The Onocor ONO Retrieval device is indicated for use in the cardiovascular system to retrieve objects using minimally invasive surgical procedures include retrieval of intravascular foreign objects such as coils, balloons, catheters, guidewires and/or filters within the cardiovascular system. This device is not intended for use in the coronary arteries or neurovasculature.
Device Story
ONO Retrieval Device is a percutaneous catheter system for retrieving intravascular foreign objects. Device consists of 12 Fr stainless steel reinforced catheter, braided nitinol loop basket, peel-away sheath, inner catheter, and stop-cocks. Operated by clinicians in cardiovascular settings. Peel-away sheath compresses nitinol basket for insertion; once positioned, sheath is removed to deploy basket for object capture. Compatible with up to 7 Fr manipulation devices/snares and 12-26 Fr retrieval sheaths. Working length 86 cm. Provides mechanical retrieval capability; assists in removing misplaced or embolized foreign bodies from cardiovascular system, potentially avoiding more invasive surgical interventions.
Clinical Evidence
No human clinical data. Evidence includes bench testing (dimensional, tensile, corrosion, kink, torque, radiopacity, fatigue, simulated use) and biocompatibility testing (ISO 10993-5, -10, -11, -4). Pre-clinical porcine model study compared subject and predicate devices for retrieval performance, surface tissue interaction, and thrombogenicity, demonstrating equivalent safety and efficacy.
Technological Characteristics
12 Fr stainless steel reinforced catheter; braided nitinol loop basket; peel-away sheath; inner catheter; stop-cocks. Working length 86 cm. Compatible with 7 Fr snares and 12-26 Fr sheaths. Biocompatibility per ISO 10993-4, -5, -10, -11. Sterilization method not specified.
Indications for Use
Indicated for patients requiring retrieval of intravascular foreign objects (coils, balloons, catheters, guidewires, filters) within the cardiovascular system via minimally invasive procedures. Contraindicated for use in coronary arteries or neurovasculature.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
Avantec Vascular Corporation Captus Vascular Retrieval System (K170987)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
May 23, 2022
Onocor Vascular LLC Tiffini Wittwer Regulatory Affairs 808 General Sterling Drive West Chester, Pennsylvania 19382
Re: K212988
Trade/Device Name: ONO Retrieval Device Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: MMX Dated: April 14, 2022 Received: April 19, 2022
Dear Tiffini Wittwer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212988
Device Name ONO Retrieval Device
Indications for Use (Describe)
The Onocor ONO Retrieval device is indicated for use in the cardiovascular system to retrieve objects using minimally invasive surgical procedures include retrieval of intravascular foreign objects such as coils, balloons, catheters, guidewires and/or filters within the cardiovascular system. This device is not intended for use in the coronary arteries or neurovasculature.
| Type of Use (Select one or both, as applicable) | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <div> <span style="font-size: 10px;"></span> <span style="font-size: 10px;"> <table style="border: 0px; margin: 0px; padding: 0px;"><tr style="border: 0px; margin: 0px; padding: 0px;"><td style="border: 0px; margin: 0px; padding: 0px;"><b> <span style="font-size: 10px;"></span> <span style="font-size: 10px;"> <span style="font-family: ZapfDingbats; font-size: 14px;"> </span> </span> </b></td><td style="border: 0px; margin: 0px; padding: 0px;">Prescription Use (Part 21 CFR 801 Subpart D)</td></tr></table> <table style="border: 0px; margin: 0px; padding: 0px;"><tr style="border: 0px; margin: 0px; padding: 0px;"><td style="border: 0px; margin: 0px; padding: 0px;"><b> <span style="font-size: 10px;"></span> <span style="font-size: 10px;"> <span style="font-family: ZapfDingbats; font-size: 14px;"> </span> </span> </b></td><td style="border: 0px; margin: 0px; padding: 0px;">Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> </span> </div> | <b> <span style="font-size: 10px;"></span> <span style="font-size: 10px;"> <span style="font-family: ZapfDingbats; font-size: 14px;"> </span> </span> </b> | Prescription Use (Part 21 CFR 801 Subpart D) | <b> <span style="font-size: 10px;"></span> <span style="font-size: 10px;"> <span style="font-family: ZapfDingbats; font-size: 14px;"> </span> </span> </b> | Over-The-Counter Use (21 CFR 801 Subpart C) |
