DM-400 HOLTER ECG CASSETTE RECORDER

K970298 · Diagnostic Monitoring · DSH · Aug 20, 1997 · Cardiovascular

Device Facts

Record IDK970298
Device NameDM-400 HOLTER ECG CASSETTE RECORDER
ApplicantDiagnostic Monitoring
Product CodeDSH · Cardiovascular
Decision DateAug 20, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2800
Device ClassClass 2

Indications for Use

The DM-400 Holter ECG Cassette Recorder is a device designed for ECG recording of an ambulatory patient.

Device Story

DM-400 Holter ECG Cassette Recorder is a battery-operated ambulatory device; records three independent channels of ECG plus one timing track onto standard C60 cassette tape. Patient connects via 5 or 7-wire lead set with silver/silver chloride electrodes. Features include patient event signal button, self-calibration, and baseline centering. Used in clinical or ambulatory settings to capture cardiac signals over 24 or 48-hour periods. Healthcare providers retrieve the cassette for subsequent scanning and analysis to aid in cardiac rhythm assessment. Device benefits include portable, continuous monitoring of ambulatory patients to detect transient cardiac events.

Clinical Evidence

Bench testing only. Testing verified functionality, power supply, low battery features, circuitry, calibration, ECG amplifier performance, event signal detection, baseline centering, frequency response, common mode rejection, and motor controller operation. Dynamic testing confirmed electronic and mechanical package integration. 24-hour calibration and scanning verification demonstrated accuracy and repeatability.

Technological Characteristics

Portable ambulatory ECG recorder; 9V alkaline battery power; records to standard C60 cassette tape; 3 ECG channels + 1 timing track; 1.0mm/sec tape speed; 0.05-100 Hz frequency response; 60 dB minimum common mode rejection; 5 or 7-wire lead set; silver/silver chloride electrodes; includes defibrillator input protection.

Indications for Use

Indicated for ECG recording of ambulatory patients requiring 24 or 48-hour monitoring.

