K201057 · Medtronic, Inc. · QHW · Jan 14, 2022 · Cardiovascular
Device Facts
Record ID
K201057
Device Name
Bio-Medicus Life Support Catheter and Introducer
Applicant
Medtronic, Inc.
Product Code
QHW · Cardiovascular
Decision Date
Jan 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4100
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Bio-Medicus Life Support catheters and introducers are single-lumen drainage or reinfusion peripheral-access catheters to be used in ECMO or ECLS with an extracorporeal circuit intended for use in adult and pediatric patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
Device Story
Single-lumen peripheral-access catheters and introducers; used for blood drainage or reinfusion in ECMO/ECLS circuits. Introducer facilitates placement over guidewire. Used in clinical settings for patients with acute respiratory/cardiopulmonary failure. Catheters available in 18 cm, 50 cm, or 55 cm tip lengths; some include side holes. Device provides vascular access for extracorporeal support; enables perfusion of vessels/organs. Operated by clinicians. Benefits include providing life-sustaining support for patients failing conventional therapies.
Clinical Evidence
No human clinical studies performed. Evidence based on non-clinical bench testing and a 7-day GLP in vivo animal study. Bench testing included 21-day simulated use durability, tensile strength, pressure testing, blood trauma, kink testing, and mechanical force testing (suture collar, introducer insertion/removal).
Technological Characteristics
Single-lumen catheter and introducer. Catheter body material: siloxane-urethane co-polymer. Sterile, nonpyrogenic, single-use. Compatible with extracorporeal circuits. Biocompatibility per ISO 10993-1:2009. Shelf life: 3 years. Storage: below 25°C.
Indications for Use
Indicated for adult and pediatric patients with acute respiratory or acute cardiopulmonary failure requiring ECMO or ECLS when other treatments fail and death is imminent. Contraindicated in vessels with arterial dissection or severe peripheral atherosclerosis.
Regulatory Classification
Identification
An extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary support (>6 hours) is a system of devices and accessories that provides assisted extracorporeal circulation and physiologic gas exchange of the patient's blood in patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The main devices and accessories of the system include, but are not limited to, the console (hardware), software, and disposables, including, but not limited to, an oxygenator, blood pump, heat exchanger, cannulae, tubing, filters, and other accessories (e.g., monitors, detectors, sensors, connectors).
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(1) The technological characteristics of the device must ensure that the geometry and design parameters are consistent with the intended use, and that the devices and accessories in the circuit are compatible;
(2) The devices and accessories in the circuit must be demonstrated to be biocompatible;
(3) Sterility and shelf-life testing must demonstrate the sterility of any patient-contacting devices and accessories in the circuit and the shelf life of these devices and accessories;
(4) Non-clinical performance evaluation of the devices and accessories in the circuit must demonstrate substantial equivalence of the performance characteristics on the bench, mechanical integrity, electromagnetic compatibility (where applicable), software, durability, and reliability;
(5) In vivo evaluation of the devices and accessories in the circuit must demonstrate their performance over the intended duration of use, including a detailed summary of the clinical evaluation pertinent to the use of the devices and accessories to demonstrate their effectiveness if a specific indication (patient population and/or condition) is identified; and
(6) Labeling must include a detailed summary of the non-clinical and in vivo evaluations pertinent to use of the devices and accessories in the circuit and adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.
