K182905 · Seaward Group · DRL · Dec 14, 2018 · Cardiovascular
Device Facts
Record ID
K182905
Device Name
UniPulse
Applicant
Seaward Group
Product Code
DRL · Cardiovascular
Decision Date
Dec 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5325
Device Class
Class 2
Indications for Use
The UniPulse 400 Defibrillator/Pacemaker Analyzer is used to determine that a defibrillator and a defibrillator equipped with transcutaneous pacemaker capability are performing within their performance specifications through the measurement of energy output. The UniPulse 400 is intended to be used in the laboratory environment, outside of the patient care vicinity and is not intended to be used on patients or to test devices while connected to patients.
Device Story
Portable, battery/AC-powered defibrillator/pacemaker analyzer; measures energy output delivered into standard resistive test loads. Inputs: defibrillator discharge via paddles or accessory cables; ECG signals from integrated simulator (normal sinus rhythm/arrhythmia). Outputs: energy measurements, waveform data on integrated display. Used by biomedical/clinical engineers in hospital labs to verify device performance specifications. Communicates via Bluetooth/USB to external printers or PCs for data management. Benefits: ensures defibrillator/pacemaker reliability through objective performance verification outside patient care areas.
Clinical Evidence
Bench testing only. No clinical data. Performance verified against predetermined specifications, including electrical safety (BS EN 61010-1, BS EN 61010-2-030) and electromagnetic compatibility (BS EN 61326-1, CISPR 11).
Technological Characteristics
Portable, rechargeable NiMH battery or AC-powered analyzer. Features integrated display, ECG simulator, and variable load (50-1600 ohms). Connectivity via Bluetooth and USB. Complies with BS EN 61010-1, BS EN 61010-2-030, and BS EN 61326-1 standards. Not sterile; no shelf-life.
Indications for Use
Indicated for use by biomedical or clinical engineers to verify performance of defibrillators and transcutaneous pacemakers in laboratory settings. Not for use on patients or while devices are connected to patients.
Regulatory Classification
Identification
A defibrillator tester is a device that is connected to the output of a defibrillator and is used to measure the energy delivered by the defibrillator into a standard resistive load. Some testers also provide waveform information.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 14, 2018
Seaward Group % Robert Steurer Principal Consultant Steurer Consulting Group 800 Blue Ouail Rd Keller, Texas 76248
Re: K182905
Trade/Device Name: UniPulse 400 Regulation Number: 21 CFR 870.5325 Regulation Name: Defibrillator Tester Regulatory Class: Class II Product Code: DRL Dated: October 12, 2018 Received: October 16, 2018
## Dear Robert Steurer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jessica E. Paulsen -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K 182905
Device Name
UniPulse 400
Indications for Use (Describe)
The UniPulse 400 Defibrillator/Pacemaker Analyzer is used to determine that a defibrillator equipped with transcutaneous pacemaker capability are performing within their performance specifications through the measurement of energy output. The UniPulse 400 is intended to be used in the laboratory environment, outside of the patient care vicinity and is not intended to be used on patients or to test devices while connected to patients.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submission Date: | October 12, 2018 |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Rigel Medical, a Seaward Group company<br>15-18 Bracken Hill,<br>South West Industrial Estate,<br>Peterlee, Co. Durham, SR8 2SW,<br>United Kingdom |
| Submitter Contact: | Rigel Medical, a Seaward Group company<br>15-18 Bracken Hill,<br>South West Industrial Estate,<br>Peterlee, Co. Durham, SR8 2SW,<br>United Kingdom<br>Name: Jim Wallace<br>Rigel Medical, a Seaward Group company<br>Phone: +44 (0)191 5878724<br>Fax: +44 (0) 191 586 0227<br>Email: JimW@seaward.co.uk |
| Application<br>Correspondent: | Bob Steurer<br>Principal Consultant<br>Steurer Consulting Group<br>800 Blue Quail Rd.<br>Keller, TX 76248<br>steurerbob@gmail.com<br>+1 (425) 358-1072 |
| Manufacturing Site: | Rigel Medical, a Seaward Group company<br>15-18 Bracken Hill,<br>South West Industrial Estate,<br>Peterlee, Co. Durham, SR8 2SW,<br>United Kingdom |
| Trade Name: | UniPulse 400 |
| Common Name: | Defibrillator Analyzer |
| Classification Name: | Defibrillator Tester |
{4}------------------------------------------------
Primary Classification Regulation: 21 CFR §870.5325
Primary Product Code:
DRL
## Substantially Equivalent Devices:
| New Device | Predicate 510(k)<br>Number | Predicate<br>Manufacturer/Model |
|---------------------------------------------------------|----------------------------|--------------------------------------------------------------|
| Rigel Medical<br>UniPulse 400 Defibrillator<br>Analyzer | K072114 | Fluke Biomedical<br>Impulse 7000dp<br>Defibrillator Analyzer |
Device Description: The Rigel Medical UniPulse 400 is a portable, rechargeable battery operated (with AC Adapter) defibrillator tester used to determine that a defibrillator and a defibrillator with a transcutaneous pacemaker function perform to their published performance specifications. The UniPulse 400 is connected to a defibrillator output and is used to measure the output energy delivered from the device into a standard resistive test load. It also provides waveform information on its integrated display. For pacemaker testing, a variable load from 50 to 1600 ohms is available in 50-ohm steps. Defibrillator paddles are connected directly to the UniPulse 400 for hands free operation, or to contact plates if using the paddles. When the defibrillator is discharged, the UniPulse 400 will display the energy delivered. Included in the UniPulse 400 is an ECG simulator with normal sinus rhythm and arrhythmia simulation to trigger automated defibrillation.
