The NIRO-200NX DP is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes in any individual. The clinical value of trend data has not been demonstrated in disease states. The NIRO-200NX DP should not be used as the sole basis for diagnosis or therapy.
Device Story
NIRO-200NX DP is a non-invasive oximeter using near-infrared light to monitor regional hemoglobin oxygen saturation and relative hemoglobin levels. Probes containing an LED light source and two photodiodes are applied to the skin over target tissue. The device measures light transmitted through tissue; the photodiode closer to the source is subtracted from the farther photodiode to calculate oxygen saturation. Changes in light detection from one photodiode calculate relative levels of oxygenated and deoxygenated hemoglobin. Used in hospital settings by healthcare professionals; probes can remain on patients for up to 12 hours with continuous data collection at 20Hz. Output is displayed on an LCD unit for physician monitoring. The device provides trend data to assist clinical decision-making, though it is not intended as a sole diagnostic tool. This version introduces disposable probes and compatible adapters to the previously cleared NIRO-200NX system.
Clinical Evidence
Bench testing only. Performance established via phantom study comparing subject and predicate devices in a simulated model, demonstrating substantial equivalence. Device also supported by two years of clinical use in Japan and Europe without reported adverse events.
Technological Characteristics
Near-infrared oximeter; Class 1 LED light source; two-photodiode detection; disposable probes with compatible adapter/AMP unit; LCD display; 20Hz sampling rate; non-invasive; software classified as moderate level of concern; tested to IEC 60601-1 (2005), IEC 60601-1-2 (2008), and IEC 60825-1 (2007).
Indications for Use
Indicated for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative levels of oxygenated/deoxygenated hemoglobin in brain or other tissue in any individual. Not for use as sole basis for diagnosis or therapy.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right. The profiles are stacked on top of each other and are connected by a flowing line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 25, 2016
Hamamatsu Photonics K.K. % Ms. Allyson Mullen Hyman, Phelps & Mcnamara 700 Thirteenth Street, Northwest, Suite 1200 Washington, District of Columbia 20005
Re: K162117
Trade/Device Name: Niro-200NX DP Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD, DOA Dated: July 29, 2016 Received: July 29, 2016
Dear Ms. Mullen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162117
Device Name NIRO-200NX DP
#### Indications for Use (Describe)
The NIRO-200NX DP is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes in any individual. The clinical value of trend data has not been demonstrated in disease states. The NIRO-200NX DP should not be used as the sole basis for diagnosis or therapy.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K162117 – NIRO-200NX DP
| Submitter Name: | Hamamatsu Photonics K.K. |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 812 Joko-cho, Higashi-ku, Hamamatsu City, 431-3196,<br>JAPAN |
| Contact Person: | Susumu Suzuki |
| Phone Number: | 81-53-431-0124 |
| Fax Number: | 81-53-431-0148 |
| Date Prepared: | October 24, 2016 |
| Device Trade Name: | NIRO-200NX DP |
| Device Common Name: | Oximeter |
| Product Code: | MUD |
| Subsequent Product Code: | DQA |
| Classification: | Class II per 21 C.F.R. § 870.2700 |
| Predicate Device: | Hamamatsu NIRO-200NX (K143219) |
| Device Description: | The NIRO-200NX is a piece of equipment that uses near infrared<br>light for non-invasive measurement of hemoglobin oxygen<br>saturation and relative levels of oxygenated hemoglobin and<br>deoxygenated hemoglobin of blood in brain or in other tissue<br>beneath the probes. Patient probes are applied to the skin over the<br>tissue of interest. The probes have a light source and 2<br>photodiodes, one closer to the light source and one further away<br>from the light source. The 2 photodiodes detect the light<br>transmitted through the patient's tissue. The detected light is<br>analyzed with the known light absorption characteristics of<br>oxyhemoglobin and deoxyhemoglobin. The amount of light<br>detected by the photodiode closer to the light source is subtracted<br>from the light detected by the farther photodiode. The result is |
