DATASCOPE'S 8FR. ALT B IAB; DATASCOPE'S PROFILE 8FR. ALT B IAB; DATASCOPE'S TRUE SHEATHLESS DL 9.5 FR. IAB; DATASCOPE'S
Applicant
Datascope Corp.
Product Code
DSP · Cardiovascular
Decision Date
Apr 24, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3535
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intra-aortic balloon is placed in the descending aorta just below the subclavian artery and is intended to improve cardiovascular functioning during the following situations: Refractory unstable angina . Impending infarction Acute MI . . Refractory ventricular failure Complications of acute MI (i.e. Acute MR or VSD, or papillary muscle rupture) . . Cardiogenic shock . Support for diagnostic, percutaneous revascularization, and interventional procedures. Ischemia related intractable ventricular arrhythmias ● . Septic shock . Intraoperative pulsatile flow generation ● . Weaning from bypass Cardiac support for non-cardiac surgery . Prophylactic support in preparation for cardiac surgery . Post surgical myocardial dysfunction/low cardiac output syndrome ● . Myocardial contusion Mechanical bridge to other assist devices . . Cardiac support following correction of anatomical defects
Device Story
Intra-aortic balloon (IAB) catheter placed in descending aorta; improves cardiovascular function via counterpulsation. Used in clinical settings (OR, ICU) by physicians. Device provides mechanical circulatory support; assists in managing cardiac failure, shock, and surgical recovery. Output is hemodynamic support; helps stabilize patient hemodynamics; reduces myocardial workload. No software or algorithm involved.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Intra-aortic balloon catheter; mechanical circulatory support device. No software, electronics, or energy source integral to the catheter itself (catheter is a passive component of the IAB pump system).
Indications for Use
Indicated for patients requiring cardiovascular support due to refractory unstable angina, impending or acute MI, refractory ventricular failure, complications of acute MI, cardiogenic shock, septic shock, myocardial contusion, or post-surgical myocardial dysfunction. Also indicated for support during diagnostic/interventional procedures, intraoperative pulsatile flow generation, weaning from bypass, non-cardiac surgery support, prophylactic cardiac surgery support, mechanical bridging to other assist devices, and support following correction of anatomical defects.
Regulatory Classification
Identification
An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K020257
# APR 2 4 2002
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## FOR
## DATASCOPE Intra-Aortic Balloon Catheter Systems
(Prepared in accordance with 21 CFR Part 807.92)
Pursuant to Section 513(I)(3)(A) of the Food, Drug, and Cosmetic act, Datascope Corp. is required to submit this Premarket Notification either an "…adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Datascope Corp. chooses to submit a summary of information respecting safety and effectiveness.
#### GENERAL INFORMATION A.
| Submitter: | Datascope Corp. |
|-----------------|------------------------------|
| | Cardiac Assist Division |
| Address: | 15 Law Drive |
| | Fairfield, NJ 07004 |
| Contact Person: | JoAnn Wolf |
| | Regulatory Affairs Associate |
#### B. DEVICE INFORMATION
| Generic Name: | Intra-Aortic Balloon (IAB) Catheter |
|----------------------|------------------------------------------------------------------------------|
| Trade Name: | Datascope Intra-Aortic Balloon (IAB) Catheter |
| Classification Name: | Intra-Aortic Balloon (IAB) Catheters are classified under 21 CFR<br>870.3535 |
Product Code: 74DSP
#### ﺰ PREDICATE DEVICE INFORMATION
Datascope's Intra-Aortic Balloon Catheters Indications for Use are substantially equivalent to the following marketed devices:
| K013326 - | Datascope 8Fr. 25, 34 & 40cc Alt B IAB |
|-----------|------------------------------------------------------------------------------------------------------------------------------------------|
| K003598 - | Datascope Profile 8Fr. 34 & 40cc Alt B IAB |
| K980780 - | Datascope Percor STAT-DL 9.5Fr. 25, 34 & 40cc Alt B IAB |
| K964987 - | Percor STAT-DL 9.5Fr. 25 & 40cc IABs for Optional Sheathless Insertion<br>and Percor STAT-DL 10.5Fr. 40 & 50cc IABs for Sheath Insertion |
| K965236 - | Datascope System 96 Intra-Aortic Balloon Pump |
| K002256 - | Arrow ACAT 2 Intra-Aortic Balloon Pump |
All of the above referenced premarket notifications were found substantially equivalent by FDA.
