K112660 · Zoll Medical Corp · LIX · May 10, 2012 · Cardiovascular
Device Facts
Record ID
K112660
Device Name
POCKETCPR
Applicant
Zoll Medical Corp
Product Code
LIX · Cardiovascular
Decision Date
May 10, 2012
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.5210
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
To assist users in the performance of effective CPR on a patient 8 years or older.
Device Story
PocketCPR is a portable CPR aid providing real-time feedback to users during cardiac arrest resuscitation, training, or practice. The device uses an internal accelerometer to measure chest wall movement during compressions. A microprocessor performs continuous analysis of these inputs to provide verbal and visual feedback on compression quality (depth and rate) and timing for ventilations. It operates in a 'Normal Operating Mode' (AHA 2010 guidelines) or a 'Silent Mode' for training. The device prompts the user to call for help, check responsiveness, and perform CPR. It features an orientation-sensing function that prompts 'Analysis Halted' if placed incorrectly on the patient's chest. The device provides a metronome for compression rate and LED indicators for compression quality ('Push Harder' vs 'Good Compressions'). By guiding the user to maintain proper compression depth and rate, the device aims to improve CPR quality, potentially increasing the likelihood of patient survival and neurological recovery.
Clinical Evidence
No clinical data on human patients provided. Evidence consists of bench testing (general safety, EMC, software performance, water ingress, vibration, sound levels), simulated use on a computerized manakin for data collection, usability studies, and biocompatibility testing of skin-contact materials.
Technological Characteristics
Accelerometer-based (Analog Devices ADXL322) sensing of chest wall movement; microprocessor-controlled; battery-operated; portable form factor; verbal and visual (LED) feedback; software-driven protocol based on 2010 AHA Guidelines; biocompatible skin-contact materials.
Indications for Use
Indicated for users performing CPR on patients 8 years or older. No specific contraindications listed.
Regulatory Classification
Identification
CPR Aid device with feedback is a device that provides real-time feedback to the rescuer regarding the quality of CPR being delivered to the victim, and provides either audio and/or visual information to encourage the rescuer to continue the consistent application of effective manual CPR in accordance with current accepted CPR guidelines (to include, but not be limited to, parameters such as compression rate, compression depth, ventilation, recoil, instruction for one or multiple rescuers, etc.). These devices may also perform a coaching function to aid rescuers in the sequence of steps necessary to perform effective CPR on a victim.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(i) Nonclinical performance testing under simulated physiological or use conditions must demonstrate the accuracy and reliability of the feedback to the user on specific compression rate, depth and/or respiration over the intended duration, and environment of use.
(ii) Labeling must include the clinical training, if needed, for the safe use of this device and information on the patient population for which the device has been demonstrated to be effective (including patient size and/or age limitations,
*e.g.,* adult, pediatric and/or infant).(iii) For devices that incorporate electrical components, appropriate analysis and testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(iv) For devices containing software, software verification, validation, and hazard analysis must be performed.
(v) Components of the device that come into human contact must be demonstrated to be biocompatible.
(vi) Human factors testing and analysis must validate that the device design and labeling are sufficient for effective use by the intended user.
(3)
*Premarket notification.* The CPR Aid with feedback device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter if it does not contain software (*e.g.,* is mechanical or electro-mechanical) and is in compliance with the special controls under paragraph (b)(2) of this section, subject to the limitations of exemptions in § 870.9.
{1}------------------------------------------------
# N C O R P O R
# 510(k) Summary: K112660
Rev. 3
Submitter's Name and Address: Bio-Detek, Inc. A Subsidiary of ZOLL® Medica! Corporation 525 Narragansett Park Drive Pawtucket, RI 02861 (401) 729-1400
Contact Person: Robert Morse (866) 639-0060 x.224 bmorse@bio-detek.com
Date Summary Prepared: April 4, 2012
Device Name: PocketCPR® (Part #2132)
Classification Name: Cardiopulmonary Resuscitation Aid; Class III (21 CFR 870.5200)
Device Description: Effective cardiopulmonary resuscitation (CPR), when administered quickly after the on-set of cardiac arrest, is effective at saving lives and enabling complete neurological recovery, especially when followed by early defibrillation.
The PocketCPR® was developed to provide users with real-time feedback during the administration of CPR, to assist CPR instructors during class exercises, and to provide CPR students with verbal instructions and feedback for at-home practice. When the PocketCPR is turned on, the default protocol being driven by the 2010 American Heart Association's (AHA) Guidelines, reminds the user to call for help. Once placed on the patient's chest and chest compressions have begun, PocketCPR generates a metronome at the recommended rate of chest compression, gives verbal and visual feedback on the quality of CPR being delivered and reminds the user to give two breaths after every 30 compressions. Measurement of actual chest wall movement is made by a state-of-the-art accelerometer micro-chip and microprocessor performing continuous analysis.
The predicate PocketCPR (2047) in K071321 is compliant to the AHA 2005 Guidelines while the proposed PocketCPR (2132) in this submission is compliant to the AHA 2010 Guidelines in the default "Normal Operating Mode" when tumed "On".
