The Cardiopulmonary Bypass Pump Tubing with X-Coating is intended to provide a conduit for extracorporeal blood flow through a roller pump during cardiopulmonary bypass procedures. The tubing is intended for use in procedures lasting up to 6-hours in duration.
Device Story
Pump tubing serves as conduit for extracorporeal blood flow in cardiopulmonary bypass circuits. Device consists of PVC tubing coated with biocompatible polymer (X-Coating) to reduce platelet adhesion. Used in clinical settings (e.g., OR) by perfusionists or surgeons; tubing is installed in roller pump head where it undergoes cyclic compression to drive blood flow. Coating benefits patient by minimizing thrombus formation during bypass procedures lasting up to 6 hours. Device is sterile and single-use.
Clinical Evidence
Bench testing only. Conducted dimensional/visual analysis, leakage/mechanical integrity, pull force, and 6-hour circulation testing (flow rate, durability/spallation, visual thrombus formation, hemolysis). Biocompatibility testing performed per ISO 10993. In-vivo animal study conducted on coating material. Results demonstrated performance equivalent to uncoated predicate.
Technological Characteristics
Material: Polyvinyl chloride (PVC) with biocompatible polymer (X-Coating). Dimensions: Various I.D./O.D. configurations. Principle: Cyclic compression by roller pump. Sterilization: Ethylene Oxide (SAL 10^-6). Compliance: ISO 10993 for biocompatibility.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass procedures requiring extracorporeal blood flow via a roller pump for up to 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass pump tubing is polymeric tubing which is used in the blood pump head and which is cyclically compressed by the pump to cause the blood to flow through the cardiopulmonary bypass circuit.
K012538 — TUBING AND CONNECTORS WITH TRILLIUM BIOPASSIVE SURFACE · Medtronic Vascular · Aug 23, 2001
K113845 — TUBING, CONMNECTORS, AND ACCESSORIES WIT BALANCE BIOSURFACE · Medtronic, Inc. · Jan 25, 2012
K013578 — PUMP TUBING · Terumo Cardiovascular Systems Corp. · Jan 22, 2002
K260043 — Smart Reactors Camouflage Coated Extracorporeal Cardiopulmonary Tubing Set (SR180C) · Smart Reactors · Feb 23, 2026
K030077 — GISH TUBING AND CONNECTORS WITH GBS COATING · Gish Biomedical, Inc. · Mar 14, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
1 022857
# SECTION II 510(k) SUMMARY AND CERTIFICATION
#### 510(k) Summary
| Submitter Information | 13 |
|----------------------------------------|----|
| Device Names | 13 |
| Predicate Device | 13 |
| Intended Use | 13 |
| Principles of Operation and Technology | 13 |
| Design and Materials | 14 |
| Performance Evaluations | 14 |
| Substantial Equivalence Comparison | 14 |
| Substantial Equivalence Summary | 15 |
| Additional Safety Information | 15 |
| Conclusion | 15 |
{1}------------------------------------------------
## Pump Tubing with X-Coating
#### Submitter Information:
Name and Address: Terumo Cardiovascular Systems Corporation 28 Howe Street Ashland, MA 01721
Contact Person: Kazuhito Inoue Terumo Cardiovascular Systems Corporation 125 Blue Ball Road Elkton, MD 21921 Telephone: 1-800-283-7866, Ext. 7001
Date of Preparation: August 21, 2000
#### Device Names:
Proprietary Name: Pump Tubing with X-Coating Common Name: Pump Tubing Classification Name: Tubing, Pump, Cardiopulmonary Bypass
#### Predicate Device:
The Pump Tubing with X-Coating that is the subject of this premarket notification is substantially equivalent in intended use, materials, design, technology and principles of operation, and performance to the uncoated Pump Tubing (K993189).
#### Intended Use:
The Cardiopulmonary Bypass Pump Tubing with X-Coating is intended to provide a conduit for extracorporeal blood flow through a roller pump during cardiopulmonary bypass procedures.
#### Principles of Operation and Technology:
The pump tubing that is the subject of this premarket notification is used in a pump head and becomes cyclically compressed by the pump to cause the blood to flow through the bypass circuit.
