K171979 · Medtronic, Inc. · DWF · Oct 27, 2017 · Cardiovascular
Device Facts
Record ID
K171979
Device Name
Tubing Pack
Applicant
Medtronic, Inc.
Product Code
DWF · Cardiovascular
Decision Date
Oct 27, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4210
Device Class
Class 2
Indications for Use
This product is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Device Story
Medtronic Tubing Packs are pre-assembled sets of tubing, connectors, and medical devices used to construct extracorporeal perfusion circuits for cardiopulmonary bypass (CPB) surgery. The device facilitates the connection of primary components within the bypass circuit to manage blood flow during surgery. The system is used in clinical settings by perfusionists or surgical staff. The current submission adds new tubing and connector components—both coated (Cortiva, Trillium, Balance) and uncoated—to the existing product line. These components function as passive conduits for blood. The device does not perform active processing or analysis; it serves as the physical pathway for extracorporeal circulation, supporting the patient's physiological needs during procedures where the heart and lungs are bypassed.
Clinical Evidence
Bench testing only. Verification and validation testing included pressure integrity, pressure decay, spallation (tubing), kink resistance, tubing life, pull force, and biocompatibility. Testing was performed on representative worst-case components to demonstrate substantial equivalence to the predicate.
Technological Characteristics
The device consists of medical-grade tubing and connectors, either uncoated or coated with Cortiva, Trillium, or Balance biosurfaces. It is a passive extracorporeal circuit component. Sterilization is unchanged from previously cleared components. No electronic or software components are present.
Indications for Use
Indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
Catheter, Cannula and Tubing, Vascular, Cardiopulmonary Bypass (K151523)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 27, 2017
Medtronic, Inc. Renee Cveykus Principal Regulatory Affairs Specialist 8200 Coral Sea Street NE (MVS83) Mounds View, Minnesota 55112
Re: K171979
Trade/Device Name: Tubing Pack Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing Regulatory Class: Class II Product Code: DWF Dated: October 18, 2017 Received: October 19, 2017
Dear Renee Cveykus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K171979
Device Name Tubing Pack
Indications for Use (Describe)
This product is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date Prepared: | June 29, 2017 |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name and Address: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration Number: 2184009 |
| Contact Person: | Renee L. Cveykus<br>Principal Regulatory Affairs Specialist<br>Medtronic Perfusion Systems<br>Phone: (763) 505-3059<br>Fax: (763) 367-0401<br>Email: renee.l.cveykus@medtronic.com |
| Alternate Contact: | Jake Roeller<br>Sr. Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: (763) 526-0404<br>Fax: (763) 514-9521<br>Email: jake.w.roeller@medtronic.com |
# Proprietary Name:
| Models | Description |
|----------------|-------------|
| See Appendix 3 | Tubing Pack |
# Device Name and Classification:
| Trade Name: | Tubing Pack |
|-----------------------|----------------------------------------------------------------|
| Common Name: | Catheter, cannula and tubing, vascular, cardiopulmonary bypass |
| Classification Name: | Cardiopulmonary bypass vascular catheter, cannula, or tubing. |
| Classification Panel: | Cardiovascular |
| Regulation Number: | 21 CFR 870.4210 |
| Product Code: | DWF |
| Classification: | Class II |
{4}------------------------------------------------
## Predicate Device:
K171308 Tubing Pack
#### Device Description
Medtronic Tubing Packs are used in cardiopulmonary bypass procedures for connecting the primary devices of the bypass circuit. The functionality and intended use of these devices are the same as those for the listed in the Table 2-1.
