K242076 · Baylis Medical Company · DXF · Nov 1, 2024 · Cardiovascular
Device Facts
Record ID
K242076
Device Name
VersaCross™ RF Wire
Applicant
Baylis Medical Company
Product Code
DXF · Cardiovascular
Decision Date
Nov 1, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5175
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VersaCross™ RF Wire is indicated for creation of an atrial septal defect in the heart.
Device Story
VersaCross™ RF Wire is a single-use, sterile device used to create atrial septal defects. It consists of a stainless steel core wire with an insulating coating, a floppy distal end with a rounded active electrode, and proximal markers. The device is used in clinical settings by physicians during interventional cardiac procedures. It is connected to a compatible RF generator (e.g., Baylis RF Puncture Generator) and an external dispersive patch. The operator uses fluoroscopy, echocardiography, or electroanatomic mapping to position the wire at the target puncture site. Upon activation, the device delivers monopolar RF energy to the distal tip, causing cell vaporization and dielectric breakdown of cardiac tissue to create the defect. The wire then serves as a rail for advancing ancillary therapy catheters into the left atrium. The device benefits patients by facilitating transseptal access for complex cardiac interventions.
Clinical Evidence
Bench testing only. Performance verified through mechanical testing (ISO 11070), electrical safety (IEC 60601-2-2), thermal arc integrity, biocompatibility (ISO 10993-1), sterilization (ISO 11135), pyrogen testing (ANSI/AAMI ST72), and packaging integrity (ISO 11607-1). Benchtop validation confirmed compatibility with commercial impedance-based electroanatomical mapping systems and performance in representative anatomical models.
Technological Characteristics
Stainless steel core wire with PTFE insulation. Monopolar RF energy delivery. Distal active electrode, marker coil, and radio-opaque band. Dimensions include variable usable length, shaft diameter, and taper. Connectivity: wired to external RF generator. Sterilization: Ethylene Oxide. Standards: ISO 11070, IEC 60601-2-2, ISO 10993, ISO 11135, ISO 11607.
Indications for Use
Indicated for creation of an atrial septal defect in the heart. No specific patient population age or gender restrictions provided.
Regulatory Classification
Identification
A septostomy catheter is a special balloon catheter that is used to create or enlarge the atrial septal defect found in the heart of certain infants.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
November 1, 2024
Baylis Medical Company Neethi Murali Senior Regulatory Affairs Specialist 5825 Explorer Drive Mississauga, ON L4W5P6 Canada
Re: K242076
Trade/Device Name: VersaCross™ RF Wire Regulation Number: 21 CFR 870.5175 Regulation Name: Septostomy Catheter Regulatory Class: Class II Product Code: DXF Dated: October 3, 2024 Received: October 4, 2024
Dear Neethi Murali:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Kalkidan Molla -S
Kalkidan Molla Acting Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K242076
Device Name VersaCross RF Wire
Indications for Use (Describe)
The VersaCross™ RF Wire is indicated for creation of an atrial septal defect in the heart.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 12. 510(K) SUMMARY
#### Submitter Information
| A. Company Name: | Baylis Medical Company Inc. |
|---------------------------|---------------------------------------------------------------|
| B. Company Address: | 5825 Explorer Drive<br>Mississauga, Ontario L4W 5P6<br>Canada |
| C. Company Phone: | +1 (905) 602-4875 |
| D. Contact Person: | Neethi Murali<br>Senior Regulatory Affairs Specialist |
| E. Date Summary Prepared: | 15-July-2024 |
# Device Identification
| A. Device Trade Name: | VersaCross™ RF Wire |
|-------------------------|------------------------------------|
| B. Device Common Name: | Septostomy catheter |
| C. Classification Name: | CFR 870.5175 – Septostomy Catheter |
| D. Product Code: | DXF |
| E. Device Class: | Class II |
{5}------------------------------------------------
#### Identification of Legally Marketed Device
| Predicate Device | Manufacturer | 510(k) | Indications for Use |
|-----------------------------|--------------------------------|---------|----------------------------------------------------------------------------------------------------------|
| ProTrack™ RF<br>Anchor Wire | Baylis Medical<br>Company Inc. | K150709 | The ProTrack™ RF Anchor<br>Wire is indicated for<br>creation of an atrial septal<br>defect in the heart. |
#### Table 12.1: Predicate Device
## Intended Use/ Indications for Use
The VersaCross™ RF Wire is indicated for creation of an atrial septal defect in the heart.
