The T-105 Series Vital Signs Monitor is intended to monitor a single patient's vital signs in the hospital, acute care settings, outpatient surgery, healthcare practitioner facilities or in an environment where patient care is provided by qualified healthcare personnel who will determine when use of this device is indicated, based upon their professional assessment of the patient's medical condition. The patient populations include adult, pediatric and neonatal. The device is capable of monitoring: . Pulse rate (via oximetry data) . Non-invasive pressure (systolic, diastolic and mean oscillometric (NIBP) . Temperature . Blood Oxygen Saturation (SpO2 via finger oximeter) This device is intended for use by qualified healthcare personnel trained in its use.
Device Story
HBP T-105 Series Vital Signs Monitor; monitors single patient vital signs including pulse rate, NIBP, temperature, and SpO2. Inputs: oscillometric pressure sensor, finger oximeter, electronic predictive thermometer. Operation: oscillometric NIBP measurement; SpO2 via Nellcor or Masimo modules; predictive temperature sensing. Output: LED display of vital sign data. Used in hospitals/clinics by qualified healthcare personnel. T-105S model performs manual BP activation; T-105 allows interval settings. Output informs clinical assessment of patient condition. Benefits: provides real-time physiological monitoring to support clinical decision-making.
Clinical Evidence
Bench testing only. Comparison of performance specifications (accuracy of pressure, pulse, SpO2, and temperature) against predicate devices. No clinical study data provided.
Technological Characteristics
Oscillometric NIBP measurement; semiconductor pressure sensor; DC rolling diaphragm pump; electromagnetic solenoid valve. SpO2 via Nellcor (NELL-1) or Masimo (MS-11) modules. Predictive electronic thermometer. LED display. Power: 12V 3.2Ah battery or AC adapter. Type BF defibrillator-protected. Dimensions: 239x239x150mm. Weight: 7.7 lbs.
Indications for Use
Indicated for monitoring vital signs (pulse rate, NIBP, temperature, SpO2) in adult, pediatric, and neonatal patients in clinical settings (hospitals, acute care, outpatient surgery, practitioner facilities) by qualified healthcare personnel. No contraindications.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
ィ
Omron
# Non-Confidential Summary of Safety and Effectiveness
Page 1 of 4 12-Jun-07
| Omron Healthcare, Inc.<br>1200 Lakeside Dr.<br>Bannockburn, IL 60015 | Tel - 847-247-5609<br>Fax - 847-680-6269 |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Ranndy Kellogg - VP Marketing & Product Development |
| Proprietary or Trade Name: | HBP T-105 Series Vital Signs monitor |
| Common/Usual Name: | Monitor, Physiological, Patient (without arrhythmia detection or<br>alarms) |
| Classification Name/Code: | MWI - Monitor, Physiological, Patient (without arrhythmia<br>detection or alarms) |
| Device: | Models HBP T-105 and T-105S |
| Predicate Devices: | Colin - Press-Mate PM-2100 - K022537<br>Omron - HEM 907 - K001848<br>Omron - HEM 757 - K001670<br>Nellcor - N595 – K012891<br>Nellcor - Oximax - K052186<br>Masimo - RAD 5 - K033296<br>Masimo - LNCS – K041815 |
### Device Description:
The Omron HBP T-105 Series is a modification of the Colin Press-Mate PM-2100 and is intended to monitor a single patient's vital signs. The device is capable of monitoring:
- . Pulse rate (via oximetry data):
- Non-invasive pressure (systolic, diastolic and mean oscillometric (NIBP); .
- . Temperature; and
- . Blood Oxygen Saturation (SpO2 via finger oximeter)
### Indications for Use:
The T-105 Series Vital Signs Monitor is intended to monitor a single patient's vital signs in the hospital, acute care settings, outpatient surgery, healthcare practitioner facilities or in an environment where patient care is provided by qualified healthcare personnel who will determine when use of this device is indicated, based upon their professional assessment of the patient's medical condition. The patient populations include adult, pediatric and neonatal. The device is capable of monitoring:
{1}------------------------------------------------
## Non-Confidential Summary of Safety and Effectiveness Page 2 of 4 12-Jun-07
- Pulse rate (via oximetry data) ◆
- Non-invasive pressure (systolic, diastolic and mean oscillometric (NIBP) .
- . Temperature
- . Blood Oxygen Saturation (SpO2 via finger oximeter)
This device is intended for use by qualified healthcare personnel trained in its use.
### Patient Population: Adult, pediatric, neonate
Hospital, acute care settings, outpatient surgery, healthcare practitioner Environment of Use: facilities or in an environment where patient care is provided by qualified healthcare personnel who will determine when use of this device is indicated, based upon their professional assessment of the patient's medical condition.
