DATEX-OHMEDA S/5 WEB VIEWER

K013387 · Datex-Ohmeda · MSX · Jan 8, 2002 · Cardiovascular

Device Facts

Record IDK013387
Device NameDATEX-OHMEDA S/5 WEB VIEWER
ApplicantDatex-Ohmeda
Product CodeMSX · Cardiovascular
Decision DateJan 8, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The Datex-Ohmeda S/5 Web Viewer is intended to be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Web Viewer displays information received from other networked devices. It is comprised of an S/5 Web Server and S/5 Web Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the networked devices in Datex-Ohmeda Network and S/5 Web Viewer clients. The S/5 Web Viewer client runs on a generic computer that is connected to the hospital local area network. The Datex-Ohmeda S/5 Web Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Web Viewer will be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 Web Viewer is not a primary alarm source. The device is for use by qualified medical personnel only.

Device Story

Web-based remote monitoring system; inputs real-time patient data (curves, numeric info, visual alarms) from bedside monitors via hospital LAN; S/5 Web Server coordinates network connections; S/5 Web Viewer client runs on generic PC using standard browser software; provides graphical/numerical trends and visual data display; used by clinicians/nurses in hospital settings (e.g., doctor's office) for remote consultation; does not provide audible alarms; not a primary alarm source; benefits include distributed access to patient information.

Clinical Evidence

Bench testing only. Verification included electrical safety (EN60950), electromagnetic compatibility (EN 55022, EN 55082-1, IEC 60601-1-2), mechanical/environmental tolerance, and software validation. No clinical data presented.

Technological Characteristics

Web-browser based remote monitoring system; utilizes hospital LAN/Intranet; PC hardware; software-based data display; complies with IEC 60601-1, EN 60950, and ISO/IEC 8802-3 network standards.

