appropriate therapy can be employed. If the patient is in ventricular fibrillation and defibrillation is successful, but cardiac standstill (asystole) ensues, the pacemaker should be used.
Ventricular or supraventricular tachycardias may be interrupted with pacing but in an emergency or during circulatory collapse, synchronized cardioversion is faster and more certain.
Electromechanical dissociation may occur following prolonged cardiac arrest or in other disease states with myocardial depression. Pacing may then produce ECG responses without evvective mechanical contractions, and other treatment is required.
Pacing may evoke undesirable repetitive responses, tachycardia, or fibrillation in the presence of generalized hypoxia, myocardial ischemia, cardiac drug toxicity, electrolyte imbalance, or other cardiac diseases.
Pacing by any method tends to inhibit intrinsic rhythmicity. Abrupt cessation of pacing, particularly at rapid rates, can cause ventricular standstill and should be avoided.
Noninvasive Temporary Pacing may cause discomfort of varying intensity, which occasionally can be severe and preclude its continued in conscious patients.
Similarly, unavoidable skeletal muscle contraction may be troublesome in very sick patients and may limit continuous use to a few hours. Erythemia or hypermia of the skin under the MFE Pads often occurrs; this effect is usually enhanced along the perimeter of the electrode.
This reddening should substantially lessen within 72 hours.
There have been reports of burns under the anterior electrode when pacing adult patients with rnore have a some of the skin. Prolonged pacing should be avoided in these cases and periodic inspection of the skin is advised.
There are reports of transient inhibition of spontaneous respiration in unconscious patients with previously available units when the anterior electrode was placed too low on the abdoment.
This device must not be connected to internal pacemaker electrodes.
#### Pediatric Pacing
Pacing can be performed on pediatric patients weighing 33lbs / 15kg or less using special ZOLL pediatric MFE Pads. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic insepction of the underlying skin is recommended.
{7}------------------------------------------------
K08 od 32
#### Monitor
#### Intended-Use Multi-parameter Monitoring
This product may be used for monitoring various patient vital signs, including: electrocardiogram (ECG), Pulse Oximetry (SpO2), End Tidal CO2, 12-Lead ECG, and Non-Invasive Blood Pressure (NIBP).
ECG monitoring is indicated by connecting the patient to the unit via the 3 or 5 lead patient cable MFE Pads, or through the paddles.
SpO2 monitoring is indicated for detecting arterial oxygen saturation of blood and pulse rate for adult, pediatric and neonatal patients who are well or poorly perfusing, during both no motion and patient motion conditions.
EtCO2 monitoring is indicated for the continuous measurement of end tidal carbon dioxide (EtCO2) and respiration rate for adult, pediatric and neonatal patients.
12 Lead ECG analysis is indicated for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction.
NIBP monitoring is indicated for the measurement of arterial blood pressure for resting adult, pediatric, and neonatal patients.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.