{1}------------------------------------------------
#### D. DEVICE DESCRIPTION/INTENDED USE
The intra-aortic balloon is placed in the descending aorta just below the subclavian artery and is intended to improve cardiovascular functioning during the following situations:
- Refractory unstable angina .
- Impending infarction
- Acute MI .
- . Refractory ventricular failure
- Complications of acute MI (i.e. Acute MR or VSD, or papillary muscle rupture) .
- . Cardiogenic shock
- . Support for diagnostic, percutaneous revascularization, and interventional procedures.
- Ischemia related intractable ventricular arrhythmias ●
- . Septic shock
- . Intraoperative pulsatile flow generation
- Weaning from bypass
- Cardiac support for non-cardiac surgery
- . Prophylactic support in preparation for cardiac surgery
- Post surgical myocardial dysfunction/low cardiac output syndrome ●
- . Myocardial contusion
- Mechanical bridge to other assist devices .
- . Cardiac support following correction of anatomical defects
#### E. TECHNOLOGICAL CHARACTERISTICS
Datascope's enhanced Indications for Use for IAB Systems are substantially equivalent to the predicate devices.
Supporting published data establish that the expanded Indications for Use is substantially equivalent to the predicate devices and does not affect the safety and effectiveness of the device.
#### F. NON-CLINICAL TESTS
N/A
#### G. CLINICAL TESTS
N/A
#### H. CONCLUSIONS
Based on the information presented in this 510(k) premarket notification, Datascope's expanded Indications for Use for the IAB Catheter System are considered substantially equivalent to the predicate devices' Indications for Use for their currently marketed IAB Catheter System.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 4 2002
Ms. JoAnn Wolf Regulatory Affairs Associate Datascope Corporation Cardiac Assist Division 15 Law Drive Fairfield, NJ 07004
Re: K020257
Trade Name: Datascope's 8Fr. 25, 34 & 40cc IAB; True Sheathless DL 9.5Fr. 25, 34 & 40cc IAB; Percor STAT-DL 9.5Fr. 34 & 40cc IAB for Optional Sheathless Insertion; Percor STAT-DL 10.5Fr. 50cc IAB Regulation Number: 21 CFR 870.3535 Regulation Name: Intra-Aortic Balloon Catheter Regulatory Class: Class III (three) Product Code: DSP Dated: January 23, 2002 Received: January 24, 2002
Dear Ms. Wolf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{3}------------------------------------------------
## Page 2 - Ms. JoAnn Wolf
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loase of actived and I Dr unation that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny I other battler and states and station of limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Orth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premaince nodicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Rödl & Partner
Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# 510(k) Number (if known): Known): Known)
Datascope's 8Fr. 25, 34 & 40cc IAB; True Sheathless DL 9.5Fr., 25, 34 & 40cc IAB; Device Name: Percor STAT-DL 9.5Fr. 34 & 40cc IAB for Optional Sheathless Insertion; Percor STAT-DL 10.5Fr. 50cc IAB
Indications for Use:
## Intra-Aortic Balloon Catheter System to be used for the following:
- Refractory unstable angina .
- . Impending infarction
- Acute MI .
- . Refractory ventricular failure
- Complications of acute MI (i.e. Acute MR or VSD, or papillary muscle rupture) ●
- Cardiogenic shock ●
- Support for diagnostic, percutaneous revascularization, and interventional procedures. .
- Ischemia related intractable ventricular arrhythmias .
- . Septic shock
- Intraoperative pulsatile flow generation ●
- . Weaning from bypass
- . Cardiac support for non-cardiac surgery
- Prophylactic support in preparation for cardiac surgery .
- Post surgical myocardial dysfunction/low cardiac output syndrome .
- . Myocardial contusion
- Mechanical bridge to other assist devices .
- Cardiac support following correction of anatomical defects .
This information can be found in Section 3 - Attachment I of our Premarket Notification under the section titled "II. Indications".
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Over-the-Counter Use OR
(Optional Format 1-2-96)
Division of Cardiovascular & Respiratory Devices
510(k) Number K000257
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.