K133151 · Terumo Cardiovascular Systems Corp. · DWF · Feb 3, 2014 · Cardiovascular
Device Facts
Record ID
K133151
Device Name
SARNS HIGH FLOW AORTIC ARCH CANNULA
Applicant
Terumo Cardiovascular Systems Corp.
Product Code
DWF · Cardiovascular
Decision Date
Feb 3, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4210
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Sarns™ High-Flow Aortic Arch Cannula is indicated for perfusion of the ascending aorta during cardiopulmonary bypass surgery for up to 6 hours of use.
Device Story
Sarns™ High-Flow Aortic Arch Cannula is a tubular conduit used during cardiopulmonary bypass surgery. It functions as a component of the extracorporeal circuit, returning oxygenated blood from the circuit into the patient's ascending aorta. The device is operated by perfusionists or surgeons in an operating room setting. It facilitates the re-entry of blood into the vascular stream after it has been processed by a blood reservoir and pump (roller or centrifugal). The device is a passive conduit; it does not perform active sensing or processing. Its use supports systemic circulation during procedures lasting up to 6 hours.
Clinical Evidence
No clinical data. Safety and effectiveness were demonstrated through bench testing, including evaluations of force at break, liquid and air leak, ink adhesion, hemolysis, torque, corrosion, and pressure drop.
Technological Characteristics
Tubular conduit cannula. Materials: polyvinylchloride, PMEA polymer, polyethylene, acrylonitrile butadiene styrene (ABS), and stainless steel. Passive flow device; no energy source or software. Biocompatible construction.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass surgery requiring perfusion of the ascending aorta for up to 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
K974736 — HEARTPORT DIRECT AORTIC RETURN CANNULA WITH INTRODUCER · Heartport, Inc. · Jul 15, 1998
K024058 — MODIFICATION TO CARDEON AEGIS CATHETER · Cardeon Corp. · Dec 30, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Terumo Cardiovascular Systems. The logo consists of a circle with a T inside of it, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" is the phrase "Cardiovascular Systems" in a smaller, sans-serif font. The logo is black and white.
ﺔ ﺍﻟﺘﻲ
# FEB - 3 2014
K133151
### SECTION 5-510(K) SUMMARY Sarns™ High-Flow Aortic Arch Cannula
Date Prepared:
September 2013
Sponsor Information: Owner/Applicant/Submitter:
Terumo Cardiovascular Systems Corporation 6200 Jackson Road Ann Arbor, Michigan 48103 Phone: 1-800-262-3304 Fax: 410-398-6079 Registration No. 1828100
Contact Person:
Garry A. Courtney, MBA, RAC Corporate Director, Regulatory Affairs 125 Blue Ball Road Elkton, MD 21921 Phone: 1-800-262-3304 ext. 7486 Fax: 410-398-6079 Email: garry.courtney@terumomedical.com
#### Device Names/Classifications:
| Device Trade Name: | Sarns™ High-Flow Aortic Arch Cannula |
|----------------------|-----------------------------------------------------------|
| Device Common Name: | Aortic cannula for cardiopulmonary bypass |
| Classification Name: | Cardiopulmonary bypass vascular catheter, cannula, tubing |
| Regulation Number: | 21 CFR 807.4210 |
| Classification: | Class II |
| Product Code: | DWF |
#### Predicate Devices:
- Sarns™ High Flow Aortic Arch Cannula, K770429 .
- Sarns™ X-coated Aortic Cannula, K083301 .
#### Purpose of Submission:
Terumo Cardiovascular Systems is submitting this Premarket Notification due to cumulative changes that have been made since the time that the original submissions were cleared by FDA. Although none of the individual changes were determined to be significant, Terumo Cardiovascular Systems Corporation believes it is appropriate to submit a 510(k) that represents the devices in their current design state.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the Terumo Cardiovascular Systems logo. The logo consists of a circle with a T inside of it, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" is the phrase "Cardiovascular Systems" in a smaller, sans-serif font.
#### Intended Use/Indications for Use:
The Sarns™ High-Flow Aortic Arch Cannula is indicated for perfusion of the ascending aorta during cardiopulmonary bypass surgery for up to 6 hours of use.
#### Principles of Operation and Technology:
The subject devices are modifications of the predicate devices, and these modifications do not impact the technology of the predicate devices, the principles of operation and do not raise different or new questions of safety and effectiveness from the predicates. The Sarns™ High-Flow Aortic Arch Cannula provides a conduit for the flow of patient blood within the extracorporeal circuit during cardiopulmonary bypass surgery. Typically, blood is gravity drained from the body into a blood reservoir. From the reservoir, blood then is propelled through the balance of the extracorporeal circuit via a roller pump or centrifugal pump. The Sams™ High-Flow Aortic Arch Cannula is a component within the circuit that tunnels patient blood back into the body.
