K123041 · Forge Medical, Inc. · DXC · May 29, 2013 · Cardiovascular
Device Facts
Record ID
K123041
Device Name
VASOSTAT HEMOSTASIS DEVICE
Applicant
Forge Medical, Inc.
Product Code
DXC · Cardiovascular
Decision Date
May 29, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VasoStat™ Hemostasis Device is indicated for use by medical professionals to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's arm, including radial artery catheterization, arterial or venous line removal, hemodialysis, and in patients on anticoagulation therapy.
Device Story
VasoStat™ Hemostasis Device is an ergonomic vascular compression device used by medical professionals to achieve hemostasis at percutaneous vascular puncture sites in the arm. The device consists of a flexible thermoplastic base that conforms to patient anatomy, secured with biocompatible, hypoallergenic adhesive pads. A plunger component, operated via a one-handed ratcheting mechanism, applies adjustable, targeted pressure to the puncture site. By varying pressure, the device promotes hemostasis while minimizing thrombosis risk. Once hemostasis is achieved (typically 15-45 minutes), the device is removed and discarded. It benefits patients by providing a comfortable, adjustable method for stopping bleeding post-catheterization or dialysis.
Clinical Evidence
In vivo assessment conducted in a domestic swine model comparing VasoStat™ to the Terumo TR Band. Vascular sheaths were placed in femoral arteries/veins and jugular veins. Both devices achieved hemostasis in an average of 5.8 minutes. Biocompatibility testing per ISO 10993 (Cytotoxicity, Intracutaneous Reactivity, and Buehler Sensitization) confirmed the device is non-cytotoxic, nonreactive, and non-sensitizing.
Technological Characteristics
Molded polypropylene/polyethylene foam base with acrylic adhesive. Mechanical ratcheting plunger for targeted compression. Conforms to arm anatomy. Biocompatible per ISO 10993-5 and ISO 10993-10. Non-powered, manual operation.
Indications for Use
Indicated for medical professionals to promote hemostasis in patients following arm blood vessel puncture, including radial artery catheterization, arterial/venous line removal, hemodialysis, and patients on anticoagulation therapy.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
{0}------------------------------------------------
Submitter: Forge Medical, Inc.
VasoStat™ Hemostasis Device Premarket Notification: Traditional 510(k)
# MAY 2 9 2013
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
K123041
### 510(k) Summary
| Submitter Name: | Forge Medical, Inc |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 8516 Parkwood Lane<br>Philadelphia, PA 19128 |
| Phone Number: | 877-466-0109 |
| Fax Number: | 866-244-8988 |
| Contact Person: | Mason Diamond, DDS<br>Texel Fortis, LLC<br>150 Levinberg Lane<br>Wayne, NJ 07470<br>Phone: 508-333-0108<br>Fax: 973-305-0213 |
| Date Prepared: | 30 April, 2013 |
| Device Trade Name: | VasoStat™ Hemostasis Device |
| Common Name | Vascular compression device |
| Classification Name | Vascular clamp |
| Number and<br>Product Code: | 870.4450<br>DXC |
| Predicate Devices: | RaDAR Vascular Compression Device (Advanced Vascular<br>Dynamics) (DXC): K092503<br>TR Band (Terumo) (DXC): K070423<br>VasoStat™ Hemostasis Device (Forge Medical) (NOO): Device<br>Listing (D147574) |
| Device Description | The VasoStat™ Hemostasis Device is an ergonomic, simple<br>device that is adhered to the patient's skin directly over a<br>percutaneous vascular puncture site and allows the application of<br>adjustable compression to the puncture site to achieve<br>hemostasis. The device is composed of a flexible thermoplastic |
base that has the ability to conform to the patient's anatomy (e.g.
### Section 5.0: 510(k) Summary
i
{1}------------------------------------------------
lower arm, upper arm, etc.) and is secured to the skin with biocompatible, hypoallergenic adhesive pads. A plunger component passes through the base, applies targeted pressure to the puncture site through a ratcheting mechanism, and can be adjusted with one hand to apply the appropriate pressure necessary to promote hemostasis. VasoStat has the ability to vary the amount of pressure applied to the puncture site to promote hemostasis while minimizing the risk of thrombosis. The device design and materials ensure that the product is comfortable for the patient to wear while their puncture site stops bleeding. Once hemostasis is achieved (in approximately 15 to 45 minutes), the device is easily removed and discarded.
Statement of Indication for Use The VasoStat™ Hemostasis Device is indicated for use by medical professionals to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's arm, including radial artery catheterization, arterial or venous line removal, hemodialysis, and in patients on anticoagulation therapy.
Summary of Technological Characteristics The VasoStat™ Hemostasis Device is designed to provide local compression to the puncture sites of a dialysis access shunt of a dialysis patient. The device applies uniform pressure to the sites of needle puncture along the dialysis access shunt until hemostasis is achieved. The degree of pressure is designed to stop bleeding without causing blood clots.
