GLYCARTISSUE REPAIR PATCH FAMILY OF PRODUCTS; GLYCAR VASCULAR REPAIR PATCH ALTERNATIVE PROPRIETARY NAME POSSIBLE

K963368 · Glycar, Inc. · DXZ · Oct 31, 1997 · Cardiovascular

Device Facts

Record IDK963368
Device NameGLYCARTISSUE REPAIR PATCH FAMILY OF PRODUCTS; GLYCAR VASCULAR REPAIR PATCH ALTERNATIVE PROPRIETARY NAME POSSIBLE
ApplicantGlycar, Inc.
Product CodeDXZ · Cardiovascular
Decision DateOct 31, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3470
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Glycar Pericardial Patch and Glycar Vascular Patch of the Glycar Tissue Repair Patch line of products are intended for pericardial closure, peripheral vascular reconstruction and repair, and cardiac and great vessel reconstruction and repair.

Device Story

Glycar Vascular Repair Patch is a biological tissue patch derived from bovine pericardium. The tissue undergoes glutaraldehyde fixation, tanning, and a proprietary anti-calcification treatment (XX). The device is intended for surgical use by physicians for repair, replacement, or reinforcement of soft tissue in cardiovascular and peripheral vascular procedures. It is supplied as a sterile, pliable patch (2 x 9 cm) to be used in the OR. The patch provides a durable scaffold for tissue repair. It is used in procedures such as carotid endarterectomy and repair of femoral, iliac, profunda, or tibial arteries. The device is sterilized using formaldehyde and propylene oxide.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on material composition, physical properties (tissue strength), and sterilization validation. The FDA noted that claims regarding anti-calcification effectiveness or long-term durability require future human clinical study data.

Technological Characteristics

Material: Glutaraldehyde-treated bovine pericardium. Dimensions: 2 x 9 cm. Sterilization: Formaldehyde and propylene oxide (SAL 10^-6). Manufacturing includes proprietary anti-calcification treatment. Biocompatible, pliable, and high-strength tissue scaffold.

Indications for Use

Indicated for patients requiring pericardial closure, peripheral vascular reconstruction and repair, or cardiac and great vessel reconstruction and repair. No specific age or gender contraindications provided.

