CONNECTOR COMPONENTS WITH BALANCE BIOSURFACE
K122811 · Medtronic, Inc. · DTL · Oct 12, 2012 · Cardiovascular
Device Facts
| Record ID | K122811 |
| Device Name | CONNECTOR COMPONENTS WITH BALANCE BIOSURFACE |
| Applicant | Medtronic, Inc. |
| Product Code | DTL · Cardiovascular |
| Decision Date | Oct 12, 2012 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 870.4290 |
| Device Class | Class 2 |
Intended Use
Medtronic Comector Components with Balance Biosurface are intended for use in the extracorporeal cardiopulmonary perfusion bypass circuit to interconnect tubing and the primary devices of the bypass circuit.
Device Story
Connector components (Y-type, straight, reducer, adapter/luer) used in extracorporeal cardiopulmonary bypass circuits; function to interconnect tubing and primary bypass devices. Components feature smooth blood-contacting surfaces to provide continuous fluid flow path. Used in surgical settings during cardiopulmonary bypass procedures. Operated by perfusionists or surgical staff. Benefit provided by Balance Biosurface coating to improve hemocompatibility within the bypass circuit.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, performance testing, and comparison to the predicate device.
Technological Characteristics
Polycarbonate components with Balance Biosurface coating. Configurations include Y-type, straight, reducer, and adapter (luer). Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for use in the extracorporeal circuit during cardiopulmonary bypass surgical procedures.
Regulatory Classification
Identification
A cardiopulmonary bypass adaptor, stopcock, manifold, or fitting is a device used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
- Tubing, Connectors and Accessories with Balance® Biosurface (K113845)
Related Devices
- K113845 — TUBING, CONMNECTORS, AND ACCESSORIES WIT BALANCE BIOSURFACE · Medtronic, Inc. · Jan 25, 2012
- K081881 — GISH TUBING AND CONNECTORS WITH HA COATING · Gish Biomedical, Inc. · Jan 23, 2009
- K151523 — Uncoated Tubing and Connector Components · Medtronic, Inc. · Jul 16, 2015
- K022918 — ADAPTERS WITH X-COATING · Terumo Cardiovascular Systems Corporation · Nov 27, 2002
- K992906 — COMPONENTS FOR CARDIOVASCULAR PROCEDURE KIT · Olson Medical Sales, Inc. · Nov 10, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
K 122811
Image /page/0/Picture/1 description: The image shows the Medtronic logo. The logo consists of a circular symbol on the left and the word "Medtronic" in bold, black letters on the right. The circular symbol contains a stylized image of a person in motion.
## 510(k) Summary
OCT 1 2 2012
Date Prepared:
Submitter :
September 12, 2012
Medtronic, Inc. Medtronic Perfusion Systems 7611 Northland Drive Minneapolis, MN 55428 Establishment Registration Number: 2184009
Contact Person:
Jacqueline A Hauge Senior Regulatory Affairs Specialist Medtronic Perfusion Systems Phone: 763.514.9967 Fax: . 763.367.8360 Email: jacqueline.a.hauge@medtronic.com
Alternate Contact: Susan Fidler Senior Regulatory Affairs Manager Medtronic Perfusion Systems Phone: 763.514.9839 Fax: 763.367.8360 Email: susan.c.fidler@medtronic.com
#### Device Name and Classification
Tubing, Connectors and Accessories with Balance® Biosurface Trade Name: Common Name: Cardiopulmonary bypass adapter, tubing, connector, stopcock, manifold or fitting Regulation Number: 21 CFR 870.4290 . DTL and DWF Product Codes: Product Classification: Class II
| Predicate Device | |
|------------------|--|
|------------------|--|
K113845
Tubing, Connectors and Accessories with Balance® Biosurface
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the Medtronic logo. On the left is a circular symbol with a figure inside, and on the right is the word "Medtronic" in bold, sans-serif font. The logo is black and white.
#### Device Description
Medtronic Comector Components with Balance Biosurface are intended for use in the extracorporeal circuit during cardiopulmonary bypass surgery. These components are primarily used to interconnect tubing and the primary devices of the bypass circuit. The lumen of the Connector Components with Balance Biosurface consist of smooth blood-contacting surfaces which provide a continuous pathway for the flow of blood and/or other fluids during cardiopulmonary bypass surgical procedures. The Connector Components with Balance Biosurface are comprised of polycarbonate and Balance Biosurface materials and provided in Ytype, straight, reducer and adapter (luer) configurations.
### Intended Use
Medtronic Comector Components with Balance Biosurface are intended for use in the extracorporeal cardiopulmonary perfusion bypass circuit to interconnect tubing and the primary devices of the bypass circuit.
## Comparison to Predicate
When compared to the predicate device (K113845), the Medtronic Connector Components with Balance Biosurface presented in this submission have the same:
- Intended Use
- Technological characteristics
- Operating principle
- Design features .
- Performance
- Biological safety
- Base materials prior to coating
- Coating materials
- Packaging materials and configurations
- Method of sterilization and sterility assurance level
- Shelf life
#### Conclusion
Medtronic has demonstrated that Connector Components with Balance Biosurface for use in extracorporeal cardiopulmonary bypass perfusion systems (tubing sets/packs) are substantially equivalent to the predicate device based upon design, test results, and indications for use. The fundamental scientific principle, labeling and intended use are unchanged from the predicate device.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 1 2 2012
Medtronic, Inc.
c/o Ms. Jacqueline Hauge
Senior Regulatory Affairs Specialist
7611 Northland Dr.
Minneapolis, MN 55428
Re: K122811
Trade/Device Names: Connector Components with Balance Biosurface Regulation Number: 21 CFR 870.4290 Regulation Name: Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting Regulatory Class: Class II Product Code: DTL Dated: September 12, 2012 Received: September 13, 2012
Dear Ms. Hauge:
We have reviewed your Section 510(k) premarket notification of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Jacqueline Hauge
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Connector Components with Balance™ Biosurface
Indications For Use:
Medtronic Connector Components with Balance™ Biosurface are indicated for use in the extracorporeal circuit during cardiopulmonary bypass surgical procedures.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of _
(Division Sign-Off)
Division of Cardiovascular Devices