K032346 · Ge Medical Systems Information Technologies · MSX · Aug 5, 2003 · Cardiovascular
Device Facts
Record ID
K032346
Device Name
UNITY NETWORK IS PATIENT VIEWER
Applicant
Ge Medical Systems Information Technologies
Product Code
MSX · Cardiovascular
Decision Date
Aug 5, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The Unity Network® IS Patient Viewer is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the Unity Network® IS Patient Viewer is to provide a remote view of physiological parameter data on adult, pediatric and neonatal patients within a hospital or facility providing patient care. The Unity Network® IS Patient Viewer is NOT intended for primary monitoring but is to be used in conjunction with the bedside monitor. The Unity Network® IS Patient Viewer is intended to provide near-real-time physiological data and graphical trends for all monitors connected to the Unity Network to secure nurse and physician personal computers (local and remote).
Device Story
Unity Network® IS Patient Viewer provides remote access to physiological waveforms, parameter data, and trends from bedside monitors. System resides on hospital intranet; data accessed via standard web browser on personal computers. Displays up to nine waveforms in near real-time; excludes alarm and status messages. Used by nurses and physicians in clinical settings as secondary viewing tool. Clinicians must reference primary bedside monitor for patient care decisions. Benefits include improved mobility and performance for viewing patient data across hospital network.
Clinical Evidence
Bench testing only. No clinical data provided. Quality assurance measures included risk analysis, requirements/design reviews, module verification, system integration testing, and final validation.
Technological Characteristics
Networked information server; connects to hospital intranet. Displays data via web browser on standard PCs. Supports wireless technology. Software-based system for data relay from bedside monitors.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in hospital or care facilities. Used for remote viewing of physiological parameter data and graphical trends. Not for primary monitoring; must be used with bedside monitors under licensed practitioner supervision.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
AUG - 6 2003
# Section 2 Summary and Certification
# 510(k) Summary of Safety and Effectiveness
| Date: | July 29, 2003 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems Information Technologies<br>8200 West Tower Avenue<br>Milwaukee, WI 53223 USA |
| Contact Person: | Diana M. Thorson<br>Regulatory Affairs Specialist<br>GE Medical Systems Information Technologies<br>Phone: (714) 247-4135<br>Fax: (414) 918-8114 |
| Device: Trade Name: | Unity Network® IS Patient Viewer |
| Common/Usual Name: | Computer, Information Network Server |
| Classification Names: | 21 CFR 870.2300 System, Network and Communication, Physiological |
| Predicate Devices: | K020661 Unity Network® IS Patient Viewer |
| Device Description: | The Unity Network® IS Patient Viewer provides remote access to<br>waveform, parameter data and trend data at a Web browser on a<br>standard personal computer. The server resides on the hospital's<br>intranet and remote access is gained through secured access to the<br>hospital's intranet.<br><br>The data relayed from the patient monitors over the Unity® MC network<br>includes patient name, unit and bed name, parameter data, and<br>waveform data monitored by the bedside monitors. The user can view<br>up to nine waveforms from the Unity® MC network as well as the<br>parameter information in near real-time. Neither alarm messages nor<br>parameter status messages are displayed.<br><br>The Unity Network® IS Patient Viewer system provides a secondary<br>view of patient information, and is NOT a patient monitoring device.<br>The clinician is instructed to always reference the primary bedside<br>monitor before making any patient care decisions. In the event that<br>data is not available via the Unity Network® IS Patient Viewer, the<br>clinician is instructed to obtain the data from the primary bedside<br>monitor. |
{1}------------------------------------------------
Intended Use: The Unity Network® IS Patient Viewer is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the Unity Network® IS Patient Viewer is to provide a remote view of physiological parameter data on adult, pediatric and neonatal patients within a hospital or facility providing patient care. The Unity Network® IS Patient Viewer is NOT intended for primary monitoring but is to be used in conjunction with the bedside monitor. -
> The Unity Network® IS Patient Viewer is intended to provide near-realtime physiological data and graphical trends for all monitors connected to the Unity Network to secure nurse and physician personal computers (local and remote).
The Unity Network® IS Patient Viewer system employs the same Technology: functional technology as the predicate device. Changes made reflect the recent advancements in wireless technology, offering improved performance and mobility.
### Test Summary:
The Unity Network® IS Patient Viewer complies with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the Unity Network® IS Patient Viewer:
- Risk Analysis .
- Requirements Reviews .
- Design Reviews
- Testing on unit level (Module verification) .
- Integration testing (System verification) .
- Final acceptance testing (Validation) ●
- Performance testing
- Safety testing .
- Environmental testing .
### Conclusion:
The results of these measurements demonstrated that the Unity Network® IS Patient Viewer is as safe, as effective, and perform as well as the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle. The logo is rendered in black and white.
AUG - 5 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
GE Medical Systems Information Technologies Ms. Diane M. Thorson Corporate Regulatory Affairs Specialist 8200 West Tower Avenue Milwaukee, WI 53223
Re: K032346
Trade Name: Unity Network® IS Patient Viewer Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: MSX Dated: July 29, 2003 Received: July 30, 2003
Dear Ms. Thorson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
## Page 2 - Ms. Diane M. Thorson
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
510(k) filed on July 29, 2003
Device Name:
Unity Network® IS Patient Viewer
Indications for Use:
The Unity Network® IS Patient Viewer is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the Unity Network® IS Patient Viewer is to provide a remote view of physiological parameter data on adult, pediatric and neonatal patients within a hospital or facility providing patient care. The Unity Network® IS Patient Viewer is NOT intended for primary monitoring but is to be used in conjunction with the bedside monitor.
The Unity Network® IS Patient Viewer is intended to provide near-real-time physiological data and graphical trends for all monitors connected to the Unity Network to secure nurse and physician personal computers (local and remote).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801:109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Nole Te
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) **Number**
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.