Predicate Devices
FDA Final Order 81 FR 7451, February 12, 2016
Reference Devices
Bio-Medicus Adult Cannulae and Introducer (K142673)
Bio-Medicus Pediatric Arterial and Femoral Cannula (K143083)
Bio-Medicus Adult Cannulae and Introducer (K180453)
Related Devices
K240534 — Bio-Medicus Life Support Catheter and Introducer · Medtronic, Inc. · Apr 19, 2024
K251831 — Bio-Medicus Life Support Catheter and Introducer · Medtronic, Inc. · Jan 15, 2026
K203409 — MC3 Crescent Jugular Dual Lumen Catheter · Mc3, Inc. · May 3, 2021
K201100 — Bio-Medicus Life Support Catheter and Introducer · Medtronic, Inc. · Jul 15, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 14, 2022
Medtronic Inc. Juli Rubin Principal Regulatory Affairs Specialist 8200 Coral Sea St. NE Mounds View, Minnesota 55112
Re: K201057
Trade/Device Name: Bio-Medicus Life Support Catheter and Introducer Regulation Number: 21 CFR 21 CFR 870.4100 Regulation Name: Extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary failure Regulatory Class: Class II Product Code: OHW Dated: January 4, 2022 Received: January 5, 2022
Dear Juli Rubin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K201057
#### Device Name
Bio-Medicus Life Support™ Catheter and Introducer
#### Indications for Use (Describe)
The Bio-Medicus Life Support catheters and introducers are single-lumen drainage or reinfusion peripheral-access catheters to be used in ECMO or ECLS with an extracorporeal circuit intended for use in adult and pediatric patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
Type of Use (Select one or both, as applicable)
| <span> <b> <span style="font-size:120%">☑</span> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------------------------------------------------|
| <span> <b> <span style="font-size:120%">☐</span> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 1 of 1
Image /page/2/Picture/20 description: The image shows the text 'KXXXXXXX' on the top line and '1-40' on the second line. The text is black and the background is white. The text is centered in the image.
{3}------------------------------------------------
#### 5.0 510(k) Summary of Safety and Effectiveness
| Date Prepared: | April 20, 2020 |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establish Registration Number: 2184009 |
| Contact Person: | Juli Rubin, MBA, RAC<br>Sr. Principal Regulatory Affairs Specialist<br>Phone: (763) 526-2357<br>Fax: (763) 367-8361<br>E-mail: juli.c.rubin@medtronic.com<br><br>Kristyn Benson (Alternate)<br>Sr. Regulatory Affairs Director<br>Phone: (763) 526-2523<br>Email: kristyn.m.benson@medtronic.com |
| Trade Name: | Bio-Medicus Life Support™ Catheter and Introducer |
| Common Name: | Extracorporeal Life Support Single Lumen Catheter |
| Classification Name: | Extracorporeal circuit and accessories for long-term<br>respiratory/cardiopulmonary failure |
| Classification: | Class II (with special controls) |
| Regulation Number: | 21 CFR 870.4100 |
| Product Code: | QHW |
| Name of Predicate Device: | FDA Final Order 81 FR 7451, February 12, 2016 |
| Name of Reference Devices: | Bio-Medicus Adult Cannulae and Introducer (K142673)<br>Bio-Medicus Pediatric Arterial and Femoral Cannula (K143083)<br>Bio-Medicus Adult Cannulae and Introducer (K180453) |
{4}------------------------------------------------
## Device Description:
The Bio-Medicus Life Support™ catheter is a single-lumen catheter used to drain or reinfuse blood. The introducer facilitates proper insertion and placement of the appropriately sized catheter over a guidewire within the vessel. These devices are intended to perfuse vessels or organs in conjunction with extracorporeal support, including Extracorporeal Membrane Oxygenation (ECMO) and Extracorporeal Life Support (ECLS). Catheter models with tip lengths of 18 cm (7.09 in), 50 cm (19.7 in), or 55 cm (21.7 in), with and without additional side holes may be used as either drainage or reinfusion catheters.
These devices are sterile, nonpyrogenic, disposable, intended for single use only. Bench studies were performed after device preconditioning including exposure (21 days) to simulated in vivo use conditions to demonstrate safety and reliability. Do not store the product above 25°C (77°F).
## Indication for Use:
The Bio-Medicus Life Support catheters and introducers are single-lumen drainage or reinfusion peripheral-access catheters to be used in ECMO or ECLS with an extracorporeal circuit intended for use in adult and pediatric patients with acute respiratory or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
## Contraindications:
Alone, the catheter and introducer are not medical treatment devices. The introducer is to be used only with the appropriately sized Bio-Medicus Life Support catheter. These devices are intended for use only as indicated in these instructions for use. Do not insert the catheter in a vessel that has arterial dissection or severe peripheral atherosclerosis.
