TRUECPR COACHING DEVICE

K123597 · Physio-Control, Inc. · LIX · Apr 17, 2013 · Cardiovascular

Device Facts

Record IDK123597
Device NameTRUECPR COACHING DEVICE
ApplicantPhysio-Control, Inc.
Product CodeLIX · Cardiovascular
Decision DateApr 17, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5210
Device ClassClass 2
AttributesPediatric

Indications for Use

The TrueCPR device is intended to provide feedback to assist rescuers to perform cardiopulmonary resuscitation (CPR). Rescuers must be trained in CPR and use of the device. The TrueCPR device is intended for use on patients eight years of age and older.

Device Story

TrueCPR coaching device provides real-time feedback to rescuers performing CPR; assists in adherence to current guidelines. Device intended for use on patients 8 years and older by trained rescuers. Input data involves chest compression monitoring; device processes signals to provide feedback on compression performance. Output allows healthcare providers or rescuers to adjust compression technique in real-time, potentially improving CPR quality and patient outcomes. Used in clinical or emergency settings.

Clinical Evidence

No clinical data. Evidence based on comprehensive bench testing including requirements verification, electromagnetic compatibility, software performance, environmental testing, simulated use testing, and biocompatibility.

Technological Characteristics

External cardiac compressor/CPR aide. Provides real-time feedback on chest compressions. Includes software for performance monitoring. Verification and validation activities performed for safety and effectiveness.

Indications for Use

Indicated for use by trained rescuers to provide real-time feedback during CPR on patients 8 years of age and older.

Regulatory Classification

Identification

CPR Aid device with feedback is a device that provides real-time feedback to the rescuer regarding the quality of CPR being delivered to the victim, and provides either audio and/or visual information to encourage the rescuer to continue the consistent application of effective manual CPR in accordance with current accepted CPR guidelines (to include, but not be limited to, parameters such as compression rate, compression depth, ventilation, recoil, instruction for one or multiple rescuers, etc.). These devices may also perform a coaching function to aid rescuers in the sequence of steps necessary to perform effective CPR on a victim.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(i) Nonclinical performance testing under simulated physiological or use conditions must demonstrate the accuracy and reliability of the feedback to the user on specific compression rate, depth and/or respiration over the intended duration, and environment of use. (ii) Labeling must include the clinical training, if needed, for the safe use of this device and information on the patient population for which the device has been demonstrated to be effective (including patient size and/or age limitations, *e.g.,* adult, pediatric and/or infant).(iii) For devices that incorporate electrical components, appropriate analysis and testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment. (iv) For devices containing software, software verification, validation, and hazard analysis must be performed. (v) Components of the device that come into human contact must be demonstrated to be biocompatible. (vi) Human factors testing and analysis must validate that the device design and labeling are sufficient for effective use by the intended user. (3) *Premarket notification.* The CPR Aid with feedback device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter if it does not contain software (*e.g.,* is mechanical or electro-mechanical) and is in compliance with the special controls under paragraph (b)(2) of this section, subject to the limitations of exemptions in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ . . . . : ### SECTION E: 510(K) SUMMARY | 510(k) Summary | This summary of 510(k) safety and effectiveness information is being submitted in<br>accordance with the requirements of 21 C.F.R. § 807.92. | |-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter | Physio-Control, Inc.<br>11811 Willow Road Northeast<br>P.O. Box 97006<br>Redmond, Washington 98073-9706<br>Registration Number: 3015876 | | Contact Person | Teresa Davidson<br>Telephone: (425) 867-4733<br>Fax: (425) 867-4154<br>email: teresa.davidson@physio-control.com | | Date Summary<br>Prepared | November 20, 2012 | | Device Name | TrueCPR™ COACHING DEVICE | | Device Common<br>Name | Aid, Cardiopulmonary Resuscitation | | Regulation<br>Description | External Cardiac Compressor | | Device<br>Classification | Device Class: III<br>Product Code: LIX<br>Classification Panel: Cardiovascular Device<br>Regulation Number: 21 CFR 870.5200 | | Identification of<br>the Legally<br>Marketed Device | PocketCPR(K112660, K071321) and HeartStart MRx with Optional Q-CPR<br>(K051134) | | (Predicate<br>Device) | Device class: III<br>Product Code: LIX<br>Classification Panel: Cardiovascular Device<br>Regulation Number: 21 CFR 870.5200 | | Performance<br>Standards | This 510(k) includes documentation related to the verification and validation activities<br>supporting the safe and effective use of TrueCPR. | | Description | The TrueCPR coaching device provides rescuers with real-time feedback on chest<br>compressions during cardiopulmonary resuscitations (CPR) in accordance with<br>current guidelines. | E-1 {1}------------------------------------------------ #### SECTION E: 510(K) SUMMARY (continued) The TrueCPR device is intended to provide feedback to assist rescuers to perform Intended Use cardiopulmonary resuscitation (CPR). Rescuers must be trained in CPR and use of the device. The TrueCPR device is intended for use on patients eight years of ac e and older. Testing Types Bench Testing (Comprehensive verification testing): - Requirements Testing . - Electromagnetic Compatibility . - . Software Performance - . Environmental - Simulated Use Testing - . Biocompatibility Technological Characteristics of subject and predicate device Device intended use, feature and functional characteristics for predicate devices identified are substantially equivalent when compared to the proposed TrueCPR device. Importantly, no new issues of safety and effectiveness have been raised with the proposed device. Conclusion of The information in this 510(k) demonstrates that the TrueCPR сомсничь ремсе is testing substantially equivalent to the identified predicate devices with respect to safety, effectiveness, and performance. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 17, 2013 Physio-Control, Inc. c/o Teresa Davidson 11811 Willows Rd. NE PO Box 97006 Redmond, Washington 98073-9706 Re: K123597 Trade/Device Name: TrueCPR Coaching Device Regulation Number: 21 CFR 870.5200 Regulation Name: External Cardiac Compressor/CPR Aide Device Regulatory Class: Class III (three) Product Code: LIX Dated: April 3, 2013 Received: April 4, 2013 Dear Ms. Davidson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {3}------------------------------------------------ Page 2 - Ms. Teresa Davidson found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Matthew G. Hillebrenner tor for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Physio-Control, Inc TrueCPR Coaching Device 510(k) Pre-Market Notification # SECTION D: STATEMENT OF INDICATIONS FOR USE 510(k) Number (if known): K123597 Device Name: TrueCPR™ coaching device Indications for Use: The TrueCPR device is intended to provide feedback to assist rescuers to perform cardiopulmonary resuscitation (CPR). Rescuers must be trained in CPR and use of the device. The TrueCPR device is intended for use on patients eight years of age and older. Prescription Use _____________________________________________________________________________________________________________________________________________________________ イ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the commen (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Optional Format 3-10-98) # Matthew G. Hillebrenner
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...