T3 Platform software

K223578 · Etiometry, Inc. · PPW · Jul 7, 2023 · Cardiovascular

Device Facts

Record IDK223578
Device NameT3 Platform software
ApplicantEtiometry, Inc.
Product CodePPW · Cardiovascular
Decision DateJul 7, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2200
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Indications for Use

The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes: - To remotely consult regarding a patient's status, and - To remotely review other standard or critical near real-time patient data in order to aid in clinical decisions and deliver patient care in a timely manner.

Device Story

T3 Platform software aggregates, stores, and visualizes physiologic data from bedside devices in ICU settings. Inputs include numeric physiologic parameters (e.g., BP, HR, SpO2, respiratory data) and laboratory measurements (blood gases, CBC, lactic acid). The T3 Risk Analytics Engine processes these inputs via model-based physiological algorithms to calculate four indices: IDO2 (oxygen delivery), IVCO2 (CO2 ventilation), ACD (acidemia), and HLA (hyperlactatemia). Output is displayed via web-based interface for healthcare professionals to remotely consult and review patient status. Device acts as adjunctive clinical decision support; does not provide alarms or control bedside hardware. Benefits include centralized, near real-time visibility into patient physiological trends to aid clinical decision-making. Clinicians use output to supplement standard evaluations; indices are not sole basis for therapy or drug administration.

Clinical Evidence

Retrospective clinical validation using two cohorts for the IVCO2 Index. Cohort 1: 1108 PaCO2 measurements from 180 NICU patients. Cohort 2: 29,841 PaCO2 measurements from 2090 non-NICU patients (42% neonates, 32% infants, 26% children; 46% female, 54% male). Indices evaluated against predicate acceptance criteria: discriminatory power, range utilization, resolution/limitation, and robustness. Results met all criteria; no adverse effects noted.

Technological Characteristics

Software-only solution; web-based visualization. Operates as adjunctive cardiovascular status indicator (21 CFR 870.2200). Utilizes model-based physiological algorithms for index calculation. Data aggregated from bedside devices via network. Moderate level of concern software. Cybersecurity controls implemented per FDA guidance.

Indications for Use

Indicated for adult, pediatric, and neonatal patients in intensive care. IDO2 Index: post-surgical patients 0-12 years, >=2 kg. IVCO2 Index: invasively ventilated patients 0-12 years. ACD Index: invasively ventilated patients 0-12 years, >=2 kg. HLA Index: post-surgical patients 0-12 years, >=2 kg. Not for alarm notification or independent therapy control.