## Comparison to Predicate (Special Controls):
Substantial equivalence evaluation includes a comparison to requirements in the FDA Final Order 81 FR 7451. February 12. 2016. as well as a comparison to the Reference Devices. Per FDA Final Order (81 FR 7451) membrane lung devices for long-term pulmonary support, a preamendment class III device, as extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary failure, were redesignated and reclassified to class II (special controls) in patients with acute respiratory failure or acute cardiopulmonary failure where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
The Bio-Medicus Life Support™ Catheter and Introducer device meets all special controls identified in 21 CFR 870.4100, as follows:
- Technological Characteristics: The geometry and design parameters of the subject ● device are consistent with the device's intended use in extracorporeal support procedures, and the device is compatible with the other devices and accessories in the extracorporeal circuit.
- . Biocompatibility: The subject device is demonstrated to be biocompatible as a prolonged use device in accordance with ISO 10993-1:2009 and with FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of
{5}------------------------------------------------
medical devices - Part 1: Evaluation and testing within a risk management process" (16 June 2016).
- Sterility and Shelf-life: Sterilization adoption evaluation and shelf-life testing . demonstrate that the subject device maintains its sterility, integrity, durability, and reliability over the stated shelf life of the device.
- Non-clinical Performance: Substantial equivalence of the performance characteristics is demonstrated on bench, mechanical integrity, durability, and reliability testing.
- . In vivo Evaluation: In vivo evaluation demonstrates the subject device's performance over a 7-day duration of use.
- Labeling: The Instructions for Use include a detailed summary of the non-clinical and in vivo evaluations pertinent to use of the device in an extracorporeal circuit and adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.
# Comparison to Reference Devices:
A comparison of the Medtronic Bio-Medicus Life Support Catheters and Introducers to the reference devices indicates the following similarities:
- . Similar intended use, the only exception is the extended duration of extracorporeal support.
- Same technological characteristics ●
- Same operating principle ●
- Same design features; the only exception is a slight increase in outer diameter (OD) of small sized devices to ensure durability throughout the duration of intended use. The inner diameter (ID) is identical to the reference device.
- . Similar materials; the only exception is the catheter body blood-contacting material which has changed from polyurethane to a siloxane-urethane co-polymer, which provides a biostable surface for a longer-term use.
- . Similar shelf life; the subject device has a 3-year shelf life, which is one year shorter than the reference device (4-year shelf life).
## Summary of Performance Data
Per FDA Final Order (81 FR 7451), February 12, 2016, in vivo clinical studies involving patients are not necessary to demonstrate the safety and performance of the subject devices. Pre-clinical bench testing was used to verify the performance characteristics of this device. In vivo animal testing was also completed to establish substantial equivalence with the predicate regulation.
The following performance tests were conducted:
- 21-day Simulated use durability testing ●
- Tensile strength after life conditioning (long term use)
{6}------------------------------------------------
- Pressure test after life conditioning (long term use)
- Blood trauma testing
- Kink testing .
- Birds Mouth testing
- ID occlusion due to suture ring ●
- Suture Collar Slide Force ●
- Introducer removal and insertion forces ●
- 7-day GLP in vivo study
#### Conclusion:
The data included in this submission are sufficient to demonstrate that the Medtronic Bio-Medicus Life Support Catheter and Introducer is substantially equivalent to the Final Order 81 FR 7451, Redesignation as Extracorporeal Circuit and Accessories For Long-term Respiratory/cardiopulmonary Failure (ECMO), February 12, 2016 for the perfusion of vessels or organs in conjunction with extracorporeal support, including Extracorporeal Membrane Oxygenation (ECMO) and Extracorporeal Life Support (ECLS).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.