{2}------------------------------------------------
## Design and Materials:
Each size of the Pump Tubing with X-Coating that is included in this submission is comprised of polyvinyl chloride (PVC).
| Tubing<br>Identification | Pump Tubing with X-Coating | | | | |
|--------------------------|----------------------------|-------|-------|--------|----|
| | End | | Mid | Durom. | |
| | I.D. | O.D. | I.D. | O.D. | |
| Tubing number 1 | .25" | .375" | .375" | .50" | 68 |
| Tubing number 2 | .375" | .561" | .50" | .686" | 68 |
The tubing size transitions in the raceway section. The specifications are as follows:
Each tubing size is coated with a biocompatible polymer coating (X-Coating) that is intended to reduce the adhesion of platelets to the internal surface of the tubes as blood flows through the circuit.
## Performance Evaluations:
The performance of the Pump Tubing with X-Coating submitted in this premarket notification is substantially equivalent to the performance of the uncoated pump tubing. The following tests were conducted to demonstrate equivalence in performance:
- Dimensional/Visual Analysis ●
- Leakage and Mechanical Integrity Testing ●
- Pull Force Testing
- 6-hour Circulation Testing .
- Flow Rate .
- Durability/Spallation .
- Thrombus Formation (Visual) .
- Effects Upon Cellular Components (Hemolysis)
## Substantial Equivalence Comparison:
The Pump Tubing with X-Coating is substantially equivalent to the uncoated pump tubing as follows:
- . Intended Use: Both the Pump Tubing with X-Coating and the uncoated pump tubing are intended to provide a conduit for extracorporeal blood flow through a roller pump during cardiopulmonary procedures.
- . Principles of Operation and Technology: The Pump Tubing with X-Coating and the uncoated pump tubing each utilize the same technologies in the operation of the devices. They are each used in a pump head and become cyclically compressed by the pump rollers to cause the blood to flow through the bypass circuit.
{3}------------------------------------------------
- Design and Materials: The design and the materials of the Pump Tubing with X-. Coating and the uncoated tubing are exactly the same with the exception of the X-Coating polymer that is applied to the coated tubing.
- Performance: Comparisons of the performance of the Pump Tubing with X-Coating . and the uncoated pump tubing were conducted. The comparisons demonstrated that there were no clinically significant performance differences between the two devices.
## Substantial Equivalence Summary:
In summary, the Pump Tubing with X-Coating and the uncoated tubing are substantially equivalent in intended use, principles of operation and technology, design and materials, and performance. Any noted differences between the two devices do not raise new issues of safety and effectiveness.
## Additional Safety Information:
- Sterilization conditions have been validated in accordance with AAMI guidelines to . provide a Sterility Assurance Level (SAL) of 106.
- . Ethylene Oxide residues will not exceed the maximum residue limits proposed for Part 821 of Title 21 in the Federal Register of June 23, 1978 (or as finalized or amended).
- Terumo Cardiovascular Systems Corporation (TCVS) conducted the biocompatibility studies recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices -- Part 1: Evaluation and Testing." {External Communicating Devices, Circulating Blood, Limited Exposure (≤ 24 hours) Contact Duration]. The blood contacting materials were found to be biocompatible.
- TCVS conducted studies for materials characterization, including physico-chemical . profiles.
- The polymer coating material that is applied to the blood-contacting surfaces of the . device was also evaluated in an in-vivo animal study. No adverse conditions were noted.
## Conclusion:
In summary, the Pump Tubing with X-Coating is substantially equivalent in intended use, principles of operation and technology, design and materials, and performance to the predicate uncoated tubing (K993189).
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Inside the circle is a symbol of three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# SEP 1 9 2002
Terumo Medical Corporation c/o Mr. Kazuhito Inoue Regulatory Affairs 125 Blue Ball Road Elkton, Maryland 21921
Re: K022857
Trade Name: Pump Tubing with X-Coating Regulation Number: 21 CFR 870.4390 Regulation Name: Cardiopulmonary Bypass Pump Tubing Regulatory Class: Class II (two) Product Code: DWE Dated: August 26, 2002 Received: August 28, 2002
#### Dear Mr. Inoue:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Kazuhito Inoue
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
V.A.R. B.Zuk MD
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known):
#### Device Name: Pump Tubing with X-coating
#### Indications For Use:
The Cardiopulmonary Bypass Pump Tubing with X-Coating is intended to provide a conduit for extracorporeal blood flow through a roller pump during cardiopulmonary bypass procedures.
The tubing is intended for use in procedures lasting up to 6-hours in duration.
Kazuhito Inoue Regulatory Affairs Terumo Cardiovascular Systems
## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular & Respiratory Devices
510(k) Number K022857
OR
Prescription Use
Over-The-Counter Use
(Per 21 CFR 801.109)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.