Reprocessed Tri Pulse Compression Garment

K203847 · Renu Medical, Inc. · JOW · May 7, 2021 · Cardiovascular

Device Facts

Record IDK203847
Device NameReprocessed Tri Pulse Compression Garment
ApplicantRenu Medical, Inc.
Product CodeJOW · Cardiovascular
Decision DateMay 7, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The RENU Medical Reprocessed Tri Pulse Compression Garment are to be used as a non-invasive therapeutic method to help prevent Deep Vein Thrombosis (DVT) and resulting pulmonary embolism.

Device Story

Reprocessed Tri Pulse Compression Garment; non-invasive limb sleeve; used with Arjo Flowtron ACS800/ACS900 pumps. Device consists of three-chamber bladder inside polyester garment; wraps around patient limb; secured with hook and eye tabs. Connected to pump via single tube; pump inflates chambers sequentially to apply limb pressure. Used in clinical settings to prevent DVT and pulmonary embolism. RENU Medical reprocesses used garments via thermal disinfection (hot water pasteurization) to achieve high-level disinfection. Reprocessed garments maintain durability and functional performance (pressure/inflation timing) equivalent to OEM specifications.

Clinical Evidence

Bench testing only. Cycle Verification Testing performed on samples reprocessed 10 times. Functional testing conducted after each cycle, including pressure and inflation time measurements using OEM pumps, to ensure compliance with performance requirements.

Technological Characteristics

Three-chamber bladder enclosed in polyester garment; hook and eye closure. Sequential inflation via single connecting tube. Powered by external Arjo Flowtron ACS800/ACS900 pumps. Reprocessing via thermal disinfection (hot water pasteurization).

Indications for Use

Indicated for patients requiring non-invasive DVT prophylaxis to prevent deep vein thrombosis and pulmonary embolism.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 7, 2021 RENU Medical, Inc Darren DeMerritt Director RA/OA 830 80th ST SW Suite 100 Everett, Washington 98203 Re: K203847 Trade/Device Name: Reprocessed Tri Pulse Compression Garment Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible limb sleeve Regulatory Class: Class II Product Code: JOW Dated: April 8, 2021 Received: April 9, 2021 Dear Darren DeMerritt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Nicole Gillette Assistant Director (Acting) DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ The item numbers included in the scope of this 510(k) submission are as follows: | Item No | Sleeve Description | |-----------|--------------------------------------------| | RM-TRP10 | Flowtron Regular Sequential Calf Garment | | RM-TRP20 | Flowtron Large Sequential Calf Garment | | RM-TRP30 | Flowtron Regular Sequential Thigh Garment | | RM-TRP40 | Flowtron Large Sequential Thigh Garment | | RM-TRP60L | Flowtron Bariatric Sequential Calf Garment | {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K203847 #### Device Name Reprocessed Tri Pulse Compression Garments #### Indications for Use (Describe) The RENU Medical Reprocessed Tri Pulse Compression Garment are to be used as a non-invasive therapeutic method to help prevent Deep Vein Thrombosis (DVT) and resulting pulmonary embolism. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| [X] Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "RENU" in large, bold, teal letters. The letters are evenly spaced and fill most of the frame. The background is plain white, which makes the word stand out. The font is sans-serif and appears modern. # 510(k) Summary ## Reprocessed Tri Pulse Compression Garments | Name & Address: | RENU Medical, Inc.<br>830 80th ST SW Suite 100<br>Everett, WA 98203 | |-----------------|---------------------------------------------------------------------| | Telephone: | 425-353-1110 | | Fax: | 425-353-9116 | | Prepared: | December 29, 2020 | | Contact: | Darren DeMerritt, Director RA/QA | | Device Name: | Reprocessed Tri Pulse Compression Garment | | Common Name: | Compressible Limb Sleeve | Classification | | Class | Product | Classification | | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | | | Code | Regulation | | | | ll | JOW | 21 CFR 870.5800 | | | Classification Name: | Sleeve, Limb, Compressible | | | | | Predicate Devices: | 12, 2003 (K031559). | | RENU Medical Reprocessed Compressible Limb Sleeves, cleared November | | | | Arjo Huntleigh (K133119). | The following is a reference device: | Tri Pulse Garments cleared February 28, 2014, originally manufactured by | | | Indications for Use: | | | The RENU Medical Reprocessed Tri Pulse Compression Garment are to be<br>used as a non-invasive therapeutic method to help prevent Deep Vein<br>Thrombosis (DVT) and resulting pulmonary embolism. | | | Description: | The Reprocessed Tri Pulse Compression Garment is a compression garment<br>that is attached to a patient's limb. It is designed to work with the Arjo<br>Flowtron ACS800 Pump and Arjo Flowtron ACS900 Pump only. Each<br>garment is compressed alternately, applying pressure to the patient's limb,<br>to help prevent deep vein thrombosis. | | | | {5}------------------------------------------------ Models: | Model ID: | Device | Features | |--------------------------------------------|----------------------------------|-------------------------------------------------------| | TRP10, TRP20,<br>TRP30, TRP40<br>and TRP60 | Tri Pulse Compression<br>Garment | Sequential inflation providing active<br>compression. | | Substantial Equivalence: | The equivalence of the Tri Pulse garments is demonstrated using<br>bench testing to determine that the reprocessing methods do not<br>adversely affect the device from performing to the specifications<br>outlined by the OEM predicate device (K133119). | |--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The bench tested done on the devices was Cycle Verification Testing. This<br>test examines the effect of the ReNu reprocessing system on the<br>durability and functionality of the devices. The purpose of this test is to<br>ensure that the Tri-Pulse Compression Garments can withstand the<br>thermal temperatures that they're subject to during a reprocessing hot<br>water pasteurization cycle. | | | During this test, a sample of devices were reprocessed 10 times each and<br>subject to functional testing after each individual cycle in order to ensure<br>that each device meets its functional requirements. | | | Each device was subject to a pasteurization cycle, then placed around<br>PVC to simulate a patient use. Once situated, it was pumped up by an<br>OEM pump and pressure readings were taken to ensure that the<br>compression garment met its pressure and inflation time requirements<br>and repeated until the Reprocessed Compression Garment met its cycle<br>limit. This verification report, document ID #12915, titled, Cycle<br>Verification Report, TRP DVT, can be found in section 18.4, Supporting<br>Documents. | | | Additionally, the Reprocessed Tri Pulse Garments are substantially<br>equivalent to other reprocessed compression garments and<br>therefore are able to achieve the same cycle limits of other<br>previously cleared reprocessed compression garments (K031559). | | Technologies Summary: | Tri Pulse garments are constructed with a three-chamber bladder<br>enclosed in a polyester garment, which is wrapped around the limb<br>and secured with hook and eye tabs. When connected to the pump, | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "RENU" in large, bold, teal-colored letters. The letters are evenly spaced and appear to be a sans-serif font. The background is plain white, which makes the text stand out. the garment inflates through a single connecting tube to generate a sequential compression effect on the limb. Reprocessing the Tri-Pulse Compression Garment is conducted by achieving high-level disinfection via thermal disinfection methods. This process is conducted by holding the devices at or above a predetermined specified temperature for a predetermined specified duration. Conclusion: The data detailed within submission demonstrates that the device is as safe and effective as the predicate and performs as well as the legally marketed predicate devices, identified in this summary.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...