#### Device Description
The subject VersaCross™ RF Wire is a single-use device that is supplied sterile to the user.
The subject device is comprised of an RF Wire and separately cleared Connector Cable (K023334) that are packaged together. The RF Wire is indicated for the creation of an atrial septal defect in the heart. The RF Wire delivers radiofrequency (RF) power in a monopolar mode between its distal active electrode and a commercially available external Disposable Indifferent (Dispersive) Patch (DIP) Electrode/grounding pad. The RF Wire is used with a separately cleared compatible generator, such as the Baylis RF Puncture Generator (K12278).
The VersaCross™ RF Wire is comprised of a core stainless steel wire. The RF Wire body is covered with an insulating material to facilitate smooth advancement of the device in conjunction with providing electrical insulation. The floppy distal end of the RF Wire is curved and the active tip is rounded to be atraumatic to cardiac tissue unless RF energy is applied. A
{6}------------------------------------------------
marker coil and radio-opaque band is positioned on the distal curve and distal tip, respectively, to facilitate RF Wire placement at the target puncture site during procedures under appropriate imaging quidance such as fluoroscopy, echocardiography and/or electroanatomic mapping (EAM) quidance. The main body of the RF Wire provides a rail for advancing ancillary devices into the left atrium following creation of an atrial septal defect. The RF Wire features visible proximal markers along its length to assist with aligning the wire tip in compatible transseptal sheath and/or dilator assemblies. The proximal end of the RF Wire is bare metal to facilitate connection with the included connector cable. The other end of the included connector cable connects to the compatible Baylis RF Puncture Generator.
## Comparison of Characteristics with Predicate Device
The intended use and indications for use of the proposed VersaCross™ RF Wire remains unchanged from the predicate ProTrack™ RF Anchor Wire (K150709). The subject and predicate device also share the same fundamental scientific technology, including principles of operation and mechanism of action and sterilization method. The subject device incorporates technological modifications to support the use of different devices, such as larger bore therapy catheters, to access the left atrium following transseptal puncture.
The technological differences between the subject and predicate devices do not raise new or different questions of safety and effectiveness. The results of verification and validation testing support substantial equivalence of the proposed VersaCross™ RF Wire with the predicate device. Table 13.2 below provides a comparison of the subject and predicate devices.
{7}------------------------------------------------
# Table 13.2: Comparison of Subject and Predicate Devices
| Characteristic | Subject Device<br>Compared to Predicate<br>ProTrack™ RF Anchor<br>Wire (K150709) | Comment |
|------------------------------------------------------------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for Use | Identical | Both subject and predicate device are<br>indicated for creation of an atrial septal<br>defect in the heart. |
| Fundamental<br>scientific<br>technology | Identical | Both subject and predicate device rely on<br>use of controlled RF energy by the user for<br>transseptal puncture. |
| Operating principle | Identical | Both subject and predicate device are<br>operator controlled. The RF energy is<br>delivered via the BMC RF Puncture<br>Generator to the Wire's distal tip by the<br>operator. |
| Mechanism of<br>action | Identical | The mechanism of action for both subject<br>and predicate device is cell vaporization<br>and dielectric breakdown by RF energy. |
| Environment of Use | Identical | Both the subject and predicate device is<br>used in facilities equipped with staff to<br>perform diagnostic and interventional<br>procedures. |
| Technological<br>characteristics<br>(Dimensions,<br>design, materials) | Similar | Both the predicate and subject device<br>share the same fundamental design.<br>The subject device differs from the<br>predicate device as follows:<br>• Device usable length<br>• Core wire shaft outer diameter<br>• Distal taper length<br>• Heat shrink insulation wall thickness<br>• Presence of proximal markers |
| | | • Clear instead of blue PTFE insulation |
| Packaging<br>configuration | Similar | The predicate device is supplied in a<br>dispenser coil with a tip straightener,<br>placed in a double Tyvek®/Nylon pouch.<br>The subject device is supplied in a<br>dispenser coil with tip straightener in a<br>tray with a lid, placed in a Tyvek®/Nylon<br>pouch. |
| Sterilization<br>method | Identical | Both subject and predicate device are<br>Single Use, Ethylene Oxide sterilized |
{8}------------------------------------------------
#### Summary of Non-Clinical Performance Testing
Non-clinical performance testing was completed to demonstrate safety and effectiveness and substantial equivalence of the subject device to the predicate device. All test requirements were met as specified by applicable standards and test protocols. The following verification and validation activities were completed to demonstrate the safety and effectiveness of the subject device:
#### Mechanical Testing
Mechanical verification was conducted for the subject device to ensure compliance with of of ISO 11070:2014/Amd.1:2018 and Baylis Medical Company Inc. requirements. The following mechanical tests were performed:
- Corrosion Resistance Test .
