DINAMAP PRO SERIES MONITOR, MODELS 110, 210, 310, 410
K020022 · Ge Medical Systems Information Technologies · MWI · Feb 1, 2002 · Cardiovascular
Device Facts
| Record ID | K020022 |
| Device Name | DINAMAP PRO SERIES MONITOR, MODELS 110, 210, 310, 410 |
| Applicant | Ge Medical Systems Information Technologies |
| Product Code | MWI · Cardiovascular |
| Decision Date | Feb 1, 2002 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 870.2300 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The DINAMAP® Pro Series 110-410 Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intra-hospital transport. Vital signs parameters include non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, and/or oxygen saturation (pulse oximetry) and/or temperature. The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/ surgical, labor and delivery, endoscopy, cardiac step-down. It can also be used in satellite areas, physicians' offices, or alternate care settings.
Device Story
Portable vital signs monitor; configurations include NIBP, pulse rate, temperature, and pulse oximetry. Inputs: NIBP cuff, temperature probe, SpO2 sensor. Operation: NIBP via ASAP algorithm; thermometry via Alaris IVAC Turbo; SpO2 via Masimo SET. Output: Real-time vital signs displayed on screen and printed via integrated printer. Used in hospital departments (emergency, recovery, etc.) and physician offices by healthcare professionals. Benefits: Continuous patient monitoring during bedside care or transport.
Clinical Evidence
Bench testing only. Testing included mechanical, environmental, electromagnetic compatibility, and electrical safety assessments.
Technological Characteristics
Portable monitor with integrated printer; AC or internal lead-acid battery power. Parameter technologies: NIBP (ASAP algorithm), thermometry (Alaris IVAC Turbo), SpO2 (Masimo SET).
Indications for Use
Indicated for monitoring vital signs (NIBP, pulse rate, temperature, SpO2) in adult, pediatric, or neonatal patients in clinical settings including bedside or intra-hospital transport.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
- DINAMAP® Pro Series Monitor 100-400 (K992638)
- Masimo (K992238)
- Alaris Medical System (K955846)
- DINAMAP Pro 1000 (K002248)
Related Devices
- K022193 — DINAMAP PROCARE SERIES MONITOR, MODELS 100N, 200N, 300N, 400N · Ge Medical Systems Information Technologies · Aug 1, 2002
- K014255 — DINAMAP PROCARE SERIES MONITOR; MODEL 100, 200, 300, 400 · Ge Medical Systems Information Technologies · Jul 2, 2002
- K102426 — CARESCAPE MODEL V100 VITAL SIGNS MONITOR · Ge Medical Systems Information Technologies · Oct 20, 2010
- K021435 — DINAMAP PRO SERIES MONITOR, MODELS 110N,210N,310N,410N · Ge Medical Systems Information Technologies · May 22, 2002
- K073203 — CARESCAPE V100 · Ge Medical Systems Information Technologies · Nov 29, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
KO20022
FEB 01 2002
#### SUMMARY OF SAFETY AND EFFECTIVENESS Dec 31 ,2001
# DINAMAP® Pro Series 110-410 Monitor
### A. Submitter
GE Medical Systems Information Technologies 4502 Woodland Corporate Boulevard Tampa, FL 33614
# B. Company Contact
Thomas J English Director, Regulatory Affairs Phone: 813-887-2107 Fax: 813-887-2413
### C. Common Name
Physiological or Vital Signs Monitor, Patient Monitor
| Classification Name | Product Code | 21 CFR |
|--------------------------------------------------|--------------|----------|
| System, Measurement, Blood Pressure, Noninvasive | DXN | 870.1130 |
| Computer, Blood Pressure | DSK | 870.1110 |
| Alarm, Blood Pressure | DSJ | 870.1100 |
| Oximeter | DQA | 870.2700 |
| Oximeter, Ear | DPZ | 870.2710 |
| Thermometer, Clinical Electronic | FLL | 880.2910 |
| Recorder, Paper Chart | DSF | 870.2810 |
#### D. Predicate/Legally Marketed Devices
DINAMAP® Pro Series Monitor 100-400-K992638 Critikon Company, LLC
# E. Device Description
The DINAMAP Pro Series 110-410 Monitor is a prescription device intended for use only by health care professionals. Four configurations of the monitor-all with integrated printer-will offer the following vital signs parameters:
- DINAMAP 110: Non-Invasive Blood Pressure and Pulse Rate
- DINAMAP 210: Non-Invasive Blood Pressure and Pulse Rate, Temperature .