| | then used to calculate the hemoglobin oxygen saturation. Also, by measuring the changes in light detected from one of the photodiodes, the relative levels of oxygenated hemoglobin and deoxygenated hemoglobin are calculated. |
| | The predicate NIRO-200NX (K143219) utilized reusable patient probes. The purpose of this premarket notification is to obtain clearance for use of disposable probes and compatible connectors with the cleared display unit. Use of the disposable probes and compatible connectors with the cleared display unit is referred to as the NIRO-200NX DP. |
| Intended Use: | The NIRO-200NX DP is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes in any individual. The clinical value of trend data has not been demonstrated in disease states. The NIRO-200NX DP should not be used as the sole basis for diagnosis or therapy. |
| Performance Data: | The following electrical and performance data have been generated using the NIRO-200NX DP and are described in this 510(k) submission. All tests demonstrate that the device functions as intended. |
| 1. | Electrical Safety was established by testing in accordance with IEC 60601-1 Edition 3.0, Medical Electrical Equipment - Part 1: General requirements for Safety (2005) |
| 2. | Electromagnetic Compatibility was established by testing in accordance with IEC 60601-1-2 3rd Edition, Medical Electrical Equipment - Part 1-2: Electromagnetic Compatibility – Requirements and Tests (2008). |
| 3. | Laser Safety was established by testing in accordance with IEC 60825-1 Ed. 2.0 (2007). The light source in the NIRO-200NX DP is a Class 1 Light Emitting Diode (LED) Product. |
| 4. | Report on the Tissue Phantom Measurements with NIRO-200NX Reusable and Disposable Probes. Hamamatsu |
{4}------------------------------------------------
{5}------------------------------------------------
performed a phantom study to compare performance of the proposed and predicate device side by side in a simulated model. The results of the study demonstrate that performance of the NIRO-200NX DP is substantially equivalent to the performance of the NIRO-200NX (K143219).
- 5. The NIRO-200NX DP has been sold and used clinically for more than two years in Japan and Europe without any reported adverse events.
- 6. Hamamatsu followed the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices May 11, 2005, ' to classify the NIRO-200NX software as a "moderate level of concern." The software was verified and validated, and the software verification and validation documents were prepared and presented in accordance with FDA's guidance document.
Substantial Equivalence:
The predicate device is the Hamamatsu NIRO-200NX (K143219).
The NIRO-200NX DP has the same intended use, indications for use, principle of operation, and measurement method as the predicate device. The NIRO-200NX DP and the predicate device have similar technological characteristics. The only difference is the inclusion of a disposable probe option, which includes a compatible adapter and AMPs for connection of the disposable probes to the NIRO-200NX display unit. This minor difference does not raise different questions of safety or efficacy, as confirmed by Hamamatsu's testing and validation activities described in this submission, including EMC, electrical safety, and laser safety testing in accordance with IEC 60601-1-2 (2008), IEC 60601-1 (2005), and IEC 60825-1 Ed. 2.0 (2007). Hamamatsu followed the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices May 11, 2005," to classify the NIRO-200NX software as a "moderate level of concern." The software was verified and validated, and the software verification and validation documents were prepared and presented in accordance with FDA's guidance document.
Further. NIRO-200NX DP is at least as safe and effective as the predicate devices as demonstrated by the results of a phantom study. Hamamatsu performed a phantom study to compare performance of the proposed and predicate devices side by side in a simulated model. The results of the study demonstrated that performance of the NIRO-200NX DP is substantially equivalent to the performance of NIRO-200NX at measuring regional hemoglobin oxygen saturation and relative levels of oxygenated hemoglobin and deoxygenated hemoglobin.