| 510(k)<br>Number | Date of<br>Clearance | Device |
|------------------|----------------------|----------------------------------------------------------------|
| K800178 | 02/21/1980 | Uncoated Tubing Packs |
| K883956 | 10/28/1988 | Biomedicus Tubing Packs |
| K924529 | 06/23/1993 | Signature Tubing Packs |
| K891687 | 05/31/1989 | Carmeda Coated Tubing Packs |
| K012538 | 23/08/2001 | Trillium Coated Tubing Packs |
| K113845 | 01/25/2012 | Tubing, Connectors, and Accessories with Balance™ Biosurface |
| K122811 | 10/12/2012 | Balance Coated Connector Components |
| K151523 | 07/16/2015 | Catheter, Cannula and Tubing, Vascular, Cardiopulmonary Bypass |
| K171308 | 06/02/2017 | Tubing Pack – New Packaging Trays |
Table 2-1: Tubing Packs Cleared
#### Indications for Use
There is no change to the intended use of the devices/components. The current Indications for Use statement for these devices is listed below:
The Medtronic Tubing Pack is indicated for use in the extracorporeal circuit during cardiopulmonary bypass (CPB) surgical procedures.
#### Comparison to Predicate Devices
When compared to the predicate device, the Tubing Packs have the same:
- Intended Use
- Performance ●
- Technological characteristics ●
- Base materials ●
- Operating principle
- Method of sterilization and sterility assurance level .
- Design features
- Shelf life
{5}------------------------------------------------
- Coatings
- Biocompatibility
## Addition of Components (Tubing and Connectors)
Medtronic Tubing Packs consist of coated and uncoated tubing, connectors and various medical devices that are pre-assembled into user-specified extracorporeal cardiopulmonary bypass perfusion circuits. Tubing Packs are intended for use within the extracorporeal blood pathway during cardiopulmonary bypass surgical procedures. Medtronic Tubing Packs are currently packaged into qualified clear tray and sterile packaging systems. This submission is to add additional tubing and connectors for use in Tubing Pack and is changing as part of continual process improvement efforts. The component additions include tubing and connectors that may be coated or uncoated. The coating types include: Cortiva, Trillium and Balance and these three types of coatings are identical to the cleared existing components for Tubing Packs (K891687, K012538 and K122811). The components being added do not have any changes in materials, technology or labeling than previously cleared components for Tubing Packs. These additional components are representative of currently cleared components and do not add any additional risk to patients therefore, Medtronic believes these components may be added to the Tubing Packs. All packaging components remain unchanged (peel pouches, ties, bands, trays, lids and shipper). There is no change to the sterile barrier, sterilization and test methods used for these components. The type of components being added are outlined in Table 2-2.
| Component | Quantity | Uncoated | Coated | | |
|------------|----------|------------------------------------------------------------------------------------------------|---------|----------|---------|
| | | | Cortiva | Trillium | Balance |
| Connectors | 121 | 39<br>Type: Cap, luer, stopcock, monitoring line, connectors | 28 | 24 | 30 |
| Tubing | 86 | 35<br>Size: .085x.062x, .107x.040x, 1/2X1/8, 1/2x3/32x, 1/4X1/16, 3/16x1/16, and<br>5/16x3/32x | 0 | 23 | 28 |
| Total | 207 | | | | |
Table 2-2: Types of Components being Added
A detailed list of the new components can be found within Appendix 2.
## Summary of Performance and Biological Testing
Verification and validation testing demonstrated that the tubing and connectors are substantially equivalent to the predicate. Medtronic conducted the following testing for the tubing and connector components:
- . Pressure Integrity
Pressure Decay
●
- Spallation (Tubing) ●
- Kink Resistance (Tubing) ●
- ●
- Tubing Life Biocompatibility ●
Pull Force
{6}------------------------------------------------
The testing completed was a representative of worse case testing for tubing and connectors and the verification results were leveraged based on most challenging components. Each of the components in scope of this submission were reviewed against the product specification and the following rationalization documentation was completed:
- . Mechanical Requirements (10561942DOC)
- Coating Requirements (10552508DOC) ●
- . Design Verification for Non-stacked conditioning and Pressure Integrity (10571289DOC)
- Biocompatibility per product specifications (10560838DOC) ●
Based on this evaluation Medtronic considers these tubing and connectors are substantially equivalent to the predicate.
# Conclusion
Medtronic has demonstrated that the Tubing Pack with the additional components used in extracorporeal cardiopulmonary bypass perfusion systems is substantially equivalent to the predicate device based upon design, testing, and indications for use. The fundamental scientific principle, labeling and the intended use are unchanged because of this device modification.