| <b> <span style="font-size: 10px;"></span> <span style="font-size: 10px;"> <span style="font-family: ZapfDingbats; font-size: 14px;"> </span> </span> </b> | Prescription Use (Part 21 CFR 801 Subpart D) | | | |
| <b> <span style="font-size: 10px;"></span> <span style="font-size: 10px;"> <span style="font-family: ZapfDingbats; font-size: 14px;"> </span> </span> </b> | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
CONTINUE ON SEPARATE SHEET
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
## K212988
| Date Prepared: | May 20, 2022 |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company: | Onocor LLC<br>808 General Sterling Drive<br>West Chester, PA 19382 |
| Contact: | Tiffini Wittwer<br>Regulatory Affairs<br>Phone: 707-799-6732<br>Email: twittwer@mededge.io |
| Device Trade Name: | ONO Retrieval Device |
| Device Common Name: | Percutaneous Retrieval Device |
| Device Classification: | Device, Percutaneous Retrieval<br>Product Code: MMX<br>21 CFR 870.5150<br>Class II<br>Review Panel: Cardiovascular Devices |
| Predicate Device: | K170987 Avantec Vascular Corporation Captus Vascular Retrieval<br>System |
| Description of the Device: | The ONO Retrieval device is constructed of 12 Fr stainless steel<br>reinforced catheter, a series of braided nitinol loops forming a basket, a<br>peel away sheath, inner catheter and stop-cocks. The peel away sheath<br>is advanced to compress the nitinol basket for introduction and then<br>removed. The ONO is compatible with up to a 7 Fr manipulation device<br>or snare and with 12Fr to 26 Fr retrieval sheaths. The ONO has a<br>working length of 86 cm. |
| Indication for Use: | The ONO retrieval device is intended for use in the cardiovascular<br>system to manipulate and retrieve foreign objects, including, but not<br>limited to, guidewires, coils balloons, catheters and filters. |
| Technological<br>Characteristics: | A comparison of the technological characteristics of the subject device<br>and the predicate device shows the ONO Retrieval device to be<br>substantially equivalent to the current marketed predicate device.<br>Equivalence is based upon the product performance, design, and<br>intended use. The ONO Retrieval device and the predicate devices have |
{4}------------------------------------------------
similar materials of construction, dimensional specifications, designs, and sterilization process.
Performance Tests No performance standards have been established under section (Non-Clinical): 514, performance standards, of the Food, Drug and Cosmetic Act for these devices. A series of testing was conducted in accordance with protocols based on requirements outlined in guidance and industry standards and the below were shown to meet the acceptance criteria that were determined to demonstrate substantial equivalence.
> The following tests were performed under the specified testing parameters to support the ONO Retrieval device substantial equivalence.
Performance Testing, including:
- Dimensional Testing ●
- Tensile Strength ●
- Corrosion Resistance ●
- Catheter Kink Testing ●
- Torque Strength ●
- Simulated Use
- Radiopacity ●
- Design Validation Testing and Summative Usability Testing
- Distribution Testing ●
- Fatigue
Biocompatibility Testing, including:
- Cytotoxicity (ISO 10993-5)
- Sensitization (ISO 10993-10)
- Intracutaneous Reactivity (ISO 10993-10) ●
- Acute Systemic Toxicity (ISO 10993-11) ●
- Material Mediated Pyrogenicity (ISO 10993-11)
- Hemolysis (ISO 10993-4)
- Thrombogenicity (ISO 10993-4)
- Complement Activation (ISO 10993-4)
A pre-clinical study was performed in a porcine model comparing the subject and predicate devices. Retrieval of various foreign objects from the vascular system was performed to demonstrate that the device is appropriate with the stated indications.
{5}------------------------------------------------
| | Furthermore, histology of the surface tissue interaction and<br>thrombogenicity were assessed and compared to the predicate. |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial Equivalence: | Based on the Indication for Use, design, safety and performance<br>testing, the ONO Retrieval device met the requirements for its<br>intended use and are substantially equivalent to the predicate<br>device. |
| Conclusion: | The result of all testing demonstrates that the ONO Retrieval<br>device is substantially equivalent to the predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.