{6}------------------------------------------------
The data presented demonstrate that the NIRO-200NX DP is at least as safe and effective as the predicate NIRO-200NX. Therefore, the NIRO-200NX DP is substantially equivalent to the predicate device. The Substantial Equivalence comparison chart is found below in Table 1.
| | Subject Device: | Predicate Device: |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | NIRO-200NX DP | NIRO-200NX (K143219) |
| Intended Use | Oximeter (MUD, DQA) | Oximeter (MUD, DQA) |
| Indications for Use | The NIRO-200NX DP is<br>intended for use as an adjunct<br>trend monitor of regional<br>hemoglobin oxygen saturation<br>and relative level of oxygenated<br>hemoglobin and deoxygenated<br>hemoglobin of blood in brain or<br>in other tissue beneath the<br>probes in any individual. The<br>clinical value of trend data has<br>not been demonstrated in<br>disease states. The NIRO-<br>200NX DP should not be used<br>as the sole basis for diagnosis or<br>therapy. | The NIRO-200NX is intended<br>for use as an adjunct trend<br>monitor of regional hemoglobin<br>oxygen saturation and relative<br>level of oxygenated hemoglobin<br>and deoxygenated hemoglobin<br>of blood in brain or in other<br>tissue beneath the probes in any<br>individual. The clinical value of<br>trend data has not been<br>demonstrated in disease states.<br>The NIRO-200NX should not be<br>used as the sole basis for<br>diagnosis or therapy. |
| | Subject Device: | Predicate Device: |
| | NIRO-200NX DP | NIRO-200NX (K143219) |
| Principle of<br>Operation | The NIRO-200NX DP is<br>intended for use in a hospital<br>setting by healthcare<br>professionals. To use the<br>NIRO-200NX DP, the probes<br>are placed on any healthy tissue<br>of the patient. The probes can<br>remain on the patient for up to<br>12 hours, and measurement data<br>can be continuously collected (at<br>a maximum sampling rate of<br>20Hz (0.05s)) over that time<br>period. The data recorded from<br>the probe is calculated as<br>discussed below and reported on<br>the LCD on the DU for<br>monitoring by the physician. | The NIRO-200NX is intended<br>for use in a hospital setting by<br>healthcare professionals. To use<br>the NIRO-200NX, the probes<br>are placed on any healthy tissue<br>of the patient. The probes can<br>remain on the patient for up to<br>12 hours, and measurement data<br>can be continuously collected (at<br>a maximum sampling rate of<br>20Hz (0.05s)) over that time<br>period. The data recorded from<br>the probe is calculated as<br>discussed below and reported on<br>the LCD on the DU for<br>monitoring by the physician. |
| Technological Characteristics | | |
| Display Unit (DU) | Liquid crystal display (LCD)<br>and instrument control panel | Liquid crystal display (LCD)<br>and instrument control panel |
| Probes | Disposable | Reusable |
| Adaptor | Disposable Probe Adaptor<br>(DPA) | Pulse 2ch Adaptor (P2A) |
| AMP Unit | DPA AMP Unit | AMP Unit |
| | Subject Device: | Predicate Device: |
| | NIRO-200NX DP | NIRO-200NX (K143219) |
| Light Source<br>Device | LED (Light Emitting Diode) | LED (Light Emitting Diode) |
| Wavelength Safety<br>Class | 3 wavelengths<br>Class I | 3 wavelengths<br>Class I |
| Light Detector | Photodiode | Photodiode |
| Measurement<br>Method | 2 Point Detection Method for<br>Hemoglobin Oxygen Saturation<br>(TOI) and Relative value of the<br>total hemoglobin (nTHI)<br><br>1 Point Detection Method for<br>Relative Levels of Hemoglobins | 2 Point Detection Method for<br>Hemoglobin Oxygen Saturation<br>(TOI) and Relative value of the<br>total hemoglobin (nTHI)<br><br>1 Point Detection Method for<br>Relative Levels of Hemoglobins |
| Patient Contact | Non-Invasive | Non-Invasive |
| EMC and<br>Electrical Safety | Passed applicable safety testing | Passed applicable safety testing |
# Table 1: Comparison to Predicate
{7}------------------------------------------------
{8}------------------------------------------------
Conclusion: Based on the indications for use, performance testing, and technological characteristics, the NIRO-200NX DP has been shown to be as safe and effective for its stated intended use as the NIRO-200NX (K143219).
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.