SLEUTH IMPLANTABLE ECG MONITORING SYSTEM

K063035 · Transnoma Medical, Inc. · MXC · Oct 1, 2007 · Cardiovascular

Device Facts

Record IDK063035
Device NameSLEUTH IMPLANTABLE ECG MONITORING SYSTEM
ApplicantTransnoma Medical, Inc.
Product CodeMXC · Cardiovascular
Decision DateOct 1, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2800
Device ClassClass 2

Indications for Use

The Transoma Medical Sleuth 2010 System is an implantable, patient- and automatically-activated monitoring system that records subcutaneous ECG and is indicated for: - patients with clinical syndromes or situations at increased risk of cardiac arrhythmias or . - patients who experience transient symptoms that may suggest a cardiac arrhythmia.

Device Story

Sleuth Implantable ECG Monitoring System consists of Model 2010 Implantable Monitoring Device (IMD), Model 4000 Activator, and Model 5000 Base Station. IMD is a titanium structure implanted subcutaneously in left pectoral region. Device continuously monitors R-R intervals; automatically stores ECG segments when intervals exceed pre-defined limits. Patients use hand-held Activator to manually trigger ECG storage during symptomatic events. Data transfers via telemetry from IMD to Activator, then via Base Station to a service center for physician review. System enables remote monitoring of arrhythmic events; assists physicians in diagnosing unexplained syncope or pre-syncope. Benefits include continuous 24/7 monitoring and capture of both symptomatic and asymptomatic events.

Clinical Evidence

Clinical study conducted at 4 medical centers with 28 patients presenting with unexplained syncope/pre-syncope or arrhythmia risk. Patients monitored 24/7 for 7-10 days, 1, 3, and 6 months. Study demonstrated diagnostic viability; 13 patients were diagnosed with arrhythmias using Sleuth records. Two adverse events reported over 4,700 patient days, deemed not serious or definitively linked to the device.

Technological Characteristics

Titanium IMD structure; subcutaneous implantation. Sensing principle: R-R interval monitoring. Connectivity: Telemetry (IMD to Activator), Bluetooth (Base Station to service center). System includes IMD, Activator, and Base Station. Software-based event detection and storage.

Indications for Use

Indicated for patients at increased risk of cardiac arrhythmias or those experiencing transient symptoms suggestive of cardiac arrhythmia.

