K250079 · Penumbra, Inc. · KRD · Mar 14, 2025 · Cardiovascular
Device Facts
Record ID
K250079
Device Name
Ruby XL System
Applicant
Penumbra, Inc.
Product Code
KRD · Cardiovascular
Decision Date
Mar 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic, PCCP
Indications for Use
The Ruby XL System is indicated for arterial and venous embolizations in the peripheral vasculature.
Device Story
Ruby XL System is a vascular embolization device consisting of a platinum embolization coil attached to a composite delivery pusher, used with a Penumbra Detachment Handle. The device is used by physicians in clinical settings to perform arterial and venous embolizations in the peripheral vasculature. The coil is delivered via a catheter to the target site; once positioned, the physician uses the mechanical detachment handle to release the coil, which serves to occlude the vessel. The system utilizes mechanical detachment technology consistent with predicate devices. The device is supplied sterile for single use. By providing a controlled method for vessel occlusion, the device assists in managing peripheral vascular abnormalities.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by bench-top performance testing, including dimensional/visual inspection, packaging integrity, radiopacity, friction, fatigue, torsional resistance, stiffness, simulated use, corrosion, tensile, and MRI compatibility testing. Biocompatibility was evaluated per ISO 10993-1, covering cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation, hemocompatibility, and chronic toxicity. Shelf-life of 12 months was established via accelerated aging.
Technological Characteristics
Materials: Platinum/tungsten coil, adhesive, titanium, polymer delivery pusher. Sensing/Actuation: Mechanical detachment mechanism. Energy: None (manual operation). Form factor: Embolization coil with delivery pusher. Connectivity: None. Sterilization: Ethylene Oxide (SAL >10^-6). Software: None. Standards: ISO 10993-1 for biocompatibility.
Indications for Use
Indicated for arterial and venous embolizations in the peripheral vasculature for patients requiring vascular occlusion.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
Predicate Devices
Penumbra Coil System (Penumbra Coil 400 and Ruby Coil System) and POD System (K173614)
Related Devices
K992121 — EMBOLIZATION COIL SYSTEM · Cook, Inc. · Dec 20, 1999
K162524 — AZUR CX Detachable 18 Peripheral Coil System · MicroVention, Inc. · Mar 3, 2017
K253511 — Concerto Versa Detachable Coil · Medtronic, Inc. · Jan 14, 2026
K173614 — Penumbra Coil 400, Ruby Coil System, POD System · Penumbra, Inc. · Apr 17, 2018
K103305 — PENUMBRA EMBOLIZATION COIL SYSTEM · Penumbra, Inc. · Jan 26, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 14, 2025
Penumbra, Inc. Samyukta Rangachari Regulatory Affairs Specialist One Penumbra Place Alameda, California 94502
Re: K250079
Trade/Device Name: Ruby XL System Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: January 10, 2025 Received: January 13, 2025
Dear Samyukta Rangachari:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new
{1}------------------------------------------------
premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part
{2}------------------------------------------------
803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Finn E. Finn E. Donaldson -S Donaldson -S Pate: 2025.03.14 09:04:35 -04'00'
Finn Donaldson Acting Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K250079
Device Name Ruby XL System
Indications for Use (Describe)
The Ruby XL System is indicated for arterial and venous embolizations in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary
#### 1. Submitter
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
Contact Person: Samyukta Rangachari Regulatory Affairs Specialist III Phone: (763) 742-9306 Email: srangachari@penumbrainc.com
Date of Preparation: January 10, 2025
## 2. Subject Device
Ruby XL System
Regulatory Class: ll Classification Panel: Cardiovascular Classification Name: Vascular embolization device Regulation Number: 21 CFR §870.3300 Product Code: KRD
## 3. Predicate/Reference Devices
| 510(k) Number | Name of Device |
|---------------|------------------------------------------------------------------------------|
| Predicate | |
| K173614 | Penumbra Coil System (Penumbra Coil 400 and Ruby Coil System) and POD System |
{5}------------------------------------------------
# 4. Predicate Comparison
# Table 1: Predicate Comparison
| Attributes | Predicate Device | Subject Device |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| General | | |
| Trade Name | Penumbra Coil System (Penumbra Coil 400 and Ruby Coil System);<br>POD System | Ruby XL System |
| 510(k) No. | K173614 | K250079 |
| Indications for use | Indicated for the embolization of:<br>• Intracranial aneurysms<br>• Other neurovascular abnormalities such as arteriovenous<br>malformations and arteriovenous fistulae<br>• Arterial and venous embolizations in the peripheral vasculature | Indicated for arterial and<br>venous embolizations in the<br>peripheral vasculature. |
| Classification | Class II, HCG & KRD | Class II, KRD |
| Materials/Construction | | |
| Coil Implant | Platinum/Tungsten, Adhesive, Titanium, Polymer | SAME |
| Introducer Sheath | Polymer | SAME |
| Dimensions/Shape | | |
| Coil Secondary Shape | Complex, Finish | SAME |
| Coil Length | Appropriately Sized For Target Vasculature | SAME |
| Coil Primary Diameter | | |
| Coil Secondary Diameter | | |
| Detachment Pusher Length | | |
| Detachment Mechanism | Mechanical | SAME |
| Other | | |
| Sterilization Method, SAL | Ethylene Oxide, >10-6 | SAME |
| Method of Supply | Sterile, single use only | SAME |
| Device Packaging Materials and<br>Dimensions | As specified in K173614 | SAME |
{6}------------------------------------------------
# 5. Device Description
The Ruby XL System is comprised of a platinum embolization coil attached to a composite delivery pusher, and the Penumbra Detachment Handle. The coil/delivery pusher is packaged separately from the Penumbra Detachment Handle.
# 6. Indications for Use
The Ruby XL System is indicated for arterial and venous embolizations in the peripheral vasculature.
## 7. Summary of Non-Clinical Data/Performance Data
Included in this section are summary descriptions of the testing which substantiates the performance of the subject Ruby XL System.
## 7.1 Summary of Performance Testing – (Bench-Top) Testing
The following bench-top performance (design verification) tests were used in support of the subject device to determine substantial equivalence:
- Dimensional/Visual
- Packaging Inspection, Leak and Seal Strength Testing
- Radiopacity Testing
- Friction Testing
- Fatigue Resistance Testing ●
- Torsional Resistance Testing
- Stiffness Testing
- Simulated Use and Compatibility Testing
- Corrosion Testing
- Tensile Testing
- MRI Compatibility Testing
#### 7.2 Summary of Biocompatibility
Biocompatibility was evaluated for the subject device in accordance with ISO 10993-1 and the FDA Guidance for Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
The following endpoints were evaluated in support of the subject device to determine substantial equivalence:
{7}------------------------------------------------
- Cytotoxicity Testing ●
- . Sensitization Testing
- Irritation Testing
- Systemic Toxicity: Acute Systemic Injection Testing & Material Mediated Pyrogenicity ● Testing
- Subacute/Subchronic Toxicity Testing ●
- . Genotoxicity Testing
- Implantation Testing ●
- Hemocompatibility Testing ●
- Chronic Toxicity Testing ●
- Carcinogenicity Assessment
# 7.3 Summary of Shelf-Life
The shelf-life of the subject device is established for 12 months based on accelerated aging data.
# 8. Summary of Performance Data – Animal, Clinical
No animal or clinical study was conducted as bench testing was determined sufficient for verification and validation purposes.
# 9. Summary of Substantial Equivalence
The subject Ruby XL System is substantially equivalent to the predicate device with regards to intended use, operating principle, design concept, fundamental technology, and device performance.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.