Published clinical literature; Clinical setting performance data
Clinical literature was used to demonstrate substantial equivalence by comparing the performance of the NIRO-200NX to the predicate INVOS device and blood draw results in a clinical setting, as well as to support the device's successful clinical history.
23 individuals (including light, intermediate, and dark skin); Sample Size: 23 individuals; 181 measurements for INVOS, 179 for NIRO-200NX
INVOS 5100C and blood draw results
Regional hemoglobin oxygen saturation
Indications for Use
The NIRO-200NX is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes in any individual. The clinical value of trend data has not been demonstrated in disease states. The NIRO-200NX should not be used as the sole basis for diagnosis or therapy.
Device Story
NIRO-200NX is a reusable near-infrared oximeter for non-invasive monitoring of regional hemoglobin oxygen saturation and relative hemoglobin levels. Probes applied to skin contain a light source and two photodiodes at varying distances. Device measures light transmitted through tissue; subtracts signal from closer photodiode from farther photodiode to calculate hemoglobin oxygen saturation; calculates relative levels of oxygenated/deoxygenated hemoglobin via changes in detected light. Used in clinical settings; provides trend data to healthcare providers as an adjunct to other diagnostic information. Benefits include non-invasive tissue oxygenation monitoring.
Clinical Evidence
Clinical evidence includes published literature (Bickler et al.) comparing NIRO-200NX and INVOS 5100C in 23 individuals (181 INVOS measurements, 179 NIRO-200NX measurements) across varying skin tones. Results showed substantially equivalent performance in measuring regional hemoglobin oxygen saturation. Additional evidence includes 4 years of clinical use in Japan and Europe without reported adverse events and literature confirming successful use in infants.
Technological Characteristics
Non-invasive near-infrared oximeter. Light source: 3-wavelength LED. Detector: Photodiode. Measurement method: 2-point detection for hemoglobin oxygen saturation (TOI) and relative total hemoglobin (nTHI); 1-point detection for relative hemoglobin levels. Safety: Class I LED product. Compliance: IEC 60601-1 (electrical), IEC 60601-1-2 (EMC), IEC 60825-1 (laser/light safety).
Indications for Use
Indicated for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative levels of oxygenated and deoxygenated hemoglobin in brain or other tissue in any individual. Not for use as sole basis for diagnosis or therapy.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an emblem consisting of three stylized human profiles facing to the right, with a design above them that resembles a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Hamamatsu Photonics K.K. % Mr. Jeffrey K. Shapiro Hyman, Phelps and McNamara 700 Thirteenth Street, North West, Suite 1200 Washington, District of Columbia 20005
Re: K143219
Trade/Device Name: NIRO-200NX Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD, DQA Dated: July 6, 2015 Received: July 6, 2015
Dear Mr. Shapiro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
July 31, 2015
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
For
Sincerely yours.
Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143219
Device Name Hamamatsu NIR0-200NX
#### Indications for Use (Describe)
The NIRO-200NX is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes in any individual. The clinical value of trend data has not been demonstrated in disease states. The NR0-200NX should not be used as the sole basis for diagnosis or therapy.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K143219
Page 1 of 6
### 510(k) Summary NIRO-200NX
Submitter Name: Hamamatsu Photonics K.K.
Submitter Address: 812 Joko-cho, Higashi-ku, Hamamatsu City, 431-3196, JAPAN
Contact Person: Susumu Suzuki
Phone Number: 81-53-431-0124
Fax Number: 81-53- 431-0148
Date Prepared: July 24, 2015
Device Trade Name: NIRO-200NX
Device Common Name: Oximeter
Product Code: MUD
Subsequent Product Code: DQA
Classification: Class II per 21 C.F.R. § 870.2700
Predicate Device:
Somanetics INVOS 5100C (K082327), Hitachi ETG-100 (K011320) and ETG-4000 (K042501)
The NIRO-200NX is a reusable piece of equipment that uses near Device Description: infrared light for non-invasive measurement of hemoglobin oxygen saturation and relative levels of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes. The patient probes are applied to the skin over the tissue of interest. The probes have a light source and 2 photodiodes, one closer to the light source and one further away from the light source. The 2 photodiodes detect the light transmitted through the patient's tissue. The detected light is analyzed with the known light absorption characteristics of oxyhemoglobin and deoxyhemoglobin. The amount of light detected by the photodiode closer to the light source is subtracted from the light detected by the farther photodiode. The result is then used to calculate the hemoglobin oxygen saturation. Also, by measuring the changes in light detected from one of the
{4}------------------------------------------------
photodiodes, the relative levels of oxygenated hemoglobin and deoxygenated hemoglobin are calculated.
