BIOTRONIK's endocardial leads are designed for use with implantable pulse generators that require pacing leads with a bipolar IS-1 connector configuration they may be used with single or dual chamber pacing systems. The leads are designed for use in patients for whom single or dual chamber pasing of pasing therapy is medically indicated. This indication for them single of the Class I definition of the ACC/AHANASPE This Indication This Indication inal recommended in Title Class I demind of the of Cardiac Pacemakers Reports, Aladoa Aloon More 2 2002 Ghuelle Opdate of Inplantabon CordialemyAmadon Mond Anad Anada Arabarhythmia Devices: A Report of the American Colleg CardiologyAmerican Heart Association Task Force on Practice Guidelines (ACCAHANASPE Committee on Pacemaker Implantation)" (Gregoratos et al. 2002). The straight lead models are intended for placement in either the right atrium or right ventricle. The JBP lead models have a pre-formed distal end to facilitate lead placement in the ight atrial appendage.
Device Story
Endocardial pacing leads for bradycardia management; connect to implantable pulse generators via bipolar IS-1 interface. Straight models placed in right atrium/ventricle; JBP models feature pre-formed J-shape for right atrial appendage placement. Device modification limited to packaging process and containers; no changes to lead design, materials, or clinical function. Used by physicians in clinical/surgical settings. Packaging ensures sterility and integrity of leads until implantation.
Clinical Evidence
No clinical data provided. The submission relies on bench testing related to the packaging modifications to ensure sterility and device integrity.
Technological Characteristics
Endocardial pacing leads with bipolar IS-1 connector configuration. Straight and J-shaped (JBP) form factors. No changes to lead materials or sensing/actuation principles. Packaging modifications only.
Indications for Use
Indicated for patients requiring single or dual chamber pacing therapy via implantable pulse generators with bipolar IS-1 connector configuration. Straight leads for right atrium or ventricle; JBP models for right atrial appendage.
Regulatory Classification
Identification
Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.
Predicate Devices
Arox 53-BP
Arox 60-BP
Arox 45-JBP
Arox 53-JBP
Synox-SX 45-JBP
Synox-SX 53-JBP
Synox SX 53-BP
Synox SX 60-BP
Synox SX 53/15-BP
Synox SX 60/15-BP
PX 53/15-BP
PX 60/15-BP
PX 45-JBP
PX 53-BP
PX 53-JBP
PX 60-BP
YP 45/15-BP
YP 53/15-BP
YP 60/15-BP
Related Devices
K990483 — RETROX RX 45-JBP 124774, 53-JBP 124775, 53-BP 124776, 60-124777 · Biotronik, Inc. · Jun 3, 1999
K023099 — AROX 45-JBP; AROX 53-JBP; MEROX 45-JBP; MEROX 53-JBP · Biotronik, Inc. · Oct 18, 2002
K993139 — YP 45/15-BP; YP 53/15-BP; YP 60/15-BP · Biotronik, Inc. · Dec 17, 1999
K955122 — INTERMEDICS MODEL 436-07 BIPOLAR & MODEL 437-07 UNIPOLAR IMPLANTABLE ENDOCARDIAL PACING LEAD · Intermedics, Inc. · Nov 4, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
Ko61212
April 28, 2006
# Packaging Changes for Endocardial Pacing Leads
Special 510(k) Premarket Notification
# 1. 510(k) SUMMARY
Name and Address of Sponsor:
BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035 JUL 1 0 2000
Establishment Registration Number:
1028232
### Device Name:
| Product | Proprietary Name (Multiple) | FDA Product Code | Classification Name | Device Classification |
|-----------------------------------------|-----------------------------|------------------|---------------------------------------|--------------------------|
| Arox | Arox 53-BP | DTB | Electrode,<br>Pacemaker,<br>Permanent | III (21 CFR<br>870.3680) |
| | Arox 60-BP | | | |
| | Arox 45-JBP | | | |
| | Arox 53-JBP | | | |
| Synox | Synox-SX 45-JBP | DTB | Electrode,<br>Pacemaker,<br>Permanent | III (21 CFR<br>870.3680) |
| | Synox-SX 53-JBP | | | |
| | Synox SX 53-BP | | | |
| | Synox SX 60-BP | | | |
| Synox with 15 mm tip<br>to ring spacing | Synox SX 53/15-BP | DTB | Electrode,<br>Pacemaker,<br>Permanent | III (21 CFR<br>870.3680) |
| | Synox SX 60/15-BP | | | |
| Polyrox w/ 15 mm<br>Spacing | PX 53/15-BP | DTB | Electrode,<br>Pacemaker,<br>Permanent | III (21 CFR<br>870.3680) |
| | PX 60/15-BP | | | |
| Polyrox | PX 45-JBP | DTB | Electrode,<br>Pacemaker,<br>Permanent | III (21 CFR<br>870.3680) |
| | PX 53-BP | | | |
| | PX 53-JBP | | | |
