K052834 · Cook, Inc. · KRD · Nov 23, 2005 · Cardiovascular
Device Facts
Record ID
K052834
Device Name
MREYE EMBOLIZATION COIL
Applicant
Cook, Inc.
Product Code
KRD · Cardiovascular
Decision Date
Nov 23, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Used for peripheral arterial and venous vessel embolization procedures.
Device Story
MReye® Embolization Coil is a sterile, single-use device for peripheral arterial and venous vessel embolization. Constructed of Inconel wire (0.025, 0.035, 0.038 inch diameters) with synthetic fibers; available in curled or straight shapes (0-15 mm). Device is pre-loaded in a shipping cannula. Physician introduces coil into target vessel via angiographic catheter; deployment achieved by pushing coil out of catheter using a wire guide. Device provides mechanical occlusion of vessels. Clinical benefit derived from targeted embolization in peripheral vasculature.
Clinical Evidence
Bench testing only. Testing included delivery friction, coil connection, coil deformation, wire tensile strength, fiber pull-out, magnetic resonance (MR) compatibility, and biocompatibility. No clinical data provided.
Technological Characteristics
Materials: Inconel wire, synthetic fibers. Dimensions: 0.025, 0.035, 0.038 inch wire diameters; 0-15 mm emboli size. Form factor: Curled or straight coil pre-loaded in shipping cannula. Energy source: None (mechanical). Connectivity: None. Sterilization: Sterile, single-use.
Indications for Use
Indicated for patients requiring peripheral arterial and venous vessel embolization.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
Predicate Devices
Cook Embolization Coil
Submission Summary (Full Text)
{0}------------------------------------------------
K052834
-6-
Traditional 510(k) Submission MReve® Embolization Coil COOK INCORPORATED 5 October 2005
NOV 2 3 2005
# 510(k) SUMMARY
Submitted By: Lisa Webb, MBA, RAC Senior Regulatory Affairs Specialist Cook Incorporated 750 Daniels Way P.O. Box 489 Bloomington, IN 47402 (812) 339-2235 x 2643 4 October 2004
Device: Trade Name: MReye® Embolization Coil
Proposed Classification Name: Arterial Embolization Device
## Predicate Devices:
The MReye® Embolization Coil is similar in terms of intended use, materials of construction, and technological characteristics to the predicate Cook Embolization Coil.
### Device Description
The MReye® Embolization Coil is supplied sterile and is intended for one time use. The embolization coil is pre-loaded in a shipping cannula. The coil is constructed of Inconel and has wire diameters of 0.025, 0.035, and 0.038 inches. It is available in curled and straight shapes. The emboli size range is 0 mm (straight) to 15 mm. The embolization coil includes synthetic fibers evenly placed down the length of the coil. The device is introduced to the target vessel using an angiographic catheter (sold separately) and deployment is achieved by a wire guide (sold separately) pushing the coil out of the catheter.
### Substantial Equivalence
Cook currently markets the Cook Embolization Coil which is substantially equivalent to the MReye® Embolization Coil. The similar indications for use and technological characteristics of the MReye® Embolization Coil as compared to the predicate device support a determination of substantial equivalency.
{1}------------------------------------------------
Traditional 510(k) Submission MReye® Embolization Coil COOK INCORPORATED 5 October 2005
### Test Data
The MReye® Embolization Coil was subjected to the following tests to ensure reliable design and performance under the specified testing parameters.
- Delivery Friction Testing ■
- I Coil Connection Testing
- . Coil Deformation Testing
- Wire Tensile Strength
- L Fiber Pull Out Testing
- 을 Magnetic Resonance (MR) Testing
- 피 Biocompatibility Testing
The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its intended use in peripheral arterial and venous vessel embolization procedures.
{2}------------------------------------------------
Image /page/2/Picture/10 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 3 2005
Cook Incorporated c/o Ms. Lisa Webb Senior Regulatory Affairs Specialist P.O. Box 489 Bloomington, IN 47402-0489
Re: K052834 MReye® Embolization Coil Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II (Two) Product Code: KRD Dated: October 5, 2005 Received: October 6, 2005
Dear Ms. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Ms. Lisa Webb
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Duina R. Viihner
Image /page/3/Picture/5 description: The image shows the text "Bram D. Zuckerman, M.D. Director". The text is left-aligned and in a simple, sans-serif font. The text is likely a name and title, possibly indicating a person's name, credentials, and position within an organization.
Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Traditional 510(k) Submission MReye® Embolization Coil COOK INCORPORATED 5 October 2005
#### Indications for Use
510(k) Number (if known): __ Ko 52834
Device Name:
MReye® Embolization Coil
Indications for Use:
Used for peripheral arterial and venous vessel embolization procedures.
Prescription Use XX (Per 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use_ (21 CRF 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
me Munica
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices
510(k) Number_Ko