K153259 · Forge Medical, Inc. · DXC · Jan 6, 2016 · Cardiovascular
Device Facts
Record ID
K153259
Device Name
VasoStat Hemostasis Device
Applicant
Forge Medical, Inc.
Product Code
DXC · Cardiovascular
Decision Date
Jan 6, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K153259 · Jan 6, 2016
VasoStat Hemostasis Device
Forge Medical, Inc.
Retrospective clinical patient cohort
Retrospective analysis of 314 patients to evaluate the safety and effectiveness of the device for achieving hemostasis following transpedal catheterization.
Hemostasis achievement, rates of pseudo-aneurysm, and access site occlusion
Indications for Use
The VasoStat™ Hemostasis Device is indicated for use by medical professionals to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's arm or lower leg, including radial artery catheterization, pedal or tibial artery catheterization, arterial or venous line removal, hemodialysis, and in patients on anticoagulation therapy.
Device Story
VasoStat™ Hemostasis Device is a vascular compression device used by medical professionals to achieve hemostasis post-catheterization or vascular puncture. The device consists of a flexible thermoplastic base secured to the patient's skin via biocompatible adhesive pads. A plunger component, operated via a one-handed ratcheting mechanism, applies adjustable, targeted pressure to the puncture site. This compression promotes hemostasis while minimizing thrombosis risk. Used in clinical settings, the device is applied directly over the puncture site on the arm or lower leg. Once hemostasis is achieved (typically 15-45 minutes), the device is removed and discarded. The device benefits patients by providing controlled, adjustable pressure to stop bleeding, particularly in patients on anticoagulation therapy.
Clinical Evidence
Retrospective evaluation of 314 patients following transpedal access for peripheral vascular disease intervention. Results demonstrated the device successfully achieved hemostasis with low rates of pseudo-aneurysm and access site occlusion. Benchtop and animal testing from K123041 were leveraged to support the expanded indications.
Technological Characteristics
Flexible thermoplastic base; biocompatible adhesive pads; ratcheting plunger mechanism for adjustable compression. Materials comply with ISO 10993 biocompatibility standards. Standalone, non-powered, single-use device.
Indications for Use
Indicated for patients undergoing catheterization or vascular puncture in the arm or lower leg, including radial, pedal, or tibial artery access, arterial/venous line removal, and hemodialysis. Suitable for patients on anticoagulation therapy.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a stylized caduceus symbol. The caduceus is represented by a triple-stranded helix, resembling a stylized bird in flight. The text reads 'DEPARTMENT OF HEALTH & HUMAN SERVICES - USA'.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 06, 2016
Forge Medical, Inc. c/o Mason Diamond, DDS Texel Fortis, LLC 150 Levinberg Lane Wayne, NJ 07470
Re: K153259
Trade/Device Name: VasoStat™ Hemostasis Device Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: DXC Dated: November 7, 2015 Received: November 10, 2015
Dear Dr. Diamond:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address:
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
M.A. Hillebrand
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153259
Device Name VasoStat™ Hemostasis Device
#### Indications for Use (Describe)
The VasoStat™ Hemostasis Device is indicated for use by medical professionals to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's arm or lower leg, including radial artery catheterization, pedal or tibial artery catheterization, arterial or venous line removal, hemodialysis, and in patients on anticoagulation therapy.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------|
| <div> <span> X Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
| Submitter Name: | Forge Medical, Inc |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 791 West Broad Street, Suite 102<br>Bethlehem, PA 18018 |
| Phone Number:<br>Fax Number: | 215-605-3225<br>866-244-8988 |
| Contact Person: | Mason Diamond, DDS<br>Texel Fortis, LLC<br>150 Levinberg Lane<br>Wayne, NJ 07470<br>Phone: 508-333-0108<br>Fax: 973-305-0213 |
| Date Prepared: | 10 December, 2015 |
| Device Trade Name: | VasoStatTM Hemostasis Device |
| Common Name | Vascular compression device |
| Classification Name<br>Number and<br>Product Code: | Vascular clamp<br>870.4450<br>DXC |
| Predicate Devices: | VasoStatTM Hemostasis Device (Forge Medical): K123041<br>RadARTM Vascular Compression Assist Devices (Advanced<br>Vascular Dynamics): K142122 |
| Device Description | VasoStatTM Hemostasis Device is a device that is adhered to the<br>patient's skin directly over a percutaneous vascular puncture site<br>and allows the application of adjustable compression to the<br>puncture site to achieve hemostasis. The device is composed of<br>a flexible thermoplastic base that has the ability to conform to the<br>patient's anatomy (e.g. lower arm, upper arm, foot) and is secured<br>to the skin with biocompatible adhesive pads. A plunger<br>component passes through the base, applies targeted pressure to<br>the puncture site through a ratcheting mechanism, and can be<br>adjusted with one hand to apply the appropriate pressure<br>necessary to promote hemostasis. VasoStat has the ability to vary<br>the amount of pressure applied to the puncture site to promote<br>hemostasis while minimizing the risk of thrombosis. The device |