> UniPulse 400 is used by biomedical and/or clinical engineers to verify performance of defibrillators used in their hospital. The UniPulse 400 provides statistical data measured from the device and can communicate to an external Bluetooth printer and a PC / Laptop via Bluetooth and USB.
{5}------------------------------------------------
Intended Use: The UniPulse 400 Defibrillator/Pacemaker Analyzer is used to determine that a defibrillator and a defibrillator equipped with transcutaneous pacemaker capability are performing within their performance specifications through the measurement of energy output. The UniPulse 400 is intended to be used in the laboratory environment, outside of the patient care vicinity and is not intended to be used on patients or to test devices while connected to patients.
## Technology Comparison:
The Rigel UniPulse 400 employs the same technological characteristics as the predicate device.
| Characteristic | Predicate Device –<br>Impulse 7000dp | UniPulse 400 |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Load resistance | 50 ohm | Same |
| Battery Operation | Internal<br>rechargeable NiMH<br>batteries with AC<br>adapter | Same |
| Defibrillator<br>waveform<br>compatibility | Compatible with<br>mono-phasic, bi-<br>phasic, and pulsed<br>waveforms | Same |
| Defibrillator Test<br>Connections | Through integrated<br>paddle connection<br>plates or you can<br>connect directly<br>through accessory<br>cable. | Same – Connect<br>directly through<br>accessory cable or<br>paddle connections<br>available through<br>accessory paddle box |
| ECG Simulator with<br>Arrhythmia output<br>capability | Yes | Same |
| Data download | USB connection to<br>PC | Same |
| Waveform Display | Through connected<br>PC | Same, can also show<br>on built-in display |
| Pacemaker Analysis | Demand and<br>asynchronous test | Same |
| Pacemaker variable | 50 to 1500 ohms in | Different - 50 to 1600 |
| load | 50-ohm steps | ohms in 50-ohm steps |
{6}------------------------------------------------
## Summary of Performance Testing:
| Sterilization and<br>Shelf-life | The Rigel Medical UniPulse 400 is not provided sterile. The Rigel Medical<br>UniPulse 400 does not have a shelf-life.<br>Therefore, this section is not applicable. |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | The Rigel Medical UniPulse 400 does not directly nor indirectly contact the<br>patient. Therefore, this section is not applicable. |
| Software Testing | The Rigel Medical UniPulse 400 was designed and developed according to Rigel<br>Medical's internal software development process. The Rigel Medical UniPulse<br>400 was tested using verification and validation methods, the results of which<br>indicate the Rigel Medical UniPulse 400 complies with its specifications. |
| Electrical Safety | The Rigel Medical UniPulse 400 was tested for performance in accordance with<br>the following standard:<br>BS EN 61010-1-2010 Safety requirements for electrical equipment for<br>measurement, control, and laboratory use - Part 1: General<br>requirements, BS EN 61010-2-030:2010 Safety requirements for electrical equipment<br>for measurement, control, and laboratory use - Part 2-30 Particular<br>requirements for testing and measuring circuits Test results indicate the Rigel Medical UniPulse 400 complies with its<br>specifications. |
| Electromagnetic<br>Compatibility<br>Testing | The Rigel Medical UniPulse 400 was tested for performance in accordance with<br>the following standard:<br>BS EN 61326-1:2013 Electrical equipment for measurement, control and<br>laboratory use - EMC requirements - Part 1: General requirements BS EN 61000-3-2:2006+A2:2009 Harmonics BS EN 61000-3-2:2013 Flicker BS EN 61000-4-2:2009 Electrostatic Discharge BS EN 61000-4-2:2006+A2:2010 Radiated Immunity BS EN 61000-4-2:2012 Fast Burst Transients BS EN 61000-4-5:2006 Surge Transients BS EN 61000-4-6:2009 Conducted Immunity BS EN 61000-4-11:2004 Voltage Dips and Short Interruptions CISPR 11/BS EN 55011 Radiated Emissions CISPR 11/BS EN 55011 Conducted Emissions Test results indicate that the Rigel Medical UniPulse 400 complies with the<br>applicable standards. |
| Performance Testing<br>– Bench | The Rigel Medical UniPulse 400 was tested for performance in accordance with<br>its predetermined specifications. |
| | The results of the performance testing show the Rigel Medical UniPulse 400<br>complies with these specifications. |
| Conclusion | A comparison of the predicate device and a review of the testing results shows<br>that the Rigel Medical UniPulse 400 is substantially equivalent to the predicate<br>device. |
{7}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.