OMRON MODEL HBP-2070 VITAL SIGNS MONITOR

K082812 · Omron Healthcare, Inc. · MWI · Oct 22, 2008 · Cardiovascular

Device Facts

Record IDK082812
Device NameOMRON MODEL HBP-2070 VITAL SIGNS MONITOR
ApplicantOmron Healthcare, Inc.
Product CodeMWI · Cardiovascular
Decision DateOct 22, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesPediatric

Indications for Use

The HBP-2070 is intended to be used to monitor electrocardiography (ECG),heart rate (HR), noninvasive blood pressure (systolic, diastolic and mean arterial pressure) (NIBP), functional arterial oxygen saturation (SpO2), pulse rate (PR), respiration (RR) temperature (Temp) for adult, pediatric, and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse, or medical specialist.

Device Story

HBP-2070 is a multi-parameter vital signs monitor for hospital use. It collects physiological data via ECG leads, NIBP cuff, SpO2 sensor, and temperature probe. The device processes these inputs to display real-time waveforms and numerical values for HR, NIBP (systolic, diastolic, MAP), SpO2, pulse rate, respiration rate, and temperature. Operated by clinicians (doctors, nurses, technicians), the device provides continuous monitoring to assist in clinical assessment and patient management. It features a 5.6-inch TFT color LCD, thermal recorder, and battery backup. The device utilizes established NIBP and SpO2 modules from previously cleared Omron devices. Output is visualized on-screen for immediate clinical decision-making. The system is designed for hospital environments to support patient care through non-invasive monitoring.

Clinical Evidence

Bench testing only. The device relies on the substantial equivalence of its integrated modules (NIBP M3500 and SpO2 Nellcor NELL-3) to previously cleared devices. Performance specifications for NIBP (accuracy ±3 mmHg), ECG, SpO2, and temperature monitoring were verified against predicate standards.

Technological Characteristics

Multi-parameter monitor with 5.6-inch TFT color LCD (400x320). NIBP: Oscillometric method, semiconductor pressure sensor, dynamic linear deflation. ECG: 3/5 lead, 0.05-40 Hz frequency response. SpO2: Nellcor NELL-3 module, 2-wavelength pulse wave. Temp: Thermistor probe (YSI 400/700). Power: AC mains or internal lead-acid battery. Connectivity: Thermal recorder. Software: Embedded firmware with algorithms identical to K071645. Defibrillator-protected Type CF/BF applied parts.

