K180904 · Embolx, Inc. · MJN · Jun 8, 2018 · Cardiovascular
Device Facts
Record ID
K180904
Device Name
Sniper Infusion Catheter with Balloon Occlusion
Applicant
Embolx, Inc.
Product Code
MJN · Cardiovascular
Decision Date
Jun 8, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Sniper Infusion Catheter with Balloon Occlusion is intended for use in the peripheral vasculature where temporary occlusion is desired and offers vessel selective technique of temporary vascular occlusion for selectively stopping or controlling blood flow. The Sniper Infusion Catheter with Balloon Occlusion is also intended to assist in the delivery of diagnostic agents such as contrast media and therapeutic agents into the peripheral vasculature.
Device Story
Coaxial dual-lumen catheter with distal occlusion balloon; used in peripheral vasculature. Clinician navigates device to target vessel via guidewire through introducer sheath; balloon inflated/deflated via handheld syringe to control blood flow. Radiopaque bands provide fluoroscopic visualization. Once positioned, clinician delivers diagnostic/therapeutic agents (e.g., contrast media, embolization particles) through infusion lumen. After procedure, balloon deflated to restore flow. Single-use device; sterile (ETO). Benefits include vessel-selective flow control and targeted agent delivery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including dimensional verification, balloon fatigue/burst/compliance, bond strength, flexibility/kink, torque, radiopacity, coating integrity, particulate analysis, leakage/burst, flow rate, corrosion, and shelf-life/sterilization testing. Biocompatibility testing per ISO 10993-1 performed.
Technological Characteristics
Coaxial dual-lumen catheter; nylon hub; compliant balloon material; hydrophilic coating. Radiopaque bands for fluoroscopy. Lengths: 110/130/150 cm. Infusion lumen ID: 0.020"; rated OD: 0.038". Max balloon diameter: 6mm. Sterilization: ETO. Standards: ISO 10993-1 (biocompatibility), ISO 10555-1/4 (catheter requirements).
Indications for Use
Indicated for patients requiring temporary vascular occlusion or delivery of diagnostic/therapeutic agents in the peripheral vasculature.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
June 8, 2018
Embolx, Inc % Dave Yungvirt Third Party Review Group, LLC The Old Station House 24 Lackawanna Place Millburn, New Jersey 07041
Re: K180904
Trade/Device Name: Sniper Infusion Catheter with Balloon Occlusion Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular clamp Regulatory Class: Class II Product Code: MJN Dated: May 22, 2018 Received: May 24, 2018
Dear Dave Yungvirt:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{2}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180904
#### Device Name
Sniper Infusion Catheter with Balloon Occlusion
#### Indications for Use (Describe)
Sniper Infusion Catheter with Balloon Occlusion is intended for use in the peripheral vasculature where temporary occlusion is desired and offers vessel selective technique of temporary vascular occlusion for selectively stopping or controlling blood flow. The Sniper Infusion Catheter with Balloon Occlusion is also intended to assist in the delivery of diagnostic agents such as contrast media and therapeutic agents into the peripheral vasculature.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:100%">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for EMBOLX. The logo consists of a spiral of black and red dots on the left, followed by the word "EMBOLX" in a bold, sans-serif font. The dots are arranged in a circular pattern, with the black dots on the outside and the red dots on the inside. The word "EMBOLX" is in all capital letters.
# 510(k) Summary
# I. SUBMITTER
Embolx, Inc. 530 Lakeside Dr, Suite 200 Sunnyvale, CA, 94085 Date Prepared: June 4, 2018
Tel: 408.990.2949 Fax: 408.990.2949
Sponsor Contact: Greg Halstead, Director Engineering and Operations Email: ghalstead@embolx.com
Correspondent Contact Information: John Talarico, Regulatory Consulting 47 Rector Place, Suite B Red Bank, NJ 07701 Tel: 510-371-1143 Email: johntal@att.net
# II. DEVICE
Name of Device: Sniper Infusion Catheter with Balloon Occlusion Common or Usual Name: Catheter, Intravascular Occluding, Temporary Classification Name: Vascular Clamp Regulatory Class: Class II Product Code: MJN
# III. PREDICATE DEVICE
Embolx Occlusion Balloon Catheter, K142692 This predicate has not been subject of a design related recall.
# IV. DEVICE DESCRIPTION
The Sniper" Infusion Catheter with Balloon Occlusion is a coaxial dual lumen catheter with an occlusion balloon located at the distal end to provide temporary complete occlusion or limitation of blood flow in peripheral blood vessels as well as provide a delivery conduit for diagnostic agents such as contrast media and therapeutic agents such as embolization particles into the peripheral vasculature. One lumen is dedicated to balloon inflation and deflation, the other lumen is intended to accommodate the guide wire and delivery of fluids and other diagnostic and therapeutic agents. Device placement in the target location is achieved using a guiding catheter that has been placed through an introducer. A guidewire is threaded through the device's lumen for navigating the device to the selected vessel location. Device visualization is provided by two radiopaque bands using fluoroscopy. One radiopaque band is located on the distal tip of the device, and one on the distal end of the balloon, which is 1 cm proximal to the distal tip. Through the guidewire (perfusion) lumen, the clinician can deliver diagnostic or therapeutic agents into the selected blood vessel. After treatment, the balloon is deflated and blood flow resumes.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for EMBOLX. The logo features a spiral design made up of small circles that transition from black to red. To the right of the spiral is the word "EMBOLX" in a bold, sans-serif font. The overall design is clean and modern.
