DEVON CIRONA 6200 DVT

K141578 · Devon Medical · JOW · Jun 27, 2014 · Cardiovascular

Device Facts

Record IDK141578
Device NameDEVON CIRONA 6200 DVT
ApplicantDevon Medical
Product CodeJOW · Cardiovascular
Decision DateJun 27, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The Cirona™ 6200 Series system is a prescription device intended to be used preventatively to increase venous blood flow in patients at risk of deep vein thrombosis due to the associated risk factors for thrombus formation during: trauma, critical care, general medicine, general surgery, as well as neurological, orthopedic, urologic, obstetric conditions and treatments.

Device Story

Cirona 6200 is a pneumatic compression system for DVT prevention; consists of control device, extension tubing, and soft compression sleeves (calf, calf-thigh, or foot). Device uses electrically generated compressed air to cyclically inflate/deflate garments; mimics natural walking pace to enhance venous flow and fibrinolytic activity. Operates on 60-second cycle: 12-second inflation, 48-second relaxation. Pressure settings: 40mmHg for calf/calf-thigh; 120mmHg for foot. Used in clinical settings; operated by healthcare providers. Output is mechanical compression of limbs; helps prevent early blood clotting. Benefits patient by reducing DVT incidence.

Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included system-level software testing, pressure accuracy, cycle timing, alarm functionality, foot bladder burst testing, and battery depreciation testing.

Technological Characteristics

Electrically powered pneumatic compression device. Components: control unit, tubing, soft compression sleeves. Pressure: 40mmHg (calf/thigh), 120mmHg (foot). Cycle: 60s (12s inflation, 48s relaxation). Standards: IEC 60601-1:2005, EN 60601-1-2:2007, AAMI ES 60601-1:2005, EN ISO 14971:2012. Software-controlled operation.