#### Design and Materials:
The design of the Sams™ High-Flow Aortic Arch Cannula allows patient blood to re-enter the body after the blood has routed through an extracorporeal circuit for blood oxygenation. The design of the cannula is simple in that it consists of a tubular conduit with a tip that can be positioned into the patient's anatomy (ascending aorta) to allow the blood to re-enter the vascular blood stream.
The materials of construction for the Sarns™ High-Flow Aortic Arch Cannula can include polyvinylchloride, PMEA polymer, polyethylene, acrylonitrile butadiene styrene and stainless steel.
#### Performance Evaluations:
Clinical studies involving patients are not necessary to demonstrate the safety and effectiveness of the subject devices. Performance assessments for safety and effectiveness were accomplished through bench studies that included the following evaluations:
- force at break
- liquid and air leak .
- ink adhesion .
- hemolysis .
- torque .
- corrosion .
- pressure drop .
#### Substantial Equivalence Comparison:
The information presented in this section depicts a comparison between the subject device, the modified Sarns™ High-Flow Aortic Arch Cannula and the predicate device, the unmodified Sarns™ High-Flow Aortic Arch Cannula.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Terumo Cardiovascular Systems. The logo consists of a circle with the letter 'T' inside, followed by the word 'TERUMO' in bold, sans-serif font. Below the word 'TERUMO' is the text 'Cardiovascular Systems' in a smaller, sans-serif font.
#### Comparison of Intended Use:
The modified Sarns™ High-Flow Aortic Arch Cannula and the predicate Sarns™ High-Flow Aortic Arch Cannula have the exact same intended use statements:
The modified and the unmodified Sarns™ High-Flow Aortic Arch Cannulae are each indicated for perfusion of the ascending aorta during cardiopulmonary bypass surgery for up to 6 hours of use.
#### Duration of Use:
The modified Sarns™ High-Flow Aortic Arch Cannula and the predicate Sarns™ High-Flow Aortic Arch Cannula can both be used in procedures lasting up to 6 hours in duration.
#### Comparison of Labeling: .
The labeling that will be used for the modified device is similar to the labeling used with the predicate device, although some labeling is revised to provide greater clarity for the user. The revised instructions labeling accurately presents the directions that are necessary for the end-user to employ the device in a safe and effective manner. Terumo submits that the labeling complies with applicable regulations in those regions where the device is to be distributed.
#### Comparison of Operation and Technology:
The modified Sarns™ High-Flow Aortic Arch Cannula and the predicate Sams™ High-Flow Aortic Arch Cannula utilize the exact same technologies and principles of operation. The technology of the modified Cannula is not impacted by the modifications made to the subject device.
#### Comparison of Design:
With respect to the design of the modified Sarns™ High-Flow Aortic Arch Cannula, there have been no significant design changes implemented. Terumo has made non-significant changes since the original submission of these devices - each of which has been determined to maintain the integrity of the product without adversely impacting safety and/or performance.
#### Comparison of Materials:
The materials of construction used in the modified Sarns™ High-Flow Aortic Arch Cannula are the same generic materials that are used in the predicate device with the addition of an luer cap made of ABS plastic. Both the modified Sarns™ High-Flow Aortic Arch Cannula and the predicate cannula have been demonstrated to meet biocompatibility requirements.
#### Comparison of Performance:
Terumo Cardiovascular has conducted performance studies with the modified devices to ensure that they all continue to satisfy appropriate device performance specifications and to ensure that they satisfy customer needs. There are no appreciable differences between the subject devices and the predicate devices.
#### Conclusion:
The information and data included in the 510(k) notice demonstrate the Sams™ High-Flow Aortic Arch Cannula is substantially equivalent to the predicate devices for perfusion of the ascending aorta during cardiopulmonary bypass for up to 6 hours of use.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 3, 2014
Terumo Cardiovascular Systems Corporation . Garry Courtney, MBA, RAC 6200 Jackson Road Ann Arbor, MI 48103
Re: K133151
Trade/Device Name: Sarns™ High-Flow Aortic Arch Cannula Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiovascular (74) Regulatory Class: II Product Code: DWF Dated: September 27, 2013 Received: October 17, 2013
Dear Mr. Courtney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device.in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Garry Courtney
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
M. A. Holliem
for Bram Zuckerman, MD Director, Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Terumo Cardiovascular Systems. The logo consists of a circle with the letter T inside, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" is the phrase "Cardiovascular Systems" in a smaller, regular font. The logo is simple and professional, and it is likely used to represent the company on its products and marketing materials.
## SECTION 4 - INDICATION FOR USE Sarns™ High-Flow Extended Aortic Arch Cannula
510(k) Number (if known):
K133151 Unknown at time of submission
Device Name:
Sarns™ High-Flow Aortic Arch Cannula
#### Indications for Use:
Device Intended Use - The device is indicated for perfusion of the ascending aorta during cardiopulmonary bypass surgery for up to six hours of use.
Prescription Use X (Part 21 CFR 801 Subpart D)
or
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. Z. Hillerman
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.