SARNS SOFT-FLOW EXTENDED AORTIC CANNULA

K132811 · Terumo Cardiovascular Systems Corporation · DWF · Feb 27, 2014 · Cardiovascular

Device Facts

Record IDK132811
Device NameSARNS SOFT-FLOW EXTENDED AORTIC CANNULA
ApplicantTerumo Cardiovascular Systems Corporation
Product CodeDWF · Cardiovascular
Decision DateFeb 27, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4210
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The device is indicated for perfusion of the assending aorta during cardiopulmonary bypass surgery for up to six hours of use.

Device Story

Sarns™ Soft-Flow® Extended Aortic Cannula serves as conduit for blood flow within extracorporeal circuit during cardiopulmonary bypass surgery. Device consists of tubular conduit with straight tip positioned in ascending aorta to return oxygenated blood to vascular system. Features include positional suture ring and centimeter depth markings for insertion guidance and securement. Used in surgical settings by clinicians. Straight tip design reduces pressure drop compared to angled tips, minimizing flow resistance and potential for hemolysis. Longer tubing (14.5 inches) improves surgical field visualization. Luer-threaded port cap utilizes ABS plastic. Device facilitates cardiopulmonary bypass for up to 6 hours.

Clinical Evidence

No clinical data. Safety and effectiveness demonstrated through bench testing including force at break, liquid and air leak, ink adhesion, hemolysis, pressure drop, exit velocity, and biocompatibility (ISO 10993).

Technological Characteristics

Tubular cannula with straight tip; 14.5-inch length; ABS plastic luer-threaded port cap; positional suture ring and centimeter depth markings. Biocompatibility per ISO 10993. No electronic components or software.

Indications for Use

Indicated for perfusion of the ascending aorta during cardiopulmonary bypass surgery for up to 6 hours of use.