- DINAMAP 310M: Non-invasive Blood Pressure and Pulse Rate, Pulse Oximetry
- DINAMAP 410M: Non-Invasive Blood Pressure and Pulse Rate, Pulse Oximetry . and Temperature.
This portable device includes an integrated printer and is capable of operation from an external AC mains power source or an internal lead-acid rechargeable battery.
{1}------------------------------------------------
K020022
P.2/2
# F. Intended Use
The DINAMAP® Pro Series 110-410 Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intra-hospital transport. Vital signs parameters include non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, and/or oxygen saturation (pulse oximetry) and/or temperature. The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/ surgical, labor and delivery, endoscopy, cardiac step-down. It can also be used in satellite areas, physicians' offices, or alternate care settings.
# G. Technological Characteristics
The DINAMAP® Pro Series 110-410 Monitor has the same technological characteristics as the predicate device, the DINAMAP® Pro Series 100-400 Monitor. There are no new technologies used on the DINAMAP® Pro Series 110-410 Monitor.
### H. Parameter Technology
The DINAMAP® Pro Series 110-410 Monitor has the following parameter technologies:
•NIBP ASAP algorithm wholly implemented from the DINAMAP Pro 1000 Monitor
- · Wholly implemented Alaris IVAC Turbo thermometry technology
- · Wholly implemented Masimo SpO2 SET technology
## I. Testing
Several bench studies were conducted which demonstrate safety and effectiveness of the DINAMAP® Pro Series 110-410 Monitor:
- · Mechanical and Environmental
- · Electromagnetic Compatibility
- · Electrical Safety
#### K. Substantial Equivalence
| Pro Series 110-410 | Predicate Device & Model | 510(k) Numbers |
|--------------------|----------------------------|----------------|
| Monitor | DINAMAP Pro Series 100-400 | K992638 |
| Pulse Oximetry | Masimo | K992238 |
| Temperature | Alaris Medical System | K955846 |
| NIBP | DINAMAP Pro 1000 | K002248 |
Page 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three lines representing its wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# FEB 01 2002
Mr. Thomas English Director, Regulatory Affairs GE Medical Systems Information Technologies 4502 Woodland Corporate Blvd. Tampa, FL 33614
Re: K020022
Trade Name: DINAMAP® Pro Series 110-410 Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (two) Product Code: MWI Dated: December 31, 2001 Received: January 3, 2002
Dear Mr. English:
We have reviewed your Section 510(k) premarket notification of intent to market the indication we have reviewed your Section - Forty promation is substantially equivalent (for the indications felerenced above and nave determined the asketed predicate devices marketed in interstate for use stated in the encrosule, to regally manative personal perice Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food Drug commerce prior to May 20, 1978, the excordance with the provisions of the Federal Food, Drug, devices that have been recarsined in accerdance what a proval application (PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act. The and Cosment Act (Act) that do not require approvial of the general controls provisions of the Act. The You may, therefore, market me devreet, belyer to the or general controls provisions of the 110 labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it II your device is classified (see above) into Existing major regulations affecting your device can
may be subject to such additional controls. Existing major may be subject to such additional controller "Entisang may - " " " " oc found in the Oous overning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Thomas English
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
y
Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): K020022
Device Name: DINAMAP® Pro Series 110-410 Monitor
Indications for Use:
The DINAMAP® Pro Series 110-410 Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intra-hospital transport. Vital signs parameters include non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate and/or temperature and/or oxygen saturation (pulse oximetry). The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/surgical, labor and delivery, endoscopy, cardiac step-down. It can also be used in satellite areas, physicians' offices, or alternate care settings.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
wwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwwww
Concurrence of CDRH, Office of Device Evaluation (ODE) OR Over-The Counter Use Prescription Use (per 21 CFR 801.109 (Optional Format 1-2-96)
4.
00009