Regulatory Classification

Identification

The adjunctive cardiovascular status indicator is a prescription device based on sensor technology for the measurement of a physical parameter(s). This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Software description, verification, and validation based on comprehensive hazard analysis must be provided, including: (i) Full characterization of technical parameters of the software, including any proprietary algorithm(s); (ii) Description of the expected impact of all applicable sensor acquisition hardware characteristics on performance and any associated hardware specifications; (iii) Specification of acceptable incoming sensor data quality control measures; and (iv) Mitigation of impact of user error or failure of any subsystem components (signal detection and analysis, data display, and storage) on accuracy of patient reports. (2) Scientific justification for the validity of the status indicator algorithm(s) must be provided. Verification of algorithm calculations and validation testing of the algorithm using a data set separate from the training data must demonstrate the validity of modeling. (3) Usability assessment must be provided to demonstrate that risk of misinterpretation of the status indicator is appropriately mitigated. (4) Clinical data must be provided in support of the intended use and include the following: (i) Output measure(s) must be compared to an acceptable reference method to demonstrate that the output measure(s) represent(s) the predictive measure(s) that the device provides in an accurate and reproducible manner; (ii) The data set must be representative of the intended use population for the device. Any selection criteria or limitations of the samples must be fully described and justified; (iii) Agreement of the measure(s) with the reference measure(s) must be assessed across the full measurement range; and (iv) Data must be provided within the clinical validation study or using equivalent datasets to demonstrate the consistency of the output and be representative of the range of data sources and data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment. (5) Labeling must include the following: (i) The type of sensor data used, including specification of compatible sensors for data acquisition; (ii) A description of what the device measures and outputs to the user; (iii) Warnings identifying sensor reading acquisition factors that may impact measurement results; (iv) Guidance for interpretation of the measurements, including warning(s) specifying adjunctive use of the measurements; (v) Key assumptions made in the calculation and determination of measurements; (vi) The measurement performance of the device for all presented parameters, with appropriate confidence intervals, and the supporting evidence for this performance; and (vii) A detailed description of the patients studied in the clinical validation ( *e.g.,* age, gender, race/ethnicity, clinical stability) as well as procedural details of the clinical study.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 7, 2023 Etiometry, Inc. Timothy Hanson VP of Regulatory Affairs and Quality Assurance 280 Summer St., 4th Floor Boston, Massachusetts 02210 Re: K223578 Trade/Device Name: T3 Platform software Regulation Number: 21 CFR 870.2200 Regulation Name: Adjunctive Cardiovascular Status Indicator Regulatory Class: Class II Product Code: PPW Dated: June 7, 2023 Received: June 8, 2023 Dear Timothy Hanson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Robert T. Kazmierski -S for LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K223578 Device Name The T3 Platform™ software #### Indications for Use (Describe) The T3 Platform™ software features the T3 Data Aggregation & Visualization software module version 5.0 and the T3 Risk Analytics Engine software module version 9.0. The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes: - To remotely consult regarding a patient's status, and - To remotely review other standard or critical near real-time patient data in order to aid in ● clinical decisions and deliver patient care in a timely manner. The T3 Data Aggregation & Visualization software module can display numeric physiologic data captured by other medical devices: - · Airway flow, volume, and pressure - · Arterial blood pressure (invasive and non-invasive, systolic, diastolic, and mean) - Bispectral index (BIS, signal quality index, suppression ratio) . - Cardiac Index - Cardiac output . - Central venous pressure . - . Cerebral perfusion pressure - End-tidal CO2 . - · Heart rate - Heart rate variability . - Intracranial pressure . - . Left atrium pressure - Oxygen saturation (intravascular, regional, SpO2) . - Premature ventricular counted beats . - Pulmonary artery pressure (systolic, diastolic, and mean) . - Pulse pressure variation - · Pulse Rate - · Respiratory rate - Right atrium pressure . - Temperature (rectal, esophageal, tympanic, blood, core, nasopharyngeal, skin) - · Umbilical arterial pressure (systolic, diastolic, and mean) {3}------------------------------------------------ The T3 Data Aggregation & Visualization software module can display laboratory measurements including arterial and venous blood gases, complete blood count, and lactic acid. T3 Data Aggregation & Visualization software module can display information captured by the T3 Risk Analytics Engine software module. The T3 Risk Analytics Engine software module calculates four indices: the IDO2 Index™ for inadequate delivery of oxygen, the IVCO2 Index™ for inadequate ventilation of carbon dioxide, the ACD Index™ for acidemia, and the HLA Index™ for hyperlactatemia. The IDO2 Index™ is indicated for use by health care professionals with post-surgical patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The IDO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module. When the IDO2 Index™ is increasing, it means that there is an increasing risk of inadequate oxygen delivery and attention should be brought to the patient. The IDO2 Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements. The IVCO2 Index™ is indicated for use by health care professionals with invasively ventilated patients 0 to 12 years of age under intensive care. The IVCO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module. When the IVCO2 Index™ is increasing, it means that there is an increasing risk of inadequate carbon dioxide ventilation and attention should be brought to the patient. The IVCO2 Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the interpretation statements. The ACD Index™ is indicated for use by health care professionals with invasively