Torpedo Gelatin Foam is indicated for use in embolization of hypervascular tumors.
Device Story
Torpedo Gelatin Foam is a resorbable, hydrophilic porcine gelatin device used for vascular embolization. It is supplied as a compressed, cylindrical dry foam preloaded in a cartridge. During clinical use, the device is rehydrated and injected into a target vessel via an intravascular or micro-catheter. Upon delivery, the device expands to provide a mechanical barrier to blood flow, effectively occluding the vessel. The procedure is typically monitored by a physician using fluoroscopy with contrast enhancement. The device is intended for single use in a clinical setting to treat hypervascular tumors by restricting blood supply. It is provided sterile and includes a blunt stylet for delivery.
Clinical Evidence
No human clinical trials were conducted. Evidence consists of bench testing and an animal study. The animal study (n=15; 7 test, 8 control) in adult female sheep renal arteries evaluated safety and degradation over 4 weeks. Results showed successful embolization with no recovery of normal flow and no significant difference in artery patency between test and predicate devices. Bench testing included macroscopic inspection, swelling, humidity, sterility, pyrogenicity, deliverability, and resorption testing.
Technological Characteristics
Material: Resorbable porcine gelatin. Form: Compressed cylindrical foam (0.9mm or 1.7mm diameter; 10mm or 20mm length). Principle: Mechanical vascular occlusion via expansion upon hydration. Connectivity: None. Sterilization: Radiation (ISO 11137-1). Standards: ASTM F2096-11, F1140/F1140M-13, F88/F88M-15, D4332-14, D4169-16, F1980-16, ISO 11607-1:2018, ISO 7864:2016, ISO 10993-1:2010.
Indications for Use
Indicated for embolization of hypervascular tumors in patients requiring vascular occlusion.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 18, 2019
Biosphere Medical, S.A. Alix Fonlladosa Regulatory Affairs Manager Parc des Nations - Paris Nord 2, 383 rue de la Belle Etoile Roissy en France, 95700 Fr
Re: K183578
Trade/Device Name: Torpedo Gelatin Foam Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular embolization device Regulatory Class: Class II Product Code: KRD Dated: May 20, 2019 Received: May 21, 2019
Dear Alix Fonlladosa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183578
Device Name Torpedo Gelatin Foam
Indications for Use (Describe) Intended Use: Torpedo Gelatin Foam is indicated for use in embolization of hypervascular tumors
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
## 510(k) Summary K183578
June 12, 2019
| General<br>Provisions | Submitter Name: | Biosphere Medical, S.A. | | |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------|-------------------|
| | Address: | Parc des Nations - Paris Nord 2<br>383, rue de la Belle Etoile<br>95700 Roissy-en-France, France | | |
| | Telephone Number: | +33 (0)1 48 17 25 25 | | |
| | Fax Number: | +33 (0)1 49 38 02 68 | | |
| | Contact Person: | Rosène Amossé | | |
| | Registration Number: | 9615728 | | |
| Subject<br>Device | Trade Name: | Torpedo Gelatin Foam | | |
| | Common/Usual Name: | Embolization device | | |
| | Classification Name: | 21 CFR § 870.3300 Vascular Embolization Device | | |
| Predicate<br>Device | Trade Name: | EmboCube Embolization Gelatin | | |
| | Classification Name: | 21 CFR § 870.3300 Vascular Embolization Device | | |
| | Premarket Notification: | K181021 | | |
| | Manufacturer: | Biosphere Medical, S.A. | | |
| Classification | Class II<br>21 CFR § 870.3300<br>FDA Product Code: KRD<br>Review Panel: Division of Cardiovascular Devices | | | |
| Intended Use | Torpedo Gelatin Foam is indicated for use in embolization of hypervascular<br>tumors. | | | |
| | Torpedo Gelatin Foam is a hydrophilic medical device which consists of<br>resorbable dry gelatin foam that is compressed into a cylindrical shape and<br>preloaded into a cartridge with a standard female and male luer fittings. The<br>device is available in two sizes (0.9 mm and 1.7 mm) and two lengths<br>configurations (10 mm and 20 mm). | | | |
| Device<br>Description | Reference code | Compressed /<br>Dehydrated<br>Torpedo Diameter | Uncompressed /<br>Hydrated Torpedo<br>Section | Torpedo<br>Length |
| | TOR2510 | 0.9 mm | 2.5 mm x 2.5 mm | 10 mm |
| | TOR2520 | 0.9 mm | 2.5 mm x 2.5 mm | 20 mm |
| | TOR5010 | 1.7 mm | 5.0 mm x 5.0 mm | 10 mm |
| | TOR5020 | 1.7 mm | 5.0 mm x 5.0 mm | 20 mm |
{4}------------------------------------------------
Safetv &
Performance
Tests
Once rehydrated, the deformable torpedoes can be injected into the target vessel with an intravascular catheter or a micro-catheter (depending on the size range) to provide a mechanical barrier to blood flow. Contrast enhancement may be used to monitor the embolization procedure using fluoroscopy. The device is intended for single use and is provided sterile.
