The ZOLL E Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing, and multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, End Tidal CO2, 12-Lead ECG Monitoring, Non-Invasive Blood Pressure measurement and data printing and recording for resting patients in critical care and transport. The ZOLL E Series intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment.
Device Story
ZOLL E Series is a multi-parameter monitor and defibrillator for critical care and transport. Inputs include ECG (3/5 lead, pads, paddles), SpO2, EtCO2, and NIBP. Device performs manual or semiautomatic (AED) defibrillation, noninvasive transcutaneous pacing, and vital sign monitoring. Operated by trained medical personnel. ECG analysis algorithm provides shock advisory determinations. Output displayed on-screen and printed/recorded. Clinical decision-making supported by real-time physiological data and automated rhythm analysis; facilitates rapid intervention for cardiac arrest and hemodynamic instability.
Clinical Evidence
Bench testing only. Performance and safety testing verified functional requirements and compliance with industry standards. No clinical trial data presented.
Technological Characteristics
Multi-parameter monitor/defibrillator. Energy source: DC defibrillator (up to 200J). Sensing: ECG, SpO2 (Masimo SET), EtCO2 (Respironics Novametrix/Capnostat), NIBP (SunTech). Connectivity: Data printing/recording. Software: Rule-based/algorithmic analysis (GE/Marquette 12SL algorithm).
Indications for Use
Indicated for defibrillation, noninvasive pacing, and multi-parameter monitoring (ECG, SpO2, EtCO2, NIBP) in adult, pediatric, and neonatal patients. Defibrillation indicated for ventricular fibrillation and rapid ventricular tachycardia in cardiac arrest (unconscious, apneic, pulseless). Pacing indicated for asystole, profound bradycardia, or standby use. Monitoring indicated for critical care and transport environments. Contraindicated for connection to internal pacemaker electrodes.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
APR 7 2005
Image /page/0/Picture/1 description: The image shows the word "ZOLL" in large, bold, black letters on the left side of the image. To the right of the word "ZOLL", there is a handwritten text "K042" in black ink. The text "K042" is smaller in size compared to the word "ZOLL". The background of the image is white.
ZOLL Medical Corporation
Worldwide Headquarters 269 Mill Road Chelmsford, Massachusetts 01824-4105 U.S.A.
(978) 421-9655 (978) 421-0025 Main Fax
# 510(k) Summary:
Submitter's Name and Address:
ZOLL Medical Corporation Worldwide Headquarters 269 Mill Road Chelmsford, MA 01824-4105 (978) 421-9655
Contact Person:
Sean Reynolds (978) 421-9655, Ext. 9386
Date Summary Prepared:
July 23, 2004
Device:
ZOLL E Series
Classification:
Defibrillator, Low-energy - DC: Class II (21 CFR 870.5300)
Automated External Defibrillator: Class III (21 CFR 870.5310)
Cardiac Monitors (including Cardiotachometers and Rate Alarms): Class II (21 CFR 870.2300
Electrocardiograph: Class II (21 CFR 870.2340)
External Transcutaneous Cardiac Pacemakers (Non-invasive): Class II (21CFR 870.5550)
Analyzer, Gas, Carbon Dioxide; Class II (21 CFR 868.1400)
Oximeters: Class II (21 CFR 870.2700)
System, Measurement, Blood-Pressure, Non-Invasive: Class II (21 CFR 870.1130)
Description:
The ZOLL E Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing, and multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, End Tidal CO2, 12-Lead ECG Monitoring, Non-Invasive Blood Pressure measurement and data printing and recording for resting patients in critical care and transport. The ZOLL E Series intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment.
{1}------------------------------------------------
# Substantial Equivalence:
K04J007
The features and functions of the ZOLL E SERIES are substantially equivalent to The Toutaroo and Rinces Biphasic: 510(k) No. k990762, cleared 9/03/1999, the M SERIES SpO2 Option 510(k) No. K982992, cleared 3/11/1999, the M the M OERIES End Tidal CO2 Option: 510(k) No. K993036 cleared 2/28/2000, the M SERIES 12SL Analysis Option 510(k) No. K991556 cleared 10/21/1999, the M SERIES Non-Invasive Blood Pressure measurement Option: 510(k) No. K032363, cleared 10/31/2003, and the ZOLL AED Plus: 510(k) No. K033474, cleared 5/21/2004.