Intended Use: The NIRO-200NX is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes in any individual. The clinical value of trend data has not been demonstrated in disease states. The NIRO-200NX should not be used as the sole basis for diagnosis or therapy.
Performance data: The following electrical, performance, and clinical literature has been conducted with and reported using the NIRO-200NX and is described in the 510(k) submission. All tests and literature demonstrate that the device functions as intended and performs equivalently to the predicate devices.
- 1. Electrical per IEC 60601-1
- 2. Electromagnetic Compatibility per IEC 60601-1-2
- 3. Light Emitting LED Product per IEC 60825-1 (Class1 LED product)
4. Bench data demonstrating that NIRO-200NX performs equivalently in tissue phantoms compared to the predicate devices.
- 5. The NIRO-200NX has been sold and used clinically for more than 4 years in Japan and Europe without any reported adverse events. A review of the published literature concludes that the device worked as intended by safely being used as an adjunct monitor of regional hemoglobin oxygen saturation and relative level of oxygenated hemoglobin and deoxygenated hemoglobin of blood in brain or in other tissue beneath the probes. Specifically, clinical literature demonstrated that NIRO-200NX and the predicate INVOS performed equivalently in a clinical setting. In addition, clinical literature was provided to show that NIRO-200NX has been used successfully in infants.
Substantial Equivalence:
The predicate devices are ETG-100 (K011320), ETG-4000 (K042501), and INVOS 5100C (K082327).
The proposed NIRO-200NX and the predicate devices all have the same intended use: as Oximeters. The proposed and predicate devices also have similar indications for use. The proposed and predicate devices all have the same principles of operation: they all measure levels of oxygen based on light measurements taken from the surface of intact tissue on the outside of a patient's body. The NIRO-200NX and the predicate
{5}------------------------------------------------
devices have similar technological characteristics, and any minor differences do not raise different questions of safety or efficacy, as confirmed by Hamamatsu's testing and validation activities described in this submission, including EMC, electrical safety, and laser safey testing in accordance with IEC 60601-1-2 (2008), IEC 60601-1 (2005), and IEC 60825-1 (1993 + A1 (1997) + A2 (2001). respectively. Hamamatsu followed the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices May 11, 2005," to classify the NIRO-200NX software as a "moderate level of concern." The software was verified and validated, and the software verification and validation documents were prepared and presented in accordance with FDA's guidance document.
Further, NIRO-200NX is at least as safe and effective as the predicate devices as demonstrated by the results of bench and clinical data reported in literature, including the phantom study and the published clinical literature described below.
#### Phantom Study
Hamamatsu performed a phantom study to compare performance of the proposed and predicate devices side by side in a simulated model. The results of the study demonstrated that performance of the NIRO-200NX is substantially equivalent to the performance of INVOS at measuring regional hemoglobin oxygen saturation; and ETG at measuring relative levels of oxygenated hemoglobin and deoxygenated hemoglobin.
### Published Clinical Literature
Bickler et al reported on a clinical study that directly compared the performance of NIRO-200NX and INVOS to each other (along with several other cleared devices) and to results of a blood draw in a clinical setting. The evaluation was performed in 23 individuals with 181 individual measurements taken for INVOS and 179 individual measurements taken for NIRO-200NX. The study subjects included those with light, intermediate and dark skin. The study results showed substantially equivalent performance between NIRO-200NX and INVOS in a clinical setting at measuring regional hemoglobin oxygen saturation.
The above information leads to the conclusion that the NIRO-200NX is substantially equivalent to the predicate devices. The Substantial Equivalence comparison chart is found below in Table 1.