Indications for Use

Indicated for hospital patients requiring remote consultation and monitoring by qualified medical personnel. Not for use as a primary alarm source.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a black circle with a white shape inside that resembles the letter 'D'. There is also a small black square located above and to the right of the circle. The black circle and square are solid, while the white 'D' shape is created by the absence of color within the circle. The image appears to be a simple graphic or logo. Page 1 of 3 K013387 JAN # Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5 Web Viewer # GENERAL COMPANY INFORMATION as required by 807.92(a)(1) ## COMPANY NAME/ADDRESS/PHONE/FAX: Datex-Ohmeda 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344 ### NAME OF CONTACT: Mr. Joel Kent ### DATE: October 10, 2001 #### DEVICE NAME as required by 807.92(a)(2) #### TRADE NAME: Datex-Ohmeda S/5 Web Viewer #### COMMON NAME: Remote monitoring device #### CLASSIFICATION NAME: ### The following Class II classification appears applicable: System, network and communication, physiological monitors 870.2910 ### NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3) The Datex-Ohmeda S/5 Web Viewer is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda S/5 Network and Central and Information Center (510(k) number: K000647). {1}------------------------------------------------ #### DEVICE DESCRIPTION as required by 807.92(a)(4) The Web Viewer is a supplementary monitor based on the WWW-browser technology. It is meant the wed for remote viewing of real-time patient information and trends from a generic personal to be used for relieve volunial of the PC uses standard browser software to gain access to the Web Server which contains the Datex-Ohmeda specific Web Viewer software. The Web Server in turn receives the patient data from Datex-Ohmeda S/5 Central serving the Datex-Ohmeda S/5 Network. The hospital is responsible for ensuring a secure and functional interface between the Datex-Ohmeda S/5 Network and the Hospital LAN, by utilizing for example a gateway, router, switch or firewall. If the Web Viewer clients are not connected to a hospital Intranet, a regular hub can be used instead. ## INTENDED USE as required by 807.92(a)(5) The Datex-Ohmeda S/5 Web Viewer is intended to be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Web Viewer displays information received from other networked devices. It is comprised of an S/5 Web Server and S/5 Web Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the networked devices in Datex-Ohmeda Network and S/5 Web Viewer clients. The S/5 Web Viewer client runs on a generic computer that is connected to the hospital local area network. The Datex-Ohmeda S/5 Web Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Web Viewer will be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 Web Viewer is not a primary alarm source. The device is for use by qualified medical personnel only. ## SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6) The Datex-Ohmeda S/5 Web Viewer is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda S/5 Network and Central (K000647). The structure and functionality of the Datex-Ohmeda S/5 Web Viewer corresponds to the structure and functionality of the Datex-Ohmeda S/5 Network and Central (predicate) as follows: The Datex-Ohmeda S/5 Web Viewer can show real-time curves, numeric information and visual alarms from bedside monitors just like the predicate Datex-Ohmeda S/5 Network and Central (K000647). In contrast to the predicate Datex-Ohmeda Network and Central the Datex-Ohmeda S/5 Web Viewer also offers graphical and numerical trends but does not provide audible alarms. Displayed information is always the same as on the source monitor. The S/5 Web Viewer can display data from one monitor at a time only, whereas the S/5 Central can provide centralized monitoring for up to 32 monitors at a time. The Datex-Ohmeda S/5 Web Viewer is an extension of the Datex-Ohmeda S/5 Network and Central. It offers distributed monitoring in form of an additional viewing tool. The Datex-Ohmeda S/5 Web Viewer is not intended as a primary alarm source. It does not adversely effect the safety and effectiveness of the system. {2}------------------------------------------------ The indications for use of the Datex-Ohmeda S/5 Web Viewer are nearly the same as for the Datex-Ohmeda Network and Central (K000647). Both are used by a clinician, a nurse or doctor, to monitor patients but the environment for use is different. The Central is used near the care area but the S/5 Web Viewer can be used anywhere that is connected to the hospital LAN, for example in a doctors office. The information shown in both the S/5 Web Viewer and the predicate are copies from corresponding bedside monitors. The Datex-Ohmeda S/5 Web Viewer and the predicate Network and Central (K000647) can both be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The physical network and the PC hardware components used by the S/5 Web Viewer are the same as in Datex-Ohmeda Network and Central. # SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3) The Datex-Ohmeda S/5 Web Viewer complies with the safety standards below and is therefore safe and effective for the intended use. The device has been thoroughly tested including electrical safety, electromagnetic compatibility, mechanical and environmental tolerance, software validation and verification of specifications. Verification of compliance with the following mandatory and voluntary standards has been made: - EN60950: 1995 (IEC950 2nd edition) Safety of information technology equipment, . including electrical business equipment - EN 55022: 1994 (IEC-CISPR 22) Information technology equipment Radio . disturbance characteristics. Limits and methods of measurement - EN 55082-1: 1991 (IEC 801-2, IEC 801-4) Electromagnetic compatibility . - Generic immunity standard - IEC 60601-1-2. Radiated Emission (IEC-CISPRII) . - IEC 60601-1-2, Immunity to Electrostatic Discharges (IEC-61000-4-2) ● - EN 1441. Medical devices Risk analysis ● - IEC 60601-1-4 . - CAN/CSA-C22.2 No 950: Information Technology Equipment Including Electrical . Business Equipment - UL 1950: Information Technology Equipment Including Electrical Business Equipment . - ISO/IEC 8802-3 (ANSI/IEEE 802.3), EIA/TIA-568, EIA/TIA-TSB40, international . network cabling standards ### Conclusion: The summary above shows that there are no new questions of safety and effectiveness for the Datex-Ohmeda S/5 Web Viewer as compared to the predicate device. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird with three curved lines representing its wings or feathers. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 8 2002 JAN Mr. Joel C. Kent Manager, Quality and Regulatory Affairs Datex-Ohmeda 86 Pilgrim Road Needham, MA 02492 Re: K013387 Trade Name: S/5 Web Viewer Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: MSX Dated: October 11, 2002 Received: October 12, 2002 Dear Mr. Kent: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Joel C. Kent Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4645. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Kally Tell am D. Zuckerman, M.D Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): _KOI 338 Device Name: Datex-Ohmeda S/5 Web Viewer Indications For Use: The Datex-Ohmeda S/5 Web Viewer displays information received from other networked devices. It is comprised of an S/5 Web Server and S/5 Web Viewer clients. The Datex-Ohmeda S/5 Web Server maintains and coordinates the network connections between the networked devices in Datex-Ohmeda Network and S/5 Web Viewer clients. The S/5 Web Viewer client runs on a generic computer that is connected to the hospital local area network. The Datex-Ohmeda S/5 Web Viewer can be used for viewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda S/5 Web Viewer will be used for patients in the hospital and it is meant for consultation and remote monitoring use. The Datex-Ohmeda S/5 Web Viewer is not a primary alarm source. The device is for use by qualified medical personnel only. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use . (Per 21 CFR 801.109) OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) Division of Cardiovascular & Respiratory Devices §10(k) Number 12015387
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...