K142673 · Medtronic, Inc. · DWF · Nov 18, 2014 · Cardiovascular
Device Facts
Record ID
K142673
Device Name
Bio-Medicus Adult Cannulae and Introducer
Applicant
Medtronic, Inc.
Product Code
DWF · Cardiovascular
Decision Date
Nov 18, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4210
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These devices are to be used by a trained physician only. Cannulae are used to cannulate vessels, perfuse vessels or organs, and/or connect with accessory extracorporeal equipment. The cannula introducer is intended to facilitate proper insertion and placement of the appropriately-sized cannula within the vessel for cardiopulmonary bypass. The Bio-Medicus™ cannula (18-cm [7.09in] tip length models) may be used in either the femoral position as an arterial delivery cannula or in the jugular position as a venous return cannula. This product is intended for use up to 6 hours.
Device Story
Device consists of polyurethane wirewound cannulae with multiport distal tips and non-phthalate PVC introducers; used for cardiopulmonary bypass. Cannulae facilitate vessel cannulation, organ perfusion, and connection to extracorporeal circuits. Introducers provide elongated taper for vessel dilation and placement. Features include radiopaque repositionable suture rings, insertion depth marks, and securement clips. Large venous models include tubing adapters for circuit compatibility. Used in clinical settings (e.g., OR) by trained physicians. Provides physical conduit for blood flow; output is the established extracorporeal circuit connection. Benefits include facilitated insertion, improved vessel access, and securement during bypass procedures up to 6 hours.
Clinical Evidence
No human clinical data. Evidence based on pre-clinical bench testing and animal studies. Bench testing included blood trauma, sterilization, biocompatibility, packaging, pressure drop, cannula life, and securement clip testing. Results demonstrated performance equivalent or superior to predicate devices.
Technological Characteristics
Polyurethane wirewound cannula body; stainless steel wirewinding/reinforced baskets (venous); non-phthalate PVC introducer; polycarbonate/PVC connectors; silicone hemostasis cap. Radiopaque suture ring. Sterile, nonpyrogenic, single-use. Dimensions: 12.5-in to 30-in lengths. Mechanical device; no software or energy source.
Indications for Use
Indicated for patients requiring cardiopulmonary bypass; used by trained physicians to cannulate vessels, perfuse vessels/organs, or connect to extracorporeal equipment. Contraindicated in patients with severe peripheral atherosclerosis or severe arterial dissection.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
Predicate Devices
Medtronic Bio-Medicus Femoral Cannula and Introducer (K884129)
{1}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2014
Medtronic, Inc. Bruce Backlund Principal Regulatory Affairs Specialist 8200 Coral Sea St. NE Mounds View, Minnesota 55112
Re: K142673
Trade/Device Name: Bio-Medicus Adult Cannulae and Introducer Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing Regulatory Class: Class II Product Code: DWF Dated: September 17, 2014 Received: September 19, 2014
Dear Mr. Backlund.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrenner
for
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K142673
Device Name Bio-Medicus Adult Cannula and Introducer
#### Indications for Use (Describe)
These devices are to be used by a trained physician only. Cannulate vessels, perfuse vessels or organs, and/or connect with accessory extracorporeal equipment. The cannula introducer is intended to faciliate proper insertion and placement of the appropriately-sized cannula within the vessel for cardiopulmonary bypass. The Bio-Medicus" cannula (18-cm [7.09-in] tip length models) may be used in either the femoral position as an arterial delivery cannula or in the jugular position as a venous return cannula. This product is intended for use up to 6 hours.
Type of Use (Select one or both, as applicable)
■ Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFFEMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.•
FORMFDA 3881(1114t
Page 1 of 1
I SC Publishier Sers ices (32) | J =J 14 | 14 | 14 | 14 |- | 7-411 利
Form Approved; OMB No_0910-0120 Expiration Date= January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
| Date Prepared: | September 17, 2014 |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establish Registration Number: 2184009 |
| Contact Person: | Bruce Backlund<br>Principal Regulatory Affairs Specialist<br>Phone: (763) 526-2357<br>Fax:<br>(763) 367-8361<br>E-mail:bruce.j.backlund@medtronic.com |
| Trade Name:<br>Common Name:<br>Classification Name: | Bio-Medicus™ Adult Cannulae and Introducer<br>Cardiopulmonary bypass vascular catheter, cannula, or tubing<br>Catheter, Cannula and Tubing, Vascular, Cardiopulmonary<br>Bypass |
| Classification: | Class II, 21 CFR 870.4210 |
| Product Code: | DWF |
| Name of Predicate Devices: | Medtronic Bio-Medicus Femoral Cannula and Introducer<br>(K884129) |
Avalon Elite Bi-Caval Dual Lumen Catheter (K081820).
