K202306 · Etiometry, Inc. · MWI · Nov 25, 2020 · Cardiovascular
Device Facts
Record ID
K202306
Device Name
T3 Platform software
Applicant
Etiometry, Inc.
Product Code
MWI · Cardiovascular
Decision Date
Nov 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes: To remotely consult regarding a patient's status, and To remotely review other standard or critical near real-time patient data in order to aid in clinical decisions and deliver patient care in a timely manner.
Device Story
T3 Platform software aggregates, stores, and visualizes physiologic data from bedside monitors for ICU clinicians. Inputs include airway flow/pressure, blood pressure (invasive/non-invasive), BIS, cardiac output, heart rate, SpO2, respiratory rate, temperature, and lab results (blood gases, CBC, lactic acid). The Risk Analytics Engine processes these inputs to calculate IDO2 Index (inadequate oxygen delivery) and IVCO2 Index (inadequate CO2 ventilation). The system provides a web-based interface for remote patient status review, trend analysis, and trajectory assessment. It features a Tele-medicine Census view for multi-patient monitoring and EMR integration via Single Sign-on. The device does not provide alarms or waveform displays and is not for active patient monitoring. It aids clinical decision-making by providing quantitative indices of cardiovascular and respiratory status. Benefits include improved visibility into patient trends and streamlined clinical workflows.
Clinical Evidence
Performance validated using clinical data sets. Software verification and validation conducted per FDA guidance. Testing confirmed that the updated hemodynamic model for IDO2 and the inclusion of regional oximetry as a noise proxy for mixed venous oxygen saturation maintain safety and effectiveness comparable to the predicate. No clinical trial outcomes (e.g., sensitivity/specificity) were explicitly reported in the summary.
Technological Characteristics
Software-only solution; web-based visualization; compatible with standard browsers. Features include data aggregation, storage, and risk analytics engine. Connectivity via hospital data sources with support for encrypted feeds. Integrates with EMR via Single Sign-on. Algorithm class: Hemodynamic modeling for oxygen delivery and ventilation indices.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in intensive care. IDO2 Index™ indicated for post-surgical patients aged 0-12 years (≥2 kg). IVCO2 Index™ indicated for invasively ventilated patients aged 29 days-12 years (≥2 kg).
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
November 25, 2020
Etiometry, Inc. Tim Hanson Director of RA/QA 280 Summer St. 4th Floor Boston, Massachusetts 02210
Re: K202306
Trade/Device Name: T3 Platform software Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, PLB Dated: August 5, 2020 Received: August 14, 2020
Dear Tim Hanson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for LT Stephen Browning Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202306
#### Device Name
The T3 Platform™ software features the T3 Data Aggregation & Visualization software module version 5.0 and the T3 Risk Analytics Engine software module version 6.0.
#### Indications for Use (Describe)
The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes:
To remotely consult regarding a patient's status, and
To remotely review other standard or critical near real-time patient data in order to aid in clinical decisions and deliver patient care in a timely manner.
The T3 Data Aggregation & Visualization software module can display numeric physiologic data captured by other medical devices:
Airway flow, volume, and pressure Arterial blood pressure (invasive and non-invasive, systolic, diastolic, and mean) Bispectral index (BIS, signal quality index, suppression ratio) Cardiac Index Cardiac output Central venous pressure Cerebral perfusion pressure End-tidal CO2 Heart rate Heart rate variability Intracranial pressure Left atrium pressure Oxygen saturation (intravascular, regional, SpO2) Premature ventricular counted beats Pulmonary artery pressure (systolic, diastolic, and mean) Pulse pressure variation Pulse Rate Respiratory rate Right atrium pressure Temperature (rectal, esophageal, tympanic, blood, core, nasopharyngeal, skin) Umbilical arterial pressure (systolic, diastolic, and mean)
The T3 Data Aggregation & Visualization software module can display laboratory measurements including arterial and venous blood gases, complete blood count, and lactic acid.
T3 Data Aggregation & Visualization software module can display information captured by the T3 Risk Analytics Engine software module.
The T3 Risk Analytics Engine software module calculates two indices: the IDO2 Index™ for inadequate delivery of oxygen and the IVCO2 Index™ for inadequate ventilation of carbon dioxide.
