CARDIOPLEGIA DELIVERY SETS CARDIOPLEGIA DELIVERY SETS WITH DURAFLO COATING
Applicant
Edwards Lifesciences, LLC
Product Code
DTL · Cardiovascular
Decision Date
Jan 12, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4290
Device Class
Class 2
Intended Use
Cardioplegia Delivery Sets with and without Duraflo coating are intended for use as adapters for connecting cardioplegia administration tubing to cardioplegia delivery catheters. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Device Story
Cardioplegia delivery sets consist of clear manifold connections (straight, Y-shaped, or 4-way) and flexible tubing. Components may include vent port adapters, tapered tubing adapters, and clamps. Certain models feature Duraflo (heparin) coating on inner-luminal or exterior cannula surfaces to improve blood compatibility. Used in cardiopulmonary surgery settings by clinical staff to connect cardioplegia administration tubing to delivery catheters. The device facilitates the delivery of cardioplegic solutions to the heart during bypass procedures. No electronic or software components are involved.
Clinical Evidence
Bench testing only. Functional evaluations included pressure vs. flow, clamp closure, kink resistance, and assembly tensile testing. All data met predetermined acceptance criteria.
Technological Characteristics
Manifold connections (straight, Y-shaped, 4-way) with flexible tubing. Components include luer lock connectors, vent port adapters, and clamps. Duraflo (heparin) coating applied to specific models for blood compatibility. Sterile, non-pyrogenic, peel-pouch packaging. No electronic, software, or energy-based components.
Indications for Use
Indicated for use in patients undergoing cardiopulmonary surgery requiring cardioplegia administration via delivery catheters; specifically for those where a heparin-coated blood path is desired (for Duraflo-coated variants).
Regulatory Classification
Identification
A cardiopulmonary bypass adaptor, stopcock, manifold, or fitting is a device used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K182586 — DLP Aortic Root Cannula, DLP Aortic Root Cannula with Vent Line, DLP Pressure Monitoring Extension Line Adapter, DLP Cardioplegia Adapter with Pressure Port · Medtronic, Inc. · Oct 19, 2018
K961180 — NMI ANTEGRAD CARDIOPLEGIA CANNULA · Naltiac Medical, Inc. · Jun 21, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summarv
JAN 1 2 2012
| Submitter: | Edwards Lifesciences LLC |
|----------------------|--------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Karen Jones, Senior Manager, Regulatory Affairs<br>12050 Lone Peak Pkwy<br>Draper, UT 84020<br>(801) 565-6231 |
| Date Prepared: | December 12, 2011 |
| Trade Name: | Edwards Lifesciences™ Cardioplegia Delivery Sets |
| Classification Name: | Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting<br>21 CFR Part 870.4290, Product Code DTL, Class II |
| Predicate Device: | K892368 Cardioplegia Delivery Sets |
# Device Description:
Edwards Lifesciences cardioplegia delivery sets are comprised of various configurations of clear manifold connections in either straight, Y-shaped, or 4-way connections which are joined by appropriate lengths of clear, flexible tubing. Depending upon configuration, the cardioplegia delivery set may contain a vent port adapter with female luer lock, a tapered tubing adapter, tubing clamps, and/ or male and/ or female luer lock connectors.
The exterior and/ or inner-luminal cannula surfaces of product codes containing a "D" or "DII" are coated with Duraflo TM (heparin) coating.
When used on devices for cardiopulmonary surgery, the Duraflo coating improves the blood compatibility of non-biological surfaces in the extracorporeal circuit.
Each Edwards Lifesciences device is packaged sterile and non-pyrogenic in a sealed, peel-type pouch.
PROPRIETARY DATA: This document and the information contained herein may not be reproduced, used, or disclosed without written permission from Edwards Lifesciences, LLC.
{1}------------------------------------------------
### Indications for Use:
Cardioplegia Delivery Sets with and without Duraflo coating are intended for use as adapters for connecting cardioplegia administration tubing to cardioplegia delivery catheters.
Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
## Comparative Analysis:
The subject devices have the same intended use and technological characteristics (i.e., design, material, chemical composition) as the predicate devices. It has been demonstrated that the subject Cardioplegia Delivery Sets are comparable to the predicate devices in fundamental scientific technology, material types, principles of operation, and functional performance evaluations. No new issues of safety or efficacy have been raised.
## Functional/Safety Testing:
The functional data indicate that the Cardioplegia Delivery sets perform in a substantially equivalent manner when compared to the predicate device. The following functional tests were performed.
- Pressure vs. flow .
- Clamp closure .
- . Kink
- . Assembly tensile
All data met acceptance criteria.
i
#### Conclusion:
Testing conducted to qualify the subject changes shows that devices in this product family have met all predetermined acceptance criteria and have been shown to be safe and effective for their intended use. The subject change does not affect the intended use or alter the fundamental scientific technology of the device.
PROPRIETARY DATA: This document and the information contained herein may not be reproduced, used, or disclosed without written permission from Edwards Lifesciences, LLC.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three curved lines representing wings or feathers. The symbol is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 1 2 2012
Edwards Lifesciences LLC c/o Ms. Karen Jones Sr. Principal Project Manager, Regulatory Affairs 12050 Lone Peak Parkway Draper, UT 84020
Re: K113667 Edwards Lifesciences™ Cardioplegia Delivery Sets Regulation Number: 21 CFR 870.4290 Regulation Name: Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass Regulatory Class: Class II (two) Product Code: DTL Dated: December 12, 2011 Received: December 13, 2011
Firman to wa
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act (Act do not require approval of a provisions of the Pederal Food, Drug, You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Karen Jones
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M. A. Hillemann
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Statement of Indications for Use
510(k) Number (if known): K|| 366 7
Cardioplegia Delivery Sets Device Name:
Indications for Use:
Cardioplegia Delivery Sets are intended for use as adapters for connecting cardioplegia administration tubing to cardioplegia delivery catheters.
Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Prescription Use × (Per 21 CFR 801.109)
OR Over-The-Counter Use_
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office Of Device Evaluation (ODE)
M.S. Willelenn
rdiovascular Devices
510(k) Number k113667
PROPRIETARY DATA: This document and the information contained herein may not be reproduced, used, or disclosed without written permission from Edwards Lifesciences, LLC.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.