MBA

K993057 · Merit Medical Systems, Inc. · DTL · Oct 5, 1999 · Cardiovascular

Device Facts

Record IDK993057
Device NameMBA
ApplicantMerit Medical Systems, Inc.
Product CodeDTL · Cardiovascular
Decision DateOct 5, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4290
Device ClassClass 2

Indications for Use

The MBA Hemostasis Valves are designed for use during percutaneous transluminal angioplasty (PTA) and other intravascular therapeutic procedures that utilize a guiding catheter.

Device Story

Dual seal hemostasis valve with integral introducer; 'Y' shaped design. Attaches to external female luer of guiding catheter to control bleed-back during interventional procedures. Female port connects to manifold for contrast injection and pressure monitoring; valve allows controlled arterial access for guide wires and interventional devices. Used in clinical settings by physicians during percutaneous transluminal angioplasty and other intravascular therapeutic procedures. Benefits include maintenance of hemostasis while facilitating device delivery.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Dual seal hemostasis valve; 'Y' shaped configuration; includes integral introducer and female luer port. Materials are biocompatible. Non-powered device.

Indications for Use

Indicated for use during percutaneous transluminal angioplasty (PTA) and other intravascular therapeutic procedures requiring a guiding catheter in patients undergoing interventional procedures via peripheral arterial access.

Regulatory Classification

Identification

A cardiopulmonary bypass adaptor, stopcock, manifold, or fitting is a device used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Attachment 4 ## 510(k) Summary ## SAFETY AND EFFECTIVENESS SUMMARY This information of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. | Submitted by Name/Address: | Chester McCoy<br>Regulatory Affairs Engineer<br>Merit Medical Systems, Inc.<br>1600 West Merit Parkway | |----------------------------|--------------------------------------------------------------------------------------------------------| | | South Jordan, UT 84095<br>(801) 253-1600 ext. 404<br>(801) 253-1684 fax | | Contact Person: | Same as above | | Date Summary Prepared: | September 9, 1999 | | Device Name: | MBA Hemostasis Valves | | Common Name: | Hemostasis Valves | | Trade Name: | MBA | | Classification (if known): | Cardiopulmonary bypass adaptor, stopcock,<br>manifold, or fitting (74DTL) Class II | | Predicate Device: | Passage™ Hemostasis Valve (k925419) | {1}------------------------------------------------ | Device Description: | The MBA™ is a Dual Seal Hemostasis Valve with an integral introducer | |-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Performance<br>Standards: | Performance standards have not been established by the FDA under section 514 of the Food, Drug and Cosmetic Act. | | Intended Use: | The MBA Hemostasis Valves are designed for use during percutaneous transluminal angioplasty (PTA) and other intravascular therapeutic procedures that utilize a guiding catheter. | | Device Use: | During angioplasty and other interventional procedures, the patient's vasculature is accessed through a peripheral artery. To control bleed-back from the artery, a hemostasis valve is attached to the external female luer of a guiding catheter. This device is typically 'y' shaped with a female luer and a manual hemostasis valve. The female port is attached to a manifold system to allow for contrast injections and pressure monitoring, while the hemostasis valve allows controlled direct arterial access when using a guide wire and/or interventional therapy devices. | | Biocompatibility: | All materials used in the MBA Hemostasis Valves are biocompatible. | | Summary of<br>Substantial<br>Equivalence: | The Merit MBA Hemostasis Valves are substantially equivalent to the previously cleared Merit Passage Hemostasis Valves. | {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human profiles facing to the right, with flowing lines beneath them. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 15 1999 Mr. Chester McCoy Requlatory Affairs Engineer Merit Medical Systems, Inc. 1600 West Merit Parkway South Jordan, UT 84095 Re : K993057 > Trade Name: MBA™ Hemostasis Valves Regulatory Class: II Product Code: DTL Dated: September 9, 1999 Received: September 13, 1999 Dear Mr. McCoy: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation {3}------------------------------------------------ Page 2 - Mr. Chester McCoy you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahon Thomas J. Callaha Director Division of Cardiovascular, Respiratory and Neuroloqical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Attachment 2 | Indications for Use Statement | | |-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <hr/> | | | 510(k) Number (if Known) | | | <hr/> | | | Device Name | MBA <sup>TM</sup> Hemostasis Valves | | <hr/> | | | Indications for Use | The MBA Hemostasis Valves are designed for use during percutaneous transluminal angioplasty (PTA) and other intravascular therapeutic procedures that utilize a guiding catheter. | ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED ### Concurrence of CDRH, Office of Device Evaluation (ODE) Christopher Callahan (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices Kyy 510(k) Number Prescription Use \$\times\$ OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...