MAXIMA FORTE HOLLOW FIBER OXYGENATOR WITH PLASMA RESISTANT FIBER
Applicant
Medtronic Vascular
Product Code
DTZ · Cardiovascular
Decision Date
Dec 16, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4350
Device Class
Class 2
Attributes
Therapeutic
Intended Use
MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber is indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from blood and is designed to operate at blood flow rates between 1 and 7 liters per minute for periods up to six hours.
Device Story
Single-use, disposable, sterile, nonpyrogenic gas exchange device with integrated venous heat exchanger; used in cardiopulmonary bypass extracorporeal circuits. Input: venous blood and temperature-regulated water. Operation: blood flows through nonporous polypropylene heat exchanger tubes, then radially through a microporous polypropylene hollow fiber bundle; gas flows through the interior of the fibers. Output: oxygenated arterial blood. Operated by perfusionists/clinicians in surgical settings. Device facilitates gas exchange (O2/CO2 diffusion) and blood temperature regulation. Benefits: provides life-sustaining gas exchange during bypass procedures.
Clinical Evidence
Bench testing only. In-vitro analysis included blood pathway integrity, heat exchanger integrity, blood volumes, connector testing, oxygen/CO2 transfer rates, heat exchanger performance, blood cell damage, time-dependent performance, and pressure drop. Testing confirms substantial equivalence to predicate devices.
Indicated for patients undergoing procedures requiring extracorporeal oxygenation and carbon dioxide removal during cardiopulmonary bypass, at blood flow rates of 1-7 L/min for up to 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
Predicate Devices
MAXIMA PLUS™ PRF Hollow Fiber Oxygenator with Improved Plasma Resistant Fiber (K941473)
BARD® WILLIAM HARVEY® HF-5000 Membrane Oxygenator (K872107)
Reference Devices
Terumo Medical Corporation Capiox SX Hollow Fiber Oxygenator (K922799)
K954850 — BARD WILLIAM HARVEY HF-6000 MEMBRANE OXYGENATOR · C.R. Bard, Inc. · Oct 30, 1996
K970463 — CAPIOX E HOLLOW FIBER OXYGENATOR WITH INTEGRATED HEAT EXCHANGER/ARTERIAL RESERVOIR · Terumo Medical Corp. · May 6, 1997
K123314 — AFFINITY FUSION OXYGENATOR WITH INTEGRATED ARTERIAL FILTER AND CARMEDA BIOACTIVE SURFACE · Medtronic, Inc. · Jan 22, 2013
K991452 — COBE OPTIMIN HOLLOW FIBER MEMBRANE OXYGENATOR · Cobe Cardiovascular, Inc. · Oct 7, 1999
K143073 — Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium Biosurface, Affinity NT Oxygenator with Carmeda Biosurface · Medtronic, Inc. · Mar 25, 2015
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS
# COMPANY AND CONTACT PERSON
K. 962641
Medtronic, Inc.
Cardiopulmonary Division
4633 E. La Palma Avenue
Anaheim, CA 92807
714-779-3700
DEC 16 1996
Debra J. Kridner, Manager
Regulatory Affairs
# DEVICE NAME
MAXIMA FORTE™ Hollow Fiber Oxygenator with Plasma Resistant Fiber (Model:MAX-FTE)
# NAME OF PREDICATE OR LEGALLY MARKETED DEVICE
- MAXIMA PLUS™ PRF Hollow Fiber Oxygenator with Improved Plasma Resistant Fiber (K941473)
- BARD® WILLIAM HARVEY® HF-5000 Membrane Oxygenator (K872107).
# DESCRIPTION OF DEVICE
The Medtronic MAXIMA FORTE™ Hollow Fiber Oxygenator with Plasma Resistant Fiber is a single use, disposable, sterile, nonpyrogenic, gas exchange device with a self contained, venous side heat exchanger for regulating blood temperature. The MAXIMA FORTE™ oxygenator consists of a polycarbonate outer case which incorporates an oxygenator fiber bundle and a self contained heat exchanger. The fiber bundle is separated from the self contained heat exchanger by a polycarbonate housing. The static priming volume of the oxygenator blood path is approximately 295 ml. The oxygenator is designed to operate at blood flow rates between 1 - 7 liters per minute for periods up to 6 hours.