{6}------------------------------------------------
## Table 1
| | NIRO-200NX<br>Hamamatsu Photonics K.K.<br>(New Device) | ETG-100 and ETG-4000<br>Hitachi Medical Systems America, Inc<br>(ETG-100 for K011320)<br>(ETG-4000 for K042501) | INVOS 5100C<br>Somanetics Corporation<br>(K082327) |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INTENDED<br>USE | Oximeter (MUD, DQA) | Oximeter (DQA) | Oximeter (MUD) |
| INDICATIONS<br>FOR USE | The NIRO-200NX is intended for<br>use as an adjunct trend monitor of<br>regional hemoglobin oxygen<br>saturation and relative level of<br>oxygenated hemoglobin and<br>deoxygenated hemoglobin of blood<br>in brain or in other tissue beneath the<br>probes in any individual. The clinical<br>value of trend data has not been<br>demonstrated in disease states. The<br>NIRO-200NX should not be used as<br>the sole basis for diagnosis or<br>therapy. | A) K011320<br>The intended use of the ETG-100 is<br>the measurement of relative levels of<br>cerebral deoxy-hemoglobin and<br>oxyhemoglobin.<br><br>B) K042501<br>The intended use of the ETG-4000<br>is the measurement of relative levels<br>of cerebral deoxy-hemoglobin and<br>oxyhemoglobin. | The noninvasive INVOS 5100C is<br>intended for use as an adjunct<br>monitor or regional hemoglobin<br>oxygen saturation of blood in the<br>brain or in other tissue beneath the<br>sensor. It is intended for use in<br>individuals greater than 2.5kg at risk<br>for reduced-flow or no-flow ischemic<br>states.<br><br>It is also intended for use as an<br>adjunct trend monitor of regional<br>hemoglobin oxygen saturation of<br>blood in the brain or in other tissue<br>beneath the sensor in any individual.<br>The clinical value of trend data has<br>not been demonstrated in disease<br>states. The INVOS System should<br>not be used as the sole basis for<br>diagnosis or therapy. |
{7}------------------------------------------------
| NIRO-200NX<br>Hamamatsu Photonics K.K.<br>(New Device) | ETG-100 and ETG-4000<br>Hitachi Medical Systems America, Inc<br>(ETG-100 for K011320)<br>(ETG-4000 for K042501) | INVOS 5100C<br>Somanetics Corporation<br>(K082327) | |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PRINCIPLE<br>OF<br>OPERATION | The patient probe consists of a light<br>source (2 wavelengths) and 1<br>detector, and it is applied to the skin<br>over the tissue of interest.<br>Relative levels of oxy-hemoglobin,<br>deoxy-hemoglobin and total<br>hemoglobin are calculated from the<br>changes in the amount of the<br>detected lights by one detector. | The patient probe consists of a light<br>source (2 wavelengths) and 2<br>detectors, and it is applied to the skin<br>over the tissue of interest.<br>Hemoglobin oxygen saturation<br>(rSO2) is calculated from the<br>difference of the amount of detected<br>lights by the 2 detectors. | The patient probe consists of a light<br>source (3 wavelengths) and 2<br>detectors, and it is applied to the skin<br>over the tissue of interest.<br>Hemoglobin oxygen saturation (TOI)<br>is calculated from the differences of<br>the amount of detected lights by the 2<br>detectors. Relative value of the total<br>hemoglobin (nTHI) is calculated by<br>the same methods as that of TOI.<br>Relative levels of oxygenated,<br>deoxygenated and total<br>hemoglobines (△O2Hb, △HHb and<br>△cHb) are calculated from the<br>changes in the amount of the<br>detected lights by one detector. |
{8}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS
| | NIRO-200NX<br>Hamamatsu Photonics K.K.<br>(New Device) | ETG-100 and ETG-4000<br>Hitachi Medical Systems America, Inc<br>(ETG-100 for K011320)<br>(ETG-4000 for K042501) | INVOS 5100C<br>Somanetics Corporation<br>(K082327) |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Light Source<br>Device<br>Wavelength<br>Safety Class | LED (Light Emitting Diode)<br>3 wavelengths<br>Class I | LD (Laser Diode)<br>2 wavelengths<br>Class I | LED<br>2 wavelengths<br>Class I |
| Light Detector | Photodiode | Photomultiplier Tube | Photodiode |
| Measurement<br>Method | 2 Point Detection Method for<br>Hemoglobin Oxygen Saturation<br>(TOI) and Relative value of the total<br>hemoglobin (nTHI)<br><br>1 Point Detection Method for<br>Relative Levels of Hemoglobins | Not applicable. Not measured by<br>this device.<br><br>1 Point Detection Method for<br>Relative Levels of Hemoglobins | 2 Point Detection Method for<br>Hemoglobin Oxygen Saturation<br>(rSO2)<br><br>Not applicable. Not measured by<br>this device. |
| Patient<br>Contact | Non-Invasive | Non-Invasive | Non-Invasive |
| EMC and<br>Electrical<br>Safety | Passed applicable safety testing | Passed applicable safety testing | Passed applicable safety testing |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.