K112219 · Thermopeutix, Inc. · MJN · Nov 18, 2011 · Cardiovascular
Device Facts
Record ID
K112219
Device Name
TAPAS CATHETER
Applicant
Thermopeutix, Inc.
Product Code
MJN · Cardiovascular
Decision Date
Nov 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ThermopeutiX TAPAS™ Catheter is intended for general intravascular use in the peripheral vasculature in arteries 1.8 mm and larger. Once placed in the selected region, the catheter can be used for infusion of diagnostic and/or therapeutic agents, and for controlling blood flow to the selected region. The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the manufacturer. The TAPAS™ Catheter is contraindicated for use in the coronary and intracranial arteries. The TAPAS™ Catheter is not intended for embolic protection or as an aspiration catheter.
Device Story
Coaxial catheter with inflow and outflow lumens; features two occlusion balloons for flow control; includes additional lumens for balloon inflation and optional pressure monitoring. Used in peripheral vasculature; operated by clinicians. Enables targeted delivery of diagnostic/therapeutic agents and regional blood flow modulation. Benefits include localized treatment delivery and flow management. Substantially equivalent to DuoFlo catheter; modified by adding second balloon and removing extracorporeal connection.
Clinical Evidence
Bench testing only. Compliance with ISO 10993-1 for biocompatibility. Physical testing included tip/shaft dimensions, balloon compliance, inflation/deflation times, fatigue, burst, bond strengths, shaft kink/torque, and joint strength. All results met specifications.
Technological Characteristics
Coaxial tubular shafts; two occlusion balloons; materials consistent with predicate; ethylene oxide sterilization (SAL 10^-6); ISO 10993-1 compliant. No software or electronic components.
Indications for Use
Indicated for general intravascular use in peripheral vasculature (arteries ≥1.8 mm) for infusion of diagnostic/therapeutic agents and blood flow control. Contraindicated for coronary and intracranial arteries. Not for embolic protection or aspiration.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
K180904 — Sniper Infusion Catheter with Balloon Occlusion · Embolx, Inc. · Jun 8, 2018
K212324 — RenovoCath · Renovorx, Inc. · Aug 23, 2021
K191606 — RenovoCath · Renovorx, Inc. · Aug 7, 2019
K023914 — COAXIA FLOCONTROL CATHETER · Coaxia, Inc. · Feb 21, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K112219/KS
P. 1 of 2
# 510 (k) SUMMARY
NOV 1 8 2011
#### Applicant and Manufacturer
ThermopeutiX, Inc. 9951B Businesspark Avenue San Diego, California 92131 Phone: (858) 549-1760 Fax: (858) 549-1717
#### Contact Person
Thomas Schroeder, Director, RA/QA
#### Device Classification
| Common Names: | Vascular Clamp, Temporary Occlusion Catheter |
|----------------------|----------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Devices of this type are classified as Class II under 21 CFR Part CFR Part<br>870.4450, Vascular Clamp (Product Code MJN). |
| Proprietary Name: | TAPASTM Catheters |
#### Predicate Device
The Predicate device is the ThermopeutiX DuoFlo™ catheter cleared via 510k's K080700 and K093647.
### TAPAS™ Catheter
The ThermopeutiX TAPAS™ Catheter utilizes the identical materials, design and construction methods that are used in the fabrication of DuoFlo™ catheter. These materials and construction methods have met applicable biocompatibility and shelf life requirements. The ThermopeutiX TAPAS™ Catheter has the same intended use as the DuoFlo™ catheter except that the extracorporeal connection option was deleted and a second balloon was added. The same contraindications that apply to the DuoFlo™ catheter apply to the TAPAS™ catheter.
### Indications for Use
The ThermopeutiX TAPAS™ Catheter is intended for general intravascular use in the peripheral vasculature in arteries 1.8 mm and larger. Once placed in the selected region, the catheter can be used for infusion of diagnostic and/or therapeutic agents, and for controlling blood flow to the selected region. The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the manufacturer.
The TAPAS™ Catheter is contraindicated for use in the coronary and intracranial arteries. The TAPAS™ Catheter is not intended for embolic protection or as an aspiration catheter.
{1}------------------------------------------------
### Device Description
Other than deletion of an optional extracorporeal circuit connection and the addition of a second balloon, the TAPAS™ catheter that is the subject of this submission is a substantially equivalent to the predicate ThermopeutiX DuoFlo catheter. The TAPAS™ catheter, as does the DuoFlo™ catheter, consists of two coaxial tubular shafts (inflow and outflow lumens) that can infuse diagnostic and/or therapeutic agents to selected regions, as well as direct arterial blood or other solutions to a specific region or organ. The lumens may be accessed using external devices connected to these lumens. There are three additional smaller lumen is used for inflation of the distal balloon on the inflow lumen. one is for optional pressure monitoring and the third lumen permits inflation of the balloon on the outflow lumen.
## Technological Characteristics Comparison
The TAPAS™ Catheter utilizes the same design, construction, materials and processes as the DuoFlo catheter with no technological differences. The TAPAS™ catheter features smaller shaft sizes and a second occlusion balloon on the outflow lumen (shaft).
### Performance and Safety
The biological safety of the device has been demonstrated through biocompatibility studies of patient contact materials in accordance with the standards outlined in ISO 10993-1. Physical testing was performed and results were in compliance with relevant standards and the product specification. The following non-clinical tests were performed and results were acceptable;
- Tip and shaft dimensions
- Balloon compliance and inflation volume .
- Balloon Inflation and deflation times .
- . Balloon fatigue
- . Balloon burst
- Balloon bond strength ●
- Shaft/lumen joint bond strength ●
- Sheath compatibility
- Shaft kink .
- Shaft torque .
- . Shaft burst pressure
- Tip bond strength .
- Concentric shaft movement .
- . Manifold lumen joint strength
- Stopcock lumen joint strength ♥
The device is supplied sterile and was adopted into an existing validated ethylene oxide sterilization cycle. Sterility conforms to a Sterility Assurance Level (SAL) of 10 °. The supplied instructions for use provide the user with the applicable warnings and cautions during use. There are no new safety or effectiveness issues related to this device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
ThermopeutiX, Inc c/o Mr. Thomas P. Schroeder Director, RA/QA 9951B Businesspark Avenue San Diego, CA 92131
Re: K112219
Trade Name: TAPAS Catheters Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: MJN Dated: October 24. 2011 Received: October 27, 2011
Dear Mr. Schroeder:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
NOV 1 8 2011
{3}------------------------------------------------
Page 2 - Mr. Thomas P. Schroeder
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Bram Zuckerman, M.D.
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K112219
Device Name: TAPAS™ Catheter
Indications for use: The ThermopeutiX TAPAS™ Catheter is intended for general intravascular use in the peripheral vasculature in arteries 1.8 mm and larger. Once placed in the selected region, the catheter can be used for infusion of diagnostic and/or therapeutic agents, and for controlling blood flow to the selected region. The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the manufacturer.
The TAPAS™ Catheter is contraindicated for use in the coronary and intracranial arteries.
The TAPAS™ Catheter is not intended for embolic protection or as an aspiration catheter.
Prescription Use: × (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-counter Use: (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|------------------------------------|
| e 1 of 1 | (Division Signature) |
| | Division of Cardiovascular Devices |
| | 510(k) Number |
Page
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.