## Device Description:
The adult venous and arterial cannulae consist of a polyurethane wirewound body with a multiport distal tip. The adult venous cannula also has a multi-port body built with stainless steel wirewinding and reinforced baskets. All adult arterial cannulae come with a vented 3/8-in vented connector and all of the adult venous come with a non-vented 3/8-in connector. The overall length of the cannulae is 12.5-in for the adult arterial/jugular and 25.6-in, 26.5-in and 30in for the adult venous. The Adult cannulae come with a repositionable suture ring that is radiopaque. Insertion depth marks aid in positioning the cannula.
{5}------------------------------------------------
The one-piece non-phthalate PVC introducers included with these cannulae have an elongated taper for dilation over a longer distance.
The Adult Cannula and Introducer package contains a securement clip which can be used near the insertion site for securement of the cannula body which aides in securement as well as an aide for physician organization. The large venous models (Sizes 23 - 29 Fr) also contain a 3/8-in x ½-in tubing adapter with an attached length of 3/8-in tubing to allow connection to a ½-in cardio pulmonary bypass circuit.
Sterile, nonpyrogenic, single use.
# Intended Use:
These devices are to be used by a trained physician only. Cannulae are used to cannulate vessels, perfuse vessels or organs, and/or connect with accessory extracorporeal equipment. The cannula introducer is intended to facilitate proper insertion and placement of the appropriately-sized cannula within the vessel for cardiopulmonary bypass. The Bio-Medicus™ cannula (18-cm [7.09in] tip length models) may be used in either the femoral position as an arterial delivery cannula or in the jugular position as a venous return cannula. This product is intended for use up to 6 hours.
# Contraindications:
Alone, the cannula and introducer are not medical treatment devices. The cannula introducer is only to be use with the appropriately-sized Bio-Medicus cannula. These devices are not intended for use, except as indicated above. Do not use if the patient has severe peripheral atherosclerosis or severe arterial dissection.
# Comparison to Predicate Devices:
A comparison of the Medtronic Bio-Medicus Adult Arterial and Venous Cannulae to the predicate device indicates the following similarities:
- . Same intended use
- Same technological characteristics
- . Same operating principle
- . Similar design features, exceptions include improved transition between the introducer body and cannula tip to increase ease of cannula insertion, venous models include additional drainage opening along the cannula body, reduction in the overall cannula body length within the non-wirewound section and added a radiopaque suture ring.
- . Similar materials, exceptions include the introducer material which has changed from Polyurethane to Polyvinyl Chloride (PVC), connector material from Polycarbonate to PVC and hemostasis cap material from Styrene to Silicone.
- . Same shelf life
{6}------------------------------------------------
### Summary of Performance Data
Pre-clinical bench testing was used to verify the performance characteristics of this device. Animal testing was also completed to establish substantial equivalence with the predicate devices.
The following performance tests were conducted:
- . Blood Trauma Testing
- . Sterilization Testing
- Biocompatibility Testing .
- Packaging Testing ●
- Pressure Drop Testing .
- . Cannula Life Testing
- Cannula and Introducer Testing ●
- Cannula Testing ●
- Introducer Testing ●
- Securement Clip Testing .
#### Conclusion:
The primary predicate device, Bio-Medicus™ Percutaneous Cannula & Introducer Set, was used to compare the fundamental technological characteristics of the proposed Bio-Medicus Adult Cannula and Introducer. When comparative testing was completed, the proposed Bio-Medicus Adult Cannulae and Introducer performed as well or better than the legally marketed devices.
The data included in this submission is sufficient to provide reasonable assurance that the proposed Bio-Medicus Adult Arterial and Venous Cannulae are substantially equivalent to the legally marketed predicate devices, Bio-Medicus Percutaneous Cannula and Introducer Set (K884129) and Avalon Elite Bi-Caval Dual Lumen Catheter (K081820).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.