| Submitter:<br>Forge Medical, Inc. | VasoStat™ Hemostasis Device<br>Premarket Notification: Traditional 510(k) |
| | design and materials ensure that the product is comfortable for<br>the patient to wear while their puncture site stops bleeding. Once<br>hemostasis is achieved (in approximately 15 to 45 minutes), the<br>device is easily removed and discarded. |
| Statement of<br>Indications for Use | The VasoStat™ Hemostasis Device is indicated for use by<br>medical professionals to promote hemostasis following a<br>catheterization or other puncture into a blood vessel in a patient's<br>arm or lower leg, including radial artery catheterization, pedal or<br>tibial artery catheterization, arterial or venous line removal,<br>hemodialysis, and in patients on anticoagulation therapy. |
| Summary of<br>Technological<br>Characteristics | The VasoStat™ Hemostasis Device is designed to provide local<br>compression to the puncture site into a blood vessel of the arm or<br>lower leg, including arterial sheath placement and dialysis shunt<br>access. The degree of pressure is designed to stop bleeding<br>without causing vessel thrombosis. The VasoStat™ Hemostasis<br>Device is the same as that previously cleared by the FDA<br>(K123041), and functions in the same way. Though different in<br>design, the VasoStat™ Hemostasis Device provides sufficient<br>compression to stop bleeding in pedal arterial catheterizations<br>similar to the RadAR™ Vascular Compression Assist Devices,<br>which also utilize compression to stop bleeding (K142122). |
| Performance<br>Testing | Since this submission applies to the VasoStat™ Hemostasis<br>Device to revise the indications for use without any changes to<br>device design, the benchtop and animal testing reported in<br>K123041 has been leveraged to demonstrate that the submitted<br>device is appropriate for the new indications for use. Similarly,<br>biocompatibility for this device has already been established per<br>ISO-10993 (K123041).<br><br>Furthermore, the VasoStat™ Hemostasis Device was<br>retrospectively evaluated in three-hundred fourteen (314) patients<br>following transpedal access for peripheral vascular disease<br>intervention. The analysis concluded that the VasoStat™<br>Hemostasis Device was capable of achieving hemostasis<br>following transpedal catheterization with low rates of pseudo-<br>aneurysm and access site occlusion. |
| Substantial<br>Equivalence | The VasoStat™ Hemostasis Device is Substantially Equivalent to<br>predicate VasoStat™ Hemostasis Device (K123041), with respect<br>to technical and design features. The Indications for Use and<br>Mode of Action is Substantively Equivalent to the RadAR™<br>Vascular Compression Assist Devices (K142122). |
| Conclusion | The information discussed above demonstrates that the<br>VasoStat™ Hemostasis Device is substantially equivalent to the<br>predicate device. |
{4}------------------------------------------------
{5}------------------------------------------------
# Summary of Substantial Equivalence
| Feature | VasoStat™ Hemostasis<br>Device | VasoStat™ Hemostasis<br>Device | RadAR™ Vascular<br>Compression Assist<br>Devices |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K153259 | K123041 | K142122 |
| Class | II | II | II |
| Classification | Vascular clamp | Vascular clamp | Vascular clamp |
| Regulation | 870.4450 | 870.4450 | 870.4450 |
| Product Code(s) | DXC | DXC | DXC |
| Indications for<br>Use | ...indicated for use by<br>medical professionals to<br>promote hemostasis<br>following a catheterization<br>or other puncture into a<br>blood vessel in a patient's<br>arm or lower leg, including<br>radial artery<br>catheterization, pedal or<br>tibial artery catheterization,<br>arterial or venous line<br>removal, hemodialysis, and<br>in patients on<br>anticoagulation therapy. | ...indicated for use by<br>medical professionals to<br>promote hemostasis<br>following a catheterization<br>or other puncture into a<br>blood vessel in a patient's<br>arm, including radial artery<br>catheterization, arterial or<br>venous line removal,<br>hemodialysis, and in<br>patients on anticoagulation<br>therapy. | ...indicated for use by<br>medical professionals to<br>promote hemostasis<br>following a catheterization<br>or other puncture into a<br>blood vessel in a patient's<br>arm or leg, including radial<br>brachial, dorsalis, pedis, or<br>tibial blood vessels, arterial<br>or venous line or sheath<br>removal, hemodialysis, and<br>in patients on<br>anticoagulation therapy. |
| Mode of Action | Compression | Compression | Compression |
| Composition | Same a predicate | Molded<br>Polypropylene/Polyethylen<br>e Foam/Acrylic Adhesive | Not available |
| Biocompatibility | Established biocompatible<br>materials | ISO 10993 - Part 5<br>ISO 10993 - Part 10 | Established biocompatible<br>materials |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.