IDO2 Index™ is indicated for use by health care professionals with post-surgical patients aged zero days to twelve vears
{3}------------------------------------------------
weighing 2 kg or more under intensive care. IDO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module, version 5.0 or higher. When the IDO2 Index™ is increasing, it means that there is an increasing risk of inadequate oxygen delivery and attention should be brought to the IDO2 Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements.
IVCO2 Index™ is indicated for use by health care professionals with invasively ventilated patients aged 29 days to 12 years weighing 2 kg or more under intensive care. IVCO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module, version 5.0 or higher. When the IVCO2 Index™ is increasing, it means that there is an increasing risk of inadequate carbon dioxide ventilation and attention should be brought to the IVCO2 Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the interpretation statements.
# WARNINGS:
Do not use the T3 Platform software as an active patient monitoring system. Do not use the T3 Platform software to replace any part of the hospital's device monitoring. Do not rely on the T3 Platform software as the sole source of patient status information. Do not use the IVCO2 Index as a substitute for arterial blood gases.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "etiometry" in a sans-serif font. The letters are a gradient of blue, with the left side of the word being a darker blue and the right side being a lighter blue. Above the word is a curved line that forms an arc, with a blue triangle at the top of the arc.
October 28, 2020
This 510(k) summary has been prepared in accordance with Title 21 CFR §807.92 and FDA's guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications 510(k)" July 28, 2014
4.1 510(k) Submitter
Timothy Hanson, Director of Regulatory Affairs and Quality Assurance 280 Summer St., 4th Floor Boston, MA 02210 Tel: 857.366.9333 ext. 2020 Email: THanson@etiometry.com
#### 4.2 Device
| Item | Description |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name | T3 Platform™ software (T3 Data Aggregation & Visualization software<br>module version 5.0 and T3 Risk Analytics Engine software module<br>version 6.0) |
| Device Common/Usual Name | Data Management Software (without alarms) |
| Classification Name | Cardiac monitor (including cardiotachometer and rate alarm) |
| Classification Number | 870.2300 |
{5}------------------------------------------------
| Item | Description |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Class | Class II<br>The primary code is MWI: monitor, physiological, patient (without<br>arrhythmia detection or alarms).<br>The secondary code is PLB: automated calculation of a summary index<br>(or indices) based on several individual measured vital sign inputs. |
## 4.3 Predicate Device
The T3 Platform™ software featuring the T3 Data Aggregation & Visualization software module version 3.3 and the T3 Risk Analytics Engine software module version 5.0, cleared under K190273.
### 4.4 Device Description
The Tracking, Trajectory, Trigger (T3) intensive care unit software solution allows clinicians and quality improvement teams in the ICU to aggregate data from multiple sources, store it in a database for analysis, and view the streaming data. System features include:
- . Customizable display of physiologic parameters over entire patient stay
- Configurable annotation
- . Web-based visualization that may be used on any standard browser
- . Minimal IT footprint
- Software-only solution - no new bedside hardware required
- . Highly reliable and robust operation
- Auditable data storage
### 4.5 Indications for Use
The T3 Platform™ software features the T3 Data Aggregation & Visualization software module version 5.0 and the T3 Risk Analytics Engine software module version 6.0.
The T3 Data Aggregation & Visualization software module is intended for the recording and display of multiple physiological parameters of the adult, pediatric, and neonatal patients from supported bedside devices. The software module is not intended for alarm notification or waveform display, nor is it intended to control any of the independent bedside devices to which it is connected. The software module is intended to be used by healthcare professionals for the following purposes:
- To remotely consult regarding a patient's status, and
- . To remotely review other standard or critical near real-time patient data in order to utilize this information to aid in clinical decisions and deliver patient care in a timelymanner.