Indications for Use

Indicated for adult, pediatric, and neonatal patients requiring monitoring of ECG, HR, NIBP, SpO2, pulse rate, respiration, and temperature in hospital or hospital-type settings. No contraindications.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## Omron HBP-2070 ## Non-Confidential Summary of Safety and Effectiveness | | Page 1 of 5<br>23-Sep-08 | OCT 22 2008 | | |----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Omron Healthcare, Inc.<br>1200 Lakeside Dr.<br>Bannockburn, IL 60015 | Tel - 847-247-5609<br>Fax - 847-680-6269 | | | | Official Contact: | Ranndy Kellogg - VP Marketing & Product Development | | | | Proprietary or Trade Name: | HBP-2070 | | | | Common/Usual Name: | Monitor, Physiological, Patient (without arrhythmia detection or<br>alarms) | | | | Classification Name/Code: | DXN - System, Measurement, Blood-pressure, Non-invasive | | | | Device: | Model - HBP-2070 | | | | Predicate Devices: | Omron - Press-Mate Advantage - K973637 | | | | Patient Population: | Adult, Pediatric, Neonate | | | | Environment of Use: | All areas of a hospital and hospital-type facilities | | | | Contraindications: | None | | | | | Summary of substantial equivalence | | | | Indications for Use | Monitors: | | | | | Non-invasive blood pressure (NIBP), | same as predicate device | | | | Oxygen saturation(SpO2), | | | | | ECG, | | | | | Respiration rate, | | | | | Heart rate (HR), and | | | | | Body Temperature (Temp) | | | | Patient Population | Adult, Pediatric or Neonatal | | | | Environment of use | All areas of a hospital and hospital-type | | | | | facilities | | | | Physical Characteristics | | | | | Dimensions (mm) | 172 (W) × 216 (D) × 228 (H) mm | 257 (W) × 210 (D) × 152 (H) mm | | | Weight | Approximately 6.0kg<br>(including internal battery) | Approximately 3.5kg<br>(including internal battery) | | | Environmental Condition | | | | | Operating conditions | 0 to 40 °C,<br>30 to 85 %RH non-condensing | 5 to 40 °C,<br>30 to 85 %RH non-condensing | | | Storage conditions | -20 to 70 °C,<br>10 to 100 %RH non-condensing | -20 to 60 °C,<br>10 to 95 %RH non-condensing | | | Display | | | | | Screen size | 5.6inch TFT color LCD | 7inch TFT color LCD | | | Resolution | 400 × 320 pixel | 800 × 480 pixel | | | Number of Traces | 1 or 2 waveforms | 1 or 3 waveforms | | | Electrical | | | | | Power source | AC Mains<br>Battery (lead acid) | AC Mains<br>Battery (Lithium ion) | | | Power range | AC: 100-120V 50 / 60Hz<br>Battery: 12V 1.8Ah | AC: 100-240V 50 / 60Hz<br>Battery: 11.1V 2.4Ah or 7.2Ah | | | Battery operation time | A battery typically provides operating<br>time of 30 minutes when fully charged<br>with no printing, no external<br>communication, no audible alarm sound<br>and one NIBP measurement per 5<br>minutes at 25°C. | A battery typically provides operating<br>time of 1 hour and 6 hour when fully<br>charged with no printing, no external<br>communication, no audible alarm sound<br>and one NIBP measurement per 10<br>minutes at 25°C. | | ### Device Description: The proposed vital signs monitor is a modification of a previously cleared Colin Press Mate Advantage (K973637). The proposed model, HBP-2070, measures and monitors: - . Noninvasive blood pressure (NIBP) - . Oxygen saturation (SpO2) - . ECG - Respiration rate (RR) . - . Heart Rate (HR), and - Temperature . ## Indications for Use: The HBP-2070 is intended to be used to monitor electrocardiography (ECG),heart rate (HR), noninvasive blood pressure (systolic, diastolic and mean arterial pressure) (NIBP), functional arterial oxygen saturation (SpO2), pulse rate (PR), respiration (RR) temperature (Temp) for adult, pediatric, and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse, or medical specialist. {1}------------------------------------------------ Omron HBP-2070 # Non-Confidential Summary of Safety and Effectiveness Page 2 of 5 23-Sep-08 {2}------------------------------------------------ Omron HBP-2070 ## Non-Confidential Summary of Safety and Effectiveness Page 3 of 5 23-Sep-08 Predicate Press-Mate Advantage K973637 New Model HBP-2070 series Recorder Type Thermal same as predicate device Resolution 8 dot/min same as predicate device 1 to 2 channels Number of Channel same as predicate device Paper width 58 min 50 mm Paper speed 6.25,25 and 50mm/sec 25 and 50mm/sec 【NIBP】 Measurement method Oscillometric method same as predicate device Patient target Adult, Pediatric, Neonatal same as predicate device SYS - Adult) 60 - 250 mmHg NIBP same as predicate device Measurement range Pediatric / Neonatal) 40 - 120 mmHg MAP - Adult) 45 -- 235 mmHg Uses another NIBP module - M3500 Pediatric / Neonatal) 30 -100 mmHg which has been used in our predicate DIA - Adult) 40 - 200 mmHg HBP-T105 (K071645) Pediatric / Neonatal) 20 -90 mmHg Pressure display range 10 - 300 mmHg 0-300mmHg ( Adult ) 0-150mmHg(Neo) Accuracy of pressure indicator Within ±3 mmHg or 1 % of reading Within ±3 mmHg Pressure sensor Semiconductor pressure sensor same as predicate device Pulse rate: 40 to 240 beats/min. Pulse rate range same as predicate device Accuracy of pulse rate Within ±2 beats/min or ±2% of reading same as predicate device DC Rolling diaphragm pump Inflation method same as predicate device Deflation