#### Contraindications: None
| Specification | Predicate<br>PM2100 - K022537 | New Models<br>HBP-T105 T-105S |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Power source | Battery (Tin acid) or AC | Battery(Lead-acid) or AC adapter |
| Power range | AC : 100-120V,220-240V,50/60Hz<br>Battery6V 5Ah | AC : 100-115V,220-240V,50/60Hz<br>Battery12V 3.2Ah |
| Display | Monocolor LCD and LED | LED |
| Operating conditions | 0 to 50 ℃, 30 to 85 %RH | 0 to 40 °C, 30 to 85 %RH |
| Storage conditions | -20 to 70 °C, 30 to 85 %RH | -20 to 60 ℃, 10 to 95 %RH |
| Dimensions (mm) | 240 (W) × 250 (D) × 238 (H) mm | 239 (W) × 239 (D) × 150 (H) mm |
| Weight (without batteries) | 7.5 lbs | 7.7 lbs |
| BP module | M2600 | M3600 |
| Measurement method | Oscillometric method | same as predicate device |
| Patient target | Adult/Pediatric/Neonatal | same as predicate device |
| Measurement range | Pressure: 0 to 300 mmHg<br>Pulse rate: 40 to 240 beats/min. | Pressure: 0 to 299 mmHg<br>Pulse rate: 40 to 240 beats/min. |
| Pressure sensor | Semiconductor pressure sensor | same as predicate device |
| Air Control Valve | Electromagnetic solenoid | Omron HEM907 - K001848 |
| Applicable cuff | Regular sized cuff | Omron HEM907 - K001848 |
| Specification | Predicate<br>PM2100 - K022537 | New Models<br>HBP-T-105 / T-105S |
| Accuracy of pressure indicator | Within ±3 mmHg or 1 % of reading | same as predicate device |
| Accuracy of pulse rate | Within ±2 beats/min or ±2% of reading | same as predicate device |
| Inflation method | DC Rolling diaphragm pump | same as predicate device |
| Deflation method | Dynamic linear deflation | same as predicate device |
| Quick measurement function of<br>Blood Pressure | No algorithm | Omron algorithm<br>HEM-757 - K001670 |
| SpO2 Measurement method | 2 wave length pulse wave type | same as predicate device |
| SpO2 display range | 0 to 100 % | same as predicate device |
| SpO2 module | Nellcor NELL-3(MP-506) | Nellcor NELL-1<br>N-595 - K012891<br>same technology as NELL-1.<br>Masimo MS-11<br>RAD-5 - K033296<br>same technology as MS-11.<br>User can select which one |
| Accuracy of SpO2 | D-25 disposable sensor<br>SpO2:70-100% ±2%<br>other ranges unspecified<br>N-25 disposable sensor<br>SpO2:70-95% ±2%<br>other ranges<br>unspecified | Nellcor - Same as predicate<br>OXIMAX Sensors - K052186<br>Add option of - Masimo<br>SpO2:70-100% ±2%<br>LNCS sensors - K041815 |
| Pulse rate display range | 20 - 250 beats/min | Nellcor -same as predicate device<br>Masimo -25 - 240 beats/min |
| Accuracy of pulse rate | Within ±3 beats/min | Nellcor - same as predicate device<br>Masimo - Within ±3 beats/min |
| Temperature<br>Measurement method | TURBO TEMP electronic<br>predictive thermometer | same as predicate device |
| TEMP display range | Predictive mode: 35.6 - 41.1℃ / 96-<br>106°F<br>Monitor mode: 26.7 - 41.1º / 80 - 106ºF | same as predicate device |
| Accuracy of TEMP | ±0.1°C / ±0.2°F | same as predicate device |
| Scale | Selectable from °C to °F | same as predicate device |
| Shock protection | Type BF(Defibrillator protected) | same as predicate device |
### Summary of substantial equivalence
{2}------------------------------------------------
# Non-Confidential Summary of Safety and Effectiveness Page 3 of 4
12-Jun-07
{3}------------------------------------------------
K071645
Omron p4/4
# Non-Confidential Summary of Safety and Effectiveness Page 4 of 4 12-Jun-07
### Differences Between Other Legally Marketed Predicate Devices
The Model HBP T-105 Series is viewed as substantially equivalent to the following predicate device - Colin Press-Mate PM-2100 - K022537.
The difference between the Model T-105 and T-105S is that the T-105S model does not have the ability to set the time interval of the BP measurement. The T-105S model BP is done on an as needed / manual activation basis. All other features, specifications, etc. are otherwise identical.
There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 2 2007
Omron Healthcare, Inc. c/o Mr. Paul Dryden, ProMedic, Inc. 3460 Pointe Creek Court # 102 Bonita Springs, FL 34134-2015
Re: K071645
HBP-105 and T-105S Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: MWI Dated: June 12, 2007 Received: June 15, 2007
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Mr. Paul Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Ashley B. Brown
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Omron
Page 1 of 1
Device Name: HBP T-105 and T-105S
## Indications for Use:
The T-105 Series Vital Signs Monitor is intended to monitor a single patient's vital signs in the hospital, acute care settings, outpatient surgery, healthcare practitioner facilities or in an environment where patient care is provided by qualified healthcare personnel who will determine when use of this device is indicated, based upon their professional assessment of the patient's medical condition. The patient populations include adult, pediatric and neonatal. The device is capable of monitoring:
- . Pulse rate (via oximetry data)
- . Non-invasive pressure (systolic, diastolic and mean oscillometric (NIBP)
- . Temperature
- . Blood Oxygen Saturation (SpO2 via finger oximeter)
This device is intended for use by qualified healthcare personnel trained in its use.
Prescription Use XX (Part 21 CFR 801 Subpart D) Or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Antilu Bvanu An BDZ
**510(k) Number** K071645
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.