Regulatory Classification

Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Terumo Cardiovascular Systems. The logo consists of a circle with a "T" inside, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" is the text "Cardiovascular Systems" in a smaller font size. The logo is simple and professional, and it is likely used to represent the company's brand. K132811 # FEB 2 7 2014 # 510(K) SUMMARY Sarns™ Soft-Flow® Extended Aortic Cannula Date Prepared: August 2013 Sponsor Information: Owner/Applicant/Submitter: Terumo Cardiovascular Systems Corporation 6200 Jackson Road Ann Arbor, Michigan 48103 Phone: 1-800-262-3304 Fax: 410-398-6079 Registration No. 1828100 Contact Person: Garry A. Courtney, MBA, RAC Corporate Director, Regulatory Affairs Phone: 1-800-262-3304 ext. 7486 Fax: 410-398-6079 Email: garry.courtney@terumomedical.com #### Device Names/Classifications: | Device Trade Name: | Sarns™ Soft-Flow® Extended Aortic Cannula | |----------------------|-----------------------------------------------------------| | Device Common Name: | Aortic cannula for cardiopulmonary bypass | | Classification Name: | Cardiopulmonary bypass vascular catheter, cannula, tubing | | Regulation Number: | 21 CFR 870.4210 | | Classification: | Class II | | Product Code: | DWF | #### Predicate Devices: - SarnsTM Low Jetting Aortic Arch Cannula, K934127 . - Sarns™ Xcoated Cannula, K083301 #### Device Description: The Sams™ Soft-Flow® Extended Aortic Cannula provides a conduit for the flow of patient blood within the extracorporeal circuit during cardiopulmonary bypass surgery. The or parent of the Sams™ Soft-Flow® Extended Aortic Cannula allows patient blood to re-enter the body after the blood has routed through an extracorporeal circuit for blood oxygenetion. The connila consists of a tubular conduit with a tip that can be positioned into the patient's anatomy (ascending aorta) to allow the blood to re-enter the vascular blood stream. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the Terumo Cardiovascular Systems logo. The logo consists of a circle with the letter T inside, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" is the phrase "Cardiovascular Systems" in a smaller font size. The logo is simple and professional, and it is likely used to represent the company's brand. # Intended Use/Indications for Use: The Sams™ Soft-Flow Extended Aortic Cannula is indicated for perfusion of the assending aorta during cardiopulmonary bypass surgery for up to 6 hours of use. # Performance Evaluations: Clinical studies involving patients are not necessary to demonstrate the safety and effectiveness. of the subject devices. Performance assessments for safety and effectiveness were eccomplished through bench studies that included the following evaluations: - force at break . - liquid and air leak . - ink adhesion # - hemolysis . - pressure drop 4 - exit velocity . # Substantial Equivalence Comparison: The information presented in this section depicts a comparison between the subject device, the modified Sarns™ Soft-Flow® Aortic Extended Cannula and the predicate device: - Sarns™ Low Jetting Aortic Arch Cannula, K934127 . - SarnsTM Xcoated Cannula, K083301 . #### Comparison of Intended Use: . The modified Sarns™ Soft-Flow® Extended Aortic Cannula and the predicate have the exact same intended use statements: The modified and the unmodified devices are each indicated for perfusion of the ascending aorta during cardiopulmonary bypass surgery for up to 6 hours of use. #### Duration of Use: The modified Sarns™ Soft-Flow® Extended Aortic Cannula and the predicate can both be used in procedures lasting up to 6 hours in duration. ## Comparison of Labeling: The labeling used for the modified device is similar to the labeling used with the predicate device, although some labeling is revised to provide greater clarity for the use. The revised instructions labeling accurately presents the directions that are necessary for the end-user to employ the device in a safe and effective manner. Terumo submits that the labeling complies with applicable regulations in those regions where the device is to be distributed. # Comparison of Operation and Technology: The modified Sarns™ Soft-Flow® Extended Aortic Cannula and the predicate utilize the cxact same technologies and principles of operation. The technology of the modified ince is not impacted by the modifications made to the subject device. # Comparison of Design: With respect to the design of the modified Sains™ Soft-Flow Extended Aortic Cannula, changes to the device have been made since the original submission - each of which Callina determined to maintain the integrity of the product without impacting the safety and effectiveness of the device. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Terumo Cardiovascular Systems. The logo consists of a circle with the letter "T" inside, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" are the words "Cardiovascular Systems" in a smaller, sans-serif font. The logo is black and white. The Sarns™ Soft-Flow® Extended Cannula is 14.5 inches long, one-half inch longer than the 14 inch predicate device. This half inch difference in length is not clinically significant. Longer tubing allows clearance from and, thus improved visualization of, the surgical field, particularly in larger patients. The Sarns™ Soft-Flow® Extended Cannula has a positional suture ring and depth . markings on the tubing. The positional suture ring and centimeter depth markings are used to gauge insertion depth of the cannula and provide a location to tighten the sutures around the cannula to hold it in place. These features are comparable in function to the predicate device which has a suture bulb to gauge insertion depth of the cannula and provide a location to tighten the sutures around the cannula to hold it inc place. The Sarns" Soft-Flow "Extended Cannula has a straight tip rather than an angled tip like the predicate devices. A straight tip design generally produces a slightly lower pressure drop (the difference between the pressure entering the camula and the pressure leaving the cannula) compared to the same size angled tip. A lower pressure drop is desirable because there is less resistance to flow and thus less potential for hemolysis. The material that is used for the luer-threaded port cap is changed from polypropylene to ABS plastic. Terumo Cardiovascular Systems maintains a complete profile of all biocompatibility data that fully supports the suggested elements of FDA General Program Memorandum #G95-1: Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part I: Evaluation and Testing," ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ #### Comparison of Materials : ---------------------------------------------------------------------------------------------------------------------------------------------------- The materials of construction used in the modified Sarns™ Soft-Flow® Extended Aortic Cannula are the same generic materials that are used in the predicate device with the exception of a cap material change from polypropylene to ABS plastic. # Comparison of Performance: Terunio Cardiovascular has conducted performance studies with the modified devices to ensure that they all continue to satisfy appropriate device performance specifications and to ensure that they satisfy customer needs. Specifically, performance studies were conduced to ensure force at break, liquid and air leak, hemolysis, pressure drop, exit velodity, ink adhesion, and biocompatibility testing passed pre-determined acceptance criteria. There are no appreciable differences in performance between the subject devices and the predicate devices. The nonclinical tests demonstrate the subject device is as safe and effective as the predicate devices. #### Conclusion: The information and data included in the 510(k) notice demonstrate the Sams™ Soft-Flow Extended Aortic Cannula is substantially equivalent to the predicate devices for perfusion of the ascending aorta during cardiopulmonary bypass for up to 6 hours of use. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Image /page/3/Picture/3 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three overlapping wings, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The logo is simple and recognizable, conveying a sense of authority and trustworthiness. ### February 27, 2014 Terumo Cardiovascular Systems Corporation Mr. Garry A. Courtney Corporate Director, Regulatory Affairs 6200 Jackson Rd. Ann Arbor, MI 48103 Re: K132811 > Trade/Device Name: Sams Soft-Flow Extended Aortic Cannula Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula or tubing Regulatory Class: Class II Product Code: DWF Dated: January 29, 2014 Received: January 30, 2014 Dear Mr. Courtney: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subiect to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply {4}------------------------------------------------ Page 2 - Mr. Garry A. Courtney with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. MADA for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Terumo Cardiovascular Systems. The logo consists of a circle with the letter "T" inside, followed by the word "TERUMO" in bold, sans-serif font. Below the word "TERUMO" are the words "Cardiovascular Systems" in a smaller, sans-serif font. # SECTION 4 - INDICATION FOR USE Sarns™ Soft-Flow® Extended Aortic Cannula 510(k) Number: K132811 Sams™ Soft-Flow® Extended Aortic Cannula Device Name: Indications for Use: Device Intended Use - The device is indicated for perfusion of the assending aorta during cardiopulmonary bypass surgery for up to six hours of use. Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (Part 21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) or Concurrence of CDRH, Office of Device Evaluation (ODE) M. FDA 15 :
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...