DATASCOPE PERCOR STAT - DL 9.5 FR. 25 AND 40CC INTRA-AORTIC BALLOON FOR OPTIONAL SHEATHLESS INSERTION AND PERCOR STAT -0

K964987 · Datascope Corp. · DSP · Mar 17, 1997 · Cardiovascular

Device Facts

Record IDK964987
Device NameDATASCOPE PERCOR STAT - DL 9.5 FR. 25 AND 40CC INTRA-AORTIC BALLOON FOR OPTIONAL SHEATHLESS INSERTION AND PERCOR STAT -0
ApplicantDatascope Corp.
Product CodeDSP · Cardiovascular
Decision DateMar 17, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3535
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intra-aortic balloon is placed in the descending aorta just below the subclavian artery and is intended to improve cardiovascular functioning during the following situations: Refractory ventricular failure Cardiogenic shock Unstable refractory angina Impending infarction Mechanical complications due to acute myocardial infarction Ischemic related intractable ventricular arrhythmias Cardiac support for high risk surgical patients and coronary angiography or angioplasty patients Septic shock Weaning from cardiopulmonary bypass Interoperative pulsatile flow generation Support for failed angioplasty and valvuloplasty

Device Story

Intra-aortic balloon (IAB) catheter; placed in descending aorta below subclavian artery; provides mechanical circulatory support. Device functions by inflating/deflating balloon to improve cardiovascular hemodynamics. Used in clinical settings (OR, cath lab) by physicians. Output is hemodynamic support; assists cardiac output; reduces myocardial oxygen demand. Benefits include stabilization of patients in shock or during high-risk cardiac procedures.

Clinical Evidence

No clinical evaluation performed in the U.S. Substantial equivalence supported by in-vitro bench testing comparing functionality and performance to predicate devices.

Technological Characteristics

Intra-aortic balloon catheter. Dimensions and form factor identical to predicate devices. Materials modified in grade and chemical composition compared to predicates; verified via in-vitro testing. No software or electronic components described.

Indications for Use

Indicated for patients requiring cardiovascular support due to refractory ventricular failure, cardiogenic shock, unstable refractory angina, impending infarction, mechanical complications of acute myocardial infarction, ischemic-related intractable ventricular arrhythmias, high-risk surgical/angiography/angioplasty procedures, septic shock, weaning from cardiopulmonary bypass, intraoperative pulsatile flow generation, or failed angioplasty/valvuloplasty.

Regulatory Classification

Identification

An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.

Special Controls

*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device; (ii) Software verification, validation, and hazard analysis must be performed; (iii) The device must be demonstrated to be biocompatible; (iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; (v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and (vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device. (2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation. (c) *Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0} D K964987 MAR 17 1997 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS FOR ## DATASCOPE PERCOR STAT-DL® 9.5Fr. & 10.5 Fr. INTRA-AORTIC BALLOON (IAB) (Prepared in accordance with 21 CFR Part 807.92) Pursuant to Section 513(I)(3)(A) of the Food, Drug, and Cosmetic act, datascope corporation is required to submit this Premarket Notification either an “...adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person.” Datascope Corporation chooses to submit a summary of information respecting safety and effectiveness. ## A. GENERAL INFORMATION Submitter: Datascope Corp. Cardiac Assist Division Address: 15 Law Drive Fairfield, NJ 07004 Contact Person: Kevin Crossen Director, Regulatory Affairs and Quality Assurance ## B. DEVICE INFORMATION Generic Name: Intra-Aortic Balloon (IAB) Trade Name: Datascope Percor STAT-DL® Intra-Aortic Balloon (IAB) Classification Name: Intra-Aortic Balloons (IABs) are classified under 21 CFR 870.3535 Product Code: 74DSP ## C. PREDICATE DEVICE INFORMATION Datascope’s Percor STAT-DL® Intra-Aortic Balloon is substantially equivalent to the following marketed devices: - Datascope Percor STAT-DL® 9.5Fr. & 10.5Fr. Intra-Aortic Balloons, K940231 & K940178. - K905663 - Pediatric IAB - K790775 - Datascope Type “S” IAB - K952221 - 30cc/40cc Sensation™ IAB; Model 930/940 IAB’s (Boston Scientific Corp.) 23 {1} D Summary of Safety & Effectiveness/Datascope Percor STAT-DL® 9.5 & 10.5Fr. IAB Page 2 - K943919 - 9.5 Fr. Sensation™ IAB (Boston Scientific Corp.) - K961358 - 8.0 Fr. 30/40cc, 10.0 Fr. 50cc Sheathless IAB (Arrow International) ## D. DEVICE DESCRIPTION/INTENDED USE The intra-aortic balloon is placed in the descending aorta just below the subclavian artery and is intended to improve cardiovascular functioning during the following situations: - Refractory ventricular failure - Cardiogenic shock - Unstable refractory angina - Impending infarction - Mechanical complications due to acute myocardial infarction - Ischemic related intractable ventricular arrhythmias - Cardiac support for high risk surgical patients and coronary angiography or angioplasty patients - Septic shock - Weaning from cardiopulmonary bypass - Interoperative pulsatile flow generation - Support for failed angioplasty and valvuloplasty ## E. TECHNOLOGICAL CHARACTERISTICS Datascope’s Percor STAT-DL® 9.5Fr. & 10.5Fr. IAB are identical to the predicate devices with regard to its indications for use and dimensional specifications. They differ technologically respecting material grade and chemical composition of the components. The difference in material grade and chemical composition has been demonstrated not to affect safety or efficacy of the device. ## F. NON-CLINICAL TESTS The results of in-vitro tests conducted demonstrate that the functionality and performance characteristics of the device are comparable to the currently marketed devices. {2} Page 3 Summary of Safety & Effectiveness/Datascope Percor STAT-DL® 9.5 & 10.5Fr. IAB ## G. CLINICAL TESTS There has been no clinical evaluation of the new device in the U.S. ## H. CONCLUSIONS Based on the information presented in this 510(k) premarket notification, Datascope’s Percor STAT-DL® 9.5Fr. & 10.5Fr. IABs are considered substantially equivalent to Datascope’s currently marketed IABs.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...