K130382 · Boston Scientific Corporation · DXC · Jun 5, 2013 · Cardiovascular
Device Facts
Record ID
K130382
Device Name
HEARTSTRING III PROXIMAL SEAL SYSTEM
Applicant
Boston Scientific Corporation
Product Code
DXC · Cardiovascular
Decision Date
Jun 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HEARTSTRING III Proximal Seal System is intended for use by Physicians during CABG procedures to maintain hemostasis and to facilitate the completion of a proximal anastomosis to the aorta without application of an aortic clamp.
Device Story
System enables proximal anastomosis of vessel graft to aorta without aortic clamp during CABG surgery. Components include Proximal Seal, Delivery Device, Loading device, and Aortic Cutter. Aortic Cutter creates aortotomy; Loading device rolls and loads Seal into Delivery Device; Delivery Device (syringe-like) deploys Seal inside aorta to provide sealed region for anastomosis. Used in OR by physicians. Output is a sealed aortic site facilitating graft attachment. Benefits include avoidance of aortic clamping, potentially reducing trauma to aorta.
Clinical Evidence
Bench testing only. In-vitro performance and biocompatibility testing demonstrated functionality and performance characteristics comparable to the predicate device.
Technological Characteristics
System comprises a seal, delivery device, loading device, and aortic cutter. Mechanical operation; no energy source. Sterilized using established processes. Modifications include added bridge on loading device and visual cues. No software or electronic components.
Indications for Use
Indicated for use by physicians during coronary artery bypass graft (CABG) surgery to maintain hemostasis and facilitate proximal anastomosis to the aorta without an aortic clamp.
Regulatory Classification
Identification
A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.
{0}------------------------------------------------
# HEARTSTRING III Proximal Seal System 510(k) Summary
Prepared in accordance with 21 CFR Part 807.92
. -
| 510(k) Number: | K130382 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | 6 March 2013 |
| Device Owner: | MAQUET Cardiovascular LLC<br>45 Barbour Pond Drive<br>Wayne, New Jersey 07470 |
| Contact Personnel: | Mark Dinger |
| Title: | Regulatory Affairs Specialist II |
| Email: | mark.dinger@maquet.com |
| Phone: | 973-709-7691 Fax: 973-807-1658 |
| Trade Name: | HEARTSTRING III Proximal Seal System: |
| Device Generic Name: | Vascular Clamp |
| Classification: | According to 21 CFR 870.4450 of the Federal Food, Drug<br>and Cosmetic Act, the device classification is Class II,<br>Product code DXC. |
| Predicate Device: | K080169 HEARTSTRING III Proximal Seal System<br>(SE: 8 Feb 2008) |
| Device Description: | The HEARTSTRING III Proximal Seal System is a device<br>designed to enable the creation of a proximal anastomosis<br>of a vessel graft to the aorta without the need for an aortic<br>clamp during coronary artery bypass graft (CABG)<br>surgery.<br><br>The HEARTSTRING III Proximal Seal System is<br>comprised of the Proximal Seal, Delivery Device, Loading<br>device and Aortic Cutter. The HEARTSTRING III Seal is<br>a device that is delivered into the aorta via an aortotomy<br>created by the Aortic Cutter and provides a sealed region to<br>facilitate the proximal anastomosis. The Delivery Device is<br>a syringe-like tube with plunger that is used to deploy the<br>HEARTSTRING III Seal inside of the aorta. The Loading |
{1}------------------------------------------------
Device is a mechanism that rolls the HEARTSTRING III Seal and loads the HEARTSTRING III Seal into the Delivery Device. The Aortic Cutter is a single use (one aortotomy) device that consists of a Grip, a Cutter, Aortic Stop, a Cap, a Needle, a Safety Lock and an Actuation Button. It is used to create the aortotomy for the anastomosis.
#### Indications for Use:
#### Technological Characteristics
The HEARTSTRING III Proximal Seal System is intended for use by Physicians during CABG procedures to maintain hemostasis and to facilitate the completion of a proximal anastomosis to the aorta without application of an aortic clamp.
The Proposed HEARTSTRING III Proximal Seal System and the predicate devices have the following similarities:
- the same but clarified intended use, ಂ
- the same operating principles, o
- ం incorporates the same seal device.
- sterilized using the same materials and processes. 0
- has same packaging. O
The Proposed HEARTSTRING III Proximal Seal System and the predicate devices have the following differences:
- Addition of a bridge on the loading device. o
- Addition of visual cues on the loading and delivery 0 device.
- Change branding colors on loading and delivery o device.
This difference is not considered a technological difference and is substantially equivalent to the predicate devices.
MAQUET Cardiovascular's development process required that the following activities be completed during the development of the Proposed HEARTSTRING III Proximal Seal System:
- . Performance testing
- . Biocompatibility Testing
The results of the in-vitro tests conducted demonstrate that the functionality and performance characteristics of the device are comparable to the currently marketed HEARTSTRING III Proximal Seal System.
#### Safety and Performance:
{2}------------------------------------------------
#### Conclusion:
Based upon the information submitted in this Traditional 510(k) premarket notification, MAQUET's HEARTSTRING III Proximal Seal System is substantially equivalent to the currently marketed HEARTSTRING III Proximal Seal System. The HEARTSTRING III Proximal Seal System is similar to the predicate devices in the intended use and the fundamental scientific technology of the device. The performance testing established that the proposed HEARTSTRING III Proximal Seal System is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 5, 2013
Maquet Cardiovascular LLC c/o Mr. Mark Dinger Regulatory Affairs Specialist. II 45 Barbour Pond Drive Wayne, NJ 07470
Re: K130382
Trade/Device Name: Heartstring III Proximal Seal System Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular clamp Regulatory Class: Class II (two) Product Code: DXC Dated: March 6, 2013 Received: March 7, 2013
Dear Mr. Dinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications, for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Mark Dinger
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
| | Matthew-G.-Hillebrenner |
|-----|--------------------------------------------|
| for | Bram D. Zuckerman, M.D. |
| | Director |
| | Division of Cardiovascular Devices |
| | Office of Device Evaluation |
| | Center for Devices and Radiological Health |
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K130382
Device Name: HEARTSTRING III Proximal Seal System
Indications For Use:
The HEARTSTRING III Proximal Seal System is intended for use by Physicians during CABG procedures to maintain hemostasis and to facilitate the completion of a proximal anastomosis to the aorta without application of an aortic clamp.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR:
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of
Matthew G. Hillebrenner
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.