The proximal hub consists of two ports: one port for use by the guidewire and delivery of fluids and the second port for inflation and deflation of the balloon. The low profile balloon is manufactured of a compliant material that allows ease of insertion and withdrawal from the vasculature and conforms to the vessel wall. The balloon is inflated and deflated with a hand held syringe. The device is supplied sterile by ETO and is intended for single use.
| Attribute | Sniper® Infusion Catheter with Balloon Occlusion |
|----------------------|--------------------------------------------------|
| Infusion Lumen | 0.020" ID |
| Infusion Pressure | Up to 900psi |
| Functional Length | 110 cm or 130 cm or 150 cm |
| Rated Outer Diameter | 0.038" |
| Shelf life | 1 year |
| Balloon diameter | Up to 6mm |
The following accessories are included in the sterile package with the device:
- Balloon aspiration syringe (10mL)
- . Balloon inflation syringe (0.25mL)
- 1-way valve
# V. INDICATIONS FOR USE
Sniper" Infusion Catheter with Balloon Occlusion is intended for use in the blood vessels of the peripheral vasculature where temporary occlusion is desired and offers vessel selective technique of temporary vascular occlusion for selectively stopping or controlling blood flow. The Sniper" Infusion Catheter with Balloon Occlusion is also intended to assist in the delivery of diagnostic agents such as contrast media and therapeutic agents into the peripheral vasculature.
The indications for use statement for the Sniper Infusion Catheter with Balloon Occlusion is identical to the predicate device. Both devices have the same intended use in the blood vessels of the peripheral vasculature for temporary vascular occlusion and delivery of diagnostic agents.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject and the predicate devices are based on the same technological principles for providing temporary complete vascular occlusion and delivery of therapeutic and diagnostic agents in the peripheral vasculature. Both devices have the following identical technological characteristics:
- Guidewire guided catheter advanced through an introducer sheath in the peripheral vasculature.
- Dual lumen coaxial design with proximal hub consisting of two ports; one port for use by the guidewire and delivery of fluids and the second port for inflation and deflation of the balloon.
- Occlusion balloon at the distal end with two embedded radiopaque bands to allow for visualization under fluoroscopic guidance.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for EMBOLX. The logo consists of a spiral of red and black dots on the left, followed by the word "EMBOLX" in a bold, sans-serif font. The dots are arranged in a circular pattern, with the red dots concentrated towards the center of the spiral.
- Therapeutic and diagnostic agents can be advanced through the lumen that holds the ● guidewire.
The following technological differences exist between the subject and predicate devices:
- . The inflation and burst volume of the occlusion balloon has increased to accommodate slightly larger vessel sizes.
- . Strain relief is longer to provide added flexibility and protection to the proximal aspect of the catheter.
- . Additional functional lengths (130cm and 150cm) have been added to accommodate current clinical practices.
- Hub material changed to nylon to improve chemical stability over long-term use.
- The occlusion balloon material has been changed to provide higher tensile strength and elongation percentage.
- Hydrophilic coating has been changed to improve lubricity.
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing
The biocompatibility evaluation for the Sniper Infusion Catheter with Balloon Occlusion demonstrates that the device is in compliance with ISO 10993-1:2009 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- ISO Intracutaneous Study - Extract
- ISO Systemic Toxicity
- Sensitization
- Pyrogen Study - Material Mediated
- ASTM Hemolysis
- SC5b-9 Complement Activation Assay
- In Vivo Thromboresistance Study - Jugular Vein
The Sniper" Infusion Catheter with Balloon Occlusion is considered tissue contacting for a duration of less than 24 hours.
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing does not apply to the subject device.
#### Software Verification and Validation Testing
Software verification and validation testing does not apply to the subject device.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for EMBOLX. The logo consists of a spiral of black and red dots on the left, followed by the word "EMBOLX" in a bold, sans-serif font. The dots are arranged in a circular pattern, with the black dots on the outside and the red dots on the inside. The word "EMBOLX" is in all caps and is slightly larger than the dots.
# Performance Testing
The performance test evaluation for the Sniper Infusion Catheter with Balloon Occlusion demonstrates that the device is in compliance with ISO 10555-1:2013 "Sterile, single-use intravascular catheters - Part 1: General requirements" and ISO 10555-4:2013 "Intravascular catheters - Sterile and single-use catheters - Part 4: Balloon dilatation catheters." The battery of testing included the following tests:
- Dimensional Verification
- Balloon Rated Burst Volume
- Balloon Fatigue (Repeat Balloon Inflations)
- Balloon Compliance (Diameter v. Volume)
- Balloon Preparation, Deployment and Retraction (Time/Position)
- Catheter Bond Strength and Tip Pull Test
- Flexibility and Kink Test (Bend/Buckle)
- Torque Strength Test (Torsional stress)
- Radiopacity
- . Coating Integrity
- Particulate Matter Analysis
- Catheter Body Leakage & Burst Test
- Power Injection and Contrast Media Flow Rate
- Corrosion Resistance
- . Lipiodol Compatibility
- Shipping
- Shelf-life & Accelerated Aging
- Sterilization
The non-clinical data support the safety of the device and demonstrate that the Sniper Infusion Catheter with Balloon Occlusion performs comparably to the predicate device that is currently marketed for the same intended use.
# Animal Study
No animal testing was required to support this 510(k) as the indications for use and technology are equivalent to those of the predicate device.
# Clinical Studies
No clinical testing was required to support this 510(k) as the indications for use and technology are equivalent to those of the predicate device.
# VIII. CONCLUSIONS
The Sniper Infusion Catheter with Balloon Occlusion was found to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.