Indications for Use

Indicated for patients at risk of deep vein thrombosis (DVT) requiring increased venous blood flow during trauma, critical care, general medicine, general surgery, neurological, orthopedic, urologic, or obstetric conditions/treatments. Contraindicated in patients with severe atherosclerosis, acute DVT, severe congestive cardiac failure, pulmonary edema, pulmonary embolisms, extreme limb deformity, local skin/tissue conditions (gangrene, untreated/infected wounds, recent skin graft, dermatitis), malignancy in legs, limb infections (cellulitis), or lymphangiosarcoma.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(K) SUMMARY #### Submitter: Devon Medical Products ## Contact Person: Ruth Wu, CCO 1100 First Avenue, Suite 202 King of Prussia, PA 19406 Phone: 610.757.4103 Fax: 610.930.4035 ### Common Classification & Proprietary Names: Common Names: Proprietary Name: Intermittent Pneumatic Compression Device Cirona™ 6200 ### Date Prepared: Apr 16th, 2014 #### Classification The classification name, 21 CFR Part and Paragraph number, product code and classification of the Cirona™ 6200. | Classification Name | 21 CFR<br>Section | Product<br>Code | Class | |-----------------------------|-------------------|-----------------|-------| | Compressible Limb<br>Sleeve | 870.5800 | JOW | II | ### Predicate Devices: The Cirona 6200 Deep Vein Thrombosis Prevention Therapy System is substantially equivalent to the following. | Predicate Device | Manufacturer | 510(k)# | |-------------------------------------|--------------------------|---------| | Cirona 6100 | Devon Medical Products | K130571 | | RESTEP DVT SYSTEM, MODEL<br>RSP-101 | STORTFORD MEDICAL<br>LLC | K090308 | | Flowtron Universal AC600 | HNE HEALTHCARE, INC | K010744 | #### Device Description The Cirona™ 6200 Series deep vein thrombosis prevention system is a pneumatic compression device that noninvasively helps reduce the incidence of deep vein thrombosis, a potentially life threatening condition. The Cirona™ 6200 Series system consists of a device and a pair of soft compression garment(s) (sleeves) and the extension tubing set for the calf, calf-thigh, and foot. The device will alternatively inflate the two garments and mimic the natural walking pace in order to enhance circulation. The device-supplied compression provides a 60-second automatically timed cycle consisting of an approximately 12-second inflation period followed by a 48-second period of relaxation. A pressure of 40mmHg is used for the calf and calf-thigh treatments, where a pressure of 120mmHg is used for the foot treatment. JUN 2 7 2014 {1}------------------------------------------------ This pressurization enhances venous flow and fibrinolytic activity in order to ultimately prevent early blood clotting. #### Intended Use: The Cirona™ 6200 Series system is a prescription device intended to be used preventatively to increase venous blood flow in patients at risk of deep vein thrombosis due to the associated risk factors for thrombus formation during: trauma, critical care, general medicine, general surgery, as well as neurological, orthopedic, urologic, obstetric conditions and treatments. ## Contraindications: The Cirona™6200 system should NOT be used in the following conditions: - · Severe atherosclerosis or other ischemic vascular diseases - · Suspected or known acute deep vein thrombosis - · Severe congestive cardiac failure - · Existing pulmonary edema - · Existing pulmonary embolisms - · Extreme deformity of the limbs - · Any local skin or tissue condition in which the garments would interfere: - · Gangrene - Untreated or infected wounds - · Recent skin graft - · Dermatitis - · Known presence of malignancy in the legs - · Limb infections, including cellulitis, that have not received antibiotic coverage - Presence of lymphangiosarcoma #### Technological Characteristics: The manufacturer believes that the technological characteristics of the Cirona™ 6200 are substantially equivalent to those of the predicate devices. The Cirona™ 6200 has very similar components to its predicate devices and has very similar principles of operation. The device consists of an electrically generated source of compressed air; tubing to convey the pressurized air to the sleeve and, like the predicates. pressure is applied cyclically for 12 seconds inflated and 48 seconds deflated. A pressure of 40mmHg is used for the calf and calf-thigh treatments, where a pressure of 120mmHg is used for the foot treatment. ### Performance Testing Bench and laboratory testing was performed and assures that the product meets its specifications. The manufacturer believes that the technological characteristics of the Cirona'™ 6200 are substantially equivalent to those of the predicate devices. The performance tests are included in section 18 Bench Test. {2}------------------------------------------------ | | List of Performance Tests | |--------|----------------------------------------| | Test 1 | Cirona 6200 System Level Software Test | | Test 2 | Cirona 6200 Pressure Accuracy | | Test 3 | Cirona 6200 Cycle, Time Test | | Test 4 | Cirona 6200 Alarm Function Test | | Test 5 | Cirona 6200 Foot Bladder Burst | | Test 6 | Cirona 6200 Battery Depreciation Test | ### Standards The Cirona™ 6200 conforms to the following standards: IEC 60601-1:2005 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance EN 60601-1-2:2007 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: AAMI ES 60601-1: 2005 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance EN ISO 14971:2012 Medical devices - Application of risk management to medical devices ## Statement of Substantial Equivalence The Cirona™ 6200 is substantially equivalent in technology, function, operating parameters, and intended use to predicate devices that are currently commercially available and in distribution, and does not raise any new questions of safety or effectiveness. ### Conclusions In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, Devon Medical Products, believes that the Cirona 6200, is substantially equivalent to the predicate devices as described herein. 3 {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO60-Cinn Silver Spring, MD 20993-0002 #### June 27, 2014 Devon Medical, Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313 Re: K141578 Trade/Device Name: Cirona 6200 Deep Vein Thrombosis Prevention System Regulation Number: 21 CFR 870.5800 Regulation Name: Compression Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: June 11, 2014 Received: June 13, 2014 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act, The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Mr. Mark Job forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, M. A. Dallem for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. ## 510(k) Number (if known) K141578 #### Device Name Cirona 6200 Deep Vein Thrombosis Prevention Therapy System #### Indications for Use (Describe) The Cirona 6200 Series system is a prescription device intented to be used preventatively to increase venous blood flow in patients at risk of deep vein thrombosis due to the associated risk factors for thrombus formation during: trauma, critical care, general medicine, general surgery, as well as neurological, urologic, obstetric conditions and treatments. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. THE CONTRACTOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR CONSTITUTION OF CONSECTION Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) M.A. Hillerman This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." 4.1 FORM FDA 3881 (1/14)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...