| | PX 60-BP | | | |
| YP | YP 45/15-BP | DTB | Electrode,<br>Pacemaker,<br>Permanent | III (21 CFR<br>870.3680) |
| | YP 53/15-BP | | | |
| | YP 60/15-BP | | | |
### General Description and Predicate Devices:
BIOTRONIK proposes modifications to the packaging process and containers for bradycardia pacing leads, at BIOTRONIK's manufacturing and sterilization facility (BIOTRONIK GmbH & Co. KG) in Pertin, Germany and contract manufacturing facility (BIOTRONIK AG) in Bülach Switzerland. BIOTRONIK believes that the previously approved products cleared through 510(k) notifications and manufactured and sterilized at BIOTRONIK GmbH & Co. KG in Berlin, Germany or BIOTRONIK AG in Bulach Switzerland are appropriate as predicate devices for the products affected by the aforementionaed packaging modifications. The above table provides a list of the products that are manufacturned nad/or sterlized at BIOTRONIK GmbH & Co. KG in Berlin, Germany or BIOTRONIK & G in Billadoh Switzrdand and the 510(k) notifications under which the products were cleared.
{1}------------------------------------------------
#### Indication for Use:
BIOTRONIK's endocardial leads are designed for use with implantable pulse generators that require pacing leads with a bipolar IS-1 connector configuration; they may be used with single or dual chamber pacing systems. The leads are designed for use in patients for whom single or dual chamber pulse generator therapy is medically indicated. This indication follows that recommended in the Class t definition of the ACC/AHA/NASPE Task Force Report, entitled "ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee on Pacemaker Implantation)" (Gregoratos et al. 2002).
The straight lead models are intended for placement in either the right atrium or right ventricle. The JBP lead models have a pre-formed J-shaped distal end to facilitate lead placement in the right atrial appendage.
| Name and Address of Manufacturer: | BIOTRONIK GmbH & Co. KG (reg. no. 9610139)<br>Woermannkehre 1,<br>12359 Berlin, Germany<br>011-49-30-689-05-1210 |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Name and Address of Contract Manufacturer: | BIOTRONIK AG (reg. no. 8043892)<br>Ackerstrasse 6<br>8180 Bülach,<br>Switzerland 011-41-44-864-5169 |
| Contact Person(s) and Phone Number: | Jon Brumbaugh<br>Director, Regulatory Affairs and Compliance<br>Phone (888) 345-0374<br>Fax (503) 635-9936<br>jon.brumbaugh@biotronik.com |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 0 2006
Biotronik, Inc. C/O Jon Brumbaugh 6024 Jean Rd. Lake Oswego, OR 97035
Re: K061212
Trade/Device Name: Biotronik Endocardial Pacing Leads Regulation Number: 21 CFR 870.3680 Regulation Name: Cardiovascular permanent or temporary pacemaker electrode. Regulatory Class: Class III Product Code: DTB Dated: June 21, 2006 Received: June 22, 2006
Dear Mr. Brumbaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Mr. Jon Brumbaugh
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely vours.
Ashley B. Bvan
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
### 510(k) Number (if known):
### Device Name: Bradycardia Pacing Leads, (see attached list)
#### Indications for Use:
#### Pacing Leads -
BIOTRONIK's endocardial leads are designed for use with implantable pulse generators that require pacing leads with a bipolar IS-1 connector configuration they may be used with single or dual chamber pacing systems. The leads are designed for use in patients for whom single or dual chamber pasing of pasing therapy is medically indicated. This indication for them single of the Class I definition of the ACC/AHANASPE This Indication This Indication inal recommended in Title Class I demind of the of Cardiac Pacemakers Reports, Aladoa Aloon More 2 2002 Ghuelle Opdate of Inplantabon
CordialemyAmadon Mond Anad Anada Arabarhythmia Devices: A Report of the American Colleg CardiologyAmerican Heart Association Task Force on Practice Guidelines (ACCAHANASPE Committee on Pacemaker Implantation)" (Gregoratos et al. 2002).
The straight lead models are intended for placement in either the right atrium or right ventricle. The JBP lead models have a pre-formed distal end to facilitate lead placement in the ight atrial appendage.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arilla B. Beane
510(k) Number LC61212
Page 1 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.