Regulatory Classification

Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} AUG 20 1997 K970298 # SECTION 19: SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92. ## 19.1 SUBMITTER INFORMATION a. Company Name: Diagnostic Monitoring b. Company Address: 1176 Main Street, Bldg. C Irvine, CA 92614 c. Company Phone: (714) 568-1695 d. Contact Person: Raymond Cohen President e. Date Summary Prepared: May 21, 1997 ## 19.2 DEVICE IDENTIFICATION a. Trade/Proprietary Name: DM-400™ Holter ECG Cassette Recorder b. Classification Name: Medical Magnetic Tape Recorder ## 19.3 IDENTIFICATION OF PREDICATE DEVICE(S) | Company | Device | 510(k) No. | Date Cleared | | --- | --- | --- | --- | | Datrix | XR-300 Holter ECG Recorder | K921068 | March 30,1992 | | Zymed | MultiTrak Ambulatory ECG Recorder | K930894 | November 11, 1993 | {1} ## 19.4 DEVICE DESCRIPTION The Diagnostic Monitoring DM-400 Holter ECG Cassette Recorder is a battery operated ambulatory ECG recording device. The DM-400 Holter Recorder is capable of recording three independent channels of ECG from an ambulatory patient. The patient is connected to the recorder by means of a 5 or 7 wire bonded lead set and silver/silver chloride self adhesive electrodes. The DM-400 Holter Recorder can perform either a 24 or 48 hour ECG recording. The recorder is equipped with self calibration and a patient event signal button. ## 19.5 SUBSTANTIAL EQUIVALENCE The Diagnostic Monitoring DM-400 Holter ECG Cassette Recorder is substantially equivalent to other ambulatory ECG recorders currently in commercial distribution by Datrix and Zymed in terms of the intended use of achieving a safe and accurate ECG recording of an ambulatory patient. The fundamental technical characteristics are similar to those of the predicate devices and are listed on the comparison charts provided in Section 19.7. ## 19.6 INTENDED USE The DM-400 Holter ECG Cassette Recorder is a device designed for ECG recording of an ambulatory patient. ## 19.7 TECHNOLOGICAL CHARACTERISTICS A comparison of the technological characteristics of the predicate and legally marketed devices is provided in the following device comparison charts. Finished product {2} specifications, schematic drawings and detailed description of the device have been provided. The Device Comparison Chart #1 provides a side by side comparison of the characteristics of the Diagnostic Monitoring DM-400 Holter ECG Cassette Recorder and the Datrix XR-300 Holter ECG Recorder. Device Comparison Chart #2 provides a side by side comparison of the characteristics of the Diagnostic Monitoring DM-400 Holter ECG Cassette Recorder and the Zymed MultiTrak Ambulatory ECG Recorder. As evidenced by the comparison charts the DM-400 Holter Recorder and the predicate devices are similar in their function and characteristics. There are no differences between the technologies which require further evaluation. {3} # DEVICE COMPARISON CHART #1 Diagnostic Monitoring DM-400 Holter ECG Cassette Recorder & Datrix XR-300 Holter ECG Recorder | Manufacturer | Diagnostic Monitoring | Datrix | | --- | --- | --- | | Model No. | DM-400 Holter ECG Recorder | XR-300 Holter ECG Recorder | | Intended use | ECG recording of ambulatory patients. | ECG recording of ambulatory patients. | | Design | | | | -portable | Yes | Yes | | -power source | 9V alkaline battery | 9V alkaline battery | | -cassette tape | standard C60 | standard C60 | | -number of heads | 1 | 1 | | -channels | 4 (3 ECG, 1 timing track) | 4 (3 ECG, 1 timing track) | | -tape speed | 1.0mm/sec | 1.0mm/sec | | -recording time | 24 or 48 hours | 25 hours, nominal | | -frequency response | 0.05 to 100 Hz (-3dB) | 0.05 to 100 Hz (-3dB) | | -common mode rejection | 60 dB minimum 1 volt peak-to-peak 60 Hz common mode signal | 60 dB minimum 1 volt peak-to-peak 60 Hz common mode signal | | -electrolead cables | 5 or 7 lead | 5 or 7 lead | {4} # DEVICE COMPARISON CHART #1 Diagnostic Monitoring DM-400 Holter ECG Cassette Recorder & Datrix XR-300 Holter ECG Recorder | Manufacturer | Diagnostic Monitoring | Datrix | | --- | --- | --- | | Features | | | | -patient event button | Yes | Yes | | -built-in calibration | Yes | Yes | | -carrying case | Yes | Yes | | -color coded ECG outputs | Yes | Yes | {5} # DEVICE COMPARISON CHART #2 Diagnostic Monitoring DM-400 Holter ECG Cassette Recorder & Zymed MultiTrak Ambulatory ECG Record | Manufacturer | Diagnostic Monitoring | Zymed | | --- | --- | --- | | Model No. | DM-400 Holter ECG Recorder | MultiTrak Ambulatory ECG Recorder (3 channel configuration) | | Intended use | ECG recording of ambulatory patients. | ECG recording of ambulatory patients. | | Design | | | | -portable | Yes | Yes | | -power source | 9V alkaline battery | 9V alkaline battery | | -cassette tape | standard C60 | Phillips C60 or C90 | | -number of heads | 1 | 1 | | -recording channels | 4 (3 ECG, 1 timing track) | 4 (3 ECG, 1 timing track) | | -tape speed | 1.0mm/sec | 1.0mm/sec | | -recording time | 24 or 48 hours | 36 hours | | -frequency response | 0.05 to 100 Hz (-3dB) | 0.05 to 100 Hz (-3dB) | | -common mode rejection | 60 dB minimum 1 volt peak-to-peak 60 Hz common mode signal | 60 dB minimum | | -electrolead cable | 5 or 7 lead | 5 or 7 lead | {6} # DEVICE COMPARISON CHART #2 Diagnostic Monitoring DM-400 Holter ECG Cassette Recorder & Zymed MultiTrak Ambulatory ECG Recorder | Manufacturer | Diagnostic Monitoring | Zymed | | --- | --- | --- | | Features | | | | - patient event button | Yes | Yes | | - built-in calibration | Yes | Yes | | - input protection for defibrillator | Yes | Yes | | - carrying case | Yes | Yes | {7} ## 19.8 PERFORMANCE DATA The DM-400 Holter ECG Cassette Recorder has been demonstrated to perform as intended with accuracy and repeatability. A summary of the performance testing of the DM-400 Holter Recorder are presented in the following. The DM-400 Holter ECG Cassette Recorder was tested for functionality and adherence to design specifications. The testing encompassed the major functional and specialized features of the DM-400 Recorder. Testing included verification of the power supply and low battery features, complete circuitry testing, verification of the unit calibration, ECG amplifier verification, testing of the event signals, baseline centering, frequency response testing, common mode rejection and Motor Controller Circuit operation. In addition, dynamic testing of the units was performed to assure that the electronic package functions correctly with the supplied mechanical package. A 24 hour Calibration test was performed, as well as a visual inspection of the unit and a scanning verification. Results of the testing showed that the DM-400 Holter ECG Recorder units met the design specifications and performed as intended. Repeatability testing of the units showed that the DM-400 Recorder provides consistent results. Testing of the DM-400 Recorder verifies that the device is designed for accurate and safe ECG recording of an ambulatory patient. The testing also confirms that the features and characteristics of the DM-400 Recorder device perform within the expected parameters of the substantially equivalent devices with the same features and characteristics. ## 19.9 510(K) CHECKLIST This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer’s Checklist is provided in this submission. {8} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 Mr. Raymond Cohen Diagnostic Monitoring™ 1176 Main Street, Building C Irvine, California 92614 AUG 20 1997 Re: K970298 DM-400 Holter ECG Cassette Recorder Regulatory Class: II (two) Product Code: 74 DSH Dated: May 22, 1997 Received: May 23, 1997 Dear Mr. Cohen: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {9} Page 2 - Mr. Raymond Cohen This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {10} # INDICATIONS FOR USE 510(k) Number: To Be Assigned By FDA Device Name: Diagnostic Monitoring DM-400 Holter ECG Recorder Indications For Use: The DM-400 Holter ECG Cassette Recorder is a device designed for ECG recording of an ambulatory patient. PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) [Signature] (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number: k970298 Prescription Use ☑ OR Over-The-Counter Use (Per 21 CFR 801.109) 26
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...