K955554 · North American Instrument Corp. · DTL · Mar 4, 1996 · Cardiovascular
Device Facts
Record ID
K955554
Device Name
HEMO-VALVE
Applicant
North American Instrument Corp.
Product Code
DTL · Cardiovascular
Decision Date
Mar 4, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4290
Device Class
Class 2
Intended Use
The HEMO-Valve is recommended for use in angiographic procedures involving blood or other potentially infectious materials, so that the procedure can be performed in such a manner as to minimize blood loss and patient/user exposure to these substances. The HEMO-Valve is intended for use in intra-arterial and intravenous administration of fluids, including radiographic contrast media.
Device Story
HEMO-Valve is an elastomeric valve adaptor with male and female luer lock fittings; used in angiographic procedures. Operation: valve opens via male luer connection or guidewire insertion (forward/backloading). Function: provides hemostasis during guidewire exchange; allows syringe aspiration/injection; withstands high-pressure injections. Used in clinical settings by physicians/technicians to manage fluid delivery and minimize blood loss/exposure to infectious materials.
Clinical Evidence
Bench testing only. Testing included hemostasis verification during and after wire exchanges, vacant valve hemostasis, and hydrostatic/dynamic pressure testing to confirm pressure ratings. Biocompatibility supported by existing data on predicate devices.
Technological Characteristics
Elastomeric valve with male/female luer lock fittings. Materials identical to predicate NAMIC devices. Designed for high-pressure injection. Sterilization identical to existing NAMIC product line.
Indications for Use
Indicated for patients undergoing angiographic procedures requiring intra-arterial or intravenous fluid administration, including radiographic contrast media, while minimizing blood loss and exposure to infectious materials.
Regulatory Classification
Identification
A cardiopulmonary bypass adaptor, stopcock, manifold, or fitting is a device used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
NAMIC High Pressure Stopcock
NAMIC PTCA Y-Adaptor
Braun Safsite [Valve]
Reference Devices
NAMIC Hemostatic Introducer Sheath
Related Devices
K034043 — H-FLOW VALVE · Elcam Medical Acal · Jun 30, 2004
K031903 — RAPIDO CUT-AWAY BLEEDBACK CONTROL VALVE (BBCV), MODEL 7568 · Guidant Corporation · Mar 3, 2004
K091634 — NMI FC · Navilyst Medical, Inc. · Mar 1, 2010
K990975 — IN-LINE HEMOSTASIS VALVE · Merit Medical Systems, Inc. · Mar 15, 2000
K993057 — MBA · Merit Medical Systems, Inc. · Oct 5, 1999
Submission Summary (Full Text)
{0}
K95554
# PREMARKET NOTIFICATION
| | VII: 510(K) SUMMARY | MAR - 4 1996 |
| --- | --- | --- |
| Manufacturer: | NAMIC U.S.A. Corporation
Glens Falls, New York 12801 | |
| Contact Person: | Mary Meagher Rubin
Regulatory Affairs Specialist
Telephone Number (518) 798-0067
Facsimile Number (518) 798-5475 | |
| Date Prepared: | December 1, 1995 | |
| Trade Name: | NAMIC HEMO-Valve | |
| Common Name: | Adaptor | |
| Classification Name: | Cardiopulmonary Bypass adaptor, stopcock, manifold, or fitting. | |
| Predicate Devices: | NAMIC High Pressure Stopcock
NAMIC PTCA Y-Adaptor
Braun Safsite [Valve] | |
## Product Description:
The HEMO-Valve consists of an elastomeric valve located between male and female luer lock fittings. The valve may be opened by connecting a male luer lock (the luer fitting pushes the gland out of the way) or may be opened by introducing a guidewire in a forward or backloading manner.
## Intended Use:
The HEMO-Valve is recommended for use in angiographic procedures involving blood or other potentially infectious materials, so that the procedure can be performed in such a manner as to minimize blood loss and patient/user exposure to these substances. The HEMO-Valve is intended for use in intra-arterial and intravenous administration of fluids, including radiographic contrast media.
The HEMO-Valve is identical to currently marketed NAMIC devices NAMIC High Pressure Stopcock, NAMIC PTCA Y-Adaptor, and the NAMIC Hemostatic Introducer Sheath with respect to materials of construction. The manufacturing process for the HEMO-Valve is common to other products manufactured by NAMIC. Packaging, sealing, and sterilization for the HEMO-Valve is identical to that of all NAMIC devices.
HEMO-Valve
NAMIC® USA CORPORATION
Page 42
{1}
PREMARKET NOTIFICATION
VII. 510(K) Summary (Continued).
The manner in which the normally closed valve is actuated (by insertion of a male luer lock) is the same as the predicate device.
The intended uses of the HEMO-Valve are consistent with those of the predicate devices.
As with all three predicate devices, the HEMO-Valve allows for syringe aspiration and Injection, its primary function, being an adaptor. Insertion and backloading of a guidewire in addition to hemostasis around a guidewire are functions the HEMO-Valve has in common with the NAMIC PTCA Y-Adaptor. Finally, as with the NAMIC High Pressure Stopcock, the HEMO-Valve is designed to withstand high pressure injections.
The HEMO-Valve has been subjected to non-clinical performance testing to provide data supporting the functional claims of the device. The test protocol included testing to confirm hemostasis during and after wire exchanges as well as with the valve in the vacant position. Hydrostatic and Dynamic pressure testing were performed to confirm the pressure rating of the device.
Biocompatibility testing has been performed on the NAMIC predicate devices. There are no changes regarding device material.
HEMO-Valve
NAMIC® USA CORPORATION
Page 43
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.