525 Narragansett Park Drive, Pawtucket, Rhode Island 02861-4323, Tel. (401) 729-1400 or (800) 729-1408
MAY 1 0 2012
{2}------------------------------------------------
A plus "4" feature has been added to the Normal Operating Mode to address the caregiver that tums the unit "On", places the device on the patient's chest and commences directly into chest compressions. During the initial power-on protocol the device intuitively monitors for compression recognition of 1.2" or greater. The verbal prompt "Call for Help" will be dropped to allow advancement to the "Start CPR" prompt. This "+" feature will automatically go into chest compression assistance, same as predicate, giving instructions for "Push Harder" as needed and "Good Compressions" when done to correct depth. It also maintains communicating "Open Airway - Give Two Breaths" after each sequence of 30 compressions, same as predicate.
Another plus "+" in the proposed device is an additional guidance source to assist the caregiver in selecting proper orientation of the device when placed on a patient's chest. To best take advantage of the LED's visual cadence and prompting of one LED flashing for "Push Harder" and all four flashing for "Good Compressions" the proper or upright orientation of the device allows the caregiver maximum visual assistance. We have added a "Patient Side" label to the center of the bottom outer surface to direct the caregiver to apply that surface on the patient's chest. The device will also prompt "Analysis Halted" if improper orientation has occurred and will stay silent awaiting proper orientation, hence "Patient Side" label, which when proper orientation occurs the device will then prompt "Start CPR and proceed into CPR assistance sequences.
"Silent Mode" is an operational optional mode both in the predicate PocketCPR and the proposed in this submission for use in training sessions where multiple units are in-play.
# NORMAL OPERATING MODE
Start-Up
- 1. Power On the device by pressing the rubber power button on the side of PocketCPR (marked with a circle and vertical line in the middle).
- 2. Allow device to Power Up and perform its system checks.
- 3. Place device in the center of the chest as shown on label.
- 4. Follow the verbal prompts to:
- · "STAY CALM"
- · "CHECK RESPONSIVENESS"
- · "CALL FOR HELP"
- "OPEN AIRWAY" "+" .
- "CHECK BREATHING" "+" .
- "START CPR" .
- 5. Place the heel of one hand on the device and the second hand on top so both hands are overlapped.
- 6. Begin chest compressions, allow for "Full Release" on "UP" stroke.
- 7. In the Normal Operating Mode "+" feature, the verbal prompts identified with a "+" are over-ridden to advance into chest compression assistance earlier for those caregivers commencing directly into chest compressions.
{3}------------------------------------------------
- 8. Begin chest compressions, allow for "Full Release" on "UP" stroke.
- 9. After 18 seconds, representing 30 compressions, follow the verbal prompts to:
- "OPEN AIRWAY" .
- "GIVE TWO BREATHS" .
- 10. Continue CPR chest compressions following the metronome and light prompts to ensure proper depth and rate. Cycles of 30 compressions and 2 ventilations are recommended or continuous chest compressions if you are unwilling or unable to provide ventilations, or if an advanced airway has been secured.
#### SILENT MODE Start-Up
- 1. Power On the device by pressing and holding in the power button on the side of the device. After hearing the verbal prompt, "STAY CALM.", wait a full second before releasing the power button.
- 2. The device metronome will begin to flash, indicating that the device is ready for CPR to begin, allow for "Full Release" on "UP" stroke.
### Substantial Equivalence:
Part Number (P/N) 2132 for PocketCPR, which represents the version in this submission, is substantially equivalent to its predecessor, PocketCPR P/N 2047, cleared on 510/k) K071321. which incorporated the 2005 AHA Guidelines. The 2132 version of PocketCPR is programmed to meet the 2010 AHA Guidelines as the default mode. The core technology is the same as its predecessor utilizing an Analog Devices accelerometer, presently an ADXL322 versus the earlier submission which had an ADXL311 accelerometer from the same manufacturer. The Indications for Use are the same as the Predicate.
#### Indications for Use
# PocketCPR Intended Use:
To assist users in the performance of effective CPR on a patient 8 years or older.
# PocketCPR Types of Testing
- . Bench Testing
- General Safety o
- Electromagnetic Compatibility o
- Software Performance o
- o Water Ingress
- o Vibration Effects
- Sound Levels of Verbal Prompts & Metronome o
- Simulated Use on Computerized Manakin for Data Collection .
- Usability Studies performing CPR using PocketCPR .
- Biocompatibility of Skin Contact Materials .
{4}------------------------------------------------
### Conclusion
The PocketCPR® was developed to provide users with real-time feedback during the administration of CPR, to assist CPR instructors during class exercises, and to provide CPR students with verbal instructions and feedback for at-home practice. When the PocketCPR is turned on, the device reminds the user to call for help. Once placed on the patient's chest and chest compressions have begun, PocketCPR generates a metronome at the recommended rate of chest compression, gives verbal and visual feedback on the quality of CPR being delivered and reminds the user to give two breaths after every 30 compressions. Measurement of actual chest wall movement is made by a state-of-the-art accelerometer micro-chip and microprocessor performing continuous analysis.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized image of an eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The text is in all capital letters.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 1 0 2012
Bio-Detek Incorporated c/o Mr. Robert Morse Director, Regulatory Affairs 525 Narragansett Park Drive Pawtucket, RI 02861-4323
Re: K112660
PocketCPR Regulation Number: 21 CFR 870.5200 Regulation Name: Cardiopulmonary Resuscitation Aid Regulatory Class: Class III Product Code: LIX Dated: May 1, 2012 Received: May 2, 2012
Dear Mr. Morse:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{6}------------------------------------------------
Page 2 - Mr. Robert Morse
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director L Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known): __K112660
Rev 3
X
Device Name: PocketCPR™
PocketCPR Intended Use:
To assist users in the performance of effective CPR on a patient 8 years or older.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of DRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_KII2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.