DLP Aortic Root Cannula, DLP Aortic Root Cannula with Vent Line, DLP Pressure Monitoring Extension Line Adapter, DLP Cardioplegia Adapter with Pressure Port
K182586 · Medtronic, Inc. · DWF · Oct 19, 2018 · Cardiovascular
Device Facts
Record ID
K182586
Device Name
DLP Aortic Root Cannula, DLP Aortic Root Cannula with Vent Line, DLP Pressure Monitoring Extension Line Adapter, DLP Cardioplegia Adapter with Pressure Port
Applicant
Medtronic, Inc.
Product Code
DWF · Cardiovascular
Decision Date
Oct 19, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4210
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DLP Aortic Root Cannula (Model 23009) is intended for short term use (six hours or less) in conjunction with cardiopulmonary bypass surgery for delivering cardions. The cannula may also be used to aspirate air from the aorta at the conclusion of the bypass procedure. The Model 23009 cannula may also be used to monitor pressure in the aorta. The DLP Aortic Root Cannula with Vent Line (Model 24009) is intended for use during cardiopulmonary bypass for delivering cardioplegia solutions and venting of the left heart for up to 6 hours. The cannula may be used to aspirate air from the aorta at the conclusion of the bypass procedure. The Model 24009 cannula may also be used to monitor pressure in the aorta. The DLP Cardioplegia Adapter with Pressure Port (Model 15004) is intended for use in conjunction with the delivery of cardioplegia solution and to provide access for pressure monitoring for up to 6 hours. The DLP Pressure Monitoring Extension Line Adapter (Models 25009 and 25010) is intended for extending the pressure line in order to reach the pressure monitoring unit during cardiopulmonary bypass surgery up to 6 hours or less.
Device Story
DLP Aortic Root Cannula and Adapter family facilitates cardioplegia delivery, aortic venting, and pressure monitoring during cardiopulmonary bypass surgery. Devices consist of flexible tubing, luer fittings, and Y-type adapters; Model 15004 includes a latex-free rubber septum for pressure monitoring access. Used in clinical settings by perfusionists or surgeons; devices connect to cardioplegia administration sets and pressure monitoring units. Output is physical delivery of cardioplegia solution and transmission of pressure signals to external monitors. Single-use, sterile, nonpyrogenic components.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological and design comparisons to the predicate device.
Technological Characteristics
Flexible tubing, luer connections, and latex-free rubber septa. Components are sterile, nonpyrogenic, and single-use. Dimensions include 1.8m extension lines and various tubing diameters (0.48 cm to 0.64 cm). No electronic or software components.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass surgery requiring cardioplegia delivery, aortic air aspiration, left heart venting, or aortic pressure monitoring for durations of 6 hours or less.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
Predicate Devices
DLP™ Aortic Root Cannula with Integral Pressure Monitoring Line and Extension Set (K831591)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 19, 2018
Medtronic. Inc. Linda Julia Davila Regulatory Affairs Specialist 8200 Coral Street NE Mounds View, Minnesota 55125
### Re: K182586
Trade/Device Name: DLP™ Aortic Root Cannula (Model 23009), DLP™ Aortic Root Cannula with Vent Line (Model 24009), DLP™ Cardioplegia Adapter with Pressure Port (Model 15004), DLPTM Pressure Monitoring Extension Line Adapter (Models 25009 and 25010) Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, Or Tubing Regulatory Class: Class II Product Code: DWF Dated: September 19, 2018 Received: September 20, 2018
Dear Linda Julia Davila:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Fernando Aguel -
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182586
#### Device Name
DLP Aortic Root Cannula (Model 23009); DLP Aortic Root Cannula with Vent Line (Model 24009); DLP Cardioplegia Adapter with Pressure Port (Model 15004); DLP Pressure Monitoring Extension Line Adapter (Models 25009 and 25010)
#### Indications for Use (Describe)
The DLP Aortic Root Cannula (Model 23009) is intended for short term use (six hours or less) in conjunction with cardiopulmonary bypass surgery for delivering cardions. The cannula may also be used to aspirate air from the aorta at the conclusion of the bypass procedure. The Model 23009 cannula may also be used to monitor pressure in the aorta.
The DLP Aortic Root Cannula with Vent Line (Model 24009) is intended for use during cardiopulmonary bypass for delivering cardioplegia solutions and venting of the left heart for up to 6 hours. The cannula may be used to aspirate air from the aorta at the conclusion of the bypass procedure. The Model 24009 cannula may also be used to monitor pressure in the aorta.
The DLP Cardioplegia Adapter with Pressure Port (Model 15004) is intended for use in conjunction with the delivery of cardioplegia solution and to provide access for pressure monitoring for up to 6 hours.