Regulatory Classification

Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary K063035 page 1/3 #### General Provisions Submitter's Name and Address Transoma Medical, Inc. 4358 West Round Lake Road Arden Hills, Minnesota 55112 Phone: 651-481-7410 Fax: 651-481-7415 Contact Person Angela Raun Phone: 651-481-7460 Fax: 651-481-7415 Fmail: araun(@TransomaMedical.com OCT 1 * 2007 Date of Summary September 18, 2007 | Proprietary Name of Device | Sleuth Implantable ECG Monitoring System | |------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------| | Common/Usual Name | Implantable ECG Monitoring System | | Classification Name | Cardiac Implantable Event Recorder<br>Product Code - MXC<br>Regulation Number 21 CFR Part 870.2800<br>Device Class II | | Legally Marketed Device(s) to<br>Which Substantial Equivalence<br>is Claimed | Medtronic, Inc., Reveal® Plus Insertable<br>Loop Recorder (ILR) System (K003667,<br>K994331 and K972242) | | | Brentwood Medical Technology Corporation<br>- IOmark™ Digital Holter, (K031466) | #### Device Description The Transoma Medical Sleuth Implantable ECG Monitoring System is designed to measure and store the patient s electrocardiogram (ECG) during symptomatic and asymptomatic events. The Sleuth System consists of the Model 2010 Implantable Monitoring Device (IMD), the Model 4000 Activator, and the Model 5000 Base Station (Bluetooth Modem). The IMD is an anatomically-shaped Litanium structure which is implanted subcutancously, typically in the left pectoral region. The Activator is a hand-held device carried by the patient. The IMD continuously measures the R-R interval and automatically stores ECG segments when this interval falls outside of pre-defined limits. Using the Activator, the patient can also request storage of ECG segments when symptoms are experienced. The IMD stores a limited quantity of events, which are · ansferred to the Activator via telemetry throughout the day. On a daily basis, the Activator transfers events through the Base Station to a service center for review and analysis. {1}------------------------------------------------ #### Intended Use The Transoma Medical Sleuth 2010 System is an implantable, patient- and automatically-activated monitoring system that records subcutaneous ECG and is indicated for: KOG: 3035 page 2/3 - patients with clinical syndromes or situations at increased risk of cardiac arrhythmias or . - patients who experience transient symptoms that may suggest a cardiac arrhythmia. 133035 . ## Summary of Technological Characteristics The Transoma Medical Sleuth 2010 System incorporates substantially equivalent technology. comparable features, labeling, and intended use, and is similar to the predicate devices currently available on the market. ### Non-clinical Test Summary The substantial equivalence of the Transoma Medical Sleuth 2010 System has been demonstrated via bench and animal testing. Testing included: In vitro/ bench testing of the IMD, the Activator, the Base Station and the system In vivo canine testing Biocompatibility testing Software testing Electromagnetic compatibility testing Electrical safety testing Sterilization validation Shelf-life testing ### Clinical Study Summary A clinical study was conducted at 4 major medical centers in Panama. The primary objectives were to evaluate the diagnostic viability of the ECG signals and the performance of the system. Twenty-eight (28) patients were enrolled in the study. The patients had unexplained syncope/ pre-syncope or were at risk of arrhythmias. Patients were monitored continuously (24/7) by the Sleuth device for arrhythmic events. Clinic follow-ups were 7-10 days. 1, 3, and 6 months after implantation. The cardiac monitoring center has been able to analyze and categorize the ECG waveform data. The Sleuth system has successfully captured events mitiated manually, as well as asymptomatic events which are detected and recorded automatically by the system. As designed, the Sleuth system has also successfully stored and transferred trending data. The Sleuth ECG system has performed well following implantation in the initial 28 patients. The system has diagnostic viability, since thirtcen of the patients were discovered during monitoring to have arrhythmus, and the Sleuth records were important in helping the physicians arrive at a firm diagnosis. The occurrence of two adverse events within > 4,700 collective patient days of use is within expectation, and these events were neither serious, nor definitively shown to derive from the Sleuth system. # Performance Standards Performance standards have not been established by the Food and Drug Administration for these devices under Section 514. # Conclusion (Statement of Equivalence) Extensive bench and animal testing have demonstrated the Sleuth Model 2010 System functions in accordance with product specifications. Additionally, testing demonstrated the Sleuth System functions equivalently to comparative systems, i.e., the Brentwood IQmark Digital Holter (K031466) and the Medtronic Reveal Plus Insertable Loop Recorder (K003667, K994331 and K972242). These data {2}------------------------------------------------ support a determination of substantial equivalence and subsequently market clearance of the Transoma Medical Sleuth System, comprised of the Model 2010 IMD, Model 4000 Activator, and Model 5000 Base Station. Kax 3035 page 3/3 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three overlapping wing segments, creating a sense of movement or flight. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Transoma Medical, Inc. c/o Ms. Angie Raun Vice President Regulatory Affairs and Quality Assurance 4358 West Round Lake Road Arden Hills, MN 55112 2007 Re: K063035 Trade/Device Name: Sleuth Implantable ECG Monitoring System Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MXC Dated: June 29, 2007 Received: July 2, 2007 Dear Ms. Raun: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bfimmimorfor Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 1 of 1 #### Section 4: Indications for Use Statement 510(k) Number_ K 063035 Device Name: Transoma Medical Sleuth Implantable ECG Monitoring System #### Indications for Use: The Transoma Medical Sleuth Implantable ECG Monitoring System is an implantable, patient- and automatically-activated monitoring system that records subcutaneous ECG and is indicated for: - Patients with clinical syndromes or situations at increased risk of cardiac . arrhythmias - Patients who experience transient symptoms that may suggest a cardiac . arrhythmia Prescription Use X (21 CFR 801 Subpart D) OR Over The Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) .. .-------------------------------------------------------------------------------------------------------------------------------------------------------------------------- Concurrence of CDRH, Office of Device Evaluation (ODE) Blymlamor ardiovascular Devic
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...