The MAXIMA FORTE™ oxygenator is placed in the cardiopulmonary bypass extracorporeal circuit for gas exchange. Venous blood enters the bottom of the oxygenator and flows through the nonporous polypropylene hollow tubes of the heat exchanger. These nonporous polypropylene hollow tubes are supported on their exterior by a polypropylene screen and contained within the polycarbonate heat exchanger housing. The polypropylene tubes run vertically through the polycarbonate chamber of the heat exchanger. The walls of these tubes provide a barrier between the venous blood and the cooling/warming water.
510(k) Notification
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTE™ Hollow Fiber Oxygenator
with Plasma Resistant Fiber
July 3, 1996
Confidential
Appendix IV
Page 2
{1}
The flow of venous blood and water through the MAXIMA FORTÉ™ Heat Exchanger portion of the oxygenator is as follows:
## Venous Blood
Venous blood enters the blood inlet port on the bottom/center of the oxygenator. This blood flows up through the polypropylene nonporous hollow tubes, exits the top of the tubes and flows radially into the outer polycarbonate housing containing the microporous fiber bundle.
## Water
The cooling/warming water enters the water inlet port. The water is directed towards the top of the heat exchanger, then flows downward passing around the outside walls of the nonporous polypropylene tubes, then exits through the water outlet port. The water outlet and inlet ports are located on the side of the oxygenator parallel to each other.
Heat exchange in the oxygenator occurs as the venous blood passes through the nonporous hollow tubes, while the temperature regulated cooling/warming water passes around the outside walls of the same tubes. The venous blood is cooled/ warmed as it flows through the heat exchanger prior to entering the outer polycarbonate case containing the microporous hollow fiber bundle (oxygenation portion of the oxygenator).
The flow of blood and gas through the MAXIMA FORTÉ™ oxygenation portion of the oxygenator is as follows:
## Blood
Venous blood (oxygen depleted) exits from the top of the heat exchanger chamber and flows radially outward while passing into the outer polycarbonate case, which contains microporous polypropylene hollow fibers. These fibers have been wound around a plastic core to produce a fiber bundle. This fiber bundle is encased in a clear polycarbonate case. The venous blood enters the fiber bundle and flows around the microporous polypropylene fibers. As blood flows over the microporous hollow fibers carbon dioxide/oxygen gas exchange occurs. Oxygenated arterial blood exits through the blood outlet port on the side of the oxygenator.
## Gas
Gas enters through the gas inlet port on the top of the oxygenator and passes through the inside of the microporous hollow fibers. As the gas flows through the microporous hollow fibers gas transfer occurs. The gas exits from the gas outlet port on the side near the bottom of the oxygenator. Gas vents are located on the bottom of the oxygenator to prevent over-pressurization and under-pressurization.
510(k) Notification
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTÉ™ Hollow Fiber Oxygenator
with Plasma Resistant Fiber
July 3, 1996
Confidential
Appendix IV
Page 3
{2}
Gas exchange (transfer) occurs by diffusion of carbon dioxide and oxygen, across the wall of the microporous hollow fiber membrane between the blood and gas phases. When blood passes over the exterior of the microporous hollow fibers, while gas passes through these same fibers, gas exchange occurs. The blood is oxygenated and carbon dioxide removed as the blood flows over the fibers prior to exiting from the oxygenator to the patient.
## STATEMENT OF INTENDED USE
MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber is indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from blood and is designed to operate at blood flow rates between 1 and 7 liters per minute for periods up to six hours.
## STATEMENT OF INTENDED USE OF PREDICATE/MARKETED DEVICES
The MAXIMA PLUS™ PRF Hollow Fiber Oxygenator with Improved Plasma Resistant Fiber (Model Number: MAX-PRF) is indicated for use in procedures requiring the extracorporeal oxygenation of and carbon dioxide removal from blood during cardiopulmonary bypass and is designed to operate at blood flow rates between 1 and 7 LPM for periods up to six hours.
The BARD® WILLIAM HARVEY® HF-5000 Membrane Oxygenator provides blood oxygen and carbon dioxide gas transfer. In addition, it provides heat exchange in order to maintain or adjust the blood temperature as desired.
## STATEMENT OF TECHNOLOGICAL CHARACTERISTICS COMPARISON
A table comparing the intended use and technological characteristics of the Medtronic Cardiopulmonary MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber with the two noted substantially equivalent devices is provided in Attachment 1. In addition, information regarding technological characteristic comparison is provided in the following section, “Determination of Substantial Equivalence”.
