The Impulse 6000D/7000DP is used to determine that defibrillators and transcutancous pacemekers are performing within their performance specifications through the measurement of energy output.
Device Story
Portable, battery-operated defibrillator/pacemaker tester; simulates patient ECG signals to trigger automated defibrillators or test for non-shock advisories; measures energy output of defibrillators and pulse rates/amplitudes of transcutaneous pacemakers. Input: defibrillator discharge pulse or pacemaker output (50-1500 Ohm impedance loads). Operation: high-voltage signal attenuation; analog-to-digital conversion; digital signal processor (DSP) calculates measurements and applies mathematical calibration constants for gain/offset errors. Output: measured energy, peak voltage, current, timing, and refractory period. Used by technicians/clinicians in clinical settings to verify device performance. Includes USB interface for PC data download and BNC connectors for oscilloscope waveform viewing.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Portable, battery-operated, rechargeable. Defibrillator input provides 50-Ohm test load. Pacemaker testing supports 50-1500 Ohm impedance loads. Features 10 ECG outputs for patient monitor simulation. Includes USB type B interface and BNC connectors for oscilloscope output. Digital signal processing used for measurement calculation and calibration. Electrically isolated measurement circuitry.
Indications for Use
Indicated for testing performance specifications of defibrillators and transcutaneous pacemakers via energy output measurement. No specific patient population; intended for use by clinical/technical personnel to verify medical device functionality.
Regulatory Classification
Identification
A defibrillator tester is a device that is connected to the output of a defibrillator and is used to measure the energy delivered by the defibrillator into a standard resistive load. Some testers also provide waveform information.
Predicate Devices
Fluke Biomedical, Impulse 4000
Related Devices
K062099 — PHASE 3, MODEL DT-1 · Datrend Systems, Inc. · Jan 25, 2007
K110192 — DEFIBRILLATOR ANALYZER - DA-2006P · Bc Group International, Inc. · Feb 8, 2011
K963190 — QA-40M DEFIBRILLATOR TESTER · Metron U.S., Inc. · Jul 1, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
K072114
JAN 2 8 2008
Image /page/0/Picture/1 description: The image shows a logo with the letters 'sds' in a stylized font. A curved line extends from the left, arching over the letters, adding a dynamic element to the design. The letters are bold and black, creating a strong visual impact against the white background. The logo appears to be for a company or organization with the initials 'SDS'.
SYBRON DENTAL SPECIALTIES Section III - 510(k) Summary of Safety and Effectiveness
Submitter:
Sybron Dental Specialties, Inc. 1717 West Collins Avenue Orange, California 92867 (714) 516-7484 - Phone (714) 516-7488 - Facsimile Colleen Boswell - Contact Person
Date Summary Prepared: January 2008
## Device Name:
- Trade Name -Impulse 6000D/7000DP છ
- . Common Name - Defibrillator Tester
- Classification Name Defibrillator Tester, per 21 CFR § 870.5325 e
Devices for Which Substantial Equivalence is Claimed:
- g Fluke Biomedical, Impulse 4000
## Device Description:
The Impulse 600013/7000DP is a portable, rechargeable, battery-operated defibrillator tester. The Impulse 7000DP also functions as a transcutaneous pacemaker tester. The device's defibrillator input is connected to the output of a defibrillator under test which provides a 50-Ohm test load, approximately the impedance of the human body. The Impulse 6000D/7000DP simulates patient electrocardiogram signals to trigger automated defibrillation when a ventricular fibrillation waveform is presented, and the device tests that the automated defibrillator does not advise shock when a normal sinus rhythm electrocardiogram is presented. The energy output delivered by the defibrillator under test is measured. The Impulse 7000DP also tests transcutaneous pacemaker outputs by presenting a low level electrocardiogram at various pulse rates and measures the response of the pacemaker under testing rate and amplitude. For pacemaker testing, the device incorporates inputs of 50 to 1500 Ohm impedance test loads. The Impulse 6000D is a defibrillator tester only without the pacemaker test option. Both models have 10 electrocardiogram outputs to simulate patient milli-volt level electrocardiogram signals to test combination patient monitors/defibrillators/pacemakers.
The Impulse 7000DP has a USB type "B" interface to a PC to allow data download to a PC. It is electrically isolated from the measurement circuitry and allows remote control of the test from a PC. A BNC type connector on the rear panel also allows an oscilloscope to
6
1717 West Collins Avenne, Brange, CA 92867 880-537 7824 214 316 7400
{1}------------------------------------------------
record the waveform output from the defibrillator under test, attenuated to a lower voltage level, and also electrically isolated from the measurement circuitry. Another BNC connector outputs a higher level signal (greater amplitude) to view the electrocardiogram signal on a second oscilloscope channel.
The accessories for the Impulse 6000D/7000DP include an external power supply to operate and re-charge the internal battery. Optional adapters to connect defibrillators marketed by different manufacturers to the standard 4 mm banana style input jacks are available.
Waveform analysis determines the characteristics of a defibrillator discharge pulse. Peak voltage amplitude, current, timing, overall energy and the refractory period of a pacemaker are recorded. Measurement is done by attenuating the high voltage signal to a lower voltage level, which is then input into an analog to digital converter. A digital signal processor calculates the measurements and corrects hardware error sources with mathematical calibration constants for any offset and gain errors.
### Indication for Use:
The Impulse 6000D/7000DP is used to determine that defibrillators and transcutaneous pacemakers are performing within their performance specifications though the measurement of energy output.
#### Substantial Equivalence:
The Impulse 6000D/7000DP is substantially equivalent to other legally marketed devices in the United States. The Impulse 6000D/7000DP functions in a manner similar to and is intended for the same use as the Impulse 4000 marketed by Fluke Biomedical.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
Public Health Service
JAN 28 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Fluke Biomedical c/o Ms. Colleen Boswell Vice President, Regulatory Affairs Sybron Dental Specialties, Inc. 1717 West Collins Avenue Orange, CA 92867
Re: K072114
Trade/Device Name: IMPULSE 6000D/7000DP Regulation Number: 21 CFR 870.5325 Regulation Name: Defibrillator tester Regulatory Class: Class II (two) Product Code: DRL Dated: January 15, 2008 Received: January 16, 2008
Dear Ms. Boswell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Colleen Boswell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blymmimon for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number: K072114
Device Name: Impulse 6000D/7000DP
## Indications for Use:
The Impulse 6000D/7000DP is used to determine that defibrillators and transcutancous pacemekers are performing within their performance specifications through the measurement of energy output.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blumima
200-Of sion of Cardiovascu 510(k) Number
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.