Needle's Eye Snare Retrieval Set - 54cm

K163353 · Cook Incorporated · DXE · Jul 11, 2017 · Cardiovascular

Device Facts

Record IDK163353
Device NameNeedle's Eye Snare Retrieval Set - 54cm
ApplicantCook Incorporated
Product CodeDXE · Cardiovascular
Decision DateJul 11, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.5150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Needle's Eye Snare® Retrieval Set - 54cm is intended for use in patients requiring the percutaneous retrieval of indwelling catheters, cardiac leads, fragments of catheter tubing or wire guides and other foreign objects.

Device Story

Percutaneous grasping device; used for retrieval of foreign objects (catheters, leads, fragments) from vasculature. Distal end delivered via 12Fr PTFE sheath inside 16Fr Check-Flo® Introducer Sheath. Mechanism: nitinol wire 'needle's eye' loop and 'threader' with stainless steel sleeves. Operation: plunger advancement deploys snare; object captured between threader and loop; inner sheath advanced to close snare and secure object. Used by physicians in clinical settings. Benefits: minimally invasive removal of foreign bodies, avoiding surgical intervention.

Clinical Evidence

Bench testing only. Particulate testing performed to demonstrate comparability to predicate and compliance with USP 778 requirements. Predetermined acceptance criteria were met.

Technological Characteristics

Materials: Nitinol wire, stainless steel sleeves, PTFE sheath. Principle: Mechanical grasping/snaring. Form factor: 54cm usable length, 12Fr inner sheath, 16Fr introducer sheath. Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for patients requiring percutaneous retrieval of indwelling catheters, cardiac leads, catheter tubing fragments, wire guides, or other foreign objects.

Regulatory Classification

Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name encircling a symbol. The symbol is a stylized representation of three human profiles facing right, layered on top of each other, with flowing lines beneath them. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 11, 2017 Cook Incorporated Thomas Kardos Vice President, Regulatory Affairs 750 Daniels Way, P.O. Box 489 Bloomington, Indiana 47402 Re: K163353 Trade/Device Name: Needle's Eye Snare Retrieval Set - 54cm Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: June 13, 2017 Received: June 14, 2017 Dear Thomas Kardos: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerelv. M.A. Willemsen for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K163353 Device Name Needle's Eye Snare® Retrieval Set - 54 cm Indications for Use (Describe) The Needle's Eye Snare® Retrieval Set - 54cm is intended for use in patients requiring the percutaneous retrieval of indwelling catheters, cardiac leads, fragments of catheter tubing or wire guides and other foreign objects. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The logo is set against a red background. The word "COOK" is written in white, block letters in the upper portion of the logo. Below the word "COOK" is the word "MEDICAL" in smaller, white block letters. COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM K163353 Page 1 of 2 ### 510(k) SUMMARY K163353 510(k) Number: Submitted By: Thomas J. Kardos Cook Incorporated 750 Daniels Way P.O. Box 489 Bloomington, IN 47402 Phone: (724) 845-8621 x2225 Fax: (724) 845-2848 Date Prepared: July 7, 2017 ### Device: | Trade Name: | Needle's Eye Snare® Retrieval Set - 54cm | |----------------------|------------------------------------------| | Common Name: | Snare | | Classification Name: | Embolectomy Catheter | | | DXE (21 CFR 870.5150) | | Class/Panel: | Class II, Cardiovascular | ### Indications for Use: The Needle's Eye Snare® Retrieval Set – 54cm is intended for use in patients requiring the percutaneous retrieval of indwelling catheters, cardiac leads, fragments of catheter tubing or wire guides and other foreign objects. ## Predicate Device: The devices, subject of this submission, are substantially equivalent to the predicate devices, the Needle's Eye Snare®, cleared on December 27, 1996 under 510(k) number K961992. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the Cook Medical logo. The logo consists of the word "COOK" in large, bold, white letters on a red background. Below the word "COOK" is the word "MEDICAL" in smaller, white letters, also on a red background. The logo is simple and clean, with a focus on the company name. COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTOΝ, ΙΝ 47402-0489 U.S.A. NE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM K163353 Page 2 of 2 ## Comparison to Predicate Device: It has been demonstrated that the Needle's Eye Snare® Retrieval Set - 54cm is comparable to the predicate device, the Needle's Eye Snare® (K961992) is identical in terms of intended use, principles of operation, and basic technological characteristics to the predicate devices, and also identical in materials of construction to the predicate device. The only differences between the predicate and subject device are the specified length of the proposed device, which is 54cm versus the cleared device length of 94cm, and the inclusion of a 12 Fr Curved Inner Sheath as an accessory. # Device Description: The Needle's Eye Snare® Retrieval Set- 54cm is a grasping device that forms a basketlike snare around the in-dwelling catheter, cardiac lead, fragment of catheter tubing, wire guide or other foreign object. The distal end is delivered to the vicinity of the lead through a long, flexible 12Fr. (O.D.) PTFE sheath placed coaxially within a 16Fr. (O.D.) PTFE Check-Flo® Introducer Sheath having a Check-Flo® Valve with stopcock at its proximal end. The device is comprised at the distal end of a nitinol wire needle's eye retrieval mechanism consisting of a "needle's eye" loop and "threader" with stainless steel sleeves housed within a PTFE protective cover sheath. The snare is activated through the forward advancement of the threader by fully depressing the plunger. The item is captured between the threader and needle's eye. The inner sheath is advanced forward closing the snare and securely capturing the lead or other foreign object. The Needle's Eye Snare® Retrieval Set - 54cm has nominal usable length of 54cm. ## Test Data: The following testing was provided to support a determination of substantial equivalence to the predicate device: - . Particulate Testing - Testing was performed to show that particulates from the device are comparable to the predicate and meet the requirements of USP 778. The predetermined acceptance criteria were met.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...