Intended Use:
### Defibrillation
Use of the E Series products in the manual mode for defibrillation is indicated for converting/terminating ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats. Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- Unconsciousness .
- Absence of breathing .
- . Absence of pulse.
This product should be used only by qualified medical personnel who are trained and authorized to respond to medical emergencies.
### External Pacemaker (Pacer Version Only)
This product may be used for temporary external, demand or non-demand pacing, as an alternative to endocardial stimulation. External cardiac pacing is indicated for use on conscious or unconscious patients in asystole, profound bradycardia or any other conditions determined by a clinician to require external pacing.
# ECG Monitoring
The device is indicated for monitoring a patient's electrocardiogram (ECG), via the 3 or 5 lead patient cable, MFE Pads, or through the paddles, for the purposes of identifying and diagnosing cardiac rhythms and dysrhythmias and calculating heart rate.
# SpO2 Option (if equipped)
The SpO2 Option with Masimo Set Technology is indicated for the continuous noninvasive monitoring of arterial oxygen saturation (SpO2) and pulse rate for adult, pediatric and neonatal patients, during both no motion and patient motion conditions, and for patients who are well or poorly perfusing in the hospital or pre-hospital environments.
# EtCO2 Option (if equipped)
The EtCO2 Option with Respironics Novametrix Technology and Capnostat® is indicated for the continuous noninvasive monitoring of end tidal carbon dioxide (EtCO2) and respiration rate in adult, pediatric and neonatal patient's requiring ventilatory support, transport and anesthesia.
{2}------------------------------------------------
K042007
# 12SL Analysis Option (if equipped)
The 12SL™ Analysis Option using the GE/Marquette 12SL™ Algorithm is useful in the 120E - Andrysis with acute myocardial infarction (AMI) and is useful in in the diagnosic of patiente with as other transient cardiac arrhythmias that the interprotution and Lead ECG Analysis is indicated for the recording and may oooal: - - That ECG signals acquired from adult and pediatric patients in the supine, resting position.
# NIBP Option (if equipped)
The NIBP Option with SunTech Medical Systems, Inc. Technology is indicated for the noninvasive measurement of arterial blood pressure for resting patients in critical care and transport. The NIBP Option on M Series units is designed to measure blood pressure for adult, pediatric, and neonatal patients.
# Comparison of Technological Characteristics
The ZOLL E Series design characteristics are the same as those of the indicated predicate devices; the technology is very similar to that of the ZOLL M Series. The ZOLL E Series acquires and analyzes ECG signals and provides shock advisory determinations for adult and pediatric patients. The ECG Analysis Algorithms are identical to those incorporated into the ZOLL AED Plus defibrillator. The ZOLL E Series provides multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, End Tidal CO2, 12-Lead ECG Monitoring, Non-Invasive Blood Pressure measurement, using technology very similar to those used by the ZOLL M Series.
Performance Testing:
Extensive performance testing ensures that the ZOLL E Series Defibrillator meets all of its functional requirements and performance specifications. Safety testing assures the device complies with applicable sections of recognized industry and safety standards.
### Conclusion
Performance and safety testing of the ZOLL E Series Defibrillator demonstrates that its features, functions and incorporated interpretive algorithm are substantially equivalent to that of the indicated commercially distributed predicate devices with regard to performance, safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is an abstract image of an eagle.