ventilated patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The ACD Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module. When the ACD Index™ is increasing, it means that there is an increasing risk of acidemia and attention should be brought to the patient. The ACD Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the interpretation statements. The HLA Index™ is indicated for use by health care professionals with post-surgical patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The HLA Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module. When the HLA Index™ is increasing, it means that there is an increasing risk of hyperlactatemia and attention should be brought to the patient. The HLA Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements. #### WARNINGS: - · Do not use the T3 Platform software as an active patient monitoring system. - · Do not use the T3 Platform software to replace any part of the hospital's device monitoring. - · Do not rely on the T3 Platform software as the sole source of patient status information. - · Do not use any of the T3 Platform indices as a substitute for taking blood samples. - · The indices present qualitative and potentially imperfect information of the patient's condition and in certain scenarios, the indices may contradict each other. The primary data should be reviewed as part of standard patient evaluations and no decisions should be solely based on the indices. {4}------------------------------------------------ Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) — Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the word "etiometry" in a sans-serif font, with the letters in a gradient of blue. Above the word is a curved line that starts above the "e" and ends above the "y". At the apex of the curve is a blue triangle pointing upwards. The logo is clean and modern. Jul 3, 2023 This 510(k) summary has been prepared in accordance with Title 21 CFR §807.92 and FDA's guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications 510(k)" July 28, 2014 #### 1 510(k) Submitter Timothy Hanson, VP of Regulatory Affairs and Quality Assurance 280 Summer St., 4th Floor Boston, MA 02210 Tel: 857.366.9333 ext. 2020 Email: thanson@etiometry.com #### 2 Device | Item | Description | |--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name | T3 Platform™ software | | Device Common/Usual Name | Clinical Decision Support Software (without alarms) | | Classification Name | Adjunctive cardiovascular status indicator | | Classification Number | 870.2200 | | Regulatory Class | Class II with special controls - the primary code is PPW: The adjunctive<br>cardiovascular status indicator is a prescription device based on sensor<br>technology for the measurement of a physical parameter(s). This device is<br>intended for adjunctive use with other physical vital sign parameters and<br>patient information and is not intended to independently direct therapy. | Table 1: Device Information ### 3 Predicate Devices The predicate device is the T3 Platform™ software featuring the T3 Data Aggregation & Visualization software module version 5.0 and the T3 Risk Analytics Engine software module version 8.0, cleared under K213230. The predicate has not been subject to a design-related recall. No reference devices were used in this submission. {6}------------------------------------------------ ## 4 Device Description The Tracking, Trajectory, Trigger (73) intensive care unit software solution allows clinicians and quality improvement teams in the ICU to aggregate data from multiple sources, store it in a database for analysis, and view the streaming data. System features include: - Adjunctive status indicators ● - Customizable display of physiologic parameters over the entire patient stay - . Configurable annotation - Web-based visualization that may be used on any standard browser - Minimal IT footprint ● - Software-only solution no new bedside hardware required ● - Highly reliable and robust operation - . Auditable data storage # 5 Indications for Use The T3 Platform™ software features the T3 Data Aggregation & Visualization software module version 5.0 and the T3 Risk Analytics Engine software module version 9.0. The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes: - To remotely consult regarding a patient's status, and . - . To remotely review other standard or critical near real-time patient data in order to utilize this information to aid in clinical decisions and deliver patient care in a timely manner. The T3 Data Aggregation & Visualization software module can display numeric physiologic data captured by other medical devices: - Airway flow, volume, and pressure - Arterial blood pressure (invasive and non-invasive, systolic, diastolic, and mean) ● - Bispectral index (BIS, signal quality index, suppression ratio) - Cardiac Index - Cardiac output - Central venous pressure - Cerebral perfusion pressure ● - End-tidal CO2 - Heart rate - Heart rate variability - Intracranial pressure - Left atrium pressure - Oxygen saturation (intravascular, regional, SpO2) {7}------------------------------------------------ - Premature ventricular counted beats - . Pulmonary artery pressure (systolic, diastolic, and mean) - . Pulse pressure variation - Pulse Rate - . Respiratory rate - Right atrium pressure - Temperature (rectal, esophageal, tympanic, blood, core, nasopharyngeal, skin) - Umbilical arterial pressure (systolic, diastolic, and mean) The T3 Data Aggregation & Visualization software module can display laboratory measurements including arterial and venous blood gases, complete blood count, and lactic acid. The T3 Data Aggregation & Visualization software module can display information captured by the T3 Risk Analytics Engine software module. The T3 Risk Analytics Engine software module calculates four indices: the IDO2 Index™ for inadequate delivery of oxygen, the IVCO2 Index™ for inadequate ventilation of carbon dioxide, the ACD Index™ for acidemia, and the HLA Index™ for hyperlactatemia. The IDO2 Index™ is indicated for use by health care professionals with post-surgical patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The IDO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module. When the IDO2 Index™ is increasing, it means that there is an increasing risk of inadequate oxygen delivery and attention should be brought to the patient. The IDO2 Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements. The IVCO2 Index™ is indicated for use by health care professionals with invasively ventilated patients 0 to 12 years of age under intensive care. The IVCO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module. When