Regulatory Classification

Identification

An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K963368 Oct. 31, 1997 510 (K) SUMMARY - GLYCAR VASCULAR REPAIR PATCH GLYCAR # 510(K) SUMMARY ## 1. Contact Person Dr. Dirk Frater 4537 Belclaire Ave. Dallas, TX 75205 Phone: (214) 526-5412 ## 2. Regulatory Consultant Bertram Hudson Shotwell & Carr Inc. Suite 100, 3003 LBJ Freeway Dallas, TX 75234 ## 3. Device Name a) Trade Name Glycar Tissue Repair Patch Family of products: Glycar Vascular Repair Patch alternative proprietary name possible b) Common Name Biological Tissue Patch ## 4. F.D.A. Device Classification Class II Device ## 5. Predicate Device The predicate device is the Vascu-Guard™ product (K942010, and is a tissue patch made from glutaraldehyde-treated bovine pericardium. The device is produced by Bio-Vascular Inc., 2670 Patton Rd., St. Paul, MN 55113, (800) 255-4018. For copies of Vascu-Guard™ labels, see Figures 3 - 4 in the Addenda under Predicate Device Labeling. ## 6. Product Description a) Candidate Device Composition The Glycar Vascular Repair Patches are made of the same material as the predicate device, the Peri-Guard™ tissue patches, which are made from glutaraldehyde-treated bovine pericardium. Aug.23, 1996 i {1} 510 (K) SUMMARY - GLYCAR VASCULAR REPAIR PATCH GLYCAR See Comparison Chart, Chart 1, in Addenda under Substantial Equivalence. ## b) Material Attributes Glutaraldehyde-treated bovine pericardium is a strong, pliable, biocompatible material which is easy to handle surgically. Glycar's glutaraldehyde-treated pericardium possesses tissue strength equaling that of the predicate device. It provides for a durable repair, replacement or reinforcement of natural soft tissue. See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence and Breaking Strength under Supporting Data ## c) Pyrogenicity Glycar glutaraldehyde-treated bovine pericardium has been found to have negative pyrogenicity. See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence and Pyrogenicity under Supporting Data. ## d) Manufacturing Process The manufacturing process for the *Glycar Vascular Repair Patches* involves the harvesting of bovine pericardium, fixation of the pericardium in glutaraldehyde, inspection and cleaning of the pericardium, selection of the pericardium for use in patch production, additional tanning of the bovine pericardium, and then cutting of the glutaraldehyde-treated bovine pericardium for the designated patch size, followed by sterilization in formaldehyde, followed then by treatment with Glycar's proprietary anti-calcification treatment, XX. See in Proprietary Manufacturing Material in Addenda under Confidential Information. Sterility is maintained by using aseptic fill under Clean Room 100 conditions, during transfer to anti-calcification treatment and to the final storage medium, 2% propylene oxide in water. See Outline of Manufacturing Process in Addenda under Manufacturing Information and Sterilization Information. ## e) Product Sizes Single size at present: 2 x 9 cm. ## 7. Intended Use of Product Aug.23, 1996 ii {2} 510 (K) SUMMARY - GLYCAR VASCULAR REPAIR PATCH GLYCAR a) Candidate Device Intended Use The Glycar Inc., Vascular Repair Patch has the same intended uses as the predicate device, the Bio-Vascular Inc., Vascu-Guard™ family of products, which are all tissue patches made from the same glutaraldehyde-treated bovine pericardium. Already approved intended uses for the Glycar Tissue Repair Patch family of glutaraldehyde-tanned bovine pericardial tissue patches: The Glycar Tissue Repair Patch (K942911) was deemed substantially equivalent to the Peri-Guard™ bovine pericardial tissue patch and received approval for U.S. marketing for the intended use of hernia and other intra-abdominal soft tissue defect or deficiency repair in December, 1994. An addition to the Glycar Tissue Repair Patch family of products, the Glycar Staple Strip (K954665) was deemed substantially equivalent to the Peri-Strips™ bovine pericardial surgical staple bolsters and received approval for U.S. marketing for the intended uses of surgical stapling of lung tissue, gastric staplings, rectal and vaginal prolapse, urethral sling, reconstruction of the pelvic floor, and hernias or defects of the diaphragm, thoracic and abdominal wall in October, 1995. See under Indication For Use Statement. The Glycar Staple Strip uses the identical material as the Glycar Tissue Repair Patch with the addition of a synthetic suture material to assist placement of the device on a surgical stapling arm. See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence. b) Predicate Device Intended Use The predicate device, the Vascu-Guard™ peripheral vascular patch has been used for repair of carotid arteries during carotid endarterectomy surgery, for repair of femoral, iliac, profunda, tibial arteries and for the revision of arterio-venous access shunts without any significant morbidity or mortality attributed to material failures. Aug.23, 1996 iii {3} 510 (K) SUMMARY - GLYCAR VASCULAR REPAIR PATCH GLYCAR ## 8. Sterilization a) Method The device is sterilized using exposure to formaldehyde and propylene oxide under environmental control in a class 100 Clean Room environment. See Outline of Sterilization Method in Addenda under Sterilization Information. b) Sterility Assurance Level The *Glycar Vascular Repair Patches* are labeled as sterile. They are sterilized to an SAL of 10⁻⁶ organisms. See in Addenda, Chart 1, Comparison Chart, under Substantial Equivalence and see Validation of Sterilization Processes under Sterilization Information. c) Bovine Spongiform Encephalopathy Pericardium used for the raw material for these devices is obtained from cattle from government approved abbatoirs in Transvaal, S. Africa. The cattle of the Republic of South Africa are certified free from B.S.E. (bovine spongiform encephalopathy) See Figure 5 in Addenda, Certificate of B.S.E.-free Status ## 9. Substantial Equivalence a) Comparison with Predicate Device The *Glycar Vascular Repair Patches* are substantially equivalent to the Vascu-Guard™ peripheral vascular patches in that they are: i) manufactured from the same material, glutaraldehyde-treated bovine pericardium, ii) have the same intended uses, iii) are sterilized to a minimum assurance against contamination of 10⁻⁶ organisms (Sterility Assurance Level, SAL = 10⁻⁶), and iv) have other performance features equal to those of the predicate device. See Comparison Chart, Chart 1, in Addenda under Substantial Equivalence and see Supporting Data b) Technological Differences Aug.23, 1996 iv {4} 510 (K) SUMMARY - GLYCAR VASCULAR REPAIR PATCH GLYCAR i) Implanted Material- Bovine Pericardium There is no difference in the implanted material used for the Glycar Vascular Repair Patches and the predicate device, Bio-Vascular Vascu-Guard™. Both devices are manufactured from gluraldehyde-treated bovine pericardium. See Comparison Chart, Chart 1, in Addenda under Substantial Equivalence. ii) Sterilization Method There is no difference in the chemical sterilization method used for Glycar Vascular Repair Patches versus that of the predicate device. See Supporting Data in Addenda Under Sleeve Configuration iii) Manufacturing Material One difference in the manufacturing process which Glycar Inc. considers proprietary is the use of the manufacturing material, XX. Glycar Inc. has demonstrated no adverse effects on safety or efficacy of the use of this manufacturing material. See Proprietary Manufacturing Material in Addenda under Confidential Information. Aug.23, 1996 v {5} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 31 1997 Dirk Frater, M.D. Official Contact Glycar Inc. 4504 Beverly Drive Dallas, Texas 75205 Re: K963368 and K963967 Glycar Pericardial Patch and Glycar Vascular Patch Regulatory Class: II Product Code: DXZ Dated: July 30, 1997 Received: August 4, 1997 Dear Dr. Frater: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please be advised that you may not make any claims about the effectiveness of the anticalcification treatment or durability of the device in any labeling or advertisements for this device until results of human clinical studies are submitted and found to substantiate such claims. You may, however, summarize results of animal studies performed, which should include the number and type of animal(s), location and duration of the implant, controls used, endpoints and results. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: {6} Page 2 - Dirk Frater, M.D. General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the *Federal Register*. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html." Sincerely yours, Thomas J. Callahan, Ph.D. Director, Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7} Page 1 of 1 510(k) Number (if known): K963967 and K963368 Device Name: **Glycar Pericardial Patch** and **Glycar Vascular Patch** Indications For Use: The Glycar Pericardial Patch and Glycar Vascular Patch of the Glycar Tissue Repair Patch line of products are intended for pericardial closure, peripheral vascular reconstruction and repair, and cardiac and great vessel reconstruction and repair. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number: K963967/K963368 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...