APR 7 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Zoll Medical Corporation c/o Mr. Sean Reynolds Regulatory Affairs Engineer Worldwide Headquarters 269 Mill Road Chelmsford, MA 01824-4105
Re: K042007
> Trade Name: Zoll E Series Regulation Number: 21 CFR 870.5310 Regulation Name: Automated external defibrillator Regulatory Class: Class III Product Code: MKJ Dated: March 11, 2005 Received: March 16, 2005
Dear Mr. Reynolds:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 -- Mr. Sean Reynolds
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised fill in I DA s issumes of a states complies with other requirements of the Act that FDA has made a decertimation administered by other Federal agencies. You must or ally redelal statutes and regaranents and admited to: registration and listing (21 comply with an the Act 3 requirements, moreans, and manufacturing practice requirements as set CFK Part 807), idocing (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Section your device as described in your Section 510(k) I ms letter will anow you to oegin maing of substantial equivalence of your device to a legally prematication. The PDF intalligstification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acreevier 7 at (240) 276-0120. Also, please note the regulation entitled, Connact the Office of Commarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blymmerman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Section 4 - Indications for Use
510(k) Number (if known): __________________________________________________________________________________________________________________________________________________ K042007
#### ZOLL E Series Device Name:
#### Defibrillator Function
The E Series products contain a DC defibrillator capable of delivering up to 200 joules of energy. It The L Sches products oonline a DO achern synchronized cardioversion by using the R-wave of may be used in synchionized mode to perferm of hone increase or disposable, pre-gelled, MFE Pads for defibrillation.
# Intended Use - Manual Operation
Use of the E Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- Unconsciousness •
- Absence of breathing .
- . Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
#### Intended Use - Semiautomatic Operation (AED)
The E Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.
They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.
The E Series products must be prescribed for use by a physician or medical advisor of an emergency response team.
Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- Unconsciousness ゃ
- Absence of breathing .
- . Absence of pulse.
#### (continued on next page)
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hummumuma
scular Devices
Page 1 of 2
{6}------------------------------------------------
Indications for Use (continued from previous page)
# External Pacemaker (Pacer Version Only)
# Intended Use — Pacemaker
This product may be used for temporary external cardiac pacing in conscious or unconscious of This product may be aber to endocardial stimulation.
Note: This device must not be connected to internal pacemaker electrodes.
The purposes of pacing include:
# Resuscitation from standstill or bradycardia of any etiology:
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagation standstill, drug induced standstill (due to procainamide, quiracy, angiography, and standstill, drug Induced (due (due verapamil, etc.) and unexpected circulatory artest (die to temporary and eneraliance of the more reliable, and other therapeutic or diagnosic procedures). It had also book and it is safer, more reliable, and bradycardia in Slokes-Adams discuse and olsk since bytal.
more rapidly applied in an emergency than endocardial or other temporary electrodes.
# As a standby when standstill or bradycardia might be expected:
s a standby whon btandby when cardiac arrest or symptomatic bradycardia Noninvasive pacing may be useful as a stalius, and toxicity, anethesia or surgey. It is
might be expected due to acute myocardial infarction, and surgey. It is might be expected due to acule myocaration market implants or the introduction
also useful as a temporary treatment in patients awaiting pacing may provide an also useful as a temporary treathen in pating applications, noninvasive pacing may provide an
of transvenous therapy. In standby pacing applications, infortion, he of transvenous therapy. In Standby paong applications, nown as a morrhage, and morrhage, and alternative to transverlous therapy that avoids the netto or displayers, and members of ventricular embolization, "perforution," phistical with endocardial pacing.
### Suppression of tachycardia:
uppression of table of catelling.
Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.
#### Pediatric Pacing
Pacing can be performed on pediatric patients weighing 33lbs / 15kg or less using special ZOLL Pacing can be performed on pedialic patients weighing outlouting in need as not of the mates, could pediatic Thi E T ads. T folongour pasing (as necommended.
#### Monitor
#### Intended-Use Multi-parameter Monitoring .
This product may be used for monitoring various patient vital signs, including: electrocardiogram This product may be used for monitoring valious patient vital signo, moduling on and online of the sure (NIBP).
(NDT ).
ECG monitoring is indicated by connecting the patient to the 3 or 5 lead patient cable, MFE Pads, or through the paddles.
will £ + as ; or this use the last in arterial oxygen saturation of blood and pulse rate for adult. SpO2 monitoring is indicated for ucteding archar oxygen sateralism of areas and patient motion conditions.
monon continuous .
EtCO2 monitoring is indicated for the continuous measurement of end tidal carbon dioxide (EtCO₂) and respiration rate for adult, pediatric and neonatal patients.
and respiration for adult postally posted for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction.
with actic myouthal marchiem.
NIBP monitoring is indicated for the measurement of arterial blood pressure for resting adult, pediatric, and neonatal patients.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.