The Edwards® GeoForm™ Annuloplasty Ring is indicated for the correction of mitral valve insufficiency where the lesions are not so severe as to require total valve replacement. The decision to undertake valvuloplasty can be made only after visual analysis of the lesion present. The most favorable conditions for valvuloplasty using an annuloplasty ring are a combination of a distended natural valve ring associated with supple valve cusps and normal chordae tendineae. The remodeling valvuloplasty technique with a prosthetic ring may be used in all acquired or congenital mitral insufficiencies with dilatation and deformation of the fibrous mitral annulus. For Type I mitral insufficiencies with no subvalvular lesions and normal valvular movements, this ring technique used alone is sufficient. However, the ring technique must be associated with mitral valvuloplasty repair in Type II insufficiencies with a prolapsed valve due to elongation or rupture of the chordae tendineae and in Type III insufficiencies with limitation of valvular movements due to fusion of the commissures or chordae tendineae, or chordal hypertrophy.
Device Story
GeoForm Annuloplasty Ring (Model 4200) is a prosthetic implant for mitral valve repair. Device functions as a structural support to remodel the mitral annulus in patients with mitral insufficiency. Implanted by surgeons during open-heart procedures; device provides mechanical stabilization of the valve annulus to improve leaflet coaptation and reduce regurgitation. Benefits include preservation of the natural valve and avoidance of total valve replacement. Device is a passive implant; no electronic or algorithmic components.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Constructed of titanium alloy core; sewing ring margin consists of silicone rubber covered with polyester velour cloth; single-seam construction. Passive mechanical implant.
Indications for Use
Indicated for patients with acquired or congenital mitral valve insufficiency requiring correction where lesions do not necessitate total valve replacement. Suitable for Type I (no subvalvular lesions), Type II (prolapsed valve), and Type III (limited valvular movement) insufficiencies involving mitral annulus dilatation and deformation.
Regulatory Classification
Identification
An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”
Predicate Devices
Carpentier-Edwards™ Classic Annuloplasty Ring (K831949)
K083470 — CARPENTIER- EDWARDS PHYSIO II ANNULOPLASTY, MODEL 5200 · Edwards Lifesciences, LLC · Jan 23, 2009
K232599 — TransForm McCarthy Mitral Annuloplasty Ring (TF) · Genesee Biomedical, Inc. · Apr 9, 2024
K190506 — NeoForm Annuloplasty Ring · Genesee Biomedical, Inc. · Apr 18, 2019
K202253 — TruForm Sievers Annuloplasty Ring · Genesee Biomedical, Inc. · Oct 6, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
K032250
## 510(k) Summary
| Submitter: | Edwards Lifesciences LLC |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Susan Reynolds, Regulatory Affairs Associate |
| Date Prepared: | July 21, 2003 |
| Trade Name: | GeoForm™ Annulplasty Ring |
| Classification Name: | Class II, CFR 870.3800 Annuloplasty Ring, 74 KRH |
| Predicate Device(s): | Carpentier-Edwards™ Classic Annuloplasty Ring<br>(K831949)<br>Cosgrove Edwards™ Annuloplasty System<br>(K923367)<br>MC3 Annuloplasty System (K020864) |
| Device Description: | The GeoForm Annuloplasty Ring, Model 4200, is<br>constructed of titanium alloy and has a sewing ring<br>margin that consists of a layer of silicone rubber,<br>covered with polyester velour cloth sewn with a single<br>seam. |
| Indications for Use: | The GeoForm Annuloplasty Ring, Model 4200 is<br>intended for the correction of mitral valvular<br>insufficiency where the lesions are not so severe as to<br>require total valve replacement. |
| Comparative Analysis: | It has been demonstrated that the GeoForm<br>Annuloplasty Ring is comparable to the predicate<br>device in design, intended use, materials, and principal<br>of operation. |
| Functional/Safety Testing: | The GeoForm Annuloplasty Ring has successfully<br>completed design verification testing. |
| Conclusion: | The GeoForm Annuloplasty Ring is substantially<br>equivalent to the predicate device. |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 6 2003
Edwards Lifesciences, LLC c/o Ms. Susan Reynolds Regulatory Affairs Associate One Edwards Way Irvine, CA 92614
Re: K032250
Trade Name: GeoForm™ Annuloplasty Ring, Mitral Model 4200 Regulation Number: 21 CFR 870.3800 Regulation Name: Annulplasty ring. Regulatory Class: Class II (two) Product Code: KRH Dated: July 21, 2003 Received: July 22, 2003
Dear Ms. Reynolds:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{2}------------------------------------------------
## Page 2 - Ms. Susan Reynolds
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
July 21, 2003
Page
510(k) Number (if known): K032250
Device Name: GeoForm™ Annuloplasty Ring
Indications for Use:
The Edwards® GeoForm™ Annuloplasty Ring is indicated for the correction of mitral valve insufficiency where the lesions are not so severe as to require total valve replacement.
The decision to undertake valvuloplasty can be made only after visual analysis of the lesion present. The most favorable conditions for valvuloplasty using an annuloplasty ring are a combination of a distended natural valve ring associated with supple valve cusps and normal chordae tendineae.
The remodeling valvuloplasty technique with a prosthetic ring may be used in all acquired or congenital mitral insufficiencies with dilatation and deformation of the fibrous mitral annulus.
For Type I mitral insufficiencies with no subvalvular lesions and normal valvular movements, this ring technique used alone is sufficient. However, the ring technique must be associated with mitral valvuloplasty repair in Type II insufficiencies with a prolapsed valve due to elongation or rupture of the chordae tendineae and in Type III insufficiencies with limitation of valvular movements due to fusion of the commissures or chordae tendineae, or chordal hypertrophy.
## (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Image /page/3/Figure/11 description: The image shows a document related to the Concurrence of CDRH, Office Of Device Evaluation (ODE). It indicates the Division of Cardiovascular, Respiratory, and Neurological Devices. The document includes a 51(k) Number, which is K032250. A checkmark indicates that it is for Prescription Use.
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.