Torpedo Gelatin Foam is similar in design to the predicate device, EmboCube Embolization Gelatin (K181021), as both of them provide a mechanical barrier to blood flow in the vasculature and are delivered using catheters. The changes to the device are as follows: a) A variant of the gelatin cube device has been added to the range, b) A new packaging has been added, c) A blunt stylet has been included.
The subject device is available in two sizes, as the predicate device. The subject and predicate devices differ in shape. The predicate EmboCube Embolization Gelatin (K181021) has a cubic shape, whether dry or Comparison to hydrated. At the dry state, the subject Torpedo Gelatin Foam is compressed Predicate into a cylindrical shape. In the uncompressed / hydrated state, the subject Torpedo Gelatin Foam is a solid rectangle of the same section as the predicate (2.5 mm or 5.0 mm), only the length differs. Both the subject device and the predicate device, EmboCube Embolization Gelatin, K181021 are made wholly from the identical resorbable porcine qelatin.
> The technological differences between the subject and predicate devices have been evaluated through biocompatibility, bench and animal testing to provide evidence of substantial equivalence of the Torpedo Gelatin Foam to the predicate device, EmboCube Embolization Gelatin (K181021).
> No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. However, vascular embolization devices are subiect to the special controls specified in "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices", issued on December 29, 2004. In addition, the subject device follows the FDA Draft Guidance on Medical Devices Containing Materials Derived from Animal Sources (Except for In Vitro Diagnostic Devices), issued on January 23, 2014.
> A battery of tests performed on Torpedo Gelatin Foam were designed to demonstrate that the device meets critical design specifications as well as clinical performance attributes for its intended use. Where appropriate, the tests were based on the requirements of the following documents:
- FDA Guidance Class II Special Controls Guidance Document: Vascular ● and Neurovascular Embolization Devices (December 29, 2004)
- ASTM F2096-11, Standard Test Method for Detecting Gross Leaks in ● Medical Packaging by Internal Pressurization (Bubble Test)
- ASTM F1140/F1140M-13, Standard Test Methods for Internal ● Pressurization Failure Resistance of Unrestrained Packages for Medical Applications
- ASTM F88/F88M-15, Standard Test Method for Seal Strength of . Flexible Barrier Materials
- ASTM D4332-14, Standard Practice for Conditioning Containers, ● Packages, or Packaging Components for Testing
- ASTM D4169-16, Standard Practice for Performance Testing of ● Shipping Containers and Systems
{5}------------------------------------------------
- ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile ● Barrier Systems for Medical Devices
- . NF EN ISO 11137-1:2016, Sterilization of health care products --Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- . NF EN ISO 10993-1:2010. Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process, and the FDA Modified ISO 10993 Test Profile FDA Memo G95-1.
- NF EN ISO 11607-1:2018. Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- ISO 7864:2016, Sterile hypodermic needles for single use --● Requirements and test methods
Torepedo Gelatin Foam was evaluated for various performance attributes to demonstrate the substantial equivalence of the device to the predicate. The following is a list of all testing that was successfully completed:
- Macroscopic appearance (visual inspection) ●
- Compressed (dry) torpedo size ●
- Maximum swelling (hydrated size) ●
- Humidity (H₂0 analysis)
- Sterility
- Pvrogenicity (Bacterial endotoxins) ●
- Deliverability with a stylet ●
- Deliverability with catheters and radiopaque contrast agents
- . Flush test carried out on aged and sterilized subject devices
- Resorption .
- Embolic effect .
- . Comparative FTIR analysis for any foreign material transfer
Safety & Performance Tests (continued)
Biocompatibility testing was adopted from the predicate device EmboCube Embolization Gelatin, K181021. The only difference between the subject device and the legally marketed predicate device is that the subject device will be crimped and shaped into a torpedo. FTIR spectroscopy was performed in order to verify that there is no transfer of material from the crimper to the subject device. No further biocompatibility testing was required to qualify the subject device.
An animal study was conducted on seven test (Torpedo Gelatin Foam) and eight control (EmboCube Embolization Gelatin, K181021) adult female sheep in the renal arteries over a period of 4 weeks. The animal study evaluated vascular safety, local tissue effects, and in vivo degradation of the Torpedo Gelatin Foam. Angiography and/or histopathology confirmed that the Torpedo Gelatin Foam provided an embolic effect in renal arteries for at least four weeks. Normal flow of the renal arterial vasculature was never recovered in any animal at any time point. There was no significant difference for artery patency between any articles at any time point. All test animals had a necrotized renal tissue, indicative of successful embolization. The animals were clinically normal throughout the study duration.
The difference between the modified device and the cleared device. Biosphere EmboCube Embolization Gelatin [K181021] has raised no new issues.
{6}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.