the IVCO2 Index™ is increasing, it means that there is an increasing risk of inadequate carbon dioxide ventilation and attention should be brought to the patient. The IVCO2 Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the interpretation statements. The ACD Index™ is indicated for use by health care professionals with invasively ventilated patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The ACD Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module. When the ACD Index™ is increasing, it means that there is an increasing risk of acidemia and attention should be brought to the patient. The ACD Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the interpretation statements. The HLA Index™ is indicated for use by health care professionals with post-surgical patients 0 to 12 years of age and weighing 2 kg or more under intensive care. The HLA Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module. When the HLA Index™ is increasing, it means that there is an increasing risk of hyperlactatemia and attention should be brought to the patient. The HLA Index™ presents partial {8}------------------------------------------------ quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements. #### WARNINGS: - Do not use the T3 Platform™ software as an active patient monitoring system. ● - . Do not use the T3 Platform™ software to replace any part of the hospital's device monitoring. - Do not rely on the T3 Platform™ software as the sole source of patient status information. ● - Do not use any of the T3 Platform™ indices as a substitute for taking blood samples. ● - The indices present qualitative and potentially imperfect information about the patient's condition and in certain scenarios, the indices may contradict each other. The primary data should be reviewed as part of standard patient evaluations and no decisions should be solely based on the indices. ## 6 Comparison of Technological Characteristics with the Predicate Device The subject T3 Platform™ software and predicate T3 Platform™ software have the same intended use. They differ with respect to the following technological features (see Table 2). | Feature/Improvement | Description | |-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------| | Removal of 2 kg Restriction from the IVCO2 Index<br>Indications For Use | The restriction that the IVCO2 Index™ is indicated for use on patients weighing 2 kg or more has been removed. | Table 2: Summary of Changes ### 7 Summary of Non-Clinical Performance Testing Documentation was provided, in accordance with the 2017 FDA guidance document Software as a Medical Device (SAMD): Clinical Evaluation section 5.3 Analytical / Technical Validation of a SaMD, to support the software with a moderate level of concern, and to confirm and provide objective evidence that the software was correctly constructed. Cybersecurity information was provided in accordance with the 2014 FDA guidance document Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. Evaluations were completed to demonstrate the consistency of the output, representative of the range of data sources and data quality, likely to be encountered. # 8 Summary of Clinical Performance Testing Documentation was provided, in accordance with the 2017 FDA guidance document Software as a Medical Device (SAMD): Clinical Evaluation section 5.3 Analytical / Technical Validation of a SaMD, to support the {9}------------------------------------------------ software with a moderate level of concern, and to yield a clinically meaningful output associated to the target use of the output in the target health care situation or condition identified in the definition statement. The adjunctive status indicators are produced by a model-based approach. The model-based approach is designed based on principles of physiology with parameters chosen to reflect those specified in the medical literature. Development test sets are used to evaluate the impact of the development changes during the development process. Validation sets are used after development is complete to validate performance using independent data sets. The adjunctive status indicators were validation sets that included data from different clinical sites in the US. The clinical study data were obtained by the T3 Platform software. No adverse effects or complications were noted. The adjunctive status indicators were retrospectively computed on all de-identified patients. The adjunctive status indicators were evaluated against the same acceptance criteria as the predicate device, i.e., discriminatory power, range utilization, resolution/limitation, and robustness. Two patient cohorts were used to validate the IVCO2 Index. The first performance analysis of the IVCO2 Index used a cohort consisting of neonatal ICU patients. The analysis contained data from two level IV regional NICUs. In total, 1108 PaCO2 measurements from arterial blood gases among 180 patients were included in the NICU validation data set. The second performance analysis of the IVCO2 Index used a cohort consisting of non-NICU patients to evaluate the impact of technical changes made to the software, to accommodate patients weighing < 2kg found in a NICU, on patients not found in the NICU. In total, 29,841 PaCO2 measurements from arterial blood gases among 2090 patients were included in the non-NICU validation data set. The overall demographics were 42% neonates, 32% infants, and 26% children. 46% of the patients were female and 54% of the patients were male. All results met the same acceptance criteria as the predicate device, i.e., discriminatory power, range utilization, resolution/limitation, and robustness. ### 9 Summary Based on the clinical performance, the subject T3 Platform™ software was found to have a safety and effectiveness profile that is similar to the predicate T3 Platform™ software, cleared under K213230. ### 10 Conclusions Substantial equivalence of the T3 Platform™ software is demonstrated through performance testing, clinical evaluation, and special controls. The T3 Platform™ software has the equivalent design, features, and functionality as the predicate T3 Platform™ software. No new questions of safety or effectiveness were raised as a result of the differences when compared to the predicate device. The software verification demonstrates that the T3 Platform™ software performs as intended in the specified use conditions. The clinical evaluation demonstrates that the T3 Platform™ software performs comparably to the predicate device that is currently marketed for the same intended use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...