## DETERMINATION OF SUBSTANTIAL EQUIVALENCE
In determining substantial equivalence of the MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber (Model: MAX-FTE), the decision-making process follows the 510(k) “Substantial Equivalence” flow diagram as follows:
The MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber (Model: MAX-FTE). is being “compared to the following two Marketed Devices”
- MAXIMA PLUS™ PRF Hollow Fiber Oxygenator with Improved Plasma
510(k) Notification
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber
July 3, 1996
Confidential
Appendix IV
Page 4
{3}
Resistant Fiber (K941473)
- BARD® WILLIAM HARVEY® HF-5000 Membrane Oxygenator (K872107).
The MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber (Model: MAXFTE). has the “same indications statement and intended uses” as the
- MAXIMA PLUS™ PRF Hollow Fiber Oxygenator with Improved Plasma Resistant Fiber (K941473)
The MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber (Model: MAXFTE). has “new technological characteristics (e.g., design, materials and manufacturing processes)” from the currently Medtronic adult oxygenators. These technological characteristics include;
- Polypropylene heat exchanger
- Transparent polycarbonate housing
- Lower priming volume (~295 ml)
- Blood flow bottom to top heat exchanger, then top to bottom fiber bundle
- Microporous hollow fiber has a smaller inner diameter
These technological characteristics are common to other adult oxygenators currently in commercial distribution as follows;
- Plastic heat exchanger - BARD® WILLIAM HARVEY® HF-5000 Membrane Oxygenator (K872107).
- Transparent polycarbonate housing - BARD® WILLIAM HARVEY® HF-5000 Membrane Oxygenator (K872107).
- Lower priming volume (~270 ml) - Terumo Medical Corporation Capiox SX Hollow Fiber Oxygenator (K922799).
- Blood flow bottom to top heat exchanger, then top to bottom fiber bundle - Avecor Cardiovascular Affinity Hollow Fiber Oxygenator (K932252).
- Microporous hollow fiber has a smaller inner diameter - Inner diameter of microporous hollow fiber for other non-Medtronic adult oxygenators is unknown. Data demonstrates that this new technological characteristic does not present any new types of safety or effectiveness questions.
These technological characteristics “could affect the safety and effectiveness of the device”. However these “new technological characteristics do not raise new types of safety or effectiveness questions”. In addition, “there are accepted scientific methods which exist for assessing effects of these new technological characteristics”. These scientific methods are provided in the following oxygenator standards;
- ISO/DIS 7199 (ISO/TC 150/SC 2) 1995 Draft International Standard titled “Cardiovascular implants and artificial organs - Blood-gas exchangers”
- BG7199-1996 (proposed new American National Standard) titled, “Cardiovascular implants and artificial organs - Blood-gas exchangers (oxygenators)”
510(k) Notification
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTÉ™ Hollow Fiber Oxygenator
with Plasma Resistant Fiber
July 3, 1996
Confidential
Appendix IV
Page 5
{4}
"Performance data to assess the effects of these new technological characteristics" has been performed. These "performance data demonstrate" that the MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber is substantially equivalent to marketed devices.
In addition, the modes of operation of the MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber are either identical or substantially equivalent to other hollow fiber oxygenators currently in commercial distribution. The function of the Medtronic MAXIMA FORTÉ™ Hollow Fiber Oxygenators is the extracorporeal oxygenation of and carbon dioxide removal from blood during cardiopulmonary bypass. The MAXIMA FORTÉ™ Hollow Fiber Oxygenators are designed to operate at blood flow rates between 1 and 7 LPM for periods up to six hours.
The predicate/marketed device is:
- MAXIMA PLUS® PRF Hollow Fiber Oxygenator with Improved Plasma Resistant Fiber (K941473)
A table comparing the intended use and technological characteristics of the MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber with the two noted substantially equivalent devices is provided in Attachment 1.
The biocompatibility testing and in-vitro bench testing demonstrated that when compared to the predicate devices the MAXIMA FORTÉ™ Hollow Fiber Oxygenator with Plasma Resistant Fiber do not significantly affect safety and effectiveness and are substantially equivalent to other commercially distributed hollow fiber oxygenators. The in-vitro bench testing included analysis of:
- Blood pathway integrity
- Heat exchanger pathway integrity
- Blood volumes
- Connectors
- Oxygenator and Carbon Dioxide Transfer Rates
- Heat Exchanger Performance Factor
- Blood Cell Damage
- Time Dependent Performance Changes
- Pressure Drop
510(k) Notification
Medtronic, Inc.
Medtronic Cardiopulmonary
MAXIMA FORTÉ™ Hollow Fiber Oxygenator
with Plasma Resistant Fiber
July 3, 1996
Confidential
Appendix IV
Page 5
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.