K942685 · Mortara Instrument, Inc. · DXJ · Aug 12, 1997 · Cardiovascular
Device Facts
Record ID
K942685
Device Name
ECG REVIEW STATION
Applicant
Mortara Instrument, Inc.
Product Code
DXJ · Cardiovascular
Decision Date
Aug 12, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2450
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The ST Review Station is intended to allow Physicians and/or Clinical personnel from ER, Cath Lab, ICU, and other related settings to retrospectively review/analyze stored ECG data. By being able review/analyze ECG trends, including ST segment measurements, Clinicians and/or Physicians have access to information to assist in their ability to detect changes in a patient’s physiological condition and therefore enabling them to respond accordingly to a patient’s medical needs.
Device Story
ST Review Station functions as an adjunct to the ELI 100/STM Electrocardiograph; it receives stored ECG data and ST segment measurements from the ELI 100/STM. The device utilizes a PC-based operating system to provide graphical visualization of ECG trends, median beat ECGs, and HR trends. It enables clinicians in clinical settings (ER, ICU, Cath Lab) to perform retrospective analysis, including superimposition scanning and serial comparison of averaged ST complexes. The device produces 3D color images of ST segment changes across 12 leads and generates 8.5 x 11-inch printouts. By facilitating the review of entire monitoring sessions, the device assists healthcare providers in identifying physiological changes, supporting clinical decision-making, and improving patient care management.
Clinical Evidence
Bench testing only. Verification and validation testing were performed to ensure the integrity of ECG data and ST segment measurements received and stored from the ELI 100/STM was maintained.
Technological Characteristics
PC-based system; color monitor display; 12-lead ECG data input; software-based graphical visualization and storage; 8.5 x 11-inch printing capability. No specific materials or ASTM standards cited.
Indications for Use
Indicated for clinicians and physicians in ER, Cath Lab, ICU, and related settings to retrospectively review and analyze stored ECG data and ST segment trends to assist in detecting changes in a patient's physiological condition.
Regulatory Classification
Identification
A medical cathode-ray tube display is a device designed primarily to display selected biological signals. This device often incorporates special display features unique to a specific biological signal.
K103640 — MIDMARK IQECG · Brentwood Medical Technology Corp. · Mar 22, 2011
K003940 — 2010 PLUS HOLTER FOR WINDOWS · Agilent Technologies · Feb 8, 2001
K102854 — PC ECG · Edan Instruments, Inc. · Dec 3, 2010
Submission Summary (Full Text)
{0}
AUG-11-97 MON 9:35 MORTARA INST FAX NO. 4143544760 P. 02
K94/2685
# ST Review Station
## 510(k) Summary
AUG 12 1997
| Submitter: | Mortara Instrument, Inc. |
| --- | --- |
| Address: | 7865 North 86^{th} Street
Milwaukee, WI 53224 |
| Telephone/Fax: | (414) 354-1600/354-4760 |
| Contact: | Regulatory Affairs Manager or Director |
| Prepared: | August 8, 1997 |
| Proprietary Name: | ST Review Station |
| Common/Classification Name: | ST Segment Display |
| Predicate Device: | ELI 100/STM and ST Central Station |
### New Device Description:
The ST Review Station is an adjunct to the ELI 100/STM Electrocardiograph (12-Lead Diagnostic Electrocardiograph) to provide rapidly responsive trend data originating from the ELI 100/STM Electrocardiograph and associated patient population. The ST Review Station utilizes a PC based operating system with specially developed software that provides for the means to graphically view and/or present the ECG data captured and stored by the ELI 100/STM (legally marketed predicate device – K895232), including ST segment measurements.
### Intended Use:
The ST Review Station is intended to allow Physicians and/or Clinical personnel from ER, Cath Lab, ICU, and other related settings to retrospectively review/analyze stored ECG data. By being able review/analyze ECG trends, including ST segment measurements, Clinicians and/or Physicians have access to information to assist in their ability to detect changes in a patient’s physiological condition and therefore enabling them to respond accordingly to a patient’s medical needs.