A domestic swine model was utilized for in vivo assessment of the VasoStat™ Hemostasis Device in comparison to the Terumo TR Band (predicate device). Vascular sheaths were created in the femoral arteries and veins, and in the jugular veins. The sheaths were removed and either a VasoStat™ Hemostasis Device or a Terumo TR Band placed. Both devices achieved hemostasis in the approximately 5.8 minutes (average). These results demonstrate that the VasoStat™ Hemostasis Device and the predicate device are substantially equivalent.
Biocompatibility testing (per ISO 10993) performed on the device demonstrated, that the VasoStat Hemostasis Device is safe for the indications of use, and was found to be non-cytotoxic, nonreactive and non-sensitizing by ISO Cytotoxicity, ISO Intracutaneous Reactivity and ISO Buehler Sensitization testing, respectively.
The VasoStat™ Hemostasis Device is substantially equivalent to the predicates. RaDAR Vascular Compression Device (K092503), the TR Band (K070423), and the VasoStat™ Hemostasis Device (D147574), with respect to technical and design features. The submitted device poses no new questions about safety or effectiveness as compared to the predicate devices.
Substantial Equivalence
{2}------------------------------------------------
| Submitter:<br>Forge Medical, Inc. | VasoStat™ Hemostasis Device<br>Premarket Notification: Traditional 510(k) |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | The information discussed above and presented in the following<br>table demonstrates that the VasoStat™ Hemostasis Device is<br>substantially equivalent to the predicate devices. |
| Declarations | • This summary includes only information that is also covered in<br>the body of the 510(k).<br>• This summary does not contain any unsubstantiated labeling<br>claims.<br>• This summary does not contain any raw data, i.e., contains<br>only summary data. |
o This summary does not contain any trade secret or confidential commercial, information.
o This summary does not contain any patient identification information.
{3}------------------------------------------------
Submitter:
Forge Medical, Inc.
# ummary of Substantial Equivalen
| Feature | VasoStat™ Hemostasis Device | RaDAR Vascular Compression Device | TR Band | VasoStat™ Hemostasis Device |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K123041 | K092503 | K070423 | D147574 |
| Class | II | II | II | I (Exempt) |
| Classification | Vascular clamp | Vascular clamp | Vascular clamp | Clamp, Dialysis arm |
| Regulation | 870.4450 | 870.4450 | 870.4450 | 876.5540 |
| Product Code(s) | DXC | DXC | DXC | NOO |
| Indications for Use | ...indicated for use by medical professionals to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's arm, including radial artery catheterization, arterial or venous line removal, hemodialysis, and in patients on anticoagulation therapy. | ...indicated for use by medical professionals to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's arm, including radial artery catheterization, arterial or venous line removal, hemodialysis, and in patients on anticoagulation therapy. | ...is a compression device to assist hemostasis of the radial artery after a transradial procedure. | ...to temporarily apply puncture site compression and, aid in obtaining puncture site hemostasis, following the removal of dialysis needles. |
| Mode of Action | Compression | Compression | Compression | Compression |
| In Vivo Performance | Hemostatic | Not Tested | Hemostatic | Hemostatic |
| Composition | Molded<br>Polypropylene/Polyethylene Foam/Acrylic Adhesive | Not Available | Plastic | Molded<br>Polypropylene/Polyethylene Foam/Acrylic Adhesive |
| Biocompatibility | ISO 10993- Part 5<br>ISO 10993- Part 10 | Established biocompatible materials | ISO 10993- Part 5<br>ISO 10993- Part 10 | ISO 10993- Part 5<br>ISO 10993 - Part 10 |
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 29, 2013
Forge Medical, Inc! c/o Mr. Mason Diamond 150 Levinberg Lane Wayne, NJ 07470
Re: K123041
Trade/Device Name: VasoStat Hemostasis Device Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: DXC Dated: April 10, 2013 Received: April 12, 2013
Dear Mr. Mason Diamond:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in-interstatecommerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Mr. Mason Diamond
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Bram D. Zackerman -S
| Bram D. Zuckerman, M.D. | |
|------------------------------------|--|
| Director | |
| Division of Cardiovascular Devices | |
| Office of Device Evaluation | |
| Center for Devices and | |
| Radiological Health | |
Enclosure
{6}------------------------------------------------
Submitter: Forge Medical, Inc.
VasoStat™ Hemostasis Device Premarket Notification: Traditional 510(k)
## Indications for Use
510(k) Number (if known): K123041
Device Name: VasoStat™ Hemostasis Device
Indications for Use:
The VasoStat™ Hemostasis Device is indicated for use by medical professionals to promote hemostasis following a catheterization or other puncture into a blood vessel in a patient's arm, including radial artery catheterization, arterial or venous line removal, hemodialysis, and in patients on anticoagulation therapy.
Prescription Use X (Part 21 CFR 801 Subpart D) · AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE-DO NOT-WRITE-BELOW-THIS-LINE-CONTINUE-ON-ANOTHER-PAGE-IF NEEDED)-
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
1
Image /page/6/Picture/13 description: The image shows the text "Bram D. Zuckerman -S" on the first line. The second line shows the date and time "2013.05.29 19:10:41". The third line shows the timezone offset "-04'00'". There is also a logo in the image, but it is difficult to make out the details.
Section 4.0
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.