HOTWIRE RF GUIDEWIRE (901XXX); HOTWIRE RF GUIDEWIRE (902XXX)

K252419 · Atraverse Medical, Inc. · DXF · Aug 27, 2025 · Cardiovascular

Device Facts

Record IDK252419
Device NameHOTWIRE RF GUIDEWIRE (901XXX); HOTWIRE RF GUIDEWIRE (902XXX)
ApplicantAtraverse Medical, Inc.
Product CodeDXF · Cardiovascular
Decision DateAug 27, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5175
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The HOTWIRE is indicated for creation of an atrial septal defect in the heart.

Device Story

The HOTWIRE RF Guidewire is a sterile, single-use device designed to create an atrial septal defect. It consists of a stainless steel core wire jacketed in insulating polymer, featuring a distal electrode segment (uninsulated stainless-steel coil) and a tungsten marker for radiopacity. The device is operated by a physician in a clinical setting, typically in conjunction with third-party steerable introducer sheaths and an RF electrosurgical generator. The operator advances the guidewire through the vasculature to the heart septum; RF energy is delivered in monopolar mode to the distal electrode to perform the septal puncture. The device provides support for subsequent wire-guided tools. Visual markers on the proximal end assist in aligning the electrode with sheaths. The system utilizes a patient return electrode to complete the circuit. Clinical benefit includes facilitating transseptal access for cardiac procedures.

Clinical Evidence

No human clinical trials were conducted. Safety and effectiveness were demonstrated through non-clinical bench testing and an in vivo porcine model. Bench testing included visual/dimensional inspection, simulated use, arc integrity, tensile/torque strength, fracture resistance, flexing, and tip flexibility. Handpiece testing included cable flex, tensile strength, EEPROM functionality, and activation button testing. In vivo testing validated design performance, including compatibility with introducer sheaths, intracardiac echocardiography (ICE) catheters, and fluoroscopic visualization.

Technological Characteristics

Stainless steel core wire with insulating polymer jacket; distal uninsulated stainless-steel coil electrode; tungsten marker coil for radiopacity. Monopolar RF energy delivery. Dimensions and design are analogous to predicate. Sterilized via gamma radiation (VDmax25) to SAL 10^-6. Compatible with third-party RF electrosurgical generators and patient return electrodes compliant with IEC 60601-2-2.

Indications for Use

Indicated for creation of an atrial septal defect in the heart.