### Comparison to Predicate Devices:
See Attached Table (Attachment 1)
### Non-Clinical Performance Data:
To ensure that the ST Review Station did not alter the integrity of the ECG data and ST segment measurements received and stored from the ELI 100/STM, the ST Review Station was subjected to verification and validation practices appropriate to determine that it is safety and effective for its intended use.
### Performance Data Conclusion:
The verification and validation tests performed demonstrated that the ECG data and ST segment measurements received and stored by the ST Review Station maintained their integrity as obtained from the ELI 100/STM. Therefore, supporting the determination of substantial equivalence while demonstrating its safety and effectiveness.
{1}
AUG-11-97 MON 9:36 MORTARA INST
FAX NO. 4143544760
P. 03
# ST Review Station
## 510(k) Summary
### Attachment 1
#### Tabular Comparison of Predicate Device's
| Features | ELI 100/STM (K895232) Predicate Device | ST Central Station (K922927) Predicate Device | ST Review Station (K942685) Premarket Notification |
| --- | --- | --- | --- |
| Computer Based System | No -- 12-lead EKG that provides continuous and simultaneous monitoring of the ST segment | Yes -- Central display of ECG data in real-time for up to 8 patients acquired from the ELI 100/STM | Yes -- Allows for the review and storage of ST monitoring sessions from the ELI 100/STM |
| Color Monitor Display | No -- LCD display | Yes | Yes |
| Receives 12-lead EKG Data from Electrocardiograph | No -- 12-lead EKG with continuous ST segment monitoring | Yes | Yes |
| Displays ST Segment Changes | No -- Only prints 12-lead ST segment changes | Yes -- Optional 2-hour ST trend for the displayed ECG lead | Yes -- Displays entire ST trends and templates from ELI 100/STM monitoring session |
| Storage of 12-lead ST Segment Monitoring | Yes -- 900 to 1800 ECG's, Stores a single patient's continuous 12-lead ST segment monitoring session | No | Yes -- 25,000 to 37,000 Allows for the storage and review of multiple patient ST segment monitoring sessions |
| Displays 12-lead ECG | No -- Print only | Yes -- Continuous display of real-time ECG data on up to 8 patient's (Can display either median beat ECG, ST trends, or HR trends) | Yes -- Allows for the review of entire monitoring sessions stored median beat ECG, ST trends, and HR trends for a selected patient record |
| Automatic Template Update | Yes -- Compares successive 12-lead EKG to updated template for detection of significant ST segment changes | No | No |
| Display of Scanning and Serial Comparison | No -- Print of serial comparison | No | Yes -- Allows for the review of averaged stored ST complexes using superimposition scanning and serial comparison |
{2}
AUG-11-97 MON 9:37 MORTARA INST FAX NO. 4143544760 P. 04
# ST Review Station
## 510(k) Summary
| Features | ELI 100/STM (K895232) Predicate Device | ST Central Station (K922927) Predicate Device | ST Review Station (K942685) Premarket Notification |
| --- | --- | --- | --- |
| Display and Printing of Trends and Graphic Data | No -- Print of ST trends | Yes -- Can display the most recent 2-hours of either the HR or ST trends received from the ELI 100/STM and the most recent 2-hours of both HR and ST trends of the displayed ECG can be printed | Yes -- Can display and print a 3D color image of ST segment changes for all 12-leads along with HR and ST trends from a patient's entire ELI 100/STM monitoring session |
| User Defined Alarms | Yes -- One lead and two lead ST segment alarm thresholds | Yes -- High and Low HR alarms and ST segment alarm thresholds as established at the ELI 100/STM | No |
| 12-Lead ECG Printouts | Yes -- 4 inch format | Yes -- 8.5 x 11 inch format | Yes -- 8.5 x 11 inch format |
| Patient Information | Name and ID # | Name and Room # -- (Name originates from the ELI 100/STM) | Name -- (Name originates from the ELI 100/STM) |
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
Mr. Scott J. Pease
Mortara Instrument, Inc.
7865 North 86th Street
Milwaukee, Wisconsin 53224
AUG 12 1997
Re: K942685
ST Review Station
Regulatory Class: II (two)
Product Code: 74 DXJ
Dated: May 13, 1997
Received: May 16, 1997
Dear Mr. Pease:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2 - Mr. Scott J. Pease
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.