The DLP Pressure Monitoring Extension Line Adapter (Models 25009 and 25010) is intended for extending the pressure line in order to reach the pressure monitoring unit during cardiopulmonary bypass surgery up to 6 hours or less.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px"><b> </b></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5.0 510(k) Summary
# 510(k) Summary
| Date Prepared: | Sept 18, 2018 |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration Number: 2184009 |
| Contact Person: | Linda L Julia Davila<br>Regulatory Affairs Specialist<br>Medtronic Perfusion Systems<br>Phone: 763.526.9737<br>Email: linda.l.juliadavila@medtronic.com |
| Alternate Contact: | Mike Green<br>Senior Regulatory Affairs Manager Medtronic<br>Perfusion Systems<br>Phone: 763.514.9774<br>Email: mike.green@medtronic.com |
| Device Name and Classification | |
| Trade Name: | DLPTM Aortic Root Cannula (Model 23009)<br>DLPTM Aortic Root Cannula with Vent Line (Model 24009)<br>DLPTM Cardioplegia Adapter with Pressure Port (Model 15004)<br>DLPTM Pressure Monitoring Extension Line Adapter<br>(Model 25009 and 25010) |
| Common Name: | Cardiopulmonary bypass vascular catheter, cannula, or tubing |
| Regulation Number: | 21 CFR 870.4210 |
| Product Code: | DWF |
| Product | |
| Classification: | Class II |
## Name of Predicate Device
DLP™ Aortic Root Cannula with Integral Pressure Monitoring Line and K831591 Extension Set
{4}------------------------------------------------
## Device Description
The proposed modification will affect the DLP Aortic Root Cannula with Integral Pressure Monitoring Line and Extension Set product family. A description of these devices follows below.
### Model 23009
The DLP Aortic Root Cannula consists of flexible tubing permanently attached to both the inlet and tip. The inlet fitting is a female luer fitting. The introducer is packaged within the cannula body.
### Model 24009
The DLP Aortic Root Cannula with Vent Line consists of a soft, flexible, thin-wall tip permanently attached to a flexible Y-type adapter. Flexible tubing is permanently attached to both the inlet and vent port of the Y-type adapter. The cannula inlet fitting is a female luer and the vent port fitting (outlet) is a slip-on connector for 0.48 cm (3/16 in) to 0.64 cm (1/4 in) Inner Diameter (I.D.) tubing. The introducer is packaged within the cannula body.
### Model 15004
The DLP Cardioplegia Adapter with Pressure Port consists of flexible tubing, luer connections, and a latex-free rubber septum. The end that terminates with a male luer connects to the cardioplegia cannula and the other end connects to the cardioplegia administration set. A latex-free rubber septum provides access for a pressure monitoring needle near the connection to the cardioplegia cannula.
### Model 25009 and 25010
The DLP Pressure Monitoring Extension Line Adapter is 1.8 m (6 ft) in length with a locking male luer fitting on one end, and either a locking female luer fitting (Model 25009) or a locking male luer fitting (Model 25010) on the other end.
All are provided sterile, nonpyrogenic, and for single use.
## Indications for Use
The DLP Aortic Root Cannula (Model 23009) is intended for short term use (six hours or less) in conjunction with cardiopulmonary bypass surgery for delivering cardioplegia solutions. The cannula may also be used to aspirate air from the aorta at the conclusion of the bypass procedure. The Model 23009 cannula may also be used to monitor pressure in the aorta.
The DLP Aortic Root Cannula with Vent Line (Model 24009) is intended for use during cardiopulmonary bypass for delivering cardioplegia solutions and venting of the left heart for up to 6 hours. The cannula may be used to aspirate air from the aorta at the conclusion of the bypass procedure. The Model 24009 cannula may also be used to monitor pressure in the aorta.
{5}------------------------------------------------
The DLP Cardioplegia Adapter with Pressure Port (Model 15004) is intended for use in conjunction with the delivery of cardioplegia solution and to provide access for pressure monitoring for up to 6 hours.
The DLP Pressure Monitoring Extension Line Adapter (Models 25009 and 25010) is intended for extending the pressure line in order to reach the pressure monitoring unit during cardiopulmonary bypass surgery up to 6 hours or less.
## Comparison to Predicate Devices
A comparison of the modified product to the currently marketed predicate products (K831591) indicates the following similarities:
- Intended use ●
- Technological characteristics ●
- Operating principle ●
- Design features ●
- Base materials
- Shelf life ●
The details of the changes for the DLP Aortic Root Cannula and Adapters are described in Section 12.0 of this 510(k) and the proposed modifications are similar to the predicate device.
## Conclusion
Medtronic has demonstrated that the modifications made to the DLP Aortic Root Cannula with Integral Pressure Monitoring Line and Extension Set product family described in this submission result in a substantially equivalent1 device because the fundamental scientific principle, operating principle, design features and intended use are unchanged from the predicate devices.
<sup>1</sup> The term 'substantially equivalent' as used herein is intended to be a determination of substantial equivalency under the Federal Food, Drug, and Cosmetic Act and relates to the fact that the product can be marketed without premarket approval or reclassification. Such a determination is not intended to have any bearing whatsoever on the resolution of patent infringement suits or other patent matters." (Federal Register, Vol. 42, No. 163, Aug. 23, 1977, page 42525 and 42529).