Regulatory Classification

Identification

A septostomy catheter is a special balloon catheter that is used to create or enlarge the atrial septal defect found in the heart of certain infants.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 27, 2025 Atraverse Medical, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114 Re: K252419 Trade/Device Name: Hotwire RF Guidewire Model numbers: 901000, 901001, 901002, 901003, 901004, 901005, 901006, 901007, 901008, 901009, 901010, 901011, 901012, 901013, 901014, 901015, 901016, 901017, 901018, 901019, 901020, 901021, 901022, 901023, 901024, 902001, 902002, 902003, 902004, 902005, 902006, 902007, 902008, 902009, 902010, 902011, 902012, 902013, 902014, 902015, 902016, 902017, 902018, 902019, 902020, 902021, 902022, 902023, 902024, 902025, 902026, 902027, 902028, 902029, 902030, 902031, 902032, 902033, 902034, 902035, 902036 Regulation Number: 21 CFR 870.5175 Regulation Name: Septostomy catheter Regulatory Class: Class II Product Code: DXF Dated: August 1, 2025 Received: August 1, 2025 Dear Prithul Bom: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252419 - Prithul Bom Page 2 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K252419 - Prithul Bom Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Katherine N. Trivedi Digitally signed by Katherine N. Trivedi-S Date: 2025.08.27 16:59:05 -06'00" Katherine Trivedi Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K252419 | | | Device Name HOTWIRE RF GUIDEWIRE (models 901XXX and 902XXX) | | | Indications for Use (Describe) The HOTWIRE is indicated for creation of an atrial septal defect in the heart. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Page 1 of 6 # 510(K) SUMMARY This 510(k) summary for HOTWIRE™ RADIOFREQUENCY (RF) GUIDEWIRE is submitted in accordance with the requirements of 21 CFR 807.87(h) and 807.92 and following the recommendation outlined in FDA Guidance document “The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notification [510(k)]” (issued on July 28, 2014). Atraverse Medical, Inc. is submitting this traditional 510(k) Premarket Notification for the HOTWIRE™ RF Guidewire. The HOTWIRE™ RF Guidewire (“Predicate device”) has previously been cleared under the Traditional 510(k) Program (K240900), in which Atraverse Medical, Inc. is the legal manufacturer. # SUBMITTER [807.92(a)(1)] Atraverse Medical, Inc. 2611 S Coast Hwy 101 #204 Cardiff by the Sea, CA 92007 Contact Person: Charles Yang SVP QA/RA Telephone: +1 (760) 278-8098 Email: Charles.yang@atraversemedical.com Date prepared: June 30, 2025 # DEVICE [807.92(a)(2)] Table 1: Device Information | Trade Name | HOTWIRE™ RF Guidewire | | --- | --- | | Common Name | Catheter, Septostomy | | Classification Name | Catheter, Septostomy | | Regulation | 21 CFR 870.5175 | | Product Code | DXF | | Regulatory Classification | Class II | | Device Panel: | Cardiovascular | # PREDICATE DEVICE [807.92(a)(3)] HOTWIRE™ RF Guidewire (K240900) # REFERENCE DEVICE {5} Baylis Medical ProTrack RF Anchor Wire (K150709) # DEVICE DESCRIPTION [807.92(a)(4)] # HOTWIRE™ Kit (901XXX): The HOTWIRE™ is a sterile, single-use guidewire device that delivers radiofrequency (RF) power in a monopolar mode to a distal electrode segment for the creation of an atrial septal defect in the heart. The HOTWIRE™ is intended to be used in conjunction with compatible third-party intravascular sheaths and/or dilators, such as the Agilis NXT Steerable Introducer, St. Jude Medical/Abbott (K081645), and third-party RF electrosurgical generator(s), such as the Valleylab FT10 Electrosurgical Platform (K151649) which utilizes a commercially-available Patient Return Electrode (PRE) which is in compliance with IEC 60601-2-2, such as the Valleylab E7507DB (K822572). Table 2: Devices that have undergone Compatibility Testing | Manufacturer | Device | 510(k) | | --- | --- | --- | | St Jude Medical / Abbott | Agilis NXT Steerable Introducer | K081645 | | Medtronic | FlexCath Steerable Sheath | K183174 | | Biosense Webster | CARTO VIZIGO Sheath | K170997 | | Valleylab | FT10 Electrosurgical Platform | K151649 | | Valleylab | E7507 Return Electrode | K822572 | The HOTWIRE™ is comprised of a stainless steel core wire. The main body of the wire is jacketed with an insulating polymer that provides electrical insulation and facilitates smooth movement of the device through vascular dilators and/or sheaths. The floppy distal segment of the wire has an atraumatic tip with an uninsulated stainless-steel coil, which serves as an electrode, and also provides fluoroscopic and echogenic visualization. A tungsten marker coil at the tip provides additional radiopacity. The stiff body of the HOTWIRE™ provides support for advancing wire-guided devices into the left atrium after the distal segment has traversed the septum. The proximal insulated portion of the wire has visual markers that align the electrode tip with third-party transseptal sheaths and/or dilators. A portion of the proximal wire is uninsulated for placement of an included Adapter Pin that connects to hand pieces used with compatible third-party RF electrosurgical generators. # HOTWIRE™ + Handpiece Kit (902XXX): The HOTWIRE™ + Handpiece Kit is a sterile, single-use guidewire device that delivers radiofrequency (RF) power in a monopolar mode to a distal electrode segment for the creation of an atrial septal defect in the heart. The HOTWIRE™ + Handpiece Kit is intended to be used in conjunction with compatible third-party intravascular sheaths and/or dilators, such as the Agilis NXT Steerable Introducer, St. Jude Medical/Abbott (K081645), and the HOTWIRE™ System {6} RF Generator, which utilizes a commercially-available Patient Return Electrode (PRE) which is in compliance with IEC 60601-2-2, such as the Valleylab E7507DB (K822572). Table 3: Devices that have undergone Compatibility Testing | Manufacturer | Device | 510(k) | | --- | --- | --- | | St Jude Medical / Abbott | Agilis NXT Steerable Introducer | K081645 | | Medtronic | FlexCath Steerable Sheath | K183174 | | Biosense Webster | CARTO VIZIGO Sheath | K170997 | | Atraverse Medical | HOTWIRE™ System RF Generator | | | Valleylab | E7507 Return Electrode | K822572 | | McKesson Argent | 22-ESRSC Return Electrode | K092761 | The HOTWIRE™ is comprised of a stainless steel core wire. The main body of the wire is jacketed with an insulating polymer that provides electrical insulation and facilitates smooth movement of the device through vascular dilators and/or sheaths. The floppy distal segment of the wire has an atraumatic tip with an uninsulated stainless-steel coil, which serves as an electrode, and also provides fluoroscopic and echogenic visualization. A tungsten marker coil at the tip provides additional radiopacity. The stiff body of the HOTWIRE™ provides support for advancing wire-guided devices into the left atrium after the distal segment has traversed the septum. The proximal insulated portion of the wire has visual markers that align the electrode tip with third-party