method Dynamic linear deflation same as predicate device Shock protection Type BF(Defibrillator protected) Type CF(Defibrillator protected) 【EGG】 HR measure range 30 - 300 BPM 30 - 300BPM HR Accuracy ±3 BPM or ±5% whichever is greater ±1 BPM or ±1% whichever is greater Lead 3 / 5 Lead same as predicate device LeadL, J, J, aVR,aVL,aVF,V(Chest) Lead off detection Detected and displayed same as predicate device Input dynamic range ±5mV AC,±300mV DC same as predicate device Voltage range ±0.5mV ~ same as predicate device ±5mV Signal Width same as predicate device 40 ms ~ 120 ms ( Q to S ) Frequency response Low Extend same as predicate device 0.05 Hz - 40 Hz Filter None 0.5 Hz - 30 Hz Monitor 0.32 Hz - 40 Hz 0.5 Hz - 40 Hz Respiration rejection None 1 Hz - 40 Hz ECG size ×1/2,×1,×2,×4 ×1/4,×1/2,×1.×1.5,×2 Display Sweep Size 6.25.12.5.25 mm/sec 12.5,25,50 mm/sec Shock protection Type CF(Defibrillator protected) same as predicate device {3}------------------------------------------------ Koszá 12 【RESP】 #### Omron HBP-2070 # Non-Confidential Summary of Safety and Effectiveness Page 4 of 5 23-Sep-08 #### Predicate Press-Mate Advantage K973637 New Model HBP-2070 series | Measurement method<br>Range<br>Accuracy of pulse rate<br>Lead<br>Display Sweep Size<br>Shock protection<br>【SpO2】 | Trans-thoracic impedance<br>3 to 150 breaths/min<br>$\pm$ 3 breaths/min<br>RA to LL<br>6.25, 12.5, 25 mm/sec<br>Type CF(Defibrillator protected) | same as predicate device<br>3 to 120 breaths/min<br>same as predicate device<br>RA to LA<br>same as predicate device<br>same as predicate device | |-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------| | SpO2 module | Nellcor MP-203 | Nellcor NELL-3<br>Predicate Omron BP-S510 (K063690) | | Measurement method | 2 wave length pulse wave type | same as predicate device | | SpO2 display range | 50 - 100 % | 1 - 100 % | | Accuracy | Adult - 70% - 100% $\pm$ 2digits<br>1% - 69% unspecified<br>Pediatric / Neonate<br>70% - 100% $\pm$ 3digits<br>1% - 69% unspecified | same as predicate device | | Display Sweep Size | 12.5,25,50 mm/sec | same as predicate device | | Pulse rate display range | 20 - 250 beats/min | same as predicate device | | Accuracy of pulse rate<br>Shock protection<br>【TEMP】 | Within $\pm$ 3 beats/min<br>Type BF(Defibrillator protected) | same as predicate device<br>Type CF(Defibrillator protected) | | Measurement method | Thermistor probe YSI 400 or 700 | same as predicate device | | Parameter Displayed | Temp ( 1ch ) | TI ( 1ch ) | | display range | 15.0 - 45.0°C | same as predicate device | | Accuracy | $\pm$ 0.1 °C (25 °C to 45 °C )<br>$\pm$ 0.2 °C ( 15 °C to less than 25 °C ) | $\pm$ 0.1 °C | | Scale | Selectable from C to F | same as predicate device | | Probe Accuracy | $\pm$ 0.1 °C | same as predicate device | | Shock protection | Type BF(Defibrillator protected) | Type CF(Defibrillator protected) | The base modifications are: - Change in the SpO2 module to Nell 3 which has been used and cleared in our predicate . BPS S510 (K063690) - Clange in the NIBP module to M3500 which has been used in our predicate HBP- T105 . (K071645). - . Software updates {4}------------------------------------------------ ## Non-Confidential Summary of Safety and Effectiveness Page 5 of 5 23-Sep-08 ## Differences Between Other Legally Marketed Predicate Devices The Omron HBP-2070 vital signs monitor is viewed as substantially equivalent to the predicate device because: #### Indications - - Identical to predicate K973637 . ### Technology - - Identical algorithms to predicate K071645 . ### Materials - - The materials in patient contact are identical to predicate devices as listed in Section 15. . ### Environment of Use - - Identical to predicate K973637 . There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate devices. {5}------------------------------------------------ Public Health Service Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle emblem. ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 22 2008 Omron Healthcare, Inc. c/o Mr. Paul Dryden ProMedic, Inc. 24301 Woodsage Drive Bonita Springs, FL 34134-2958 Re: K082812 Trade/Device Name: Model HBP-2070 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Codes: MWI, DXN, DQA, FLL Dated: September 23, 2008 Received: September 24, 2008 Dear Mr. Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {6}------------------------------------------------ Page 2 -Mr. Paul Dryden CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ## Indications for Use Statement Page 1 of 1 510(k) Number: KOS Z& 2 (To be assigned) HBP-2070 Device Name: ## Indications for Use: The HBP-2070 is intended to be used to monitor electrocardiography (ECG),heart rate (HR), noninvasive blood pressure (systolic, diastolic and mean arterial pressure) (NIBP), functional arterial oxygen saturation (SpO2), pulse rate (PR), respiration (RR) temperature (Temp) for adult, pediatric, and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse, or medical specialist. Prescription Use XX (Part 21 CFR 801 Subpart D) or Over-the-counter use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Atuthe for B Zuckerman sion of Cardiovascular Devices 510(k) Number K082812
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...