The T3 Data Aggregation & Visualization software module can display numeric physiologic data captured by other medical devices:
- . Airway flow, volume, and pressure
- . Arterial blood pressure (invasive and non-invasive, systolic, diastolic, and mean)
- Bispectral index (BIS, signal quality index, suppression ratio)
- Cardiac Index
- . Cardiac output
{6}------------------------------------------------
- Central venous pressure ●
- . Cerebral perfusion pressure
- . End-tidal CO2
- . Heart rate
- . Heart rate variability
- Intracranial pressure
- . Left atrium pressure
- . Oxygen saturation (intravascular, regional, SpO2)
- Premature ventricular counted beats
- . Pulmonary artery pressure (systolic, diastolic, and mean)
- . Pulse pressure variation
- . Pulse Rate
- . Respiratory rate
- . Right atrium pressure
- Temperature (rectal, esophageal, tympanic, blood, core, nasopharyngeal, skin)
- . Umbilical arterial pressure (systolic, diastolic, and mean)
The T3 Data Aggregation & Visualization software module can display laboratory measurements including arterial and venous blood gases, complete blood count, and lactic acid.
The T3 Data Aggregation & Visualization software module can display information captured by the T3 Risk Analytics Engine software module.
The T3 Risk Analytics Engine software module calculates two indices: the IDO2 Index™ for inadequate delivery of oxygen and the IVCO2 Index™ for inadequate ventilation of carbon dioxide.
IDO2 Index™ is indicated for use by health care professionals with post-surgical patients aged zero days to twelve years weighing 2 kg or more under intensive care. IDO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation & Visualization software module, version 5.0 or higher. When the IDO2 Index™ is increasing, it means that there is an increasing risk of inadequate oxygen delivery and attention should be brought to the IDO2 Index™ presents partial quantitative information about the patient's cardiovascular condition, and no therapy or drugs can be administered based solely on the interpretation statements.
IVCO2 Index™ is indicated for use by health care professionals with invasively ventilated patients aged 29 days to 12 years weighing 2 kg or more under intensive care. IVCO2 Index™ is derived by mathematical manipulations of the physiologic data and laboratory measurements received by the T3 Data Aggregation and Visualization software module, version 5.0 or higher. When the IVCO2 Index™ is increasing, it means that there is an increasing risk of inadequate carbon dioxide ventilation and attention should be brought to the patient. The IVCO2 Index™ presents partial quantitative information about the patient's respiratory condition, and no therapy or drugs can be administered based solely on the interpretation statements.
# WARNINGS:
- Do not use the T3 Platform™ software as an active patient monitoring system. ●
- . Do not use the T3 Platform™ software to replace any part of the hospital's device monitoring.
- Do not rely on the T3 Platform™ software as the sole source of patient status information. ●
- Do not use the IVCO2 Index™ as a substitute for arterial blood gases.
{7}------------------------------------------------
### 4.6 Comparison of Technological Characteristics with the Predicate Device
The subject and predicate T3 Platform™ software have the same Intended Use/Indication for Use. The subject and predicate devices differ with respect to several technological features. (see Table 2).
| Feature/Improvement | Description |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Addition of data from regional oxygen saturation<br>monitors as an input for the calculation of the<br>IDO2 Index™ | Data from regional oxygen saturation monitors, when available, have<br>been added as an input to the Risk Analytics Engine module to<br>facilitate the detection of inadequate oxygen delivery and<br>respectively improve the performance of the IDO2 Index™.<br>Specifically, data from regional oximetry are used as a noise proxy of<br>mixed venous oxygen saturation which the hemodynamic model<br>interprets in the context of other physiologic measurements to infer<br>IDO2 value. The performance of the IDO2 Index™ was evaluated with<br>and without regional oximetry to confirm the utility of the<br>measurement as an input. |
| Measure Contributions signifying measurement<br>inputs which have the highest contribution to an<br>increase in the IDO2 Index™ or the IVCO2 Index™ | To increase the transparency and understanding of the risk indices<br>computed by the platform, the computations have been<br>supplemented with an external routine which computes how either<br>removing an existing measurement as an input (if the measurement<br>is not in the minimum data set required for the computation of the<br>index) or setting a measurement to what the algorithms perceive as<br>nominal (if the measurement is part of the minimum data set) affects<br>the individual indices. This computing routine enables assessment of<br>the "contribution" of each available input in both the IDO2 Index™ or<br>the IVCO2 Index™. The three measurements with the highest<br>contribution, i.e. the measurements whose removal or setting to<br>nominal leads to the highest drop in a particular index are then<br>displayed as the major contributing measurements at the particular<br>time instance. |