transseptal sheaths and/or dilators. A portion of the proximal wire is uninsulated for placement of an included handpiece that connects with the HOTWIRE™ System RF Generator. # INDICATIONS FOR USE [807.92(a)(5)] The HOTWIRE™ is indicated for creation of an atrial septal defect in the heart. # COMPARISON OF TECHNOLOGICAL CHARACTERISTICS [807.92(a)(6)] The technological characteristics of the HOTWIRE™ is highly analogous to the technological characteristics of the Predicate device (K240900). Substantial equivalence is determined based on the following similarities: Same intended use/indications for use - Same principles of operation - Same fundamental scientific technology - Same basic guidewire design - Same guidewire construction material {7} Atraverse Medical, Inc. shall demonstrate that the device modifications made to the Predicate device remains substantially equivalent to the Predicate device. Specifics of this substantial equivalence determination is included in the discussion below. The HOTWIRE™ + Handpiece Kit has identical or similar intended use, technological characteristics, and principles of operation as the Predicate device. A device comparison table is provided in this section in Table 4. This table provides a concise comparison between the relevant features of the HOTWIRE™ + Handpiece Kit and the Predicate device. Table 4: HOTWIRE™ Comparison Table | Characteristic | Subject Device Compared to Predicate Device (K240900) | Subject Device Compared to Reference Device (K150709) | Comment | | --- | --- | --- | --- | | Indications for Use | Identical | Identical | The subject device, predicate device and reference device are indicated for creation of an atrial septal defect in the heart. | | Fundamental scientific technology | Identical | Identical | The subject device, predicate device and reference device rely on use of controlled RF energy by the user for transseptal puncture. | | Principal of Operation | Similar | Similar | The subject device, predicate device and reference device are operator controlled. RF energy is delivered via a compatible RF Generator / electrosurgical handpiece to the RF wire distal tip by the operator. The minor design differences do not raise new questions of safety or effectiveness. | | Method of Supply | Identical | Similar | Both the subject and predicate device are single use; Sterilized by Gamma radiation – Method VDmax25; Sterility Assurance Level of 10-6The Reference device similar achieves a Sterility Assurance Level of 10-6. The Reference device utilizes Ethylene Oxide to achieve Sterility Assurance Level. | | Technological characteristics (Dimensions, design, materials) | Identical | Similar | Both the predicate and subject device share the same fundamental design, in terms of: • Materials, • Insulation • Dimensions (i.e., diameter, length) • Configuration (i.e., tip shape) The Subject Device differs from the predicate as follows: • Distal taper length The Reference device includes a distal taper length similar to that of the Subject device. The minor design differences between the Subject and Predicate device do not raise new questions of safety or effectiveness. | {8} | Characteristic | Subject Device Compared to Predicate Device (K240900) | Subject Device Compared to Reference Device (K150709) | Comment | | --- | --- | --- | --- | | Energy Source | Similar | Similar | The subject device, predicate device and reference device can be used with 510(k) cleared compatible RF electrosurgical generator. Risks associated with change in energy source from predicate were adequately mitigated and verified. The addition of a compatible RF electrosurgical generator source does not raise new questions of safety or effectiveness. | | Packaging configuration | Similar | Similar | The Subject device packaging configuration differs in dimensions and accessories included as compared to the predicate device. The HOTWIRE™ packaging was validated after environmental conditioning per ASTM D4332, transportation simulation per ASTM 4169 and 1 year accelerated aged equivalent. The Subject device packaging configuration has been determined to be equivalent to that of the predicate device. The minor difference in packaging does not raise new questions of safety or effectiveness | ## PERFORMANCE DATA [807.92(b)] Non-clinical bench top and in vivo performance testing was completed to demonstrate safety and effectiveness and substantial equivalence of the subject device to the predicate device. All test requirements were met as specified by applicable standards and test protocols. The following verification and validation activities were completed to demonstrate the safety and effectiveness of the subject device: - RF Guidewire Testing - Visual and dimensional inspection - Simulated use test - Arc integrity test - Tensile strength test - Torque strength test - Torquability test - Fracture resistance test - Flexing test - Tip flexibility test - Handpiece Testing: - Cable flex test - Cable tensile strength test - EEPROM functionality test Page 5 of 6 {9} Page 6 of 6 - Cable length dimensional inspection - Retention test - Activation button test - In vivo testing **In vivo Testing:** Customer requirements were validated by *in vivo* testing utilizing a porcine model. Design validation testing evaluated subject device design performance during normal intended use, including compatibility with commercially-available introducer sheaths, intracardiac echocardiography (ICE) catheters, and fluoroscopic visualization. **Shelf life:** The accelerated shelf life testing for HOTWIRE™ has been conducted (T=1 years accelerated aging) with test results confirmed that all acceptance criteria were met. No new questions of safety or effectiveness are raised. Based on the results, we can conclude that HOTWIRE™ will perform as intended to the Design Specification. HOTWIRE™ will be labeled for 1-year shelf life. **Packaging:** Packaging validation was performed on the HOTWIRE™. The results from packaging testing conducted on HOTWIRE™ showed that the acceptance criteria were met. Therefore, we can conclude the HOTWIRE™ packaging will provide the adequate and effective protection and sterile barrier requirements. **Sterilization:** HOTWIRE™ is sterilized using gamma radiation. HOTWIRE™ is sold sterile, for single use and single patient only. The sterilization validation was performed and is documented. The sterilization validation results showed that the sterilization dose and routine sterilization process was validated to achieve an SAL of 10⁻⁶ for the HOTWIRE™. # CONCLUSION The HOTWIRE™ met all specified criteria and did not raise new safety or performance questions. Based on the 510(k) summary and information provided herein, we conclude that the subject device, HOTWIRE™, is substantially equivalent in its intended use, design, material, performance, and the underlying fundamental scientific technology used, to the Predicate device (K240900)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...