- . Distal Curve integrity test
- . Joint Bending Test
- Flex test .
- Fracture test
- Tensile test .
- Bumper stiffness .
{9}------------------------------------------------
#### Electrical Testing
Electrical testing was conducted for the subject device to ensure compliance of the device with relevant aspects of IEC 60601-2-2:2017, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.
- . HF Current leakage
- HF Dielectric Strength Test ●
- . Mains Frequency Dielectric Withstand Test
- . Electrical Resistance test
#### Thermal Testing
Arc Integrity testing to ensure compatibility of the subject device with the BMC RF Puncture Generator was conducted to verify compliance with Baylis self-enforced requirements.
#### General Physical Testing
General physical verification testing was conducted for the subject device to verify compliance with Baylis self-enforced requirements. The following general physical tests were performed:
- . J curve Retention Test
- . Pigtail Curve Retention Test
#### Biocompatibility Verification
The biological safety was evaluated for the subject device to verify compliance with the current applicable requirements of ISO 10993-1:2020 and the September 8, 2023 FDA quidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
{10}------------------------------------------------
#### Sterilization Verification
Sterilization and residual limit verification were evaluated for the subject device to verify compliance with the current applicable requirements of ISO 11135:2014/A1:2019 and ISO 10993-7:2008/A1:2022. Sterilization was performed with Ethylene Oxide to a Sterility Assurance Level (SAL) of 10-6.
#### Pyrogen Testing
The subject device is supplied non-pyrogenic. Limulus Amoebocyte Lysate (LAL) testing was evaluated using the Kinetic Chromogenic method, as per ANSI/AAMI ST72:2019 and the FDA guidance document, "Guidance for Industry – Pyrogens and Endotoxins Testing: Questions and Answers," to verify the subject device meets current FDA and USP pyrogen limit specifications.
#### Packaging Verification
Ship testing was evaluated to verify the integrity of the subject device packaging through the rigors of shipping and handling as well as storage over time. The sterile barrier integrity was also evaluated to verify compliance with the current applicable requirements of ISO 11607-1:2020 over the proposed intended shelf life of the subject device.
#### Benchtop Validation
Customer requirements were validated through benchtop validation activities. Benchtop validation testing evaluated subject device design, compatibility with commercial impedance based electroanatomical mapping (EAM) systems and performance during normal and worse case intended use and was completed using an appropriate representative anatomical model.
{11}------------------------------------------------
#### Non-clinical Performance Testing recommended per FDA guidance Coronary, Peripheral, and Neurovascular Guidewires -Performance Tests and Recommend Labeling
- Dimensional verification and Visual Inspection .
- . Repeated Simulated use
- . Tensile strength
- . Torque strength
- . Torqueability
- . Kink resistance
The proposed VersaCross™ RF Wire met all test requirements as specified by applicable standards and test protocols. The verification and validation activities for safety and effectiveness, along with the testing completed for the design changes demonstrated the subject device meets its intended use and is as safe, as effective, and performs in a manner that is substantially equivalent to the predicate device.
## Conclusion
The subject and predicate device share the same intended use, indications for use and fundamental scientific technology, including principles of operation and mechanism of action. Technological differences between the subject and predicate devices do not raise new or different questions of safety and effectiveness. The results of verification and validation activities support substantial equivalence of the proposed VersaCross™ RF Wire to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.