| User selectable thresholds for defining<br>Inadequate Oxygen Delivery and Inadequate<br>Ventilation of Carbon Dioxide | The previously cleared version of the software computed the IDO2<br>Index™ and the IVCO2 Index™ as the likelihood of SvO2 < 40% and<br>PaCO2 > 50 mmHg respectively. Using the same computation<br>framework, the current version enables several more thresholds that<br>a clinical user can configure. This facilitates individualizing the indices<br>based on patient condition or other clinical considerations. |
| Modification of minimum data required to<br>compute IVCO2 Index™ | The minimum set for the IVCO2 Index™ to exclude the arterial blood<br>gases requirement. |
| Support for encrypted data feeds into the<br>platform | To improve the cybersecurity of the platform, the software<br>introduces the ability to accept encrypted data feeds from hospital<br>data sources. |
| Display of complete lab panel information | The software introduces a detailed view of a plotted lab result to<br>provide additional clinical context to the results. This allows clinicians<br>to both contextualize lab results with physiologic data, but also with<br>other lab results within the panel. |
| Embedded Standalone Platform Manual | To further ensure labeling availability and distribution, the software<br>has embedded the complete platform manual as part of the release.<br>Users can easy access, consult, and/or download the manual within<br>the system. |
| Feature/Improvement | Description |
| Addition of Tele-medicine Census view in the<br>existing User Interface | The User Interface introduces a new feature to facilitate the viewing<br>of the status of all patients in an ICU in a single display, which can<br>serve as a starting point for a more detailed review of individual<br>patient information. The tool supports interpreting trends for<br>selected variables, understanding the trajectory of patients, and<br>supporting bedside clinicians in their decision making. For each bed<br>in the ICU, the bed number is displayed along with the last name and<br>age if there is currently a patient in that bed. This tool differs from<br>the Census Overview by displaying for each patient the most recent<br>vital signs and algorithm values in tiles containing graphs instead of a<br>table of rows |
| Integration of Single Sign-on | To facilitate access to the platform, the software has an added<br>capability to integrate with an Electronic Medical Record system<br>(EMR). A user can navigate to display T3 patient information from the<br>EMR with a single click, which automatically authenticates them in T3<br>and navigates them to the patient of interest. |
| Miscellaneous Updates of the Hemodynamic<br>Model Employed to Infer the IDO2 Index™ | The hemodynamic model powering the IDO2 Index™ has been<br>updated by improving specific modeled physiologic effects such as<br>the adaptable relationship between pulse pressure and stroke<br>volume which now utilizes a logarithmic relationship to facilitate<br>linear estimation. The updated IDO2 Index™ algorithm was subject to<br>full performance clinical testing. |
| Addition of a Retrospective Teaching Census | To facilitate software education and training, the platform introduces<br>the ability to collate, de-identify, and save a specific patient case in a<br>hospital specific teaching census for use as a training resource. |
| Repair incorporation of incorrectly labeled SvO2<br>measurements into the algorithm | The software has been modified to reject measurements which may<br>not correctly be associated with mixed or central venous oxygen<br>saturation. The software classifies venous oxygen saturation<br>measurements as either validated or non-validated and incorporates<br>the measures under different conditions based on that classification. |
{8}------------------------------------------------
Table 2: Summary of Changes
# 4.7 Performance Data
The changes to the algorithm were validated using clinical data sets. The software verification and validation testing were conducted for the subject device, and documentation was provided in accordance with FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", May 11, 2005. The results of this testing demonstrate that the safety and effectiveness of the subject T3 Platform™ software (subject device) are comparable to that of the predicate T3 Platform™ software (target device).
### 4.8 Conclusions
Substantial equivalence of the T3 Platform™ software is demonstrated through performance testing. The T3 Platform™ software has the equivalent design, features and functionality as the predicate T3 Platform™ software with few exceptions and these exceptions do not affect the safety or effectiveness of the system.
No new questions of safety or effectiveness are raised as a result of the differences when compared to the predicate device and the data provided in the submission show that the subject device is substantially equivalent to the legally-marketed predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.